Operating Table MOT-VS600 Reverse mode: Supplementary information
Table of contents 1.
Preparing and checking the reverse mode ・・・・Supplement-1
2.
Lateral tilt ・・・・・・・・・・・・・・・・・・Supplement-2
3.
Trendelenburg ・・・・・・・・・・・・・・・・Supplement-3
4.
Sliding
5.
The back plate ・・・・・・・・・・・・・・・・Supplement-5
・・・・・・・・・・・・・・・・・・・Supplement-4
MES-CK18-600-20EN-1 Ver.2
2021-04-26
1. Preparing and checking the reverse mode
WARNING
• The lateral tilt, Trendelenburg and sliding operate in the directions opposite to the switches on the control unit. The operators must make sure to check each of the switches and their operating directions prior to surgeries. If they are operated without them having been checked, the patients may get injured due to them unexpectedly operating. • Do not perform the flexing, reflexing nor moving up and down the lift-up unit (MOT-VS600DK/MOT-VS600DHK). Since the tilting positions of the patients are different from the normal positions, the patients may unexpectedly bear more loads,and thereby get injured. • The allowable patient weight when in reverse mode is 135 kg. If patients whose weights exceed 135 kg are placed on the operating table, then it may tip over, and the patients may thereby get injured.
Preparing the reverse mode For the MOT-VS600, detach the leg plate, connect the Leg-side Reversal Adapter, and attach the accessories in accordance with their applications.
Checking the reverse mode To verify that the patient is in the reverse mode, check that the reverse mode identification label is aligned with the patient's head as per the following illustration. The illustration shows the Leg-side Reversal Adapter and the 100cm Carbon Fiber Extension Ⅱ after removing the MOT-VS600's leg plate.
Reverse mode identification label Supplement-1
Reverse mode
2. Lateral tilt Tilting to the left
1. Hold down
and press
.
The tabletop tilts to the left in the view from the reverse mode identification label side. MOT-VS600DK
Left down REV. TREND.
TREND.
TILT LEFT
TILT RIGHT
BACK UP
BACK DOWN
TABLE UP
TABLE DOWN
SLIDE FOOT
SLIDE HEAD
KIDNEY UP
KIDNEY DOWN
FLEX
REFLEX
BRAKE UNLOCK
LEVEL
ENERGIZE
ON/OFF
PRESS TO LOCK BRAKES
Tilting to the right
1. Hold down
and press
.
The tabletop tilts to the right in the view from the reverse mode identification label side. MOT-VS600DK
REV. TREND.
TREND.
TILT LEFT
TILT RIGHT
BACK UP
BACK DOWN
TABLE UP
TABLE DOWN
SLIDE FOOT
SLIDE HEAD
KIDNEY UP
KIDNEY DOWN
FLEX
REFLEX
Right down
BRAKE UNLOCK
LEVEL
ENERGIZE
ON/OFF
PRESS TO LOCK BRAKES
Supplement-2
Reverse mode
3. Trendelenburg Trendelenburg (Head down)
1. Hold down
and press
.
The reverse mode identification label side of the tabletop moves down. MOT-VS600DK
REV. TREND.
TREND.
TILT LEFT
TILT RIGHT
BACK UP
BACK DOWN
TABLE UP
TABLE DOWN
SLIDE FOOT
SLIDE HEAD
KIDNEY UP
KIDNEY DOWN
FLEX
REFLEX
Head down position
BRAKE UNLOCK
LEVEL
ENERGIZE
ON/OFF
PRESS TO LOCK BRAKES
Reverse Trendelenburg (Head up)
1. Hold down
and press
.
The reverse mode identification label side of the tabletop moves up. MOT-VS600DK
REV. TREND.
TREND.
TILT LEFT
TILT RIGHT
BACK UP
BACK DOWN
TABLE UP
TABLE DOWN
SLIDE FOOT
SLIDE HEAD
KIDNEY UP
KIDNEY DOWN
FLEX
REFLEX
BRAKE UNLOCK
LEVEL
ENERGIZE
ON/OFF
PRESS TO LOCK BRAKES
Supplement-3
Head up position
Reverse mode
4. Sliding Sliding in the head direction
1. Hold down
and press
.
The tabletop slides to the reverse mode identification label side. MOT-VS600DK
REV. TREND.
TREND.
TILT LEFT
TILT RIGHT
BACK UP
BACK DOWN
TABLE UP
TABLE DOWN
SLIDE FOOT
SLIDE HEAD
KIDNEY UP
KIDNEY DOWN
FLEX
REFLEX
Head direction
BRAKE UNLOCK
LEVEL
ENERGIZE
ON/OFF
PRESS TO LOCK BRAKES
Sliding in the foot direction
1. Hold down
and press
.
The tabletop slides to the head plate side. MOT-VS600DK
REV. TREND.
TREND.
TILT LEFT
TILT RIGHT
BACK UP
BACK DOWN
TABLE UP
TABLE DOWN
SLIDE FOOT
SLIDE HEAD
KIDNEY UP
KIDNEY DOWN
FLEX
REFLEX
Foot direction
BRAKE UNLOCK
LEVEL
ENERGIZE
ON/OFF
PRESS TO LOCK BRAKES
Supplement-4
Reverse mode
5. The back plate Tilt up the patient’s leg
1. Hold down
and press
.
The back plate moves up. (The plate on the patient’s leg side moves up.) MOT-VS600DK
REV. TREND.
TREND.
TILT LEFT
TILT RIGHT
BACK UP
BACK DOWN
TABLE UP
TABLE DOWN
SLIDE FOOT
SLIDE HEAD
KIDNEY UP
KIDNEY DOWN
FLEX
REFLEX
BRAKE UNLOCK
LEVEL
ENERGIZE
ON/OFF
PRESS TO LOCK BRAKES
Tilt down the patient’s leg
1. Hold down
and press
.
The back plate moves down. (The plate on the patient’s leg side moves down.) MOT-VS600DK
REV. TREND.
TREND.
TILT LEFT
TILT RIGHT
BACK UP
BACK DOWN
TABLE UP
TABLE DOWN
SLIDE FOOT
SLIDE HEAD
KIDNEY UP
KIDNEY DOWN
FLEX
REFLEX
BRAKE UNLOCK
LEVEL
ENERGIZE
ON/OFF
PRESS TO LOCK BRAKES
Supplement-5
Sales Agent Emergo Europe B.V. Prinsessegracht 20, 2514 AP The Hague, The Netherlands
MIZUHO Corporation 3-30-13 Hongo, Bunkyo-ku Tokyo 113-0033, Japan https://www.mizuho.co.jp