MA706700i
User’s Guide
Refer to compliance chart in this manual.
003-1507-00 Rev N (2/6/14)
Product Information
(The information below is required when calling for service) Dealer [name / phone]:
Date of Purchase:
Model / Serial Number:
Midmark Authorized Service Company [name / phone]:
Model / Serial Number Location
MA706800i
2
© Midmark Corporation 2012
Table of Contents Important Information
Safety Symbols... 4 Warnings...5 Intended Use... 5 Electrical Requirements... 5 Electromagnetic Interference... 6 Disposal of Equipment... 6 Authorized CE Representative... 6 Transportation / Storage Conditions... 6
Operation
Back Section Positioning... 7 Stirrups...8 Footrest & Treatment Pan... 9 Footstep...9 Paper Roll & Tear Strap... 10 Drawer Heater (optional)... 11 Pelvic Tilt (optional)... 12 Receptacle (optional)... 13
Maintenance
Cleaning...14 Preventative Maintenance... 14 Calling For Service... 14
Specification
Weights & Dimensions... 15 Upholstery Dimensions... 16 Voltage, Current, Frequency Ratings & Compliance Chart... 17
Warranty Information
Limited Warranty... 18
3
© Midmark Corporation 2012
Important Information Safety Symbols Warning
Indicates a potentially hazardous situation which could result in serious injury.
Caution
Indicates a potentially hazardous situation which may result in minor or moderate injury. It may also be used to alert against unsafe practices.
Equipment Alert
Indicates a potentially hazardous situation which could result in equipment damage.
Note
Amplifies a procedure, practice, or condition.
Proper shipping orientation
226 kg 500 lbs
Keep dry
Fragile
Maximum stacking height (palletted units)
Type B, Applied Part (Upholstery)
Fuse rating specification
Protective earth ground
Dangerous voltage / shock hazard
Patient Weight Limit
Refer to manual
4
© Midmark Corporation 2012
Warnings Warning
No modification of this equipment is allowed.
Warning
Connecting equipment to the multiple socket outlet effectively leads to creating a Medical Electrical System and the result can be a reduced level of safety.
Warning
Table must be positioned in normal use such that the power cords can be easily accessed to unplug.
Warning
To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
Warning
Equipment is not suitable for use in the presence of a flammable anesthetic mixture. Clarification: Equipment is suitable for use in the presence of oxygen, air, or nitrous oxide.
Intended Use This product is intended to be used as a table to provide positioning and support of patients during examination procedures conducted by qualified medical professionals in an office environment.
Electrical Requirements Caution
To completely isolate the table from electrical mains supply, both power cords must be unplugged. When using high frequency surgical devices or endocardial catheters: • Use non-conductive material to insulate patient from metal portions of the table. • Consult operating instructions for the device before using in conjunction with drawer or upholstery heater. Failure to comply may result in electric shock or burns to patient.
5
© Midmark Corporation 2012
Electromagnetic Interference This product is designed and built to minimize electromagnetic interference with other devices. However, if interference is noticed between another device and this product: • • • •
Remove interfering device from room. Plug chair into an isolated circuit. Increase separation between chair and interfering device. Contact Midmark if interference persists.
Disposal of Equipment At the end of product life, the table, accessories, and other consumable goods may become contaminated from normal use. Consult local codes and ordinances for proper disposal of equipment, accessories and other consumable goods.
Authorized CE Representative Countries in the EEC should direct all questions,incidents, and complaints to Midmark’s Authorized CE representative listed here:
Midmark Europe SAS 22, Rue De St Denis 53500 Ernee, FRANCE Phone: 33-2-430-517-76 Fax: 33-2-430-569-10
Transportation / Storage Conditions Ambient Temperature Range:...-5°C to +38°C (+23°F to 100°F) Relative Humidity...10% to 90% (non-condensing) Atmospheric Pressure...500hPa to 1060hPa (0.49atm to 1.05atm)
6
© Midmark Corporation 2012
Operation Back Section Positioning To position the back section...
A) While holding the back section in place, squeeze the Back Release Handle (either side). B) Move the back section to desired position. C) Release handle to lock back section in position.
