M5 Diagnostic Ultrasound System
Operator’s Manual [Acoustic Power Data and Surface Temperature Data]
© 2007 Shenzhen Mindray Bio-medical Electronics Co., Ltd.
All rights Reserved.
Introduction
This manual gives all the transducers Acoustic Output Power data and Surface Temperature Data for M5 Diagnostic Ultrasound System. Please refer to correlative tables in use.
IMPORTANT! 1. No part of this manual may be copied or reprinted, in whole or in part, without written permission. 2. The contents of this manual are subject to change without prior notice and without our legal obligation.
I-1
Contents
How to Read the Acoustic Power Tables ...1 Transducer Maximum Surface Temperature...2 Acoustic Output Reporting Tables ...3 3C5s ...3 7L4s...10 6CV1s...17 10L4s...24 7L6s...31 6C2s ...38 6LE7s ...45 6LB7s ...52
C-1
How to Read the Acoustic Power Tables
How to Read the Acoustic Power Tables Description of symbols used in acoustic output tables Symbol
Description
α
Acoustic attenuation coefficient
Aaprt
-12 dB output beam area
Deq
Equivalent aperture diameter
deq
Equivalent beam diameter
fawf
Acoustic working frequency (MHz)
Ipa
Pulse average intensity
Ipa, α
Attenuated pulse average intensity
Ipi
Pulse intensity integral
Ipi, α
Attenuated pulse intensity integral
Ita(z)
Temporal average intensity
Ita, α(z)
Attenuated temporal average intensity
Izpta (z)
Spatial-peak temporal-average intensity
Izpta, α(z)
Attenuated Spatial-peak temporal-average intensity
MI
Mechanical Index
P
Output power
Pα
Attenuated power
P1
Bonded output power
pi
Pulse pressure squared integral
pr
Peak rarefactional acoustic pressure
pr.a
Attenuated Peak rarefactional acoustic pressure
prr
Pulse repetition frequency (Hz)
TI
Thermal index
TIS
Soft tissue thermal index
TIB
Bone thermal index
TIC
Cranial-bone thermal index
Zb
Depth for TIB
Zbp
Break point depth
Zs
Depth for TIS
NOTE: About the detail of acoustic power, please refer to the regulation of ALARA in operator’s manual (Basic Volume).
1
Transducer Maximum Surface Temperature
Transducer Maximum Surface Temperature According to the requirements of the section 42.3 in the standard IEC60601-2-37:2005, the transducer surface temperature has been tested in two kinds of conditions: the transducer suspended in still air or transducer contacting human-tissue mimicking material. The calculation of the expanded uncertainty is based on the ISO Guide to the Expression of uncertainty in measurement. Three transducer samples have been tested and the confidence coefficient is at 95%, the value of t.975 is 4.30.
* The measurement data were obtained under the test conditions employed at Mindray. Transducer model
Maximum surface temperature (°C)
Maximum surface temperature (°C)*1
Contacting TMM
Suspending in air
3C5s
40.57±0.68
36.33±2.09
6C2s
40.13±2.16
36.10±4.61
7L4s
41.80±0.57
37.43±1.04
7L6s
37.73±0.68
30.77±0.68
10L4s
40.87±0.40
32.90±1.34
6CV1s
42.07±0.81
31.87±2.03
6LE7s
41.27±0.59
31.17±0.68
6LB7s(Convex)
41.47±0.40
29.13±0.81
6LB7s(Linear)
41.27±0.59
31.17±0.68
*1
*1: Values following the "±" mark indicate the expanded uncertainty with a confidence level of 95%, t.975 = 4.30.
2
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: 3C5s Imaging Mode: M-mode TIS Index Label
MI
Maximum Index Value
Scan
1.08E+00 1.81E+00
pr.a
TIB
Non-scan
P
Nonscan
TIC
7.26E-01
1.58E+00
1.00E+00
5.41E+01
5.41E+01
5.41E+01
Aaprt≤1cm
2
Aaprt> 2 1cm
Min of [Pα(zs), Ita,α(zs)] zs Associate Acoustic Parameter
zbp zb
4.67E+00 4.75E+00
z at max.Ipi,α
X
2.83E+00 5.91E-01
3.12E-01 2.83E+00 5.91E-01
Y
7.76E-01
7.76E-01
deq(zb) fawf
2.83E+00
Dim Aaprt
of
td Other Information
prr pr at max. Ipi
Operating Control Conditions
7.76E-01
5.70E-07 2.00E+03 2.59E+00 3.03E-01
deq at max. Ipi Ipa,α at max. MI
2.83E+00 5.91E-01
2.53E+02
Control 1
Focal position: at the 8.0cm of B-mode
Control 2
Depth: 9.90cm
Control 3
Working Frequency: H6.0MHz
Control 4
PRF: 2000Hz
Control 5
Highest transmission power (100%)
Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
3
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: Imaging Mode:
3C5s B-mode TIS
Index Label
1.08E+00 1.81E+00
pr.a P Min of Ita,α(zs)] zs Associate Acoustic Parameter
Non-scan
MI
Maximum Index Value
TIB
Scan
Aaprt≤1 2 cm
Aaprt> 1 2 cm
Nonscan
TIC
8.39E-01
1.26E+00
6.04E+02
6.04E+02
2.83E+00 5.91E-01
2.83E+00 5.91E-01
7.76E-01
7.76E-01
[Pα(zs),
zbp zb z at max.Ipi,α
4.75E+00
deq(zb)
2.83E+00
fawf Dim of Aaprt
X Y
td Other Information
prr pr at max. Ipi deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
5.70E-07 7.83E+03 2.59E+00 2.53E+02
Control 1
Focal zone number: one ; Focal position: 8.0cm for B-mode
Control 2
Depth: 9.90cm
Control 3
Working Frequency: H6.0MHz
Control 4
PRF: 3900Hz
Control 5
Highest transmission power (100%)
Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
4
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: Imaging Mode:
3C5s B+M-mode TIS
Index Label
1.08E+00 1.81E+00
pr.a P Min of Ita,α(zs)] zs Associate Acoustic Parameter
Non-scan
MI
Maximum Index Value
TIB Nonscan
TIC
7.26E-01
1.58E+00
1.55E+00
5.41E+01
5.41E+01
3.22E+02
Scan
Aaprt≤1 2 cm
3.73E-01 2.69E+02
[Pα(zs),
zbp zb
4.67E+00
z at max.Ipi,α
4.75E+00
deq(zb)
2.83E+00
fawf Dim of Aaprt
X Y
td Other Information
prr pr at max. Ipi
2.83E+00 5.91E-01
2.83E+00 5.91E-01
3.12E-01 2.83E+00 5.91E-01
7.76E-01
7.76E-01
7.76E-01
5.70E-07 2.00E+03 2.59E+00
deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
Aaprt> 2 1 cm
3.03E-01 2.53E+02
Control 1
Focal position: at the 8.0cm of B-mode
Control 2
Depth: 9.90cm
Control 3
Working Frequency: H6.0MHz
Control 4
PRF: 7830Hz
Control 5
Highest transmission power (100%)
Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
5
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: Imaging Mode:
3C5s
PW-mode TIS
Index Label
MI
Non-scan Scan
1.33E+00 2.15E+00
Maximum Index Value pr.a P Min of Ita,α(zs)] zs Associate Acoustic Parameter
TIB Nonscan
TIC
1.10E+00
3.28E+00
1.53E+00
8.91E+01
8.91E+01
8.91E+01
Aaprt≤1 2 cm
Aaprt> 2 1 cm
[Pα(zs),
zbp zb
4.25E+00
z at max.Ipi,α
4.42E+00
deq(zb)
2.62E+00
fawf Dim Aaprt
of
td Other Information
prr pr at max. Ipi
Y
2.86E-01 2.62E+00 3.38E-01
2.62E+00 3.38E-01
7.78E-01
7.78E-01
7.78E-01
1.45E-06 1.65E+03 2.79E+00
deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
X
2.62E+00 3.38E-01
2.78E-01 2.76E+02
Control 1
Sampling Volume Position: at the 6.0cm of B-mode
Control 2
Depth: 7.40cm
Control 3
Working Frequency: 2.5MHz, SV=0.5mm.
