Barrx™ radiofrequency ablation system
Quick reference guide
Barrx™ radiofrequency ablation system
Quick reference guide The following information is intended as a quick reference. For full details, please refer to this product's Instructions For Use (IFU). Contact customer service or your sales representative for the most in–to–date revision of the package insert. Task
Review
Basic Troubleshooting
Problem: No RF Power Output
• Energy Generator not plugged in • Energy Generator not turned on
• Energy Generator in STANDBY mode • Energy Generator in FAULT mode
• Energy Generator still in ABLATION COMPLETE mode • Fault in accessory or Footswitch
• Output Cable not connected to the Energy Generator
• No Ablation Catheter Connected • Defective Ablation Catheter
• Energy Density Parameter Not Set • Internal Energy Generator failure
• Balloon Inflation Error. The pressure in the Balloon Based Ablation Catheter or Barrx 360 Express RFA Balloon Catheter is less than the target pressure
Operational and Fault Codes
If a fault condition occurs, the System Status display on the front panel will display an operational code, and all others will be blank. The interpretation of the operational, error and fault codes, as well as possible causes and solutions are listed in Tables 3.A, 3.B, 3.C and 3.D. In the event of an operational /error or fault code, endoscopic visualization must be used to verify complete balloon deflation if using a Balloon Based Ablation Catheter, Barrx 360 Express RFA Balloon Catheter, or Sizing Balloon. If the Energy Generator displays an E95 operational code when using a Balloon Based Ablation Catheter, Barrx 360 Express RFA Balloon Catheter, or Sizing Balloon, this is probably due to an air leak within the system. In such situations verify the following connections are secure: • The pneumatic connector between the Energy Generator and the Output Cable; • The pneumatic connector between the Output Cable and the Barrx filter; • The connection between the filter and the Ablation Catheter.
If the E95 operational code is observed again, there may be an air leak in the Ablation Catheter. Under endoscopic visualization, confirm that the balloon is fully deflated and then remove the Ablation Catheter. If the balloon is not deflated by depressing the DEFLATE control button or by depressing the Auto Inflation Foot Pedal, then manually deflate the balloon using a syringe via the pneumatic connector. Remove and replace the Ablation Catheter. Complaint Filing
1. Complete the field and technical report form (which includes the information below) 2. Details must be included:
• Date Medtronic employee was first notified of the event/problem
• Initial Reporter: Who informed the sales representative or tech support of the event
• List of all products involved
• Accurate description of the problem
• Issues that the Device problem directly resulted into 3. Retain defective product for return
Barrx™ radiofrequency ablation system
Quick reference guide Troubleshooting Codes
Table 3.A. Post Faults (Unrecoverable) Error code
Description
System status messages
Auto deflate function active on error for balloon based catheters
P01
POST CPU Configuration Operational Code
Disconnect Catheter P01 – Cycle Power
N/A
P02
POST ROM Test Operational Code
Disconnect Catheter P02 – Cycle Power
N/A
P03
POST RAM Test Operational Code
Disconnect Catheter P03 – Cycle Power
N/A
P04
POST Power Operational Code
Disconnect Catheter P04 – Cycle Power
N/A
P05
POST Impedance Operational Code
Disconnect Catheter P05 – Cycle Power
N/A
P06
POST Voltage Operational Code
Disconnect Catheter P06 – Cycle Power
N/A
P07
POST Current Operational Code
Disconnect Catheter P07 – Cycle Power
N/A
P08
POST Reference Voltage Operational Code
Disconnect Catheter P08 – Cycle Power
N/A
P09
POST Watchdog Test Operational Code
Disconnect Catheter P09 – Cycle Power
N/A
P010
POST Output Transformer Impedance Operational Code
Disconnect Catheter P10 – Cycle Power
N/A
P011
POST Pressure Sensor Offset Operational Code
Disconnect Catheter P11 – Cycle Power
N/A
P012
POST Pressure Reference Voltage Operational Code
Disconnect Catheter P12 – Cycle Power
N/A
P015
POST Real Time Clock Test Operational Code
Disconnect Catheter P15 – Cycle Power
N/A
P016
POST Real Time Clock Date Operational Code
Disconnect Catheter P16 – Cycle Power
N/A
P017
POST Switch Closed Operational Code
Disconnect Catheter P17 – Cycle Power
N/A
Barrx™ radiofrequency ablation system
Quick reference guide Table 3.B. Software/Hardware Faults (Unrecoverable) Error code
Description
System status messages
Auto deflate function active on error for balloon based catheters
H10
COP Operational Code
Disconnect Catheter H10 – Cycle Power
Yes
H11
Illegal CPU Instruction
Disconnect Catheter H11 – Cycle Power
Yes
H12
Duplicate Variable Operational Code
Disconnect Catheter H12 – Cycle Power
Yes
H13
Software Operational Code
Disconnect Catheter H13 – Cycle Power
Yes
H15
Reference Voltage Operational Code
Disconnect Catheter H15 – Cycle Power
Yes
H16
Vrms Offset Operational Code
Disconnect Catheter H16 – Cycle Power
Yes
H17
Irms Offset Operational Code
Disconnect Catheter H17 – Cycle Power
Yes
H18
Pmes Offset Operational Code
Disconnect Catheter H18 – Cycle Power
Yes
H19
RF ON Synch Operational Code
Disconnect Catheter H19 – Cycle Power
Yes
H20
Pressure Out of Range Operational Code
Disconnect Catheter H20 – Cycle Power
Yes
H22
Output Transformer Channel Set Operational Code
Disconnect Catheter H22 – Cycle Power
Yes
H23
Duplicate Pressure Operational Code
Disconnect Catheter H23 – Cycle Power
Yes
Barrx™ radiofrequency ablation system
Quick reference guide Table 3.C. Catheter Errors (Recoverable) Error code
Description
System status messages
Auto deflate function active on error for balloon based catheters
C50
EEPROM CRC Code
C50 – Reconnect or Discard Catheter
No
C51
Invalid Catheter ID Code
C51 – Reconnect or Discard Catheter
No
C52
Expired Catheter Code
C52 – Max Time Reached Discard Catheter
Yes
Sizing Balloon: Discard Catheter C53 – Max Use Reached nnmm Recommended Barrx 360 RFA Balloon Catheter: Discard Catheter C53 – Max Use Reached
C53
Catheter Use Exceeded Code
nnn% nnn% nnn% Energy Delivered Barrx 360 Express RFA Balloon Catheter: Discard Catheter C53 – Max Use Reached
Yes
Energy Delivered Non Balloon Based Ablation Catheter: Discard Catheter C53 – Max Use Reached nn% Energy Delivered C55
Catheter Type Code
C55 – Reconnect or Discard Catheter
Yes
C56
Filter Blocked Code
C56 – Filter Blocked Remove Filter Deflate with Syringe Discard Catheter
No
C57
EEPROM Setting Code
Unusable Catheter C57 – Discard Catheter
No
Barrx™ radiofrequency ablation system
Quick reference guide Table 3.D. Application Operational Codes (Recoverable) Error Code
E71
Description
System Status Messages
Handpiece Communication Loss
Barrx 360 Express RFA Balloon Catheter: E71 – Catheter Communication Error
Auto deflate function active on error for balloon based catheters
Require Reset Button
Yes
Yes
Yes
N/A, catheter must be disconnected and discarded
Yes
N/A, catheter must be disconnected and discarded
Yes
No
Press Reset
E72
E73
IO Expander Readback Error
Barrx 360 Express RFA Balloon Catheter: E72 – Catheter Electronics Fault
MOSFET Stuck Error
Barrx 360 Express RFA Balloon Catheter: E73 – Catheter Electronics Fault
Replace Catheter
Replace Catheter Barrx 360 RFA Balloon Catheter: E81 – Catheter Too Big Check Sizing Results Consider Smaller Ablation Catheter nnn% nnn% nnn% Energy Delivered E81
Low Impedance Error
Barrx 360 Express RFA Balloon Catheter: E81 – Clean Electrode Low Energy Delivered Non Balloon Based Ablation Catheter: E81 – Clean Electrode nn% Energy Delivered n Ablations
Barrx™ radiofrequency ablation system
Quick reference guide Table 3.D. Application Operational Codes (Recoverable) Error Code
Description
System Status Messages
Auto deflate function active on error for balloon based catheters
Require Reset Button
Yes
No
Yes
No
Barrx 360 RFA Balloon Catheter: E82 – Clean Electrode or Poor Elec Contact nnn% nnn% nnn% Energy Delivered
E82
High Impedance Error
Barrx 360 Express RFA Balloon Catheter: E82 – Clean Electrode or Poor Elect Contact Low Energy Delivered Non Balloon Based Ablation Catheter: E82 – Clean Electrode or Poor Elec Contact nn% Energy Delivered n Ablations Barrx 360 RFA Balloon Catheter: E83 – Clean Electrode or Poor Elec Contact nnn% nnn% nnn% Energy Delivered
E83
Voltage Limit Error
Barrx 360 Express RFA Balloon Catheter: E83 – Clean Electrode or Poor Elect Contact Low Energy Delivered Non Balloon Based Ablation Catheter: E83 – Clean Electrode or Poor Elec Contact nn% Energy Delivered n Ablations