MIDAS REX® MR7
Pneumatic High-Speed System
Instruction Manual Rx Only
Symbols The following symbols may appear within this manual, on product labeling, or on the product itself: Attention, see Instructions for Use.
Tube Control
United States federal law restricts this device to sale by or on the order of a physician.
Air Pressure Relief
Reference Number
Use with
Lot Number
Multi-Use Disposable Attachment
Serial Number Quantity
STERILE
Dissecting Tool
Approximately equal to
Attachment
Attachment
Non-Sterile
Control Unit
Control Unit
Sterilized by Gamma Irradiation
Refurbished
Refurbished
Temperature Limitations Unlock Lock
EC REP
Lubricant/Diffuser
Dissecting Tool
Date of manufacture
Lubricant/Diffuser
Instrument Case
For single patient use only. Do not re-use, re-process, or re-sterilize this product. Re-use, re-processing or re-sterilization may compromise the structural integrity of the device and/ or create a risk of contamination of the device, which could result in patient injury, illness, or death.
Use by date
|
Instrument Case
On
Accessory REGULATOR Bone Mill MOTOR Brush Adapter
Accessory Regulator Bone Mill Motor Brush Adapter
Off
USA Only
Finger-Operated Control
Manufacturer
Foot-Operated Control
Do not dispose to unsorted municipal waste.
Compliant with European Council Directive MDD 93/42/EEC.
Tool Control
Authorized Representative in the European Community
Consult Instructions for Use
Contents General Information...1 Indications for Use...1 Contraindications...1 Special Notices...1 General Safety Precautions...2 No Latex Policy...2 System Components...3 Non-Disposable Components...3 Disposable Components...8 Legend® Dissecting Tools...9 General Guidelines For Attachment and Tool Applications... 10 Setting up the Operating Room... 13 Power Source Requirements... 13 Pneumatic Connections... 13 Setting up the MR7 System... 14 Installing the Oiler Cartridge... 14 Connecting the Motor... 15 Installing an Attachment and Tool... 16 Straight Attachments... 16 Specialized Attachments... 16 Activating the Motor... 17 Disassembling the MR7 System... 18 Depressurize the System... 18 Disconnect Hoses... 18 Discard the Lubricant/Diffuser Cartridge... 18 Remove the Attachment and Tool... 18 Cleaning & Sterilizing the MR7 System... 19 MR7 Motor... 19 Legend Attachments / Tubes... 20
i
MR7 Pneumatic Control Unit / Regulator Hose / Triton Adapter / Instrument Case... 22 Transmissible Spongiform Encephalopathies (TSE) Return Policy... 22 Troubleshooting... 23 Refurbishing or Repairs... 26 Preventative Maintenance... 27 Limited Warranty... 28 Appendix A-Specialized Attachments... 29 Angled Attachments... 29 Angled Double Lock Attachments... 30 Curved Bur Attachments... 32 Variable Exposure Attachments... 33 Footed Attachments... 33 Rotating Footed Attachments... 34 Contra-Angle Attachment (16-MF)... 35 Metal Cutting Attachments... 36 Telescoping Attachments... 37 Perforator Driver Attachments... 38 Jacobs® Chuck Attachments... 39 Bone Mill Attachment... 40
ii
General Information
General Information Read and understand this manual before use of the MR7 System. The Midas Rex® MR7 system is designed for use by medical professionals familiar with powered surgical instrumentation. The surgeon is responsible for learning the proper techniques in the use of this system, as inappropriate use may potentially be harmful. It is strongly recommended that the surgeon and dedicated operating room personnel are knowledgeable with the use of this equipment by being trained in Medtronic Midas Rex Hands-On Workshops or by one of the local authorized representatives. The MR7 system consists of the following components: • MR7 or MR7 Touch Motor • MR7 Pneumatic Control Unit with Various Connectors • MR7 Regulator Hose • MR7 Lubricant/Diffuser Cartridge • MR7 Triton® Adapter (optional) • Legend® Attachments* • Legend® Dissecting Tools* *The MR7 system uses the same attachments and dissecting tools as the Legend® Pneumatic High-Speed System.
Indications for Use The Medtronic Midas Rex MR7 System is a pneumatically operated surgical instrument system. The pneumatic motors provide power to operate removable rotating surgical cutting tools and their accessories intended for use in neurosurgery, including craniotomy and spinal surgery; as well as Ear, Nose and Throat (ENT), orthopedic and general surgical applications including maxillofacial, craniofacial and sternotomy surgeries.