Control 4
PRF: 1650Hz
Control 5
Highest transmission power (100%)
Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
6
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: 3C5s Imaging Mode:
B+PW-mode TIS
Index Label
Non-scan
MI
Maximum Index Value
Scan
1.15E+00 1.86E+00
pr.a P
TIB
2
Aaprt≤1 cm
TIC
Non-scan
Aaprt> 2 1 cm
3.68E-01
9.00E-01
2.76E+00
1.76E+00
2.61E+02
7.18E+01
7.18E+01
3.30E+02
Min of [Pα(zs), Ita,α(zs)] zs Associate Acoustic Parameter
zbp
4.30E+00
zb z at max.Ipi,α
4.40E+00
deq(zb) X
2.88E+00 4.69E-01
2.62E+00 3.36E-01
2.67E-01 2.62E+00 3.36E-01
Y
7.24E-01
7.55E-01
7.55E-01
fawf
2.62E+00
Dim of Aaprt td Other Information
prr pr at max. Ipi deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
Note 1 mode.
1.45E-06 1.80E+03 2.61E+00 2.60E-01 2.32E+02
Control 1
Focal zone number: one , Focal position: 6.0cm for B-mode
Control 2
Sampling Volume Position: at the 6.0cm of B-mode
Control 3
Depth: 7.40cm
Control 4
Working Frequency:H6.0MHz for B-mode, and 2.5MHz for PW-mode
Control 5
Scan Line Mode: lower-density line and minimum scan angle for B mode
Control 6
PRF: 7830Hz for B-mode, and 1650Hz for PW-mode; SV=0.5mm.
Highest transmission power (100%) Control 7 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that
7
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: 3C5s Imaging Mode:
Color+B-Mode/Power+B-Mode TIS
Index Label
1.02E+00 1.63E+00
pr.a P Min of Ita,α(zs)] zs Associate Acoustic Parameter
Non-scan
MI
Maximum Index Value
TIB
Scan
Aaprt≤1 2 cm
Aaprt>1 2 cm
Nonscan
TIC
1.36E+00
2.22E+00
3.62E+02
3.62E+02
[Pα(zs),
zbp zb
5.03E+00
z at max.Ipi,α deq(zb) fawf
2.53E+00
Dim Aaprt
of
td Other Information
prr pr at max. Ipi
Y
1.51E-06 7.80E+03 2.18E+00
deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
X
1.44E+02
Control 1
B-Mode Focal zone number: one ; Focal position: 6.0cm for B-mode
Control 2
Color Sampling Gate Position: at the 6.0cm of B-mode; Color Sampling Gate Width: minimum.
Control 3
Depth: 7.40 cm
Control 4
Working Frequency: H6.0MHz for B-mode, and 2.5MHz for Color-mode
Control 5
Scan Line Mode: minimum scan angle for B-mode, and lower-density line for B-mode and Color-mode, and color sensitivity : 3 .
Control 6
PRF: 7830Hz for B-mode, and 7800Hz for Color-mode
Control 7
Highest transmission power (100%)
Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
8
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: Imaging Mode:
3C5s
Color+B+PW-Mode/Power+B+PW-Mode TIS
Index Label
MI
Maximum Index Value
1.05E+00 1.70E+00
pr.a P Min of Ita,α(zs)] zs Associate Acoustic Parameter
TIB Non-scan
Scan
2
Aaprt≤1 cm
Aaprt> 2 1 cm
6.57E-01
2.00E+00
2.39E+00
2.51E+02
5.23E+01
5.23E+01
3.03E+02
[Pα(zs),
zbp
4.48E+00
zb
4.72E+00
z at max.Ipi,α
X
2.63E+00 3.43E-01
2.77E-01 2.63E+00 3.43E-01
Y
7.47E-01
7.47E-01
fawf
2.63E+00
Dim of Aaprt td prr pr at max. Ipi
1.45E-06 1.65E+03 2.18E+00 2.66E-01
deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
TIC
9.10E-01
deq(zb)
Other Information
Nonscan
1.81E+02
Control 1
B mode Focal zone number : one ; Focal position: 6.0cm for B-mode
Control 2
Sampling Volume Position: at the 6.0cm of B-mode; Color Sampling Gate Position: at the 7.4cm of B-mode.
