Disassembly Instructions
1 Page
Preview
Page 1
Tracking #
Enclosure 2: Disassembly Instructions Medrad® Continuum™ MR Infusion System SAFETY WARNING: Do Not Disassemble Continuum System in MR Scan Room. Remove the Continuum system from the MR Scan Room. Do not remove the pump from the bracket or the bracket from the IV pole while in the MR Scan Room. Do not perform any disassembly of the IV pole while it is in the MR Scan Room. These pieces contain ferromagnetic material and may be attracted toward the magnet. Disassembly Procedure: 1. If using the MR compatible power supply, disconnect it from the bracket. 2. Remove the Continuum IV pole with bracket(s) and pump(s) still attached from the MR Scan Room. 3. Remove and discard any items hanging from the IV pole. 4. Ensure the Continuum Pump is turned off. 5. Loosen the telescoping pole adjustment knob and lower the inner pole to the lowest height. Re-tighten the adjustment knob. 6. Use one hand to hold the counterweight in place. Carefully lay the IV pole onto the floor. Ensure the bracket(s) do not contact the floor. 7. Continue holding the counterweight in place. Remove the bolt that secures the pole and leg assembly using a 6 mm hex wrench. Note: The bolt has thread adhesive and may require extra effort to loosen and remove it. 8. Set the IV Pole upright on the casters. 9. Carefully lift the pole off the leg assembly and place it on a horizontal surface. If the pole has more than one pump/bracket, it will be more difficult to handle when it is removed from the leg assembly. 10. If applicable, remove the fluid cup and holder from the pole. The fluid cup and holder can be discarded. 11. Loosen the mounting knob on the lowest bracket and slide it off the open end of the pole. 12. Repeat step 11 for each additional bracket. 13. If applicable, remove the syringe holder from the pole. 14. Carefully lift the counterweight off the leg assembly.
Note: The counterweight weighs approximately 22 lbs. (10 kg). Observe proper safety measures when handling the counterweight.
Removal of Continuum Accessories Any components and accessories supplied with, or purchased for, the Continuum system will also need to be returned. Accessories include the following:
Non-Wireless Ready Systems Battery Charger(s) Wireless / Wireless Ready Systems Non-MR Compatible Power Supply – OR -- MR Compatible Power Supply
Continuum Wireless Remote Display Accessory o Remote Display o Remote Display Power Supply o Remote Display Stand (VESA Mount)
Certo Wireless Network (if used exclusively for Wireless Continuum) o Control Room (CR) Module o Scan Room (SR) Module o External Cisco Access Point (Control Room)
RAD-MR-14-06441
April 2014