Millipore Corporation
Water Purification Systems
RiOs Essential x and Elix Essential x User Manual V1.0 Dec 2011
User Manual
54 Pages
Preview
Page 1
User Manual Elix® Essential 3, 5, 10, 15 (UV) system RiOs™ Essential 5, 8, 16, 24 system
Preface Congratulations! Thank you for buying our Water Purification System. For any observation or request, please contact Millipore.
By Internet The Millipore Internet Site can be used to find addresses, telephone/fax numbers and other information. Internet Site Address: www.millipore.com www.millipore.com/techservice www.millipore.com/lab_water
Manufacturing Site Millipore SAS 67120 Molsheim France
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
1
Preface
Safety information Your water system should be operated according to the instructions in this manual. In particular, the hydraulic and electrical specifications should be followed and met. It is important to use this equipment as specified in this manual; using this equipment in a different manner may impair the safety precautions of the water system.
This ATTENTION symbol is used to refer to instructions in this manual that need to be done carefully. These symbols are used to indicate that proper safety equipment has to be used. Protective glasses and gloves must be worn. This UV RADIATION sticker is used to refer to a position on the water system Cabinet or inside of it where exposure to UV light is possible. This DANGER sticker is used to refer to a position on the water system Cabinet or inside of it that could be hazardous. This ELECTRICAL GROUND sticker is used to refer to a position on the water system Cabinet or inside where an electrical ground connection is made. This ELECTRICAL DANGER sticker is used to refer to a position on the water system Cabinet or inside where an electrical danger could exist. IMPORTANT! Your water system should be installed and operated in a clean and dry area. Please refer to the environment requirements page at the end of this manual. Your water system is not designed for domestic use. Powering off the water system for a long period of time can discharge the battery used by the time keeper: time and date can be lost. If the water system is shut off for a long period of time, use the lab closed mode or go through an installation procedure of the water system. Contact your Service Representative. Save the system’s history frequently – every 3 or 6 months.
2
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
Preface
Documentation usage The instruction manual found with your water purification system is important for a good usage and the maintenance in compliance with the manufacturer specifications.
Your water system is delivered with two documents, and these are: A user manual which describes all specifications, operations and functions of the water
system. A quick guide that describes quick access to maintenance and operating information.
It may be necessary for to reference these documents for maintenance or other procedures with the water system. It is recommended to store these documents in such way that they can be quickly located and where they will not be damaged. A plastic folder provided with the water system. Use this to safety store the documents. You can stick the side of the plastic folder on the side of the water system and place the documents in there if desired.
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
3
Preface
System overview The water system and its different modules are described here.
Item Description 1 System name and type 2 Display and keypad 3 Progard® Pretreatment Pack location 4 Locking handle 5 Sanitization/cleaning port
Item Description 6 Electrical connections 7 System’s hydraulic connections 8 System ID and serial number 9 External power supply*
* For powering off the system, disconnect the power cord from the power socket. The power cord should be plugged into a wall outlet that is accessible. 4
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
Preface
Keypad usage The keypad and its usage are explained below.
Item Function 1 MODE key This button is used to switch between different system modes: Standby mode, Ready mode or in Configuration mode. From Standby, press twice to go to Ready. From Ready, press once to go to Standby. 2 LEFT or RIGHT key These buttons are used to navigate in the user interface software. They are also used to change configuration or set point values. 3 VALIDATE key This button is used to confirm a change or an action required by the water system. It is also used to save any changes made in configuration. 4 HELP INDICATOR icon There are up to 3 of these icons on the display depending upon the software mode. These indicators are here to tell the user which keypad button is available and active. See the examples below: 3 help indicators: Press on (Mode key) to switch operating mode. Press on to navigate in the software branch. Press on (Validate key) to configure ECO mode. 2 help indicators: Press on (Mode key) to switch operating mode. Press on to navigate in the software branch. 1 help indicator: Press on (Mode key) to switch operating mode.