Control 3
Color Sampling Gate Width : Minimum Gate.
Control 4
Depth: 7.40cm Working Frequency : H6.0MHz for B-mode, 2.5MHz for PW-mode, and 2.5MHz for Color-mode, and color sensitivity: 3 . Scan Line Mode: Minimum scan angle for B-mode, and lower-density line for B-mode and Color-mode, SV=0.5mm. PRF: 7830Hz for B-mode, 1650Hz for PW-mode, 1650Hz for Color-mode. Acoustic Power: Highest transmission power (100%).
Control 5 Control 6 Control 7
Control 8 Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
9
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: Imaging Mode:
7L4s M-mode TIS
Index Label
Non-scan
MI
Scan
1.10E+00 2.85E+00
Maximum Index Value pr.a
TIB
P
Nonscan
TIC
6.02E-02
1.98E-01
8.18E-02
1.85E+00
1.85E+00
1.85E+00
Aaprt≤1cm
2
Aaprt> 2 1cm
Min of [Pα(zs), Ita,α(zs)] zs Associate Acoustic Parameter
zbp
1.32E+00
zb
1.40E+00
z at max.Ipi,α deq(zb) fawf
6.76E+00
Dim Aaprt
of
td Other Information
prr pr at max. Ipi
Y
1.13E-01 6.76E+00 2.04E-01
6.76E+00 2.04E-01
2.52E-01
2.52E-01
2.52E-01
1.23E-07 2.00E+03 3.72E+00 1.09E-01
deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
X
6.76E+00 2.04E-01
4.30E+02 of
B-mode
Control 1
Focal position: at the 2.0cm
Control 2
Depth: 2.50cm
Control 3
Working Frequency: 5.0MHz
Control 4
PRF: 2000Hz
Control 5
Highest transmission power (100%)
Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
10
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: Imaging Mode:
7L4s B-mode TIS
Index Label
1.10E+00 2.85E+00
pr.a P Min of Ita,α(zs)] zs Associate Acoustic Parameter
Non-scan
MI
Maximum Index Value
TIB
Scan
Aaprt≤1 2 cm
Aaprt> 1 2 cm
Nonscan
TIC
1.86E-01
1.65E-01
1.42E+01
1.42E+01
6.76E+00 2.04E-01
6.76E+00 2.04E-01
2.52E-01
2.52E-01
[Pα(zs),
zbp zb z at max.Ipi,α
1.40E+00
deq(zb)
6.76E+00
fawf Dim of Aaprt
X Y
td Other Information
prr pr at max. Ipi deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
1.23E-07 1.57E+04 3.72E+00 4.30E+02
Control 1
Focal zone number: one ; Focal position: 2.0cm for B-mode
Control 2
Depth: 2.50cm
Control 3
Working Frequency: 5.0MHz
Control 4
PRF: 15680Hz
Control 5
Highest transmission power (100%)
Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
11
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: Imaging Mode:
7L4s B+M-mode TIS
Index Label
1.10E+00 2.85E+00
pr.a P Min of Ita,α(zs)] zs Associate Acoustic Parameter
Non-scan
MI
Maximum Index Value
TIB Nonscan
TIC
6.02E-02
1.98E-01
2.25E-01
1.85E+00
1.85E+00
1.42E+01
Scan
Aaprt≤1 2 cm
1.62E-01 1.24E+01
[Pα(zs),
zbp zb
1.32E+00
z at max.Ipi,α
1.40E+00
deq(zb)
6.76E+00
fawf Dim of Aaprt
X Y
td Other Information
prr pr at max. Ipi
6.76E+00 2.04E-01
6.76E+00 2.04E-01
1.13E-01 6.76E+00 2.04E-01
2.52E-01
2.52E-01
2.52E-01
1.23E-07 2.00E+03 3.72E+00
deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
Aaprt> 2 1 cm
1.09E-01 4.30E+02
Control 1
Focal position: at the 2.0cm of B-mode
Control 2
Depth: 2.50cm
Control 3
Working Frequency: 5.0MHz
Control 4
PRF: 15680Hz
Control 5
Highest transmission power (100%)
Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
12
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: Imaging Mode:
7L4s
PW-mode TIS
Index Label
MI
Non-scan Scan
1.17E+00 2.57E+00
Maximum Index Value pr.a P Min of Ita,α(zs)] zs Associate Acoustic Parameter
Nonscan
TIC
2.92E-01
1.38E+00
5.04E-01
1.30E+01
1.30E+01
1.30E+01
Aaprt≤1 2 cm
zbp zb
1.35E+00
z at max.Ipi,α
1.42E+00 4.81E+00
fawf Dim Aaprt
of
td prr pr at max. Ipi
X Y
4.81E+00 2.78E-01
1.22E-01 4.81E+00 2.78E-01
4.81E+00 2.78E-01
1.94E-01
1.94E-01
1.94E-01
8.59E-07 2.30E+03 3.16E+00
deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
Aaprt> 2 1 cm
[Pα(zs),
deq(zb)
Other Information
TIB
1.19E-01 3.12E+02
Control 1
Sampling Volume Position: at the 2.8cm of B-mode
Control 2
Depth: 3.70cm
Control 3
Working Frequency: 4.6MHz
Control 4
PRF: 2300Hz, SV=0.5mm.
Control 5
Highest transmission power (100%)
Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
13
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: 7L4s Imaging Mode:
B+PW-mode TIS
Index Label
9.51E-01 2.55E+00
pr.a P Min of Ita,α(zs)] zs Associate Acoustic Parameter
Non-scan
MI
Maximum Index Value
TIB
Scan
2
Aaprt≤1 cm
1.24E-01
1.61E-01
7.49E-01
3.69E-01
9.17E+00
7.06E+00
7.06E+00
1.55E+01
[Pα(zs),
zbp zb
1.37E+00
z at max.Ipi,α
1.43E+00
X
7.04E+00 2.04E-01
4.82E+00 2.06E-01
1.07E-01 4.82E+00 2.06E-01
Y
2.08E-01
1.93E-01
1.93E-01
deq(zb)
7.04E+00
fawf Dim of Aaprt td Other Information
prr pr at max. Ipi
1.11E-07 1.57E+04 3.06E+00
deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
TIC
Non-scan
Aaprt> 2 1 cm
1.05E-01 3.40E+02
Control 1
Focal zone number: one , Focal position: 2.8cm for B-mode
Control 2
Sampling Volume Position: at the 2.8cm of B-mode
Control 3
Depth: 3.70cm
Control 4
Working Frequency:5.0MHz for B-mode, and 4.6MHz for PW-mode
Control 5
Scan Line Mode: lower-density line and minimum scan angle for B mode
Control 6
PRF: 15680Hz for B-mode, and 2300Hz for PW-mode, SV=0.5mm.
Highest transmission power (100%) Control 7 Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
14
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: 7L4s Imaging Mode:
Color+B-Mode/Power+B-Mode TIS
Index Label
9.37E-01 2.52E+00
pr.a P Min of Ita,α(zs)] zs Associate Acoustic Parameter
Non-scan
MI
Maximum Index Value
TIB
Scan
Aaprt≤1 2 cm
Aaprt>1 2 cm
Nonscan
TIC
4.50E-01
5.23E-01
2.35E+01
2.35E+01
[Pα(zs),
zbp zb
1.47E+00
z at max.Ipi,α deq(zb) fawf
7.12E+00
Dim Aaprt
of
td Other Information
prr pr at max. Ipi
Y
1.09E-07 1.57E+04 2.90E+00
deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
X
3.13E+02
Control 1
B-Mode Focal zone number: one ; Focal position: 2.0cm for B-mode
Control 2
Color Sampling Gate Position: at the 2.0cm of B-mode; Color Sampling Gate Width: minimum.
Control 3
Depth: 2.50 cm
Control 4
Working Frequency: 5.0MHz for B-mode, and 4.6MHz for Color-mode
Control 5
Scan Line Mode: minimum scan angle for B-mode, and lower-density line for B-mode and Color-mode, and color sensitivity : 3 .
Control 6
PRF: 15680Hz for B-mode, and 11400Hz for Color-mode
Control 7
Highest transmission power (100%)
Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
15
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: 7L4s Imaging Mode:
Color+B+PW-Mode/Power+B+PW-Mode TIS
Index Label
MI
Maximum Index Value
9.81E-01 2.60E+00
pr.a P Min of Ita,α(zs)] zs Associate Acoustic Parameter
TIB Non-scan
Scan
Aaprt≤1 cm
2
Aaprt> 2 1 cm
9.52E-02
5.68E-01
6.90E-01
4.68E+01
4.18E+00
4.18E+00
5.02E+01
[Pα(zs),
zbp
1.38E+00
zb
1.45E+00
z at max.Ipi,α
X
4.81E+00 2.05E-01
1.01E-01 4.81E+00 2.05E-01
Y
1.95E-01
1.95E-01
fawf
7.02E+00
Dim of Aaprt td prr pr at max. Ipi
9.91E-02 3.30E+02
Control 1
B mode Focal zone number : one ; Focal position: 2.8cm for B-mode
Control 2
Sampling Volume Position: at the 2.8cm of B-mode; Color Sampling Gate Position: at the 2.8cm of B-mode.
Control 3
Color Sampling Gate Width : Minimum Gate.
Control 4
Depth: 3.70cm Working Frequency : 5.0MHz for B-mode, 4.6MHz for PW-mode, and 4.6MHz for Color-mode, and color sensitivity: 3 . Scan Line Mode: Minimum scan angle for B-mode, and lower-density line for B-mode and Color-mode, SV=0.5mm. PRF: 15680Hz for B-mode, 2300Hz for PW-mode, 2300Hz for Color-mode. Acoustic Power: Highest transmission power (100%).
Control 5 Control 6 Control 7
Note 1 mode.
1.10E-07 1.57E+04 3.07E+00
deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
TIC
4.74E-01
deq(zb)
Other Information
Nonscan
Control 8 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that
16
Acoustic Output Reporting Tables
Acoustic Output Reporting Tables Transducer Model: 6CV1s Imaging Mode: M-mode TIS Index Label
MI
Scan
7.95E-01 1.97E+00
Maximum Index Value pr.a
TIB
Non-scan
P
Nonscan
TIC
7.83E-02
1.76E-01
1.10E-01
2.79E+00
2.79E+00
2.79E+00
Aaprt≤1cm
2
Aaprt> 2 1cm
Min of [Pα(zs), Ita,α(zs)] zs Associate Acoustic Parameter
zbp
1.57E+00
zb
1.63E+00
z at max.Ipi,α
6.00E+00 2.02E-01
1.88E-01 6.00E+00 2.02E-01
6.00E+00 2.02E-01
3.23E-01
3.23E-01
3.23E-01
deq(zb)
6.00E+00
fawf Dim Aaprt
of
td Other Information
prr pr at max. Ipi
Y
2.79E-07 1.00E+03 2.47E+00 1.82E-01
deq at max. Ipi Ipa,α at max. MI
Operating Control Conditions
X
2.11E+02
Control 1
Focal position: at the 2.0cm focus of B-mode
Control 2
Depth: 3.70cm
Control 3
Working Frequency: 5.0MHz
Control 4
PRF: 2000Hz
Control 5
Highest transmission power (100%)
Note 1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
17