Thopaz Quick Card ™
This Quick Card in no way replaces the Instructions for Use REF 200.0685
Prepare Thopaz for use
2
1 check presence of sealing ring
attach tubing
attach c anister
ready for use
Turn Thopaz on Choose patient therapy number
☛
keep number
new number
☛
☛ press power button
☛
Standby-Mode
therapy number
switch on pressure
Perform functional check
seal off patient tubing (keep sterile)
Data-Mode
Data-Mode
check flow rate
when flow decreases connect patient according to hospital guidelines
Adjust pressure (when pump is running)
☛ press simultaneously
☛
☛
☛
increase or decrease
☛
☛
Data-Mode
or select Gravity mode
confirm with OK
Graph-Mode
Zoom-Mode
☛
flow scale max. 1000 mL/min
☛
Data-Mode
☛
Check therapy progress
flow scale max. 100 mL/min
Change canister Data-Mode
Standby-Mode
press longer than 3 seconds
Thopaz in Standby-Mode
☛ clamp off tubing
5
Standby-Mode
Select
leave standby-mode
☛
change canister seal canister and dispose
➞
➞
release clamp
Enter
Catheter check Zoom-Mode
- Only visible in Zoom-Mode when flow is 0 mL/min
catheter open
- Patient must inhale and exhale deeply and slowly
7
catheter clogged
3 – 24 min.
5x
Alarm management
☛
☛
error number
press simultaneously
9
follow advice on screen
Switch off Thopaz
☛ press longer than 3 seconds
press once quickly