2-8-511 HOOKUP CONNECTOR ASSEMBLY REPLACEMENT INSTRUCTIONS Using MEDIVATORS® ADVANTAGE and MEDIVATORS® ISA® Automated Endoscope Reprocessors
HOOKUP DIAGRAM
Medivators Inc. 14605 28th Ave N. Minneapolis, MN 55447 USA Medivators BV Sourethweg 11 6422PC Heerlen The Netherlands
LOT XXXXXX ID: XXXX
P/N: X-X-XXX
S
CONNECTOR ASSEMBLY IDENTIFICATION CHART Port ID
Port Name
Order Number
LEAK TEST
H28511-1
JET/AUX
H28511-26
BIOPSY
H28511-3
AIR
ELEVATOR
Connectors
DIRECTIONS FOR REPLACEMENT OF CONNECTOR ASSEMBLIES: 1. Identify if the Hookup Connector Block is the new (Figure 1 below) or previous (Figure 2 below) design. 2. Identify current Connector Assembly on the Hookup Connector Block to be replaced. 3. Remove current Connector Assembly by pulling associated tubing B from the Hookup Connector Block Barb A (NOTE: If using the previous design, unscrew the Connector Coupling F of Connector Assembly from the Port Connector C, keeping the Port Connector firmly held in place and pull the tubing off the Port Connector stem D). Discard this Connector Assembly. 4. Retrieve the enclosed new Connector Assembly. Note that it may not be identical to the Connector Assembly it is replacing. Discard Ferrule E and Connector Coupling F if using the new design. Confirm Connector Assembly labeling matches the Hookup Connector Block Label Port Name and the Port ID Number using the Connector Assembly Identification Chart. NOTE: Port ID # may not be identified on some older Hookup Connector Block labels. 5. Push tubing B of the new Connector Assembly onto the Hookup Connector Block Barb A. NOTE: If experiencing difficulty in attaching the tubing to the Hookup Connector Block Barb A, dip the tip of the tubing into hot water for 10 seconds. (NOTE: If using the previous design, push tubing of new Connector Assembly onto the Port Connector Stem C and manually screw Connector Coupling F onto Port Connector C until tight. Do not over tighten. Ensure Port Connector C does not turn. If needed, use a wrench to prevent turning.)
HOOKUP CONNECTOR BLOCK TYPES Exploded View of Port Connector, Coupling Connector, Ferrule and Tubing:
H28511-4
Figure 1 - New Design N/A
WATER
H28511-26
SUCTION
H28511-7
N/A A. Hookup Connector Block Barb
B. Hookup Tubing
Figure 2 - Previous Design C. Port Connector
EXTRA
N/A
S
N/A
78400-930
N/A
CHANNEL SEPARATOR 78400-930
If a Connector Assembly needs replacement, locate the port number/name and use the corresponding order number found in the chart above. Connector Assemblies are specific to individual hookups. © 2016 Medivators Inc. 50098-804/C
E. Ferrule
D. Stem
F. Connector Coupling
NOTE: Ferrule must be present to prevent leaking.
WARNINGS & CAUTIONS This hookup is intended for connection only with the specific endoscope models identified in the appropriate MEDIVATORS online hookup guide. Use of this hookup with endoscopes other than those specified by the online hookup guide may result in inadequate HLD. Users operating MEDIVATORS AERs and hookups must be trained and competent in the understanding of endoscope channel systems. Prior to reprocessing, users must verify that the hookup used contains connections for all appropriate channels that require a separate reprocessing connection according to the hookup instructions. All connectors and adapters must remain firmly attached and unrestricted for the entire disinfection cycle to ensure adequate HLD. Users must inspect the endoscope and hookup combination to verify proper connection and flow. Modifications or repairs to hookups which do not correspond to the manufacturer’s specifications may result in inadequate HLD and/or damage to endoscopes. Failure to properly clean and prepare endoscope for immersion prior to hookup connection may result in inadequate HLD and/or damage to the endoscope. Endoscopes must be inspected for damage and verified to be in proper working order prior to hookup connection. Connection of hookups to damaged endoscopes or endoscopes with obstructed/restricted channels could result in inadequate HLD. Manual leak testing must be carried out in accordance with manufacturer’s recommendations prior to reprocessing. Install all necessary waterproof caps, plugs and cleaning adapters prior to immersion. Do NOT autoclave MEDIVATORS hookups. MEDIVATORS makes no claim on high-level disinfection efficacy when these reprocessing instructions are not followed, or when this hookup kit is applied to endoscopes other than those specified in the online interactive hookup guide. Avoid biological contamination. Always wear personal protective equipment when handling endoscopes.
If you have any questions regarding this hookup, in the U.S., please contact Medivators Technical Support at 1-800-444-4729. Outside the U.S., please contact your Medivators representative. MEDIVATORS® is a registered trademark of Medivators Inc. ISA® is a registered trademark of Cantel Medical (Italy) S.r.l. All company or product names referenced are trademarks of their respective owners.
Medivators Inc. 14605 28th Avenue North Minneapolis, MN 55447-4822 USA Tel: +1.763.553.3300 Fax: +1.763.553.3387 www.medivators.eu
Medivators BV Sourethweg 11 6422 PC Heerlen The Netherlands Tel: +31.45.5.471.471 Fax: +31.45.5.429.595
Cantel (France) SAS 3 rue du pré fermé 31200 Toulouse France Tel: +33 (0) 5.61.245.275 Fax: +33 (0) 5.61.247.001
Cantel Medical (Italy) S.r.l. Via Laurentina, 169 00071 Pomezia (RM) Italia Tel.: +39 06 9145399 Fax: +39 06 9146099 www.cantelmedical.it
Cantel Medical (UK) Limited Wolseley House Staffordshire Technology Park Beaconside, Stafford ST18 0GA United Kingdom Tel: +44 (0) 1785.782420 Fax: +44 (0) 1785.782421
Cantel Medical Asia/Pacific Pte. Ltd. 1A International Business Park #05-01 Singapore 609933 Tel: +65.6227.9698 Fax: +65.6225.6848
Cantel Medical Devices (China) Co. Ltd. Unit 804-805, Innov Tower Block A, Hongmei Road, Xuhui 200233 Shanghai Tel: +86 21 60161380 Fax: +86 21 61210913