Operating Table Model VACB-3006B Operators Manual Ver.3 May 2019
Operators Manual
62 Pages
Preview
Page 1
Operating Table Operator’s Manual VACB-3006B
The operating table is designed to support a patient during surgical procedures. Using it for other purpose may result in damage or injury. The operator and the person in charge of the maintenance of this operating table MUST read this Operating Table Instruction Manual carefully and understand the contents before operating, inspecting, adjusting, or servicing the table. Keep this Manual where it is readily accessible for reference by those concerned in case of need.
MES-CK18-480-10EN Ver.3 2019.05
Contents 1 Introduction...1 2 Before use...5 3 Component Identification...13 4 Operation...17 Installation and battery charging...18 Switch On/Off power...24 Operating the emergency stop switch...26 Lock and unlock the operating table...27 Brake Release...28 Auxiliary brake...29 Lateral tilt...30 Trendelenburg...32 Move up/move down of the table...33 Return to level...34 Slide the tabletop...35 Accessory installation...37
5 Maintenance and inspection...39 6 Specifications...43 7 W hen the control unit can not be used...47 8 Before contacting your distributor for repair...49 Trouble shooting...50 In case of malfunction...51 Warranty...51
Appendix...53
1
Introduction
2
Before use
3
Component Identification
4
Operation
5
Maintenance and inspection
6
Specifications
7
When the control unit can not be used
8
Before contacting your distributor for repair
1
Introduction
1 Introduction
1
Introduction
3 1
Introduction
Introduction
This Manual is intended to support safe and effective operation of the operating table. Read this Manual carefully and familiarize yourself thoroughly with its contents (operation, inspection, adjustment, servicing) before operating the table. Using the table without following the instructions contained in this Manual may cause a serious accident. Safety information in this Manual and on the labels attached to the table is given in the following three forms for easy and full understanding:
WARNING
A WARNING notice indicates a situation which can result in death or serious injury to a person if the notice is ignored and the table is handled improperly.
CAUTION
A CAUTION notice indicates a situation which can result in injury to a person and/or damage to property if the notice is ignored and the table is handled improperly.
NOTE
Description under this heading gives supplementary explanation of the functions of the table.
The WARNING and CAUTION notices in this Manual relating to the operation, inspection, adjustment, or servicing of the table apply in using the table for surgical procedures. The user is solely responsible for safety if the table is operated,inspected, adjusted, or serviced in a way not specified in this Manual.
■ About this product This product is designed to support patients during surgical procedures. This product has functions to adjust the height of the tabletop and to position a patient’s body. Both a medical grade AC power source and a battery are available as a power source. This product must be used in an operating room by doctors, nurses or medical engineers who are familiar with this product.
1
Introduction
2
Before use
2 Before use
6 2
Before use
Please read carefully before use Misuse of the operating table could cause damage, electric shock or fire.
Table plate
WARNING - Do not sit or lean on the parts of the table described in the following figures. Anyone on the table may fall off or get injured.
CAUTION - Before lowering or tilting the table, check if there are any devices under the table. If the tabletop comes in contact with such devices and excessive force is applied to it, the table may be damaged.
Power cord
WARNING
Control unit / hand console
WARNING - Do not pull the cords of the control unit or hand console forcefully. - Do not subject the control unit or the hand console to forceful impacts. They may get damaged.
- Do not place any heavy object on the power cord. - Do not ride over the power cord with a castered device. - Do not pull the power cord forcefully. - Do not place any objects blocking access for the wall outlet. Doing so may prevent the power cord from being unplugged. power cord
7 Patient's body position
2
Before use
WARNING - When using accessories to secure a patient's body position, make sure to monitor the condition of the patient. Improper body position can lead to paralysis or pressure injuries. - Make sure to attach the provided mattress securely. Improper attachment may cause a patient to get injured. - Position a patient's body at least 1 cm away from the side rails. When using the electric scalpel, they may become hot and if the body touches the rails, it may cause the patient to get burned. - Do not position a patient with his/her head above the column during a Trendelenburg position or when sliding the table to the head side. The table will become unbalanced and it may fall over. - Follow the procedure to position a patient properly. 1. Install the provided mattress to the tabletop. 2. Place a patient on the mattress. Make sure the patient's head or legs are placed to the edge of the tabletop as described in the pictures below.
3. Position the table to the purpose.
Rails
WARNING - Keep your hands away from the gap of the frame during the operation of the table. Otherwise you may get injured.
Installation of the operating table
CAUTION - Do not install the operating table on an uneven floor. - Do not put a pad under the column base for raising the table. The table could fall down and cause injury.
uneven floor surface
pad
8 2
Others
Before use
WARNING Prohibition of disassembling and modification - Do not attempt to disassemble or modify the table. It may cause a malfunction. When used together with other medical electrical equipment - Before using together with other equipment, make sure to check that it is not affected by electromagnetic interference. - Electromagnetic interference from other medical electrical equipment may cause malfunction. When performing CPR, slide the tabletop toward the leg side as far as it will go. Improper body position may make the table unstable and cause the patient to get injured.
