Thopaz Quick Reference Guide March 2013
Quick Reference Guide
16 Pages
Preview
Page 1
Thopaz Quick Reference Guide
Precious life – Progressive care
How to prepare Thopaz for use single
0.3 L
Canister/ Tubing Set Up Functional Test Set Up
0.8 L
Change Pressure Access Patency Check
double
Silence Alarm
Select single or double tubing and ONE of the following canisters (0.3 L, 0.8 L)
Filter Clogged Leak in System System Clogged Standby Alarm
Open sterile tubing
Insert tubing into Thopaz ensuring tubing end does not bend and that it is tight against Thopaz.
Ensure that sealing ring is in place on side of Thopaz.
Battery Low Internal Error
Canister/ Tubing Set Up Functional Test Set Up
2
Change Pressure
1
Access Patency Check
– Attach canister to Thopaz – Place bottom pins in Thopaz tracks – Click top of canister into Thopaz ensuring it is tight against Thopaz
Thopaz ready for functional test.
Silence Alarm Filter Clogged Leak in System System Clogged Standby Alarm Battery Low Internal Error
How to perform functional check
Canister/ Tubing Set Up Functional Test Set Up
☛
☛
☛
Change Pressure
☛
therapy number
Switch Thopaz on. Ensure tubing is not connected to patient yet.
Choose if the treatment is for a new patient.
Switch on pressure.
Access Patency Check Silence Alarm Filter Clogged Leak in System System Clogged Standby Alarm
Check Thopaz is running. The flow rate should rise.
Cover patient end of tubing using your thumb.
Flow rate should decrease. If not, ensure canister is clicked in, tubing is tight against Thopaz and sealing ring is in place.
Battery Low Internal Error
Canister/ Tubing Set Up Functional Test Set Up
Press standby button for 3 seconds. Thopaz will go into standby mode.
Connect Thopaz tubing to patient chest tubing according to local hospital guidelines.
☛
☛
Continue treatment by pressing on button.
Change Pressure Access Patency Check Silence Alarm Filter Clogged Leak in System System Clogged Standby Alarm Battery Low Internal Error
How to change pressure
Canister/ Tubing Set Up
☛
Press both buttons simultaneously to highlight pressure.
Use up/down arrows to select desired pressure level or select Gravity mode.
Change Pressure
☛
☛
☛
Functional Test Set Up
Press the OK button to return to previous screen.
Access Patency Check Silence Alarm Filter Clogged Leak in System System Clogged Standby Alarm Battery Low Internal Error
How to access patency check
Canister/ Tubing Set Up
To access the patency check press the Graph button.
Change Pressure
☛
☛
Functional Test Set Up
Press down arrow to zoom.
Access Patency Check
Indicator only visible in zoommode when flow is 0 mL/min.
Silence Alarm Filter Clogged Leak in System
catheter open
catheter clogged
System Clogged Standby Alarm
Patient must inhale and exhale deeply and slowly.
Battery Low Internal Error
How to silence alarms
Canister/ Tubing Set Up
☛
☛
☛
The alarm will beep and a description will appear on screen.
To silence alarm push up/down buttons simultaneously. Thopaz will proceed when problem resolves.
Functional Test Set Up Change Pressure Access Patency Check
Turning Thopaz off by power button will not silence alarm.
Silence Alarm Filter Clogged Leak in System System Clogged Standby Alarm Battery Low Internal Error
Problem: Filter in canister is clogged
Canister/ Tubing Set Up Functional Test Set Up
☛
Change Pressure
☛
The alarm will beep and a description will appear on screen. Thopaz will automatically go into standby mode.
To silence alarm press both up/down arrows simultaneously and follow on screen instructions.
Access Patency Check
Clamp tubing as close to Thopaz as possible to prevent fluid leakage or follow prevailing hospital protocol if different than recommened here.
Silence Alarm Filter Clogged Leak in System System Clogged
☛ Press blue button to release canister.
Ensure the sealing ring is still in place.
Release clamp and continue treatment by pressing on button.