Contraindications None
Special Notices The words warning, caution and note have special meanings in this manual, and should be carefully reviewed: WARNING: A warning indicates that the personal safety of the patient or physician may be involved. Disregarding this information could result in injury to the patient or physician. CAUTION: A caution indicates that there is a risk of damaging equipment. NOTE: A note is intended to provide additional information, which may be useful, but is not essential to complete the procedure.
1
General Information
General Safety Precautions WARNINGS: •
Do not use the Midas Rex MR7 System before proper cleaning and sterilization.
•
Do not operate the Midas Rex MR7 System in the presence of Magnetic Resonance Imaging devices.
•
Do not use damaged, faulty, or modified Midas Rex MR7 System components. Inspect the Midas Rex MR7 System for damage prior to each use: • • •
Check the motor’s exhaust hose for cracks or tears. Visually inspect attachments and tools. Do not use bent or damaged tools. Install attachment and dissecting tool, then briefly run motor. * Check motor for overheating and leaking lubricant. * Check attachment for overheating. * Check dissecting tool for flail.
•
Do not operate the Midas Rex MR7 System without eye protection.
•
Motors and attachments which fail due to extended use may allow a component to detach and fall from the motor or attachment, and may cause patient injury.
•
Heavy side loads and/or long operating periods may cause the device to overheat. If overheating occurs: • • •
Never place an overheated motor on the patient or draping during surgery. Discontinue use and rest the motor by using intermittently, or wrap the motor/attachment interface with a moist sterile towel. If the motor is passed off, the receiver should grasp the motor by the proximal end close to the motor hose.
•
To avoid injury to the patient or user, do not place the handpiece on the patient or in an unsecured location, when not in use.
•
Midas Rex MR7 motors should only be operated when the attachment is in the position.
No Latex Policy Legend and MR7 products, packaging materials, labels, package inserts, and similar items manufactured by and/or for Medtronic Powered Surgical Solutions (MPSS) do not contain latex.
2
System Components
System Components Non-Disposable Components MR7 Motor The MR7 motor is a high-speed, high-torque motor used to dissect bone and biomaterials. Figure 1: Motor Components
1
3
2
4 5
6
1. Collet 2. Motor Case 3. Swivel 4. Hose 5. Finger Control Lever 6. Safety Slide (MR7 Touch Only) In addition to the components listed in Figure 1, each MR7 motor has a lubricant/diffuser housing at the end of the motor hose, as seen in Figure 2. Figure 2: Lubricant/Diffuser Housing
3
System Components
WARNING: Use only Medtronic Midas Rex Legend or MR7 devices with an MR7 motor. Use of other devices may cause injury or damage equipment, and will void the manufacturer’s warranty.
Figure 3: Pneumatic Control Unit
Pneumatic Control Unit The pneumatic control unit (Figure 3) provides variable speed motor control controls through a foot pedal. It also allows the user to switch between finger and foot control of the motor (if applicable).
Regulator The regulator (Figure 4) controls the delivery pressure of compressed gas to the pneumatic control unit. The pressure gauges monitor cylinder pressure (right gauge) and delivery pressure (left gauge). Note: Outlet pressure gauge accurate to +/- 12 psi.
Instrument Case The instrument case (Figure 5) is used to organize equipment.
Figure 4: Regulator
Regulator Hose Connects from the gas source to the pneumatic control unit to deliver compressed gas.
Figure 5: Instrument Case
4
System Components
N2 DISS to Male Schrader Adapter (PC110)
Figure 6. N2 DISS to Male Schrader Adapter
The N2 DISS to male Schrader adapter (Figure 6) allows for the regulator hose to be attached to a female Schrader in-house gas connection.
N2 DISS to Male Schrader Adapter - UK (PC111) The N2 DISS to male Schrader adapter (Figure 6a) allows for the Pneumatic Control Unit’s N2 DISS pressure hose connection to be attached to a Female Schrader (UK) in-house gas connection.
N2 DISS to Air DISS Adapter (PC120)
PC110 Figure 6a. N2 DISS to Male Schrader Adapter - UK
The N2 DISS to air DISS adapter (Figure 7) allows for the regulator hose to be attached to an Air DISS inhouse gas connection.
