Medicraft Hill-Rom
Medicraft HLE 8000 User Manual
User Manual
14 Pages
Preview
Page 1
MODEL HLE 8000 OPERATING MANUAL
Medicraft Australia PTY LTD 72-84 Shepherd St.Marrickville, Sydney, Australia 2204 Phone (02) 9 569 0255 Fax (02) 9 569 8394
TABLE OF CONTENTS PAGE PAGE PAGE PAGE PAGE PAGE PAGE PAGE PAGE PAGE PAGE PAGE
3 4 5 6 7 8 9 10 11 12 13 14
-
UNPACKING & TESTING – STORAGE/TRANSPORT CONDITIONS GENERAL DESCRIPTION OF BED FUNCTIONS DEFINITION OF SYMBOLS MARKED ON THE BED CENTRAL LOCKING CASTORS HANDSET OPERATION NURSE LOCK OUT BACKREST & KNEE BREAK OPERATION SAFETY SIDE RAILS HEIGHT RANGE AND BED POSITION ANGLES IP CODES AND WHAT THEY MEAN CLEANING AND DISINFECTION MAINTENANCE - POWER CABLE NOTICE
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
2
UNPACKING & TESTING. 1 2 3 4 5 6 7 8 9 10 11 12
Remove all bubble wrap from bed. Remove protective foam pad from the end of the bed. Remove ties from safety side rails Remove ties from the backrest, thigh section and the foot section. Remove tie, which secures the upper mattress frame to the wheelbase. Plug the handset into the handset socket. The handset will be stored under the bed frame. Remove the plastic sheath from the handset cable and make sure that the handset is fully inserted into the socket. Remove bubble wrap from the head & footboards. Insert the head & footboards into the sockets and tighten the tri wheel screws. Uncoil the power cable and plug into mains power. Check operation of all functions with the mains power turned on. Check operation of all functions with the mains power turned off. (If fitted with battery backup). Check that the safety side rails lock into all positions. Clean the bed with a general-purpose spray.
STORAGE & TRANSPORT CONDITIONS Ambient temp. range: +10°C to +40°C Rel. humidity range: 30% to 75% Atmospheric Pressure range: 700 hPa to 1060 hPa Inspect all cables particularly the mains power cable for any damage; replace where necessary. Stow the mains power cable on the storage hooks and clip the handset to the head –board frame when transporting the bed. For ease of steering, set the castors to the steering position prior to transportation. When the bed needs to moved to either side, set the castors to the free position When bed has been positioned, set the castors to the fully locked position The bed must be placed on a level surface to ensure all castors make contact with the floor. Be sure to check for obstructions both above and below the bed frame and position the bed so that it can operate through the full height range without any chance of collision.
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
3
GENERAL DESCRIPTION OF BED FUNCTIONS. The Medicraft ® HLE 8000 is an electrically operated bed with 3 basic movement functions. Hi – Lo adjustment Backrest adjustment Knee break adjustment (Optional) The electrical components run from a standard 240-volt AC outlet, which is stepped down to 24 volts DC. HI LOW Operation A single actuator system operates the Hi – Lo function. This actuator is fitted with internal limit switches to stop the piston at the fully extended and fully retracted positions. When raising the bed the actuator extends. When lowering the bed the actuator retracts. Backrest Operation A single actuator extends to raise, and retracts to lower. This actuator is freewheeling which means that it cannot pull. If an obstruction is present the actuator will free wheel. Knee Break Operation (Optional) A single actuator extends to raise, and retracts to lower. This actuator is fitted with internal limit switches to stop the piston at the fully extended and fully retracted positions. This actuator cannot pull. If an obstruction is present the actuator will free wheel.
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
4
EXPLANATION OF SYMBOLS
SAFE WORKING LOAD
Symbol placed before value of safe working load.
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
5
CENTRAL LOCKING CASTORS
TO APPLY THE BRAKES
The central locking braking system is activated by depressing the Black pedal at the foot end of the bed. The two-foot end castors and one head end castor will then be locked from swiveling and rotating.
TO RELEASE THE BRAKES Position foot under the pedal and lift the brake pedal up to the middle free position. All four wheels should rotate and swivel freely.
TO APPLY STEERING LOCK Position foot under the pedal and lift the brake pedal up to the tracking position. One castor at the head end of the bed will lock into a non-swivel mode. This allows straight steering by the attendant, when pushing from the foot end.
TO RELEASE THE STEERING LOCK Depress the pedal at the foot end of the bed until the pedal returns to the horizontal position. All four wheels should rotate freely.
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
6
HANDSET OPERATION TO RAISE AND LOWER THE BACKREST To raise the backrest, press the backrest up button until the desired posture is reached To lower the backrest, press the backrest down button until the desired posture is reached TO RAISE AND LOWER THE BED To raise the bed, press the bed up button until the desired height is reached To lower the bed, press the bed down button until the desired height is reached. TO RAISE AND LOWER THE KNEE-BREAK (Optional) To raise the knee-break, press the knee-break up button until the desired posture is reached To lower the knee-break, press the knee-break down button until the desired posture is reached
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
7
NURSE LOCK OUT
A nurse lockout switch box is fitted at the foot end of the mattress base. To disable any or all of the functions, turn the appropriate control to the locked position. This is indicated by a locked padlock symbol. Turn this control to the unlocked padlock symbol to enable any or all functions. Each channel is identified by the logos directly below each control.
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
8
BACKREST OPERATION To raise or lower the Backrest, press the appropriate button on the handset.
KNEE-BREAK OPERATION To raise or lower the knee-break, press the appropriate button on the handset. When the knee-break is raised, the lower leg section will remain parallel with the bed frame. The lower leg section can be lowered manually to bend the knee. This will be reset back to the normal parallel function when the knee break is lowered back to the flat position.
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
9
SAFETY SIDE RAILS The safety side rails have 4 positions. Fully up position.
Mid position.
Check position. When lowering the rail it will stop at this position. This allows the attendant to check for any obstructions prior to lowering to the fully down position. *NOTE*The rail must never be left in this position. If the rail is to be lowered, it must be fully lowered immediately after reaching this position.
Fully Down position. IMPORTANT This position must be set prior to patient transfer. If the rail is to be lowered, it must be fully lowered to this position.
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
10
SPECIFICATIONS Minimum & maximum bed height can vary with different castor sizes. The measurement shown below are with standard 100mm central locking castors fitted. All angles shown are with respect to horizontal. Equipment Class
Double Insulated
IP Rating
IP54
Safe Working Load
175Kg.
Mass
110Kg.
Minimum Height
240mm
Maximum Height
760mm
Maximum backrest angle
70º
Maximum upper leg section angle
25º
Lower leg section maximum angle with upper leg Section fully raised
23º
Max. Backrest Angle
Knee-Break Angles
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
11
IP Codes: IP51:
5: Protection against dust – limited degrees (No harmful deposit) 1: Protection against vertically falling drops of water Provides: Complete protection against contact with live or moving parts inside the enclosure. Protection against harmful deposits of dust. Protection against drops of condensed water. Units with IP51 rating are not normally labeled IP51.
IP54:
5: Protection against dust - – limited degrees (No harmful deposit) 4: Protection against water sprayed from all directions. Limited ingress permitted. Provides: Protection against harmful deposits of dust. Protection against water sprayed from any direction. Units are labeled IP54
IP 65:
6: Totally protected against dust. 5: Protection against low-pressure jets of water from all directions. Limited ingress permitted. Provides: Complete protection against contact with live or moving parts inside the enclosure. Protection against the ingress of dust. Protection against water jets from any direction. Units are labeled IP65
IP66:
6: Totally protected against dust 6: Protection against strong jets of water from all directions. Limited ingress permitted. Provides: Complete protection against contact with live or moving parts inside the enclosure. Protection against ingress of dust. Protection against stronger water jets and temporary submersion. Incorporating high temperature materials for wash tunnels. Tested 100% for air tightness during manufacture. Units are labeled IP66.
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
12
CLEANING AND DISINFECTION The materials used in the control boxes handsets and actuators are resistant to many detergents and disinfectants commonly used in hospital and care sectors. Units with IP51 protections ratings must not be washed down. Disconnect the mains power cable and use a slightly damp cloth to wipe over. The control box material used in IP66 protection ratings are resistant to hot and cold water and softeners. Organic solvents like Halogenide / aromatic hydrocarbons and ketones must not be used. Units rated to IP66 protection ratings are suitable for cleaning in wash tunnels which correspond to those, manufactured by KLEINDIENST, type Clean- Station ES3-387, provided that the following conditions are observed: -
All handset and actuator cable connection must be secure and be fully inserted into the control box before the bed enters the wash tunnel. The “O” ring seals on the cable plugs must be intact. Check the locking ring is in place on the CB12. The cleaning and disinfection cycle in wash tunnels must not exceed 10 minutes. Pressures on jet nozzles in wash tunnels must not exceed 10 bar. Distance between the jet nozzles and electric actuator parts must be at least 300mm. Water temperature must not exceed 85º C. Chilling with cold water is not permitted.
It is recommended to run the actuators in to their minimum stoke length before washing them to avoid degreasing of the spindle tubes. The following must be observed as regards to detergents: -
They must not be strongly alkaline or acidic (pH-value 6-8) They must not contain corrosive or caustic matters They must not contain chemicals that will change the surface structure or adhesive ability of the plastics. They must not dissolve grease.
The following disinfectants with detergents are recommended: -
NEODISCHER DEKONTA, Dose 1-3%, make: CHEMISCHE FABRIK Dr. WEIGERT CCOTRADE RW, Dose 0.5%, make: CCOTRADE GmbH. (CCOTRADE and NEODISCHER are trademarks.)
NOTE. Cleaning by means of a hand controlled jet pipe, which is for example connected to a steam cleaner, is not permitted, since it is not possible to maintain a minimum distance of 300mm to the electric parts.
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
13
RECOMMENDED MAINTENANCE SHEDULE PROCEEDURE Lubricate all moving and sliding parts and hinge points.
SHEDULE
MATERIALS REQD.
At least12months.
Lubricating oil / grease.
Inspect all fasteners to ensure proper fit, position, and tightness Cleaning Bed frame
At least12 months
Inspect all surfaces and remove any user developed sharp or burred areas, apply touch up where required Adjust castors (if adjustable)
Periodically
Correct size spanners, sockets, screwdrivers, hex keys, ect. Soap, Water, and medium strength cleaners. (Do not pressure clean) Metal file, correct match touch up paint.
Periodically
At least12months
Correct size spanner 13mm
Power Cable Notice. Be aware, that mains power cable management must be of a high priority when operating this bed. Care must be taken to ensure that the power cable is not tangled, or that the potential for the cable to be caught or severed by the bed frames moving parts.
If the mains power cable has been damaged in any way use the battery back up system (If Fitted) in place of mains power until such time that the cable has been replaced.
COPYRIGHT© MEDICRAFT AUSTRALIA ® 2004
14