Patient Interface Guide
2 Pages
Preview
Page 1
Instructions For Use Additional patient interface systems have been identified which are sufficiently non-restrictive for use with the MaxVenturi. The flow rates for these patient interface systems differ from flow rates of the Fisher & Paykel heated humidifier and Optiflow patient interfaces. The lettered scale on the flow meter allows the user to determine the flow for various alternate patient circuits according to the table on the other side of this card. The actual flow rate for each alternate patient circuit can also be found on the Flow Look up Card attached to the device. Locate the patient interface and delivery circuit in the left hand column. Find the corresponding flow rate by reading to the right under the corresponding column A through F. The lettered columns correspond to the lettered graduations on the flow meter. See section 2.2.3 of the User Manual for specific details on approved patient circuits and interfaces.
R211P23 REV A
WARNING: » Ensure the delivery circuit is not kinked or occluded. » Use only with approved humidifier, circuit and patient interface*. »W ait 15 seconds after a change in oxygen to allow oxygen signal to stabilize before reading. »D o not humidify inlet air. »D o not deliver aerosol medication to the air inlet. »D o not remove oxygen inlet valve or the attached fittings. »D o not operate device with oxygen sensor out of its port. »N ot intended for use life sustaining therapy. »N ot for use in an MRI environment. »U se only Maxtec replacement sensors. *See section 2.2.3 for a list of approved humidifier, circuit and patient interface. Maxtec Inc., 6526 South Cottonwood Street • Salt Lake City, UT 84107 Tel 801.266.5300 • Fax 801.270.5590 • Toll Free 800.748.5355 • www.maxtecinc.com
Hudson RCI Concha Humidifier with 22mm Single Heated Limb Circuit
6' - 22mm Corrugated Tube
F&P Humidifier with 22mm Single Heated Limb Circuit
Flow Rate in LPM* Patient Interface
A
B
C
D
E
F
Adult/Pediatric Aerosol Mask, Trach collar, Trach T-piece
15
27
39
50
61
72
Utah Medical - Disposa Infant Hood
15
27
38
50
61
72
Maxtec - Superdome Infant Hood
14
24
35
45
54
64
F&P Small Adult Nasal Cannula (RT051)
12
21
31
40
49
58
F&P Large Adult Cannula (RT050)
13
22
32
42
51
60
F&P Tracheostomy Direct Connection (RT052)
14
25
36
47
56
67
Adult/Pediatric Aerosol Mask, Trach collar, Trach T-piece
18
32
47
62
73
86
Utah Medical - Disposa Infant Hood
18
31
46
60
71
84
Maxtec - Superdome Infant Hood
15
27
38
50
60
69
F&P Small Adult Nasal Cannula (RT051)
14
24
34
44
53
62
F&P Large Adult Nasal Cannula (RT050)
14
25
35
45
54
64
F&P Tracheostomy Direct Connection (RT052)
17
29
41
51
62
73
Adult/Pediatric Aerosol Mask, Trach collar, Trach T-piece
16
29
42
54
66
78
Utah Medical - Disposa Infant Hood
16
29
42
54
66
78
Maxtec - Superdome Infant Hood
15
26
37
47
57
68
F&P Small Adult Nasal Cannula (RT051)
13
23
32
42
51
61
F&P Large Adult Nasal Cannula (RT050)
14
24
34
44
54
65
F&P Tracheostomy Direct Connection (RT052)
15
27
39
50
61
72
*Caution: The flow rates listed in this table are the results of bench testing of the MaxVentrui on the indicated patient circuits and interfaces. Actual flows may vary in clinical use depending on the patients physiology, breathing rate and other factors listed in section 3.0 of the operating manual.