N2 DISS to Female Schrader Adapter (PC140) The N2 DISS to female Schrader adapter (Figure 7a) allows for the Pneumatic Control Unit’s N2 DISS pressure hose connection to be attached to a Male Schrader in-house gas connection.
PC111 Figure 7. N2 DISS to Air DISS Adapter
PC120 Figure 7a. N2 DISS to Female Schrader Adapter
PC140
5
System Components
N2 DISS to Surgical Tool Air Male SIS Adapter (PC150)
Figure 7b. N2 DISS to Surgical Tool Air Male SIS Adapter
The N2 DISS to surgical tool air male SIS adapter (Figure 7b) allows for the Pneumatic Control Unit’s N2 DISS pressure hose connection to be attached to an AUS-SIS surgical tool air in-house gas connection.
N2 DISS to WF4 Adapter The N2 DISS to WF4 adapter (Figure 8) allows for the regulator hose to be attached to a Midas Rex safety valve regulator previously used for Midas Rex Classic or Midas Rex III motors. The in-line oiler must be removed from the safety valve regulator.
PC150 Figure 8. N2 DISS to W4 Adapter
Motor Wrench The motor wrench (Figure 9) is used to align arrows on motor collet flats prior to installation of a Legend attachment.
Triton Adapter The Triton adapter (Figure 10) allows the Triton handpiece to be driven by the MR7 pneumatic control unit. It functions much the same way as the Triton port on the Legend pneumatic control unit, except that it is connected between the control unit and the gas source, rather than being integrated into the control unit
Figure 9. Motor Wrench
Figure 10. Triton Adapter
6
System Components
Legend Attachments Legend motor attachments are available in various designs to facilitate a variety of surgical procedures. Attachments vary in length, diameter, and overall design. They are marked and color-coded to correspond with their associated dissecting tools. A few of the Legend attachments available are listed in the table below. Attachment
Example
Standard Straight Attachments
AS09
Standard Angled Attachments
AA14
Straight Variable Exposure Attachments
AVS07
Angled Variable Exposure Attachments
AVA07
Fixed Footed Attachments
AF01
Rotating Footed Attachments
AF01R
Telescoping Attachments
AT10 (base) TT12A (tube)
Contra-Angle Attachment
AC16
Metal Cutting Attachment
ASMC
Perforator Attachment
AD01
5/32” Jacobs Chuck Attachment
AD02
Bone Mill Attachment
BM100
Other Details/Options
The telescoping attachment requires the use of the AT10 attachment base, as well as a telescoping tube. Tubes are available in straight, curved, or hooded form.
Available in 800 RPM or 1000 RPM form.
NOTE: Angled and straight attachments with the same length, marking, and color band share the same dissecting tool. Curved and straight telescoping tubes with the same length, marking, and color band also share the same dissecting tool. Example: The 14-AM straight and 14-AM angled attachments are 14 cm long, marked 14-AM and have a green color band. All dissecting tools with the prefix 14 (14MH30) may be used in either the 14-AM straight or 14-AM angled attachment. Be sure to match the color code and nomenclature on the Legend Dissecting Tool packaging with the color band and nomenclature on the Legend Attachment.
7
System Components
Disposable Components
Figure 11: Lubricant/Diffuser Cartridge
WARNING: Use only Medtronic Midas Rex Legend or MR7 devices with an MR7 motor. Use of other devices may cause injury or damage equipment, and will void the manufacturer’s warranty.
Lubricant/Diffuser Cartridge The lubricant/diffuser cartridge (Figure 11) provides lubrication to the motor and filters oil from exhausted air.
Telescoping Tubes Telescoping tubes (Figure 12) provide support to the rotating dissecting tool. Telescoping tubes are disposable following multiple uses and should be discarded when heat or excessive vibration is noticed or insertion of tools becomes difficult.
Figure 12: Telescoping Tube
Cleaning Brushes Cleaning brushes (Figure 13) are used to clean debris from lumen of attachments and telescoping tubes. Sized for an internal bore diameter of 3.2 mm, 2.4 mm or 1.2 mm in Legend Attachments and Telescoping Tubes. NOTE: Cleaning brushes will not pass through angled, contra-angle, metal cutting, perforator, or Jacobs Chuck attachments, because they are not cannulated. Figure 13: Cleaning Brushes
3.2mm 2.4mm 1.2mm
8
System Components
Legend® Dissecting Tools Legend dissecting tools are sterile cutting tools, intended for cutting bone and biomaterials.
Dissecting Tool Nomenclature Part numbers for Legend dissecting tools follow a standard naming convention, which is described in the diagram below. A basic part number consists of the associated attachment length, the tool head shape, and the tool head diameter. Part numbers may also include a variety of prefixes to identify specific attachment types, as well as a variety of suffixes to provide additional information about the dissecting tool. Tools that use a design taken from the Mednext line are designated by an additional “-MN” suffix.
Associated Attachment Length
Tool Head Diameter (x.x millimeters)
Optional Prefix Tool Head Shape
Optional Suffix
Tool Number Prefixes (not all inclusive) F...
For use with footed attachments
MC
For use with metal cutting attachments
T
For use with telescoping attachments
Tool Head Shapes (not all inclusive) AC
Acorn
MH
Match Head
BA
Ball
OV
Oval
CY
Cylinder
RT
Reverse Taper
HM
Hole Maker
TA
Tapered
HS
Hole Saw
TD
Twist Drill
Tool Number Suffixes (note that more than one of the suffixes listed may be combined in a single part number) L
Long
S
Spiral
D
Diamond
SH
Short
X
Extra
DC
Diamond Coarse
F
Fine
DX
Diamond Extra Coarse
C
Carbide
MN
Mednext Tool Design
9
System Components
WARNINGS: •
Dissecting tools are for single-use only. Do not attempt to sterilize them. The dissecting tools are packed sterile and are not intended for repeat use. To prevent contamination, use only once.
•
Do not use an attachment and dissecting tool combination that results in tool flail or excessive vibration.
•
Do not attempt to remove a tool while the motor is running.
•
Do not attempt to remove a tool from an overheated motor or attachment.
•
Do not use the device if the package is opened or damaged.
General Guidelines For Attachment and Tool Applications These are general guidelines for dissecting tool applications and are not an all-inclusive listing. WARNING: Be sure to match the color code and nomenclature on the Legend Dissecting Tool packaging with the color band and nomenclature on the Legend Attachment. Failure to do so could result in injury to the patient or operating room staff. Surgical Application Spine
10
Commonly Used Attachments
Commonly Used Dissecting Tools
Suggested Motor(s)
MR7, MR7 8-B, 9-M, 14-AM, Match Head Touch 15-A Elongated spherical design allows controlled, delicate dissection. For entry hole, nerve decompression, osteophyte removal, sinus dissection, etc. Ball Helical cutting flutes dissect bone or cement effectively from a wide variety of approach angles. For debridement, decortication, sinus dissection, etc. Oval Helical cutting flutes and curved design blend acorn and ball styles to vary dissection efficiency with approach angle. For decortication, laminotomy, entry hole, nerve decompression, osteophyte removal, etc. Hole Maker/Saw Matched sets of Hole Makers and Hole Saws are efficient and effective for interbody fusion. Cylinder Effective bone sculpting and planing. For graft shaping, debridement, corpectomy, decortication, interbody fusion, fusion takedown, etc. Acorn Curved design varies dissection efficiency with varied approach angles. For entry hole, laminotomy, bone shaping, debridement, corpectomy, decortication, fusion takedown, etc. Telescoping
Match Head Elongated spherical design allows controlled, delicate dissection. For entry hole, nerve decompression, osteophyte removal, sinus dissection, etc.
Footed, Straight
Tapered Slender design for precise dissection with minimal bone loss. For transection, osteotomy, graft harvesting, bone shaping, entry hole, suture hole, midface advancement, etc.
MR7, MR7 Touch
System Components
Surgical Application Neurosurgical– Cranial
General Surgery and Plastic Surgery (Craniofacial/ Maxillofacial/ Sternotomy)
Commonly Used Attachments
Commonly Used Dissecting Tools
Suggested Motor(s)
7-6ST, 8-B, 9-M, 10-9ST, 14-AM, 15-A
Match Head MR7 Elongated spherical design allows controlled, delicate dissection. For entry hole, nerve decompression, osteophyte removal, sinus dissection, etc. Ball Helical cutting flutes dissect bone or cement effectively from a wide variety of approach angles. For debridement, decortication, sinus dissection, etc. Twist Drill Helical design with stop produces a hole with a precise depth. Ideal for plating. Acorn Curved design varies dissection efficiency with varied approach angles. For entry hole, laminotomy, bone shaping, debridement, corpectomy, decortication, fusion takedown, etc.
Telescoping
Match Head Elongated spherical design allows controlled, delicate dissection. For entry hole, nerve decompression, osteophyte removal, sinus dissection, etc.
Footed
Tapered Slender design for precise dissection with minimal bone loss. For transection, osteotomy, graft harvesting, bone shaping, entry hole, suture hole, midface advancement, etc.
7-6ST, 8-B, 9-M, 10-9ST, 14-AM
Match Head MR7, MR7 Elongated spherical design allows controlled, delicate Touch dissection. For entry hole, nerve decompression, osteophyte removal, sinus dissection, etc. Ball Helical cutting flutes dissect bone or cement effectively from a wide variety of approach angles. For debridement, decortication, sinus dissection, etc. Tapered Slender design for precise dissection with minimal bone loss. For transection, osteotomy, graft harvesting, bone shaping, entry hole, suture hole, midface advancement, etc. Twist Drill Helical design with stop produces a hole with a precise depth. Ideal for plating.
Ear, Nose, and 7-6ST, 10-9ST Throat (Otology, Neurootology)
Ball Helical cutting flutes dissect bone or cement effectively from a wide variety of approach angles. For debridement, decortication, sinus dissection, etc.
MR7
11
System Components
Surgical Application Orthopaedics
Biometals/ Bioceramics/ Biomaterials
12
Commonly Used Attachments
Commonly Used Dissecting Tools
Suggested Motor(s)
8-B, 9-M, 14‑AM, 21-TU, 26-R, Footed, Telescoping
Ball MR7, MR7 Helical cutting flutes dissect bone or cement effectively Touch from a wide variety of approach angles. For debridement, decortication, sinus dissection, etc. Tapered Slender design for precise dissection with minimal bone loss. For transection, osteotomy, graft harvesting, bone shaping, entry hole, suture hole, midface advancement, etc. Acorn Curved design varies dissection efficiency with varied approach angles. For entry hole, laminotomy, bone shaping, debridement, corpectomy, decortication, fusion takedown, etc. Cylinder Effective bone sculpting and planing. For graft shaping, debridement, corpectomy, decortication, interbody fusion, fusion takedown, etc.
Footed
Tapered Slender design for precise dissection with minimal bone loss. For transection, osteotomy, graft harvesting, bone shaping, entry hole, suture hole, midface advancement, etc.
MC
Metal Cutter Cutting flutes or diamond wheel design remove metals, ceramics and other biomaterials effectively from a variety of approach angles. For cutting rods, pins, plates, implants, screws, etc.
MR7, MR7 Touch
Setting up the Operating Room
Setting up the Operating Room Power Source Requirements Required Operating (Dynamic) Pressure
Nominal Operating (Dynamic) Pressure
Approximate Flow Rate Required
Gas Type
80–120 psi
100 psi
12 cubic feet/min.
5.5–8.3 bar
6.9 bar
340 liters/min.
Nitrogen or Dry-Filtered Compressed Air
CAUTION: Do not run the motor at an operating pressure above or below the required operating pressure range. Operating pressure below 80 psi (5.5 bar) may not provide proper lubrication to the motor. Operating pressure above 120 psi (8.3 bar) may damage or reduce the life of the motor.
Pneumatic Connections Figure 14: Gas Connection Options 4 5 1. Regulator Hose (N2 DISS) 2. N2 DISS to Male Schrader Adapter - UK Gas Source 3. N2 DISS to Female Schrader Adapter Gas Source 4. N2 DISS to Surgical Tool Air Male SIS Adapter Gas Source 5. N2 DISS to Air DISS Adapter Gas Source 6. N2 DISS to Male Schrader Adapter Gas Source 7. Gas Source (N2 DISS) 8. Regulator 9. DISS/WF4 Adapter 10. Regulator
3
6
2
7
1
8
9
10
WARNING: You must ensure the adapter is fully threaded onto the regulator hose and that the adapter is fully threaded/ connected to the corresponding wall outlet before operating the Pneumatic system. CAUTION: If you are using the Midas Rex Safety Valve Regulator instead of the Legend Regulator, you must replace the in‑line oiler with the DISS/WF4 adapter before use.