Back Release Handle MA696701i
Max. 70° ±5°
Min. 0° ±1°
MA697500i
7
© Midmark Corporation 2012
Stirrups Caution
Be sure the stirrups are locked in place before using. The stirrups will not support the patient’s entire weight.
To position the stirrups...
A) Pull the stirrup out, then unfold. B) Lift the stirrup slightly, then move it left or right as desired. C) Release stirrup to lock in position.
MA696300i
8
© Midmark Corporation 2012
Footrest & Treatment Pan Caution
Do not use footrest to support patient’s entire weight.
To access the treatment pan...
Lift footrest pad slightly, then slide pad back into table.
To extend footrest...
Pull footrest out to desired position.
Footstep To access the footstep...
Lift footstep slightly, then pull outward.
MA696600i
9
© Midmark Corporation 2012
Paper Roll & Tear Strap Note
Paper Roll Size (max.): Seamless Upholstery...21 in. wide x 3.5 in. diameter (53 cm x 9 cm) Soft Touch Upholstery...21 in. wide x 4.75 in. diameter (53 cm x 12 cm)
To install the paper roll...
Insert rod into sockets under head end of upholstery.
Soft Touch upholstery shown
To install the paper tear strap...
Attach strap to snaps on each side of table.
MA696900i
10
© Midmark Corporation 2012
Drawer Heater (optional) Note
The drawer heater switch illuminates when heater is ON (I). The drawer heater warms the contents of the drawer to approximately 98°F (37°C).
To turn the drawer heater ON/OFF...
Press the heater switch to desired position.
ON
OFF
I O
MA696500i
11
© Midmark Corporation 2012
Pelvic Tilt (optional) To elevate seat section...
Lift seat until pelvic support drops into place.
To return seat section to flat position... A) Lift seat section slightly. B) Pivot trigger handle upward. C) Lower seat section.
For clarity, the upholstered top is not shown.
Pelvic Support
Trigger Handle
Max. 7° ±2° Min. 1.5°
MA697500i
12
© Midmark Corporation 2012
Receptacle (optional) The duplex receptacle on the side of the table, provides power for accessories used during medical procedures. There are two circuit breakers located under the seat section. If the receptacle’s maximum load is exceeded, the circuit breakers interrupt power to the receptacle. Maximum Load... 5 amps
To reset circuit breaker(s)...
Push the circuit breaker toggle switch.
Circuit Breakers 5 A M P
Tripped (no power)
5 A M P
Reset (power)
MA697200i
13
© Midmark Corporation 2012
Maintenance Cleaning Upholstery
Equipment Alert
The upholstery is resistant to most medicinal-type stains, but may be damaged by solvents and dyes. Immediately remove any fluids spilled on the upholstery.
Wash your upholstery weekly with a mild liquid soap and water mixture, rinse with clear water and dry completely to remove disinfection cleaner build-up. Disinfect your upholstery with a solution of standard bleach and water mixed 1 in 10 (10%) or chlorine based cleaners. Follow this with a clear water rinse and thorough drying of material. See current CDC Guideline for Disinfection & Sterilization in Healthcare Facilities. To minimize damage caused by disinfectant cleaner residue build-up, do not allow disinfectants to pool on the upholstery surface. Once the approved contact time has been obtained, remove and dry excess liquid remaining on the surface. Detailed care and maintain instructions are available in your user guide located on the CD that came with your table or by referring to Upholstery Care and Maintenance. Painted Metal / Plastic Surfaces Clean the painted metal and plastic surfaces weekly using a clean soft cloth, and mild cleaner. Periodic applications of common furniture wax will ease cleaning, and maintain the finished luster of the surfaces.
Preventative Maintenance Periodically inspect the following areas: • Power cord should be free of cuts or other visible damage. • All fasteners should be present and tightened securely. • All mechanical functions should operate properly. Have a qualified service technician inspect your table every six months.