Standby Alarm Battery Low Internal Error
Problem: There is a disconnection of the tubing, the canister may not be clicked in properly, the o ring is missing between Thopaz and canister or patient is experiencing unusually large airleak
☛ To silence alarm press both up/ down arrows simultaneously and follow on screen instructions. Hold down standby button until standby mode is enabled.
Functional Test Set Up Change Pressure Access Patency Check
☛
The alarm will beep and a description will appear on screen.
Canister/ Tubing Set Up
Silence Alarm
Clamp Thopaz tubing as well as the patient catheter tube.
Filter Clogged Leak in System System Clogged Standby Alarm Battery Low Internal Error
Canister/ Tubing Set Up Functional Test Set Up Change Pressure
Check that Thopaz tubing is tight at the patient end.
Press blue button to release canister.
Ensure the sealing ring is still in place.
Access Patency Check Silence Alarm Filter Clogged Leak in System
☛
Remove and reinsert tubing into Thopaz ensuring that the tubing end does not bend and that it is tight against Thopaz.
System Clogged
– Reattach canister – Release clamp on tubing – Continue treatment by pressing on button
Standby Alarm
If problem persists replace tubing and canister.
Battery Low Internal Error
Problem: Tubing is kinked or clogged or the filter in canister is clogged
☛
☛
To silence alarm press both up/down arrows simultaneously and follow on screen instructions. Hold down standby button until standby mode is enabled.
Functional Test Set Up Change Pressure Access Patency Check Silence Alarm
The alarm will beep and a description will appear on screen.
Filter Clogged
g g
Is there an obvious kink or clot in the Thopaz tubing?
Canister/ Tubing Set Up
YES – Remove kink, Thopaz will resume treatment NO
– Clamp tubing next to canister and clamp patient catheter tube – Hold down standby for 3 seconds – Remove canister and dispose – Remove tubing and dispose – Insert new tubing – Check for sealing ring – Connect new canister – Release both tubing clamps – Continue treatment with on button
Leak in System System Clogged Standby Alarm Battery Low Internal Error
Problem: Thopaz left in standby mode for 5 minutes
Canister/ Tubing Set Up Functional Test Set Up
☛
Change Pressure
☛
The alarm will beep and a description will appear on screen.
To silence alarm press both up/ down arrows simultaneously and follow on screen instructions.
Access Patency Check
Thopaz will ask you if you want to turn the power off or resume treatment. Select “on” or power off.
Silence Alarm Filter Clogged Leak in System System Clogged Standby Alarm Battery Low Internal Error
Problem: Thopaz has approximately 30 minutes of battery remaining
Functional Test Set Up
☛
Change Pressure
☛
The alarm will beep and a description will appear on screen.
To silence alarm press both up/ down arrows simultaneously and follow on screen instructions.
Canister/ Tubing Set Up
Access Patency Check
Place Thopaz onto docking station or plug power cord directly into Thopaz to maintain charge.
Silence Alarm Filter Clogged Leak in System
Tip
System Clogged Standby Alarm
Ensure docking station is connected to the charger and is plugged into the mains.
Thopaz battery meter graphic will show charge is being applied.
Battery Low Internal Error
Problem: Internal error
Canister/ Tubing Set Up
To silence alarm press both up/ down arrows simultaneously and follow on screen instructions.
☛
☛ Switch Thopaz on. Ensure tubing is not connected to patient yet.
keep number
☛
The alarm will beep and a description will appear on screen.
new number
☛
☛
☛
Functional Test Set Up
Thopaz will instruct you to power off. Press power button once quickly.
– Keep patient therapy number by selecting “No” when Thopaz asks “new patient?” – Continue treatment with on button. If internal error alarm sounds again, turn off Thopaz and use replacement Thopaz. Label Thopaz faulty and report incident to Medela Account Manager.
Change Pressure Access Patency Check Silence Alarm Filter Clogged Leak in System System Clogged Standby Alarm Battery Low Internal Error
Medela AG Lättichstrasse 4b 6341 Baar, Switzerland www.medela.com