13
Setting up the MR7 System
Setting up the MR7 System Installing the Oiler Cartridge WARNING: Do not use the MR7 system with the Midas Rex in-line oiler. The MR7 motor is sufficiently lubricated by the lubricant/diffuser on the motor hose, and will be over-lubricated if the Midas Rex in-line oiler is used. 1. 2.
3. 4.
Figure 15: Aligning the Lubricant/Diffuser Cartridge with the Housing
Set the non-running (static) pressure to 80–120 psi (5.5–8.3 bar) at the gas source. Operating (dynamic) pressure may be adjusted later. Hold the lubricant/diffuser cartridge perpendicular to the housing (Figure 15), and press the cartridge’s circular fitting onto the housing’s circular receptacle (Figure 16), breaking the foil seal. Rotate the cartridge down until it clicks into place. Verify that the symbol on the cartridge is lined up with the notch on the housing (Figure 17).
WARNINGS: •
Failure to properly secure the lubricant/diffuser cartridge may cause injury to operator and/or operating room staff.
•
Do not attempt to remove the lubricant/diffuser cartridge while the system is pressurized.
Figure 16: Pressing the Cartridge onto the Housing
CAUTIONS:
14
•
Do not use an MR7 motor without a lubricant/ diffuser installed.
•
Do not use a lubricant/diffuser cartridge for more than one hour of drill time.
•
Do not re-use a lubricant/diffuser cartridge. It is a single-use product.
•
Do not attempt to refill a used lubricant/diffuser cartridge.
•
Do not use a lubricant/diffuser cartridge if it appears to be damaged, or if the inner foil seal is punctured.
Figure 17: Correctly Installed Lubricant/Diffuser Cartridge
Setting up the MR7 System
Connecting the Motor
Figure 18: Connecting the Motor Hose to the Motor Port
Connect the motor hose to the motor port on the top of the pneumatic control unit, by swinging the port cover to the side and pressing the end of the hose into the port (Figure18). WARNING: Do not pinch, kink, obstruct, cut, tear, or step on the motor/exhaust hose. This may cause the hose to burst, potentially injuring the patient or user. NOTES: •
If using the MR7 Touch motor, slide the control slide on the pneumatic control unit to the position (Figure 19). This will automatically depress and lock the foot pedal. The control will not lock into the position unless the motor hose is connected into the motor port. When the motor hose is removed from the motor port, the foot pedal will return to normal position.
•
If using the Triton Power Surgical Instrumentation System in conjunction with the MR7 motor, use the optional Triton adapter to connect the Triton hose. Refer to the documentation accompanying the adapter for connection instructions.
•
The motor’s exhaust hose may have an oily film on the external surface from pressure and/or temperature differentials following sterilization. Wipe the exhaust hose with a sterile cloth prior to use. If motor continues to have oil on the exhaust hose, return the motor to MPSS for refurbishing.
Figure 19: Finger/Foot Control Slide
WARNING: To avoid injury to the patient or user, do not use the pneumatic control unit to operate systems other than the MR7, Legend, and Triton systems. Prior to installation of a Legend attachment and dissecting tool, ensure that the arrows on the motor collet flats are aligned (Figure 20). If the arrows are not aligned, use the motor wrench to turn the collet flat closest to the motor case until its arrow is aligned with the arrow on the other collet flat. WARNING: To avoid injury when using the MR7 Touch motor, ensure the safety slide is in the “O” position before installing the attachment and tool.
Figure 20: Aligning the collet flats
15
Installing an Attachment and Tool
Installing an Attachment and Tool WARNING: Dissecting tool flutes are sharp and may perforate surgical gloves. Tools may be grasped with a hemostat to aid in installation and removal.
Figure 21: Sliding the Attachment over the Motor Collet
Straight Attachments Installation: 1.
2.
3.
Slide the attachment over the motor collet, aligning the triangular markers (Figure 21). You will feel and hear the attachment click into place when it is fully seated. Insert the dissecting tool into the attachment with a slight rotational motion (Figure 22). You will feel and hear the tool click into place when it is fully seated in the attachment. Turn the attachment to the position on the motor case (Figure 23). Gently pull on the shaft of the dissecting tool to verify proper installation.
Figure 22: Inserting the Dissecting Tool into the Attachment
Removal: Removal is the reverse of installation.
Specialized Attachments See Appendix A-Specialized Attachments for installation and removal instructions for other attachments.
Figure 23: Attachment in the Locked Position
16