MicroAire

PAL Systems

MicroAire PAL-650 Cleaning and Sterilization Chart Rev D

Cleaning and Sterilization Chart

1 Page

PAL-650 Cleaning & Sterilization CLEAN (CAP ON) Connect CAP-600E or Cable to keep liquid out. Scrub with warm water, a soft brush and pH neutral detergent, then rinse.* Do not immerse. Do not put nose of handpiece into water stream.  STERILIZE (CAP OFF)  Dynamic-Air-Removal  Remove CAP-600E or Cable to allow steam to penetrate handpiece.  Exposure Time (Full Cycle)  Full dry cycle is required to prevent corrosion and to ensure proper performance.  Temperature  Cool slowly at room temperature. Never run while warm. Inspect carefully to ensure all moisture is completely removed.**  Dry Time  4 minutes 270° F (132° C) 20 minutes  Inspect Dry Instrument - Okay to Use  **Residual moisture can result in error messages and functional issues. *More information can be found in the PAL-650 Instruction Manuals. Additional languages available at www.microaire.com. IM-PAL-CS Rev D | August 2017
File Type: PDF
File Size: 277 KB
File Name: MicroAire - MicroAire PAL-650 Cleaning and Sterilization Chart Rev D.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions