Scope Buddy Rinse Matrix (For serial numbers ≥ 201345) Seconds required to flush a minimum of 90mls through each channel Olympus Endoscopes Endoscope Model
Scope Buddy (DSD or universal hookup)
Bronchoscope
BF-10, 20, 30, 40, 60, 160, 180, 200 and 240 series
≤ 30 sec
Colonoscope Colonoscope Colonoscope Colonoscope Colonoscope Colonoscope (using MAJ-855 auxiliary water tube)
All models not listed below CF-140, 160 and 180 series (1 channel models) CF-140, 160 and 180 series (2 channel models) PCF-140, 160 and 180 series CF-1T140L CF-140, 160 and 180 series (1 and 2 channel models)
≤ 60 sec ≤ 30 sec ≤ 30 sec ≤ 30 sec ≤ 30 sec 40 sec
Duodenoscope (including elevator wire channel)
All models
≤ 60 sec
Gastroscope Gastroscope
All models not listed below GIF-2T100
≤ 30 sec ≤ 60 sec
Choledochoscope Cystoscope Hysteroscope Laryngoscope Pancreatoscope Rhinolaryngoscope Sigmoidoscope Ultrasound Scope (including elevator wire channel) Uretero-Renofiberscope
All models All models All models All models All models All models All models All models All models
≤ 30 sec ≤ 30 sec ≤ 30 sec ≤ 30 sec ≤ 30 sec ≤ 30 sec ≤ 60 sec ≤ 60 sec ≤ 30 sec
Endoscope Type
Auxiliary water tube (connected to endoscope by itself)) MAJ-855 (quantity one) 30 sec Auxiliary water tube (not connected to endoscope) MAJ-855 (quantity two) 30 sec * Rinse times represent the grouping of endoscopes. Rinse times may be reduced by individual flow testing of a specific endoscope. Ensure flushing volumes comply with endoscope manufacturers’ guidelines
50096-849 Rev E
Page 1 of 2
Scope Buddy Rinse Matrix
(For serial numbers ≥ 201345) Seconds required to flush a minimum of 100mls through each channel Fujinon Endoscopes
Scope Buddy (DSD or universal hookup) Bronchoscope ≤ 35 sec Choledochoscope ≤ 35 sec Colonoscope ≤ 65 sec Cystoscope ≤ 35 sec Duodenoscope (excluding elevator wire channel) ≤ 65 sec Duodenoscopes (elevator wire channel only, 15mls) ≤ 3 min (using EWC hookup) Forward Water Jet ≤ 1 min 30 sec (using EWC hookup) Gastroscope ≤ 65 sec Hysteroscope ≤ 35 sec Laryngoscope ≤ 35 sec Pancreatoscope ≤ 35 sec Rhinolaryngoscope ≤ 35 sec Sigmoidoscope ≤ 65 sec Ultrasound Scope ≤ 65 sec Uretero-Renofiberscope ≤ 35 sec Pentax Endoscopes Scope Buddy Endoscope Type Endoscope Model (DSD or universal hookup) Bronchoscope All models ≤ 35 sec Choledochoscope All models ≤ 35 sec Colonoscope (including forward water jet w/non-check valve adapter) All models ≤ 65 sec Cystoscope All models ≤ 35 sec Duodenoscope (including elevator wire channel) All models ≤ 65 sec Gastroscope (including forward water jet w/non-check valve adapter) All models ≤ 35 sec Hysteroscope All models ≤ 35 sec Laryngoscope All models ≤ 35 sec Pancreatoscope All models ≤ 35 sec Rhinolaryngoscope All models ≤ 35 sec Sigmoidoscope (including forward water jet w/non-check valve adapter) All models ≤ 35 sec Ultrasound Scope All models ≤ 35 sec Uretero-Renofiberscope All models ≤ 35 sec * Rinse times represent the grouping of endoscopes. Rinse times may be reduced by individual flow testing of a specific endoscope. Ensure flushing volumes comply with endoscope manufacturers’ guidelines Endoscope Type
50096-849 Rev E
Endoscope Model All models (including 530) All models All models (including 530) All models All models (including 530) All models All models All models not listed below (including 530) All models All models All models All models All models All models (including 530) All models
Page 2 of 2