USING SCOPE BUDDY™ PLUS ENDOSCOPE FLUSHING AID TO ASPIRATE OLYMPUS® DUAL CHANNEL ENDOSCOPES CONNECTING THE ENDOSCOPE FOR ASPIRATION •
Place endoscope in sink
•
Attach 24-Hour Multi-Use Hookup for OLYMPUS® Endoscopes to endoscope and close all pinch clips.
•
Push blanking cap for OLYMPUS® Endoscopes onto the endoscope’s air/water channels.
•
Install the channel plug over valve openings, leaving both biopsy/instrument ports open.
•
Attach pump tube to 24-Hour Multi-use Hookup for OLYMPUS® Endoscopes
•
Remove the strainer from the end of the pump tube and set it on the counter.
•
Hold end of the pump tube without the strainer over sink.
The ASPIRATION step runs the pump in reverse, drawing fluid through the biopsy channels.
ASPIRATION •
Select the desired aspiration time by using the up and down arrows. The minimum time recommended by Olympus is 30 seconds.
•
Press PLAY to begin aspirating.
PREPARING THE ENDOSCOPE FOR THE NEXT STEP AFTER ASPIRATION •
Replace the strainer on the pump tube when aspiration is complete.
•
Release pinch clips on the tubing attached to the endoscope. Leave the pinch clip closed on tubing if it is not connected to an endoscope port.
•
Install the OLYMPUS® MAJ-420 (Medivators part # DSD-110-HU0159) over the endoscope’s biopsy ports (Shown in picture) per Olympus IFU.
SCOPE BUDDY™ is a trademark of Medivators Inc. OLYMPUS® is a registered trademark of Olympus Corporation.
www.medivators.com © 2020 Medivators Inc. LIT-00188/A
Installation of OLYMPUS® MAJ-420 over endoscope’s biopsy ports after completing aspiration