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
5
Table of contents
Table of contents
Items needed...11 Items checklist ... 12 Millitrack® Software - option... 13 System ...15 Display icons... 16 Software modes... 17 Operating principle and status... 18 Maintenance information ... 19 Configure system parameters ... 20 View operation parameters ... 22 Software navigation map ... 23 Consumables replacement ...25 Replacing Progard Pack... 26 Sanitization and cleaning ...29 Sanitizing RO membranes with chlorine ... 30 Cleaning RO membranes with acid/base (pH) ... 31 Other functions ...33 Saving display energy - ECO mode ... 34 System is not used for a long time - Lab Closed ... 35 Error messages ...37 Error codes ... 38 Basic troubleshooting ... 39 Chemical and technical specifications...41 Specifications and requirements ... 42 Legal information ...45 Notice ... 46 Copyright ... 46 Trademarks... 46 Product warranty and limitation of liability ... 47 Recycling ... 47
8
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
Table of contents
Item catalog numbers... 49 Consumables catalog numbers ... 50 Optional consumables catalog numbers ... 50 Optional equipment catalog numbers... 51 System’s catalog numbers... 51
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
9
Items needed
Items needed
Items checklist Verify that all items are available before beginning the installation. See the “Content list” for more details. Water system unit (A) User manual and Quick Reference Guide (B) External power supply (C) Progard Pack (D) Accessories bag with: Tubings Fittings Power cord Elbow tubing guides
12
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
Items needed
Millitrack® Software - option Millitrack Software is an embedded GUI (graphical user interface) developed with AJAX (Asynchronous JavaScript and XML) techniques that generate interactive web pages through a point-to-point computer or direct network connection using TCP/IP Ethernet protocol. Millitrack Software allows a lab user to download or view instantaneously quality data through their internet browser for rapid monitoring of his water purification system. Millitrack Software users have also real–time remote access to interactive web pages through standard TCP/IP network describing water purification system status, configuration, instant quality parameters, settings, identification and Dashboard. Key Benefits: Native Web graphical user interface. Real-time and remote Dashboard of water systems. XML data format available for print or download to any spreadsheet database. Compatible with any lab data management system such as LIMS, ELN, SDMS/ECM.
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
13
System
System
Display icons Know how to interpret each icon displayed. Display with all symbols is represented below.
Item Description Item Description 3 Values area is used to display 1 System’s status icons are Water quality measures. Maintenance or tool: Software menu and the system is in Standby or configuration values. maintenance can be started. Error code numbers. Configuration or gears: Units are configuration parameters such 4 ‘ or minutes. as set points can be changed. MΩ.cm or megOhm centimeter. Process or hour glass: μS/cm or micro siemens by the system is busy. No action centimetre. can be done. °C or temperature degree 2 Warning maintenance or alarm (from Celsius. left to right) are displayed to indicate % or percentage. that: Consumables status icons are 5 CL2 sanitization is required: used to indicate that a perform the sanitization of the maintenance action is RO cartridge. required. For more information, refer to System sanitization is required. the Quick Guide or the Product water resistivity is Consumables Replacement below set point. section in this document. Low feed water pressure is Tank water level is represented 6 detected. by 10 bars. Each bar represents 10% of the total volume of Water is detected (if water water in the tank. sensor installed). The outline is blinking when Contact Millipore. the tank is empty. 16
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
System
Software modes The different software modes are described below with some examples. System is in Standby or maintenance. Tank is 70%. Tool icon indicates maintenance. System is in Standby or maintenance. Tool icon indicates maintenance. Values area indicates Cl2 sanitization start. Gears icon indicates parameters. Values area indicates ECO mode setting. System is in Ready or in Tank filling or in Tank full. Tank is 90%. Product water resistivity is 15 MΩ.cm. Tank is 70%. Product water resistivity is 15 MΩ.cm. The Progard pack must be changed in 15 days: order a new
Progard pack. Yellow color display and
: the Progard pack must be changed.
Product water resistivity is 15 MΩ.cm. Yellow color display and
: the CL2 cleaning must be started.