Calling For Service Note
Model / Serial Number information is required when calling for service. If service is required, contact your authorized Midmark dealer. To contact Midmark directly: 1-800-MIDMARK (1-800-643-6275) 8:00 am until 5:00 pm Monday - Friday (EST) [excluding standard U.S. holidays]
14
© Midmark Corporation 2012
Specifications Weights & Dimensions Maximum Patient Weight:
500 lbs (226 kg)
Paper Roll (maximum size):
[See Paper Roll & Tear Strap page]
Range of Motion & Dimensions:
(See illustration below)
Weight of Table:
220 lbs (99.8 kg)
Power Cord:
6 ft. (1.8 m) long
Max. 70° ±5°
20.5 in. (52 cm)
Max. 7° ±2° Min. 1.5°
Min. 0°
30.2 in. (76.7 cm)
46.4 in. (117.8 cm)
15
© Midmark Corporation 2012
Specifications (continued) Upholstery Dimensions Soft Touch Upholstery 56.25 in. (142.9 cm)
30.5 in. (77.5 cm)
Up to: 17.5 in. (44.6 cm)
28.5 in. (72.4 cm)
74 in. (188 cm)
Seamless Upholstery 59.5 in. (151.1 cm)
30 in. (76.2 cm)
Up to: 16.25 in. (41.3 cm)
28.5 in. (72.4 cm)
76 in. (193 cm)
Soft Touch: 8.5 in. (21.6 cm) Seamless: 10 in (25.4 cm)
9 in. (23 cm) 16.5 in. (42 cm)
46.4 in. (117.8 cm)
6 ft. (1.8 m)
16
© Midmark Corporation 2012
Description
Ritter Exam Table (Pass-thru Drawers) w/ Receptacle
Ritter Exam Table (Pass-thru Drawers) w/ Receptacle, Pelvic Tilt, & Drawer Heater
Ritter Exam Table (Bulk Storage) w/ Receptacle
Ritter Exam Table (Bulk Storage) w/ Receptacle, Pelvic Tilt, & Drawer Heater
Ritter Exam Table - Export (Pass-thru Drawers) w/ Pelvic Tilt
Ritter Exam Table - Export (Pass-thru Drawers) w/ Receptacle & Pelvic Tilt
Model
204-001
204-002
204-003
204-004
17
204-005
204-006
•
•
•
•
•
•
UL STD. 60601-1
•
•
•
•
•
•
ES / EN / IEC 60601-1
•
•
•
•
•
•
CAN / CSA C22.2, 601.1 / 60601-1
Complies To:
•
•
CE
230
-
115
115
115
115
VAC +/- 10%
5
-
5.5
5
5.5
5
Amps
50/60
-
50/60
50/60
50/60
50/60
Cycles (Hz)
Electrical Ratings:
Specifications (continued)
Voltage, Current, Frequency Ratings & Compliance Chart
© Midmark Corporation 2012
Warranty Information Limited Warranty SCOPE OF WARRANTY
Midmark Corporation (“Midmark”) warrants to the original purchaser its new Alternate Care products and components (except for components not warranted under “Exclusions”) manufactured by Midmark to be free from defects in material and workmanship under normal use and service. Midmark’s obligation under this warranty is limited to the repair or replacement, at Midmark’s option, of the parts or the products the defects of which are reported to Midmark within the applicable warranty period and which, upon examination by Midmark, prove to be defective.
APPLICABLE WARRANTY PERIOD
The applicable warranty period, measured from the date of delivery to the original user, shall be one (1) year for all warranted products and components.
EXCLUSIONS
This warranty does not cover and Midmark shall not be liable for the following: (1) repairs and replacements because of misuse, abuse, negligence, alteration, accident, freight damage, or tampering; (2) products which are not installed, used, and properly cleaned as required in the Midmark “Installation” and or “Installation / Operation Manual for this applicable product. (3) products considered to be of a consumable nature; (4) accessories or parts not manufactured by Midmark; (5) charges by anyone for adjustments, repairs, replacement parts, installation, or other work performed upon or in connection with such products which is not expressly authorized in writing in advance by Midmark.
EXCLUSIVE REMEDY
Midmark’s only obligation under this warranty is the repair or replacement of defective parts. Midmark shall not be liable for any direct, special, indirect, incidental, exemplary, or consequential damages or delay, including, but not limited to, damages for loss of profits or loss of use.
NO AUTHORIZATION
No person or firm is authorized to create for Midmark any other obligation or liability in connection with the products. THIS WARRANTY IS MIDMARK’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. MIDMARK MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS. SF-1487 REV. A1
18
© Midmark Corporation 2012
Notes: