User Manual
46 Pages
Preview
Page 1
SCOPE BUDDY
®
ENDOSCOPE FLUSHING AID
USER MANUAL
50096-381 Rev D
SCOPE BUDDY USER/SERVICE MANUAL
FUJINON™ is a registered trademark of Fujinon Corporation MEDIVATORS® is a registered trademark of Minntech Corporation. OLYMPUS® is a registered trademark of Olympus Corporation. PENTAX® is a registered trademark of Pentax Precision Instrument Corporation. SCOPE BUDDY® is a trademark of Minntech Corporation.
© 2009 Minntech Corporation All rights reserved. This publication is protected by copyright. Copying, disclosure to others, or the use of this publication is prohibited without the express written consent of Minntech Corporation. Minntech reserves the right to make changes in the specifications shown herein without notice or obligation. Contact your Medivators representative or Medivators customer service for more information.
TABLE OF CONTENTS Page
1 INTRODUCTION Using this Manual... 1 Product Description ... 1 Product Specifications... 2
2 SAFETY Intended Use ... 3 Electrical Compliance ... 3 Operator Safety... 4 Endoscope Safety ... 4 Notes, Cautions, and Warnings ... 5 Guidelines ... 6 Endoscope Pre-cleaning ... 6 Cleaning and Disinfection... 7
3 SETUP AND FAMILIARIZATION Device Installation ... 9 System Components...10 Control Panel ...11 Tubing Connections...12
4 DEVICE OPERATION Programming Times... 13 Elevator Wire Channel – Fujinon Endoscopes……….………………………………... 14 Forward Water Jet – Fujinon Endoscopes ... 15 Endoscope Flushing using a Medivators DSD Hook-up ... 15 Endoscope Flushing using the Endoscope Manufacturer’s All-channel Irrigator ...20 Endoscope Flushing using the Elevator Wire Channel Hook-up – Fujinon Endoscopes ... 25 Daily Quality Assurance Test ...30 Daily Decontamination Procedure ...33
5 MAINTENANCE & TROUBLESHOOTING Maintenance ...35 Troubleshooting Guide ...36
6 GLOSSARY ______________________________________________________________ 37 7 WARRANTY _____________________________________________________________________39
SCOPE BUDDY USER/SERVICE MANUAL
MEDIVATORS REPROCESSING SYSTEMS
1 INTRODUCTION Using this Manual This manual describes the features of the Medivators SCOPE BUDDY Endoscope Flushing Aid, including setup, operation, safety, maintenance, and troubleshooting procedures. It is important to follow the provided instructions to keep the device in good operating condition, delivering proper fluid flows, and to ensure long product life. Given the unique design of the electrical and mechanical components used within this product, no userserviceable parts are contained within the unit and no internal calibration is required. Annual calibration and preventative maintenance schedules are also not required as the unit is designed to be maintenancefree. It is however; strongly recommended that daily quality assurance testing and the daily decontamination procedure be performed to ensure the system is functioning properly and ready for use. Refer to Section 4 for further information. Contact the manufacturer’s representative if you have any questions regarding this product, its use, or operation.
Product Description The Medivators SCOPE BUDDY Endoscope Flushing Aid is designed to facilitate the flushing of rinse water and cleaning solutions through the channels of flexible, immersible endoscopes. This device is to be used during the manual pre-cleaning phase of endoscope reprocessing, prior to the endoscope being placed in an automated endoscope reprocessor (AER) for high level disinfection. The SCOPE BUDDY combines a diaphragm pump, digital timer, and electronic control circuitry to pump fluids through tubing connections to the channels of the endoscope. Three different tubing sets allow connection to most types of endoscopes. The Elevator Wire Channel Hookup allows connection to the Fujinon endoscope’s elevator wire channel, or to the Fujinon endoscope’s Forward Water Jet. The DSD Extension Line allows connection to the DSD hook-up which attaches to the endoscope. The Universal Cleaning Adapter Extension Line and multi-way adapters allow connection to the Universal Cleaning Adapter Hook-up provided with the endoscope. The SCOPE BUDDY is designed to be used as a flushing aid only during the manual cleaning process of a flexible endoscope. This device is not an endoscope disinfector. All endoscopes must be high-level disinfected prior to use on a patient.
1
SCOPE BUDDY USER/SERVICE MANUAL
Product Specifications Chassis Dimensions (height - width - depth)
9 ½”H x 6”W x 8”D (Inches)
Weight (approx.)
10 lb (4.5kg)
Power Cord and Connector
Medical Grade – 6 feet (2.83m) 18AWG,
24.13H x 15.24W x 20.32D (CM’s)
#3 Connector Electrical Requirements
100 – 250 VAC ±10% 47-63 HZ 1φ
Current Draw
< 1.0 Amperes
Breaker and Protection
1 Amp max. at 90 VAC input, Non-replaceable, internal primary current fuse, inrush limiting, thermal shutdown
Maximum Fluid Delivery
1 Liter per minute
Maximum Pump Pressure
29 PSI w/Safety Bypass
Cycle Time Range Setting
0 to 99 minutes/59 seconds
Cycle Completion Indicators
Audible and Visual
Electrical Safety Certifications
ETL and cETL Listed Conforms to UL 61010-1 Certified to CAN/CSA C22.2 No. 61010-1
Electromagnetic Interference
FCC Title 47, Part 15, Class A, EN 55011 and EN 50082-1
Environmental Rating
Standard
Designed for Use
Indoor Use Only
Installation/ Over Voltage Category
II
Classification
1, Ordinary protection
Pollution Degree
2
Operating Temperature Range
15° C to 35° C
Operating Humidity Range
10-75% non-condensing
Storage Temperature Range
-29° C to 57° C
Storage Humidity Range
5 – 90% non-condensing
Warranty Period
One Year
2
MEDIVATORS REPROCESSING SYSTEMS
2 SAFETY Intended Use The Medivators SCOPE BUDDY Endoscope Flushing Aid is an electro-mechanical device intended to pump fluids through channels of flexible, immersible endoscopes. The device is designed to be used as a flushing aid only during the endoscope manual cleaning process. This device is not an endoscope disinfector and does not take the place of an automatic endoscope reprocessor. Endoscopes must be high-level disinfected by an approved endoscope reprocessor or manual HLD soaking prior to use on a patient. Throughout this manual are notes, cautions, and warnings, which are also represented by specific icons. These are designed to provide important information to the user regarding the proper use and operation of this device. It is important to read and understand all notes, cautions, and warnings, prior to using this product. Should you have any questions or require further clarification, please contact your manufacturer’s representative. Examples of these icons are shown below. Note: A note refers to important information to which the user should pay special attention. Notes provide added insight and helpful information which can be useful to the operator. Caution: This information is an alert to the possibility of a problem with the device associated with its use or misuse.
!
WARNING: This information is an alert to possible injury or other serious adverse reaction associated with the use or misuse of this device.
WARNING: Do not use this device on NON-IMMERSIBLE endoscopes or on endoscopes with internal leaks.
Electrical Compliance SCOPE BUDDY has been rigorously tested to stringent electrical certifications and safety standards. This device meets the laboratory standard for Electromagnetic Interference Requirements for FCC Title 47, Part 15, Class A, EN 55011 and EN 50082-1, and the Electrical Safety Requirements of UL 61010-1, Pollution Degree 2, and Installation Category II Equipment. SCOPE BUDDY carries and is certified to ETL and cETL, the Electrical CE Mark, UL 61010-1, and CAN/CSA C22.2 No. 61010-1. Caution: This device must always be plugged into an electrical wall outlet that includes an approved Ground Fault Interrupter (GFI) prior to operation of the pump. 3
SCOPE BUDDY USER/SERVICE MANUAL
Operator Safety 1. This device is only to be operated by properly trained and qualified personnel who are fully familiar with its intended use and operation. 2. Avoid biological contamination and chemical burns. Always wear appropriate personal protective equipment (PPE) including clothing, gloves, and safety glasses when handling used endoscopes. 3. Cleaning and detergent solutions for endoscope flushing must be used in accordance with the regulations governing their use, safety, and shelf life. Decontamination solutions for performing the Scope Buddy Daily Decontamination Cycle must be used in accordance to the regulations governing their direction for use. 4. In case of contact with cleaning solutions, flush eyes or skin with copious amounts of water for a minimum of 15 minutes. Seek immediate medical attention in case of eye contact. 5. Decontamination solutions can be harmful if swallowed. In case of accidental ingestion, drink large quantities of water and call a physician immediately. Avoid contact with food. Do not store decontamination solutions near food.
Patient Safety 1. Failure to adhere to the endoscope manufacturer’s recommendations governing the proper cleaning of endoscopes after each patient examination may compromise the efficacy of the subsequent endoscope high level disinfection process. 2. All endoscopes must be high-level disinfected prior to patient use. SCOPE BUDDY is not designed to replace an automated endoscope disinfector and should never be used to disinfect an endoscope.
Endoscope Safety Scope Buddy has been designed and tested to meet or exceed endoscope manufacturer’s guidelines including pressure and flow requirements. Medivators Scope Buddy is certified and validated safe for use with Olympus, Pentax, Fujinon and Storz flexible, immersible endoscopes.
4
MEDIVATORS REPROCESSING SYSTEMS
Notes, Cautions, and Warnings General Cautions •
Use only a medical-grade detergent which has bacteriostatic properties, is low-foaming, freerinsing, and has a neutral pH.
•
Detergent and rinse water cleaning solutions can be drawn from either a clean container for a single-pass through Scope Buddy, or recirculated from the sink through Scope Budd. Scope Buddy can be used in either format depending on the facility’s endoscope cleaning procedure.
•
For the daily decontamination procedure, select a decontamination solution whose label indicates that it may be used at room temperature to achieve decontamination (see Section 6, Glossary, for definition) of medical instruments. Consult the manufacturer’s label for the appropriate directions for use and decontamination solution contact time.
•
Do not use alcohol-based products in or on the SCOPE BUDDY device. Alcohol is not compatible and should never be used to clean, flush, or rinse SCOPE BUDDY. Use of alcohol or alcohol-based products with SCOPE BUDDY will void the warranty. The tubing however can be reprocessed independently using 70% alcohol when not attached.
•
Do not use bleach solutions or hydrogen peroxide in or on Scope Buddy.
•
Do not exceed a detergent, water, decontamination solution temperature of 50°C in Scope Buddy.
•
Follow the endoscope manufacturer’s instructions and established professional guidelines for cleaning endoscopes.
•
The SCOPE BUDDY Endoscope Flushing Aid User Manual is intended to be used in conjunction with the instructions and labeling provided by the endoscope/accessory manufacturer. Endoscope/accessory manufacturers provide instructions and labeling necessary for users to understand the basic design, specifications, nomenclature, and components of specific flexible endoscopes/accessories.
•
Users should be properly trained and instructed to inspect, prepare, use, clean, disinfect, rinse, dry, and store endoscopes. Follow recommended standards as referred by ASTM, ASGE, SGNA, and APIC.
•
Read all precaution labels on the Scope Buddy device before using.
•
Never use Scope Buddy in a manner not specified by the manufacturer.
•
Ensure that fluid exiting the open end of the endoscope or tubing is not facing a direction that would allow fluid to be sprayed on personnel, electrical fixtures, or the user.
•
Scope Buddy, through its’ inherit use, becomes contaminated and requires cleaning and decontamination. Precautions must be taken to insure the device is properly decontaminated. Please refer to the Daily Decontamination Procedure for further information.
•
Duodenoscopes with elevator wire channels (EWC) may require additional cleaning and disinfection steps. It is particularly important for the elevator wire channel to be manually cleaned and fully reprocessed correctly. Extra attention should also be given to the position of the EWC throughout the cleaning process. Refer to the endoscope manufacturer’s instruction for the proper positioning of the elevator wire channel.
•
If the Scope Buddy is to be unused for an extended period of time such as overnight, or over the weekend, it should be left full of decontamination fluid or clean rinse water. This maintains pumphead decontamination, lubrication and facilitates priming. Scope Buddy should not be air-purged and store empty between uses. The tubing can be detached from the Scope Buddy unit during this period.
5
SCOPE BUDDY USER/SERVICE MANUAL
Guidelines Endoscope Pre-cleaning Proper endoscope precleaning is essential to removing microbial burden and bio-debris from the inside and the outside of an endoscope. Precleaning guidelines should always be consulted and followed, and a pre-cleaning process should always be developed and strictly adhered to, to ensure an endoscope is effectively and properly cleaned. Always follow the endoscope manufacturer’s instructions and established professional guidelines to properly preclean an endoscope. •
Meticulous manual cleaning of the endoscope or accessory must be performed using an appropriate detergent solution.
