Quick Guide
2 Pages
Preview
Page 1
For complete operational information, warnings,
cautions, please refer to the MEDRAD MRXperion CIT] and Injection System operation manual.
INJECTION STEPS
®
1. Install syringes and spikes.
Syringe A (Contrast) = Short spike Syringe B (Saline) = Long spike
2. Fill the syringes.
"' E G N RI Y S
< E G N
meDRaD® MRXperion
Double pressCID and a.to fill the syringes.
i�i i)ii: )'"B
Wait 3-5 seconds for Auto Fill to complete to avoid
3 •
5. Press
e,J
I�)
syringes facing down.
7. Connect tubing to the patient IV. 8. Press
(C)J
9. Press
RIil
Press to activate the priming function. Button illuminates to indicate that the operator can begin priming.
[I 5.
AUTO FILL Double press the A and B buttons to Auto Fill the syringes. To change the Auto Fill volume, press A or B once. Then use the + or - buttons. Press A or B again to begin Auto Fill to the adjusted volume.
+o confirm that you have
6. Rotate the injector head with the
• 4. PRIME BUTTON
2. FILLING THE SYRINGES
to prime air from tubing.
expelled all air from the syringes and tubing. {Illuminated= You have checked for air.)
■
MANUALLY FILL Enable the piston controls by pressing the Enable Piston Button. {. �� , .,. �;;,..
�
Use the Forward and Reverse Piston controls or
the Manual Knobs (shown below) to change the amount of fluid in the syringes.
■
12. START/HOLD BUTTON
10. Press
to lock the protocol from the control room.
I CHECKED FOR AIR CONFIRMATION BUTTON Press to confirm that the operator has examined the syringes and tubing for the presence of air. Illuminates to indicate that the operator has confirmed that the syringes and tubing have been inspected for the presence of air. (A confirmed check for air is required to continue.) ENABLE PISTON CONTROL BUTTON Press to activate the forward and reverse piston controls. Deactivates after ten seconds of inactivity.
[1 8.
•o perform an optional test injection. (Fast blinking= Test injection in progress.)
•o activate KVO (keep vein open). (Illuminated = KVO in progress.)
INJECTOR HEAD CONTROLS These steps correspond to the Injector head only, For complete steps see injection steps on the left.
A = Contrast B = Saline (flush)
4. Press
MEDRAD® MRXPERION INJECTION SYSTEM QUICK GUIDE
Press to initiate an injection when the system is Armed. Press to hold an injection in progress.
liJ
ABORT BUTTON Press to terminate the injection and disarm the injector.
TEST INJECT BUTTON Press to initiate a test injection. Blinks to indicate that the operator can perform a test injection.
9. KVO BUTTON Press to activate the KVO (keep vein open) feature. Blinks to indicate that the operator can start KVO to maintain patency of the patient connection.
11. ARM BUTTON
Press to arm the system after the protocol has been locked from the control room.
11. Press� on the injector head or on the display to arm the injector.
12. Press (
@
)on the injector head or
Pod to start the injection.
MANUAL KNOBS Locked during Auto Fill or when Enable Piston Control is activated. Can be used to manually fill the syringes.
meDRaD® MRXperion MR Injection System
Protocols
Create/Edit a Protocol
Find Protocol
1. Create/edit a protocol on the Home screen. 2. Save/recall protocols in Protocol Manager. 3. Organise/delete protocols in Protocol Manager Setup.
1. Select an arrow to create a phase. 2. Choose a phase type from the list. 3. Add/edit volumes and flow rates. 4. Select a pressure limit. 5. Press Lock to lock the protocol.
1. On the Home screen, select Protocol Manager. 2. Select View All to see all saved protocols. 3. Select applicable protocol from list. 4. Review all protocol values. 5. Select OK to use or edit the protocol.
System Settings 1. Shutdown/Restart Injector 2. System Setup (clock, units, display, etc.) 3. eGFR and Weight Based Dosing Calculator Setup (formulas, units, etc.) 4. Organise protocols 5. Fluid delivery settings (KVO interval, enable/disable test inject, etc.) 6. Informatics setup (optional feature) Syalem Setup
Setup
>I VlrtuaICare
System
ID1.0010.0� Hold I I (!)1.00
$
Stalul
Setup
2
Fluld Deltvery
Help
®
Shutdown
@(S)(Z)
@
®
Protocol Manager © Selup
VirtualCare
II�
FluklO.itv.,y
Help @ Shutdown
@(S)0
--·::.-·-@ ·-· -
Calculator
Calculator
Setup
(D
lfdformallcs
Setup
® ®
Status
Contacting Service If VirtualCare ® is installed: Setup
1. Select VirtualCare® from the Launch Menu on the Home screen. 2. On the Service screen, the connection status will indicate "Online." 3. Under "Request a Call From Support", select Technical. 4. When prompted, enter phone number and contact name. 5. Press Continue. Bayer Services will contact you shortly at the contact information provided.
Help
@ Shutdown
Australian Sponsor IMAXEON Pty Ltd
Status
ABORT BUTTON
Press to terminate the injection
Unit 1, 38-46 South Street Rydalmere NSW 2116 Australia Customer Service: 1800 633 723 (Toll-free within Australia) Tel:+61 2 88454999 Fax: +61 2 9391 6542 Email: [email protected]
Bayer, the Bayer Cross, VirtualCare, MEDRAD, MRXperion, and MEDRAD MRXperion are trademarks of Bayer and may be registered in the US and other countries. © 2020 Bayer. All rights reserved. PP-PF-RAD-CA-0366-1 Date of preparation: February 2020
POD BUTTONS
VirtualCare
®@0 Fabricant : Bayer inc. 2920, Matheson Blvd East, Mississauga (Ontario) L4W 5R6 Téléphone : 800-268-1432 Télécopieur : 800-567-1710
>
Calculator
and disarm the injector.
POWER BUTTON
Press to turn the system ON or OFF. Provides the operator with shutdown options or System Restart.
CE
2797
www.radiology.bayer.com.au
START/HOLD BUTTON
Press to initiate an injection when the system is Armed. Press to hold an injection in progress. HELP BUTTON
Press to launch the Help System.