Power Instruments and Accessories Sterilization Parameters Chart
1 Page
Preview
Page 1
Power Instruments and Accessories Large Bone Power
Series: 6000, 7000, Pulse Lavage
Small Bone Power Series: 1000, 2000, 5000
NiMH Batteries
REF: 7505-710, 6640-710
Gravity-Displacement Steam
Dynamic Air Removal (pre-vac)
Instrument
Minimum Time & Temperature
Min. Heated Dry Time
Instrument
Minimum Time & Temperature
Min. Heated Dry Time
Single Instrument
30 min. full cycle @ 132–135°C (270–275°F)
8 min.*
Single Instrument
4 min. full cycle @ 132–135°C (270–275°F)
8 min.*
Sterilization Tray
45 min. full cycle @ 132–135°C (270–275°F)
20 min.*
Sterilization Tray
4 min. full cycle @ 132–135°C (270–275°F)
10 min.*
Single Instrument
30 min. full cycle @ 132–135°C (270–275°F)
8 min.*
Single Instrument
4 min. full cycle @ 132–135°C (270–275°F)
8 min.*
Sterilization Tray
35 min. full cycle @ 132–135°C (270–275°F)
60 min.*
Sterilization Tray
4 min. full cycle @ 132–135°C (270–275°F)
45 min.*
Unwrapped
3 min. full cycle @ 132–135°C (270–275°F)
No drytime*
Wrapped OR Unwrapped
4 min. full cycle @ 132–135°C (270–275°F)
1 min.*
Sterrad® 100 Sterilization 44 minutes total expose time
Sterrad® NX Sterilization 28 minutes full cycle *Additional times listed in corresponding product Instructions for Use (IFU).
MicroAire Sterilization Parameters for Power Instruments and Accessories MicroAire Surgical Instruments, LLC www.microaire.com CH-PWRSTER Rev A
3590 Grand Forks Boulevard Charlottesville, VA 22911 USA
Domestic Phone: 800.722.0822 Fax: 800.648.4309 International Phone: 1.434.975.8000 Fax: 1.434.975.4134