•
Carefully brush all channels and valve ports using the detergent solution and cleaning tools designed for this purpose.
•
Avoid the use of cotton swabs, gauze pads, and other lint-producing tools.
•
Endoscopes should be leak tested before submersion as part of the overall reprocessing protocol as defined in the instrument manufacturer’s instructions. Leak testing detects damage within the interior or exterior of the endoscope.
•
Follow instructions for manual pre-cleaning contained in the endoscope manufacturer’s manual.
6
MEDIVATORS REPROCESSING SYSTEMS
Cleaning and Disinfection Always follow the endoscopes manufacturer’s recommendations pertaining to the reprocessing of endoscopes and endoscope accessories to ensure that the endoscopes and accessories will be safe for patient use. A protocol should be developed and followed based on such guidelines. The following professional organizations have published guidelines pertaining to the reprocessing of endoscopes and associated accessories: Society of Gastroenterology Nurses and Associates, Inc. 401 North Michigan Ave. Chicago, IL 60611-4267 USA Tel: 800-245-7462 Fax: 312-527-6658 www.sgna.org
Assoc. for Professionals in Infection Control and Epidemiology, Inc. 1275 K Street NW, Suite 1000 Washington, DC 20005-4006 USA Tel: 202-789-1890 Fax: 202-789-1899 www.apic.org
American Society for Gastrointestinal Endoscopy 1520 Kensington Rd., Suite 202 Oak Brook, IL 60523-2141 USA Tel: 630-573-0600 Fax: 630-573-0691 www.asge.org
American Society for Testing and Materials 100 Barr Harbor Dr. West Conshohocken, PA 19428-2959 USA Tel: 610-832-9585 Fax: 610-832-9555 www.astm.org
Association of periOperative Registered Nurses, Inc. 2170 S. Parker Rd., Suite 300 Denver, CO 80231-5711 USA Tel: 800-755-2676 Fax: 303-750-3462 www.aorn.org
Canadian Society of Gastroenterology Nurses & Associates 27 Nicholson Dr. Lakeside, NS B3T 1B3 CANADA Fax: 902-473-4406 www.csgna.com
British Society of Gastroenterology 3 St. Andrews Place Regents Park London NW1 4LB ENGLAND Tel: (+44) (0)20 7387 3534 www.bsg.org.uk
Gastroenterological Nurses College of Australia, Inc. (GENCA) PO Box 483 Boronia Vistoria 3155 Australia Tel: 1300 788 155 Fax: 03 9801 6352 www.genca.org
7
SCOPE BUDDY USER/SERVICE MANUAL
8
MEDIVATORS REPROCESSING SYSTEMS
3 SETUP & FAMILIARIZATION Device Installation •
SCOPE BUDDY can only be used with the power supply supplied with the device. Do not attempt to substitute or use a non-approved replacement power supply.
•
The cycle run time can be adjusted by the operator, but always verify the rinse time is appropriate for the endoscope being flushed.
•
Do not use alcohol or alcohol-based products, bleach solutions, or hydrogen peroxide in the SCOPE BUDDY.
•
The tubing hookups, hookup extensions, and fluid intake line are not autoclavable and must be reprocessed using low temperature decontamination methods only.
•
SCOPE BUDDY is designed to be placed on a counter top or mounted to the wall by using the supplied wall-mounting bracket. If mounting to a wall, ensure the bracket is properly secured using the correct mounting hardware, adequate to support at least ten pounds of weight.
•
SCOPE BUDDY must be located within a vertical height no greater than three feet above the incoming detergent or rinse water solution. The distance between the top of the fluid solution and SCOPE BUDDY must not exceed three feet.
•
SCOPE BUDDY is designed to operate in a damp environment, but should not be immersed in any fluids. Do not use the device if it becomes immersed in any type of fluid, or if fluid is observed leaking from the device.
9
SCOPE BUDDY USER/SERVICE MANUAL
System Components •
SCOPE BUDDY Base Unit
•
Wall-mount bracket
•
Power Supply
•
Fluid Intake Line
•
DSD Extension Line
•
Universal Cleaning Adapter Extension Line
•
Universal Cleaning Adapter 2-way adapter
•
Universal Cleaning Adapter 3-way adapter
•
Universal Cleaning 4-way adapter
•
Flow Verification Tube
•
Elevator Wire Channel (EWC) Hookup
SCOPE BUDDY Base Unit
Wall-Mount Bracket
Power Supply
Fluid Intake Line
UCA 4-way Adapter DSD Extension Line
Elevator Wire Channel Hook-up Universal Cleaning Adapter Extension Line
UCA 3-way Adapter Flow Verification Tube Figure 1 System Components
10
UCA 2-way Adapter
MEDIVATORS REPROCESSING SYSTEMS
Control Panel •
Timer Display – Displays the timer setting (when idle) or the time remaining when pumping.
•
Time Set – Used to increase or decrease the timer setting.
•
Start/Stop Switch – Press to start, stop, or pause the timer countdown and pumping action.
•
Fluid Inlet – “IN” port where fluid is drawn into the base unit.
•
Fluid Outlet – “OUT” port where fluid is discharged from the base unit.
•
Power On/Off Switch – Used to turn the device on or off.
•
Power Connection – Connection port for incoming power from the power supply to the base unit.
Time Display
Start/Stop Switch Time Set Increase/Decrease
Power On/Off Switch Power Connection
Fluid Inlet
Fluid Outlet
Figure 2 Control Panel
11
SCOPE BUDDY USER/SERVICE MANUAL
Tubing Connections The device is designed to function with a wide variety of manufacturer’s endoscopes of various configurations. The device is supplied with different types of tubing connections which allow connection to most endoscopes. The DSD Extension Line allows connection to a Medivators DSD hook-up which attaches to the endoscope. Refer to the DSD Hook-up Application Guide to identify the correct hook-up to use for the endoscope being flushed and the hook-ups’ Directions for Use Instructions for correct connection. The Universal Cleaning Adapter Extension Line and multi-way adapters allow connection to the Universal Cleaning Adapter provided with the endoscope. Refer to the endoscope manufacturer’s Instructions for Use to ensure the correct Cleaning Adapter is being used and that it is correctly connected. The Elevator Wire Channel Hook-up allows flushing of the elevator wire channel of a Fujinon Duodenoscope. The Elevator Wire Channel Hook-up is also used to flush the forward water jet of a Fujinon endoscope. Note: The Fujinon forward water jet adapter is always used in conjunction with the Elevator Wire Channel Hook-up. Note: When using a Pentax or Fujinon water jet adapter, always verify it does NOT contain a check valve. Only non-check valve forward water jet adapters can be used with Scope Buddy to achieve the correct flushing volume over the stated Rinse Matrix times. Note: The universal cleaning adapter Irrigator by some endoscope manufacturers.
may
be
referred
to
as
an
All-Channel
Reference the Medivators Hook-up Application Guide which can be downloaded at www.minntech.com/medivators to identify the appropriate DSD hook-up for the manufacturer’s endoscope being cleaned. The supplied tubing segments which facilitate endoscope connection are: •
Fluid Intake Line – Tubing set that allows fluid to be drawn up from the sink into the SCOPE BUDDY.
•
DSD Extension Line – Tubing set that allows fluid to be delivered from the machine to the DSD Hook-up.
•
Universal Cleaning Adapter Extension Line – Tubing set that allows fluid to be delivered from the machine to the Universal Cleaning Adapter supplied with the manufacturer’s flexible endoscope.
•
UCA 2-way Adapter – Tubing adapter which connects to the Universal Cleaning Adapter Extension Line and allows connection to an endoscope’s Universal Cleaning Adapter with 2 connection ports.
•
UCA 3-way Adapter – Tubing adapter which connects to the Universal Cleaning Adapter Extension Line and allows connection to an endoscope’s Universal Cleaning Adapter with 3 connection ports.
•
UCA 4-way Adapter – Tubing adapter which connects to the Universal Cleaning Adapter Extension Line and allows connection to an endoscope’s Universal Cleaning Adapter with 4 connection ports.
•
Elevator Wire Channel Hookup – Tubing set that allows fluid to be delivered from the machine to the elevator wire channel, or to the forward water jet of a Fujinon endoscope.
•
Flow Verification Tube – Tubing segment used during the Daily Quality Assurance Test to ensure SCOPE BUDDY is delivering the proper amount of fluid. Note: A complete set of tubing connections is available by contacting Medivators Customer Support. The reorder part number is 78399-669, and the tubings are only available as a complete set. 12
MEDIVATORS REPROCESSING SYSTEMS
4 DEVICE OPERATION Programming Times SCOPE BUDDY Programming Times for Olympus, Pentax & Fujinon Brand Endscope are listed below for Scope Buddy units having a serial number >201345. Programming times are different depending on which version Scope Buddy is being used. Reference the Scope Buddy serial number (found on Scope Buddy’s back panel) to identify which “Scope Buddy Rinse Matrix” is to be used. •
For Scope Buddy units having a serial number ≤ 201344, then use the Scope Buddy Rinse Matrix part number 50096-704.
•
For Scope Buddy units having a serial number ≥ 201345, then use the Scope Buddy Rinse Matrix part number 50096-849.
The Scope Buddy Rinse Matrices can be found by logging onto the Medivators website located at: www.minntech.com/medivators. Click on the “Medivators Reprocessing Systems” tab followed by the “Resource Center” dropdown and then “User Library”. Select “Scope Buddy” as the product and then scroll down to the “Users Manuals” heading and “Document Descriptions” to locate the correct Scope Buddy Rinse Matrix. Always verify the correct Scope Buddy Rinse Matrix is being used based off of the serial number of the unit in use. Reference the “Scope Buddy Rinse Matrix” to identify the programming time required to flush cleaning and rinse solutions through the endoscope channels during the manual cleaning process. The time stated assumes that the connection tubings and hookup kits from SCOPE BUDDY to the endoscope are primed and full of fluid. Program times should be extended by 10 seconds if the tubes are not pre-primed. The rinse matrix times area based on in-house validation tests that were conducted using endoscopes with channels in good working condition, free from restrictions and/or defects. These rinse times represent the grouping of endoscopes. Rinse times may be reduced by individual flow testing of a specific endoscope. Always ensure flushing volumes comply with endoscope manufacturer’s guidelines. Due to differences in individual endoscope channel performance, users should always verify that the solution volumes meet the manual cleaning requirements of the endoscope manufacturer and or the facility. Use these programming times for Scope Buddy units having serial numbers ≥ 201345.
Programming Time to Flush 90 mls through Olympus Endoscopes including those with an Elevator Wire Channel Small Caliber Endoscopes Gastroscopes Gastroscope : Model GIF-2T100 Duodenoscopes (including elevator wire channel) Colonoscopes Colonoscopes: (Model CF-140, 160 and 180 (1 and 2 channels), PCF-140, 160 and180, CF-1T140L Sigmoidoscopes Ultrasound Scope
30 seconds 30 seconds 1 minute 1 minute 2 minutes 45 seconds 30 seconds 1 minute 1 minute
13
SCOPE BUDDY USER/SERVICE MANUAL
Programming Time to Flush 100 mls through Pentax Endoscopes including those with an Elevator Wire Channel (serial number ≥201345) Note: When flushing the Forward Water Jet of a Pentax endoscope, always ensure the forward
water jet adapter does not contain a check valve. Only a forward water jet non-check valve adapter can be used. Also, the Forward Water Jet can be flushed simultaneously with the other endoscope channels, and does not require the use of the Elevator Wire Channel Hook-up. The Pentax forward water jet can be flushed using the DSD Hook-up or the All-channel Irrigator Hook-up. Small Caliber Endoscopes Gastroscopes (including forward water jet w/non-check valve adapter) Duodenoscopes (including elevator wire channel) Colonoscopes (including forward water jet w/non-check valve adapter) Sigmoidoscopes Ultrasound Scope
35 seconds 35 seconds 1 minute 5 seconds 1 minutes 5 seconds 35 seconds (using the forward water jet w/non-check valve adapter 35 seconds
Programming Time to Flush 100 mls through Fujinon Endoscopes excluding those with an Elevator Wire Channel or a Forward Water Jet (serial number ≥201345) Small Caliber Endoscopes Gastroscopes Gastroscope: Model EG-450HR Duodenoscopes (including elevator wire channel) Duodenoscopes (elevator wire channel only, 15 mls) Colonoscopes Sigmoidoscopes Ultrasound Scope Fjuinon Forward Water Jet
35 seconds 1 minute 30 seconds 35 seconds 1 minute 5 seconds 3 minutes (using EWC hook-up) 1 minutes 5 seconds 1 minute 5 seconds 1 minute 5 seconds 1 minute 30 seconds (using EWC hook-up)
Elevator Wire Channel – Fujinon Endoscopes SCOPE BUDDY can be used to flush an endoscope’s elevator wire channel (EWC). Due to the internal structure of the Fujinon EWC channel, it must be flushed independently of the other endoscope channels The Fujinon endoscope’s EWC simultaneously with other endoscope channels as the elevator wire channel will not be properly flushed per the endoscope manufacturer’s standards and guidelines. The independent flushing of the elevator wire channel using Scope Buddy’s Elevator Channel Hook-up ensures fluid is delivered directly to this channel. Go to the section labeled “Endoscope Flushing using the Elevator Wire Channel Hook-up – Fujinon Endoscopes” for further instructions. .
14
MEDIVATORS REPROCESSING SYSTEMS
Forward Water Jet – Fujinon Endoscopes The SCOPE BUDDY elevator wire channel (EWC) hook-up must be used to flush the Forward Water Jet of a Fujinon endoscope. The Fujinon endoscope’s forward water jet must be flushed independently and while using Scope Buddy’s Elevator Wire Channel hook-up. These specific endoscope’s have an internally contained check-valve and Scope Buddy’s elevator wire channel hook-up is uniquely designed for this function. The DSD Hook-up or the All-channel Irrigator Hook-up cannot be used to flush the Forward Water Jet of a Fujinon endoscope. Never flush the Fujinon endoscope’s forward water jet simultaneously with other endoscope channels as the Fujinon endoscope’s forward water jet will not be properly flushed per the endoscope manufacturer’s standards and guidelines. The independent flushing of this channel using Scope Buddy’s EWC Hook-up ensures fluid is delivered directly to this channel. Go to the section labeled “Endoscope Flushing using the Elevator Wire Channel Hook-up – Fujinon Endoscopes” for further instructions.
Endoscope Flushing using a Medivators DSD Hook-Up Use this step-by-step procedure to properly attach the device to an endoscope using the Fluid Intake Tubing, the DSD Extension Line, and the appropriate Medivators DSD Hook-up. The Fluid Intake Tubing and the DSD Extension Line are provided with the device. The Medivators DSD Hook-up is provided by current users of a Medivators DSD Endoscope Reprocessor. Reference the Medivators Hook-up Application Guide, located at www.minntech.com/medivators to identify the correct DSD Hook-up for the endoscope being cleaned. Refer to the Directions for Use Instructions supplied with the hook-up to ensure correct connection. When properly connected, fluid will be drawn up from the sink or from a clean container of detergent or rinse water solution, flow through the Fluid Intake Tubing to SCOPE BUDDY, and then out of SCOPE BUDDY and through the DSD Extension Tubing and the DSD Hook-up to the endoscope. This provides fluid delivery for flushing the internal lumens of the endoscope. Note: If flushing the Forward Water Jet or the Elevator Wire Channel of a Fujinon endoscope, then the elevator wire channel hook-up must be used. The Medivators DSD Hook-up cannot be used. Go to the section labeled “Endoscope Flushing using the Elevator Wire Channel Hook-up – Fujinon Endoscopes” for further instructions.
15
SCOPE BUDDY USER/SERVICE MANUAL
Scope Buddy Unit
Fluid Intake Line
DSD Extension Line
Figure 3 Scope Buddy Line Hook-up
Note: The endoscope leak test must be performed prior to commencing the cleaning process according to manufacturer’s guidelines. Note: Detergent and rinse water cleaning solutions can be drawn from either a clean container for a single-pass through Scope Buddy, or recirculated from the sink through Scope Buddy. Scope Buddy can be used in either format depending on the facility’s endoscope cleaning procedure. Caution: Verify fluid exiting the open end of the endoscope or tubing is not facing a direction that would allow fluid to be sprayed on personnel or electrical fixtures. The endoscope should always be fully immersed under water.
Flushing an Endoscope with Detergent: 1. Plug the unit in and then turn it on by pressing the power switch located on the side of the unit. The timer display will now illuminate. See Figure 3.
16