Product water resistivity is 15 MΩ.cm. Yellow color display, Er:20 and
. The system has detected a minor problem that does not affect the normal operation. Refer to section Error codes at page 38. Red color display and . The system has detected a dysfunction when operating in normal conditions. In this example, system has detected a low feed water pressure and has stopped. More investigation is required. Red color display, Er:02 and . The system has detected a dysfunction when operating in normal conditions. Refer to section Error codes at page 38 or contact Millipore. Red color display and
: the Progard pack has been removed.
Tool icon indicates maintenance.
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
17
System
Operating principle and status The RiOs Essential is a water system designed to produce type III water. The Elix Essential is a water system designed to produce type II water. This water can be stored in a reservoir, if installed. As a summary, the Elix Essential purifies tap water using the Progard Pack and Reverse Osmosis followed by Electrical Deionization ("EDI"). There is no EDI in a RiOs Essential system. Understanding the terminology of operating status. TANK FILLING
The tank is being filled with EDI or RO water. A sensor starts and stops this mode.
FLUSH
A periodic flush of the RO portion of the system is done. This refreshes the water in this part of the system if they are not being used to produce water.
RINSING
A sensor has determined that the RO portion needs to rinse itself before sending water to the EDI device.
SANITIZATION
There are different types of sanitization that can be done. See the maintenance section in this document for more information.
CLEANING
An acid or base cleaning can be done. A packaged chemical is available for this type of cleaning. See the Ordering information section at the end of this document.
18
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
System
Maintenance information Some information can be shown on the display. CL2 In this view, the chlorine sanitization can be performed. PH In this view, the acid or base cleaning can be performed. SAN In this view, the system’s sanitization can be performed by a qualified service representative. LF 02 In this view, the UV Lamp 254 nm lifetime can be reset once it has been changed by a qualified service representative. LF 03 In this view, the UV Lamp ASM lifetime –if installed and activated, can be reset once it has been changed by a qualified service representative. LAb C In this menu, the Lab Closed mode can be activated.
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
19
System
Configure system parameters Access and configure all system parameters. SP 01 Set point for the tank level where the water system goes in Tank Filling. SP 02 Set point for the RO % Rejection. Below this set point, an alarm is indicated. This measurement indicates the percent of ions removed by the RO relative to the amount of ions entering it. SP 03 Set point for the resistivity or conductivity for product water. Below this set point, an alarm is indicated. DATE Menu where date and time can be adjusted. IP AD – used with Millitrack software Menu where the actual IP address can be viewed. IP DF - used with Millitrack software Menu where the actual IP address can be reset and switched to the default one. ECO Menu where the display energy saving mode can be activated or deactivated. ASM Menu where Automatic Sanitization Module can be activated. C--Restricted menu accessible with a code and used for Service. 20
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
System
An example of a configuration change: the DATE (YYYY, MM, DD) and TIME (00:00).
1- System is in STANDBY mode. Press on right . arrow
4- Press menu.
to enter dAtE
7- Change day with arrows and press .
9- Set the time. Press
.
2- In maintenance menu, press .
3- Press on right arrow until the date menu is displayed.
5- Change year with arrows and press .
6- Change month with and press arrows .
8- The date is set now.
10- Change with arrows and press .
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system
11- Change with arrows and press .
21
System
View operation parameters In Ready or Tank filling, press and hold 2 sec on left or right arrow operation parameters.
to access
Tap water feed conductivity value can be displayed by pressing . RO feed water conductivity value can be displayed by pressing . RO water temperature value can be displayed by pressing . RO pump pressure value can be displayed by pressing . RO % rejection value can be displayed by pressing . On Elix Essential: RO permeate water conductivity value can be displayed by pressing .
Note: the product water quality and temperature are displayed on the main display view. On RiOs Essential, the RO permeate water conductivity is shown on the main display:
On Elix Essential, the resistivity or conductivity is shown on the main display:
Press arrow to view temperature.
Press arrow to view temperature.
22
Elix® Essential 3, 5, 10, 15 (UV) system / RiOs™ Essential 5, 8, 16, 24 system