medtronic

HYSTEROLUX System

HYSTEROLUX Canister Setup Guide April 2020

Setup Guide

4 Pages

SETUP GUIDE HYSTEROLUX™ CANISTER SETUP Four Canisters  SIDE LEVER set to split suction  REG position  2  3 4  2  REAR port  1  2  5  3  4  Yellow outflow tube dropped into patient drape†  7  †No connection to canister.  4  FRONT port  Vacuum tube from vacuum source  5  3  2  3  1  6  7  6 Blue outflow tube from patient drape  6  SUCTION REGULATOR SETTINGS Soft Tissue Shaver Plus:  150 mmHg  Soft Tissue Shaver Mini:  200-300 mmHg  Dense Tissue Shaver Plus:  150-250 mmHg  Dense Tissue Shaver Mini:  250-300 mmHg  4  White outflow tube from handpiece  6  5  Attach the tubes as indicated, in the order and locations specified by the numbered steps.  5  Use the comparison chart on page 4 as a visual reference to cross-check the finished assembly. Use the suction regulator settings chart below to determine the correct suction setting for each shaver.
File Type: PDF
File Size: 1.37 MB
File Name: medtronic - HYSTEROLUX Canister Setup Guide April 2020 - 2020-04.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions