MINDRAY
DP6600 Operation Manual [Acoustic Power Data] Ver 1.1
Operation Manual
27 Pages
Preview
Page 1
DP-6600
Digital Ultrasonic Diagnostic Imaging System
Operation Manual [Acoustic Power Data]
© 2005-2006 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
Product Information: Product Name: Digital Ultrasonic Diagnostic Imaging System Model: DP-6600 Issued Date of this manual: 2006-05 Manual Version: 1.1
Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties. Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner without the written permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution, rent, adaptation and translation of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.
IMPORTANT! 1. 2.
No part of this manual may be copied or reprinted, in whole or in part, without written permission. The contents of this manual are subject to change without prior notice and without our legal obligation.
I-1
Introduction
This manual gives all the transducers Acoustic Output Power data for DP-6600 Digital Ultrasonic Diagnostic Imaging System. Please refer to correlative tables in use.
I-2
Contents
HOW TO READ THE ACOUSTIC POWER TABLES ... 1 ACOUSTIC POWER TABLES ... 2 35C50EA ... 2 35C20EA ... 5 65EC10EA ... 8 65C15EA ... 11 75L38EA ... 14 75L60EA ... 17
C-1
How to Read the ACOUSTIC POWER TABLES
How to Read the ACOUSTIC POWER TABLES Description of symbols used in acoustic output tables
ISPTA.3
Derated spatial-peak temporal-average intensity (milliwatts per square centimeter).
ISPPA.3
Derated spatial-peak pulse average intensity (watts per square centimeter).
MI
Mechanical Index. The value of MI at the position of ISPPA.3.
Pr.3
Derated peak rarefractional pressure (megapascals) associated with the transmit pattern giving rise to the value reported under MI.
W0
Ultrasonic power (milliwatts)
fc
Center frequency (MHz)
Zsp
The axial distance at which the reported parameter is measured (centimeters)
X-6,Y-6
In-plane (azimuthal) and out-plane (elevational) -6dB dimensions in the x-y plane where Zsp is found (centimeters)
PD
Pulse duration (microseconds)
PRF
Pulse repetition frequency (Hz)
EBD
Entrance beam dimensions for the azimuthal and elevational planes (centimeters)
EDS
Entrance dimensions of the scan for the azimuthal and elevational planes (centimeters)
NOTE: About the detail of acoustic power, please refer to the regulation of ALARA in operation manual (Basic Volume).
1
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (NON-AUTO-SCANNING MODE) Transducer Model: 35C50EA Operating Mode: M-Mode Application(s): Fetal Imaging & Other, Cardiac, Adult & Pediatric
MI
Acoustic Output
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) EBD
Operating Control Conditions
Control 1
ISPPA.3
(mW/cm )
(W/cm2)
20.0
40.9
2
Global Maximum Value Associated Acoustic Parameter
ISPTA.3
0.595 1.12
4.28
4.28
3.76
3.76
3.76
1.83
1.83
1.83
0.222
0.222
0.902
0.902
x-6 (cm) y-6 (cm)
0.481
0.481
998.3
998.3
Az. (cm) Ele. (cm)
0.30 1.18
Control 4
Max Power One Focus Focus #4 Depth=4.31cm
Control 5
Working frequency=3.5MHz
Control 2 Control 3
2
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (AUTO-SCANNING MODE) Transducer Model: 35C50EA Operating Mode: B-Mode, B/B-Mode Application(s): Fetal Imaging & Other, Cardiac, Adult & Pediatric
MI
Acoustic Output
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) EDS
Operating Control Conditions
Control 1
ISPPA.3
(mW/cm )
(W/cm2)
22.6
40.9
34.7
34.7
3.76
3.76
2
Global Maximum Value Associated Acoustic Parameter
ISPTA.3
0.595 1.12
3.76
1.83
1.83 x-6 (cm) y-6 (cm)
0.222 0.902
Az. (cm) Ele. (cm)
0.481
0.481
10000
10000 0.30 1.18
Control 4
Max Power Angle #3 High density line One Focus, Focus #4
Control 5
Depth=4.31cm, Working frequency=3.5MHz
Control 2 Control 3
3
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (COMBINED MODE) Transducer Model: 35C50EA Operating Mode: B/M-Mode Application(s): Fetal Imaging & Other, Cardiac, Adult & Pediatric
MI
Acoustic Output
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) ESN
Operating Control Conditions
Control 1
ISPPA.3
(mW/cm )
(W/cm2)
20.1
40.9
38.9
38.9
3.76
3.76
2
Global Maximum Value Associated Acoustic Parameter
ISPTA.3
0.595 1.12
3.76
1.83
1.83 x-6 (cm) y-6 (cm)
0.222 0.902
Az. (cm) Ele. (cm)
0.481
0.481
10000
10000 0.30 1.18
Control 4
Max Power Angle #3 High density line One Focus, Focus #4
Control 5
Depth=4.31cm, Working frequency=3.5MHz
Control 2 Control 3
4
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (NON-AUTO-SCANNING MODE) Transducer Model: 35C20EA Operating Mode: M-Mode Application(s): FETAL IMAGING AND OTHER*, CARDIAC ADULT AND PEDIATRIC
Acoustic Output
MI
Global Maximum Value Pr.3 (MPa)
0.604 0.918
ISPTA3
ISPPA3
(mW/cm2)
(W/cm2)
32.9
38.7
8.10
8.10
Wo
(mW)
fc
(MHz)
2.65
2.65
2.65
Zsp
(cm)
2.62
2.62
2.62
0.808 1.010
0.808 1.010 0.678 935
Beam X-6 Dimensions Y-6 PD (µS) PRF (Hz) EBD Az. (cm) Ele. (cm) Operating Control Max Power Conditions
0.678 935
Focus #11
5
1.700 0.917 Depth=12.94cm
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (AUTO-SCANNING MODE) Transducer Model: 35C20EA Operating Mode: B, B/B Mode Application(s): FETAL IMAGING AND OTHER*, CARDIAC ADULT AND PEDIATRIC
Acoustic Output
MI
Global Maximum Value Pr.3 (MPa)
0.587 0.897
ISPTA3
ISPPA3
(mW/cm2)
(W/cm2)
50.4
38.1
4.16
4.16
2.63
2.63
Wo
(mW)
fc
(MHz)
2.63
Zsp
(cm)
2.58
2.58
0.678 4651
0.791 1.020 0.678 4651
Beam X-6 Dimensions Y-6 PD (µS) PRF (Hz) EDS Az. (cm) Ele. (cm) Operating Control Angle 3 Conditions
High density line
6
2.86 0.92 Max power
Focus #11 Depth=12.94cm
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (COMBINED MODE) Transducer Model: 35C20EA Operating Mode: BM-Mode Application(s): FETAL IMAGING AND OTHER*, CARDIAC ADULT AND PEDIATRIC
Acoustic Output
MI
Global Maximum Value Pr.3 (MPa)
0.587 0.897
ISPTA3
ISPPA3
(mW/cm2)
(W/cm2)
39.7
38.1
4.16
4.16
2.63
2.63
Wo
(mW)
fc
(MHz)
2.63
Zsp
(cm)
2.58
2.58
0.678 4651
0.791 1.02 0.678 4651
Beam X-6 Dimensions Y-6 PD (µS) PRF (Hz) EDS Az. (cm) Ele. (cm) Operating Control Angle 3 Conditions
High density line
7
2.800 0.823 Max power
Focus #11 Depth=12.94cm
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (NON-AUTO-SCANNING MODE) Transducer Model: 65EC10EA Operating Mode: M-Mode Application(s): Fetal Imaging & Other, Cardiac, Adult & Pediatric
MI
Acoustic Output Global Maximum Value Associated Acoustic Parameter
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) EBD
Operating Control Conditions
Control 1
ISPTA.3
ISPPA.3
(mW/cm2)
(W/cm2)
31.3
112
1.61
1.61
5.87
5.87
5.87
1.56
1.56
1.56
0.129
0.129
0.272
0.272
0.713 1.77
x-6 (cm) y-6 (cm)
0.285
0.285
998
998
Az. (cm) Ele. (cm)
0.27 0.42
Control 4
Max Power One Focus Focus 4# Depth=2.16cm
Control 5
Working frequency=5.0MHz
Control 2 Control 3
8
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (AUTO-SCANNING MODE) Transducer Model: 65EC10EA Operating Mode: B-Mode, B/B-Mode Application(s): Fetal Imaging & Other, Cardiac, Adult & Pediatric
MI
Acoustic Output Global Maximum Value Associated Acoustic Parameter
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) EDS
Operating Control Conditions
Control 1
ISPTA.3
ISPPA.3
(mW/cm2)
(W/cm2)
14.7
118
5.81
5.81
5.88
5.88
0.713 1.74
5.88
1.59
1.59 x-6 (cm) y-6 (cm)
0.129 0.272
Az. (cm) Ele. (cm)
0.285
0.285
10000
10000 0.27 0.42
Control 4
Max Power Angle 3 High density line One Focus, Focus #4
Control 5
Depth=2.16cm, Working frequency=5.0MHz
Control 2 Control 3
9
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (COMBINED MODE) Transducer Model: 65EC10EA Operating Mode: B/M-Mode Application(s): Fetal Imaging & Other, Cardiac, Adult & Pediatric
MI
Acoustic Output Global Maximum Value Associated Acoustic Parameter
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) ESN
Operating Control Conditions
Control 1
ISPTA.3
ISPPA.3
(mW/cm2)
(W/cm2)
32.6
118
5.81
5.81
5.88
5.88
0.713 1.77
5.88
1.59
1.59 x-6 (cm) y-6 (cm)
0.129 0.272 0.285
0.285
10000
10000
Az. (cm) Ele. (cm)
0.27 0.42
Control 4
Max Power Angle 3 High density line One Focus, Focus #4
Control 5
Depth=2.16cm, Working frequency=5.0MHz
Control 2 Control 3
10
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (NON-AUTO-SCANNING MODE) Transducer Model: 65C15EA Operating Mode: M-Mode Application(s): Fetal Imaging &Other*
MI
Acoustic Output Global Maximum Value Associated Acoustic Parameter
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) EBD
Operating Control Conditions
Control 1
0.884 2.17
ISPTA.3
ISPPA.3
(mW/cm2)
(W/cm2)
63.6
245
3.46
3.46
5.90
5.90
5.90
2.22
2.22
2.22
0.171
0.171
0.201
0.201
x-6 (cm) y-6 (cm)
0.288
0.288
998.3
998.3
Az. (cm) Ele. (cm)
0.25 0.37
Control 4
Max Power One Focus Focus #4 Depth=2.16cm
Control 5
Working frequency=5MHz
Control 2 Control 3
11
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (AUTO-SCANNING MODE) Transducer Model: 65C15EA Operating Mode: B-Mode, B/B-Mode Application(s): Fetal Imaging &Other*
MI
Acoustic Output Global Maximum Value Associated Acoustic Parameter
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) EDS
Operating Control Conditions
Control 1
ISPTA.3
ISPPA.3
(mW/cm2)
(W/cm2)
25.3
245
13.7
13.7
5.90
5.90
0.884 2.17
5.90
2.22
2.22 x-6 (cm) y-6 (cm)
0.171 0.201
Az. (cm) Ele. (cm)
0.288
0.288
10000
10000 0.25 0.37
Control 4
Max Power Angle #3 High density line One Focus, Focus #4
Control 5
Depth=2.16cm, Working frequency=5MHz
Control 2 Control 3
12
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (COMBINED MODE) Transducer Model: 65C15EA Operating Mode: B/M-Mode Application(s): Fetal Imaging &Other*
MI
Acoustic Output Global Maximum Value Associated Acoustic Parameter
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) ESN
Operating Control Conditions
Control 1
ISPTA.3
ISPPA.3
(mW/cm2)
(W/cm2)
63.7
245
17.1
17.1
5.90
5.90
0.884 2.17
5.90
2.22
2.22 x-6 (cm) y-6 (cm)
0.171 0.201
Az. (cm) Ele. (cm)
0.288
0.288
10000
10000 0.25 0.37
Control 4
Max Power Angle #3 High density line One Focus, Focus #4
Control 5
Depth=2.16cm, Working frequency=5MHz
Control 2 Control 3
13
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (NON-AUTO-SCANNING MODE) Transducer Model: 75L38EA Operating Mode: M-Mode Application(s): Fetal Imaging & Other, Cardiac, Adult & Pediatric, Peripheral Vessel
MI
Acoustic Output Global Maximum Value Associated Acoustic Parameter
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) EBD
Operating Control Conditions
Control 1
ISPTA.3
ISPPA.3
(mW/cm2)
(W/cm2)
75.1
85.3
2.62
2.62
5.97
5.97
5.97
2.58
2.58
2.58
0.296
0.296
0.127
0.127
0.67 1.66
x-6 (cm) y-6 (cm)
0.310
0.310
953.7
953.7
Az. (cm) Ele. (cm)
0.70 0.20
Control 4
Max Power One Focus The 12th focus Depth=6.47cm
Control 5
Working frequency=5.0MHz
Control 2 Control 3
14
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (AUTO-SCANNING MODE) Transducer Model: 75L38EA Operating Mode: B-Mode, B/B-Mode Application(s): Fetal Imaging & Other, Cardiac, Adult & Pediatric, Peripheral Vessel
MI
Acoustic Output Global Maximum Value Associated Acoustic Parameter
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) EDS
Operating Control Conditions
Control 1
0.67 1.66
5.97
ISPTA.3
ISPPA.3
(mW/cm2)
(W/cm2)
21.4
85.3
30.1
30.1
5.97
5.97 2.58
2.58 x-6 (cm) y-6 (cm)
0.296 0.127 0.310
0.310
8591
8591
Az. (cm) Ele. (cm)
0.70 0.20
Control 4
Max Power Angle #3 High density line The 12th focus
Control 5
Depth=6.47cm, Working frequency=5.0MHz
Control 2 Control 3
15
ACOUSTIC POWER TABLES
ACOUSTIC POWER TABLES ACOUSTIC OUTPUT REPORTING TABLE FOR TRACK 1 (COMBINED MODE) Transducer Model: 75L38EA Operating Mode: B/M-Mode Application(s): Fetal Imaging & Other, Cardiac, Adult & Pediatric, Peripheral Vessel
MI
Acoustic Output Global Maximum Value Associated Acoustic Parameter
pr.3 (MPa) Wo (mW) fc (MHz) zsp (cm) Beam dimensions
PD (µsec) PRF (Hz) ESN
Operating Control Conditions
Control 1
0.67 1.66
5.97
ISPTA.3
ISPPA.3
(mW/cm2)
(W/cm2)
24.6
85.3
32.4
32.4
5.97
5.97 2.58
2.58 x-6 (cm) y-6 (cm)
0.296 0.127 0.310
0.310
8591
8591
Az. (cm) Ele. (cm)
0.70 0.20
Control 4
Max Power Angle #3 High density line The 12th focus
Control 5
Depth=6.47cm, Working frequency=5.0MHz
Control 2 Control 3
16