Hookup Connector Assembly Replacement Instructions
2-8-627 / 2-8-627HAN / 2-8-627CAS
Using MEDIVATORS™ ADVANTAGE PLUS™ Reprocessors, ADVANTAGE PLUS™ Pass-Thru Reprocessors and ISA™ Reprocessors.
Directions for Replacement of Connector Assemblies
Hookup Diagram MEDIVATORS Hookup
MD
AIR
1. Identify if the Hookup Connector Block is the new (Figure 1 below) or previous (Figure 2 below) design. 2. Identify current Connector Assembly on the Hookup Connector Block to be replaced. 3. Remove current Connector Assembly by pulling associated tubing B from the Hookup Connector Block Barb A (NOTE: If using the previous design, unscrew the Connector Coupling F of Connector Assembly from the Port Connector C, keeping the Port Connector firmly held in place and pull the tubing off the Port Connector stem D). Discard this Connector Assembly. 4. Retrieve the enclosed new Connector Assembly. NOTE: that it may not be identical to the Connector Assembly it is replacing. Discard Ferrule E and Connector Coupling F if using the new design. Confirm Connector Assembly labeling matches the Hookup Connector Block Label Port Name and the Port ID Number using the Connector Assembly Identification Chart. NOTE: Port ID Number may not be identified on some older Hookup Connector Block labels. 5. Push tubing B of the new Connector Assembly onto the Hookup Connector Block Barb A. NOTE: If experiencing difficulty in attaching the tubing to the Hookup Connector Block Barb A, dip the tip of the tubing into hot water for 10 seconds. (NOTE: If using the previous design, push tubing of new Connector Assembly onto the Port Connector Stem C and manually screw Connector Coupling F onto Port Connector C until tight. Do not over tighten. Ensure Port Connector C does not turn. If needed, use a wrench to prevent turning.)
ELEVATOR
™
WATER SUCTION
BIOPSY JET / AUX
www.medivators.com
Hookup Connector Block Types Exploded view of Port Connector, Coupling Connector, Ferrule and Tubing:
EXTRA
LEAKTEST
Figure 1 - New Design
A
Connector Assembly Identification Chart Port ID
Port Name
Order No.
1
LEAK TEST
H28627-1
Hookup Connector Block Barb
Connector
B
Hookup Tubing
Figure 2 - Previous Design
2
JET/AUX
C
Port Connector
E
H28627-26
F
D Stem 3
BIOPSY
H28627-3
4
AIR
H28627-4
5
ELEVATOR
H28627-5
6
WATER
H28627-26
7
SUCTION
H28627-7
8
EXTRA
N/A
S
N/A
47049-947
Ferrule
Connector Coupling
NOTE: Ferrule must be present to prevent leaking.
Warnings & Precautions • This hookup is intended for connection only with the specific endoscope models identified in the appropriate Medivators online hookup guide. Use of this hookup with endoscopes other than those specified by the online hookup guide may result in inadequate HLD. • Users operating Medivators AERs and hookups must be trained and competent in the understanding of endoscope channel systems. • Prior to reprocessing, users must verify that the hookup used contains connections for all appropriate channels that require a separate reprocessing connection according to the hookup instructions. • All connectors and adapters must remain firmly attached and unrestricted for the entire disinfection cycle to ensure adequate HLD. Users must inspect the endoscope and hookup combination to verify proper connection and flow. • Modifications or repairs to hookups which do not correspond to the manufacturer’s specifications may result in inadequate HLD and/or damage to endoscopes. • Failure to properly clean and prepare endoscope for immersion prior to hookup connection may result in inadequate HLD and/or damage to the endoscope. • Endoscopes must be inspected for damage and verified to be in proper working order prior to hookup connection. Connection of hookups to damaged endoscopes or endoscopes with obstructed/restricted channels could result in inadequate HLD. • Manual leak testing must be carried out in accordance with manufacturer’s recommendations prior to reprocessing. • Install all necessary waterproof caps, plugs and cleaning adapters prior to immersion. • Do NOT autoclave Medivators hookups. • Medivators makes no claim on high-level disinfection efficacy when these reprocessing instructions are not followed, or when this hookup kit is applied to endoscopes other than those specified in the online interactive hookup guide. • Avoid biological contamination. Always wear personal protective equipment when handling endoscopes.
N/A
If a Connector Assembly needs replacement, locate the Port Number/Name and use the corresponding order number found in the chart above. Connector Assemblies are specific to individual hookups.
1
SYMBOLS Symbol
Title
Explanatory Text
Standard Title
Reference
Manufacturer
Indicates the medical device manufacturer
Medical devices - Symbols to be used with information to be supplied by the manufacturer
I.S. EN ISO 15223-1:202 5.1.1
Authorized representative in the European Community
Authorized representative in the European Community/European Union
Medical devices - Symbols to be used with information to be supplied by the manufacturer
I.S. EN ISO 15223-1:202 5.1.2
Catalogue number
Indicates the manufacturer’s catalogue number so that the medical device can be identified
Medical devices - Symbols to be used with information to be supplied by the manufacturer
I.S. EN ISO 15223-1:202 5.1.6
Importer
Indicates the entity importing the medical device into the locale
Medical devices - Symbols to be used with information to be supplied by the manufacturer
I.S. EN ISO 15223-1:202 5.1.8
Caution
Indicates that caution is necessary when operating the device or control close to where the symbol is placed, or that the current situation needs operator awareness or operator action in order to avoid undesirable consequences
Medical devices - Symbols to be used with information to be supplied by the manufacturer
I.S. EN ISO 15223-1:202 5.4.4
NON-STANDARD SYMBOLS Symbol
Title Australian sponsor
For additional symbol reference information, definition and translation, please reference the Symbols Glossary on the STERIS website, www.steris.com. If you have any questions regarding this hookup, in the United States, please contact Cantel Medivators Technical Support Representative at 1-800-444-4729. Outside the United States, please contact your Medivators representative.
ADVANTAGE PLUS™, ISA™ and MEDIVATORS™ are trademarks of STERIS, its affiliates and related companies. All other company or product names referenced are trademarks of their respective owners.