When using with a C-arm X-ray machine - Read any documents or instruction manuals included with the C-arm X-ray machine to fully understand any influence that the machine may have on the operation table. The transmission of X-rays may decrease because of the carbon-made tabletop. - Make sure the C-arm X-ray machine does not come in contact with the table. Improper installation may obstruct operations or cause a malfunction. Allowable load - Do not exceed the allowable load*. The table may become inoperable and cause malfunction. *Lift: 225 kg / Trendelenburg, Lateral tilt: 135 kg
Preventive maintenance and inspection - Make sure to perform inspection before and after use, and appropriate preventive maintenance inspection. The table may not survive its useful life depends on number of years and frequency of use. - For preventive maintenance inspection, please contact your distributor or us directly. Cardiac defibrillator Consult manufacturer’s instructions when using high frequency surgical equipment, cardiac defibrillator and cardiac defibrillator monitors. Antistatic pathway When an antistatic pathway is required, the table has to be used on an antistatic floor.
9 Others
2
Before use
CAUTION When used together with other medical equipment or accessories - Before using other devices or accessories, read carefully the instruction manual of the device and make sure that the operating table will not be affected adversely. Before fitting an accessory of third party company, make sure to contact your distributor or us. Some accessories cannot be fitted. - For hygiene, be sure to use sterile drapes on the areas on this product where the patient comes into contact with it. - When using the operating table, make sure to check the positions of other equipment. They may come in contact with each other during the operation, and may get damaged. Cleaning and disinfection - After the table has been used, make sure to perform cleaning and disinfection for the table. Do not use hydrogen peroxide solution to clean the hand console. Using it may cause discoloration. Procedure of cleaning and disinfection 1. Turn off power and remove the power cord from the medical grade wall outlet. 2. Remove the mattress from the table. 3. Wipe the surface, the sides and the back of the mattress with a clean cloth dampened with disinfectant. 4. In the same way as step 3, wipe the table plate and the side rails as well. 5. Wipe the parts again with a dry cloth within 15 minutes. Usable disinfectant solutions which we have already confirmed as compatible. 1 2 3 4 5 6 7 8 9
Medicine Name Sodium hypochlorite 0.1% (halogen containing compound) Sodium thiosulfate Chlorhexidine (chlorhexidine gluconate 0.5%) Benzalkonium chloride (invert soap 10%) Povidone iodine Ethanol 80% Hydrogen peroxide Saline 0.9% Isopropyl alcohol (IPA) 99.5%
Moving and transportation - Do not move the table while a patient is on it. - Follow the procedure below for moving the table. 1. Power off and remove the power cord from medical grade wall outlet. 2. Check if handles and levers are in the fixed position, and each section is fixed firmly. 3. Unlock a brake to move the table. 4. Before moving the operating table, disinfect the whole operating table in order to prevent infection. - Transportation should be carried out with the following conditions met. 1. Disinfect the whole operating table before transportation. 2. Take measures against falling down, like moving down the table to the lowest position. 3. Actuate the brake. 4. Use cushioning materials to prevent damage during transportation. 5. Protect against dust, wind and rain. Disposal - Always follow the local country regulation in disposing the operating table.
10 2
Before use
Warning label: details and position Warning labels are attached on the following area of the operating table. Before using it, make sure to check the contents of the warning label.
⑲
■ WARNING, CAUTION labels ① C653044
② C653045
③ C653046
④ C656740
⑤ C653043
⑥ C655803
CAUTION
THE AUXILIARY SWITCH IS INTENDED TO BE USED WHEN THE CONTROL UNIT IS DEFECTIVE. USE THE CONTROL UNIT WHENEVER IT IS IN NORMAL CONDITION. BECAUSE THE AUXILIARY SWITCH HAS NO FUNCTION TO RESTRICT THE OPERATION OF THE TABLE. THE TABLE MAY BE DAMAGED WHEN IT IS OPERATED WITH THE AUXILIARY SWITCH. WHEN OPERATING THE TABLE WITH THE AUXILIARY SWITCH, BE SURE TO WATCH THE MOVEMENT OF THE TABLETOP CONTINUOUSLY. IF THE TABLETOP SHOULD TOUCH ANYTHING OR GET DAMAGED IN ANY DURING THE OPERATION, STOP OPERATING THE TABLE IMMEDIATELY.
⑦ C657333
⑧ C653624
⑨ C653047
11 ■ Other labels ⑩ C653614
⑪ C610171
Before use
⑫ C653620
⑭ C653050
⑯ C653048
2
⑬ C653035
⑮ C653034
⑰ C653513
⑱ 303A6M2
PIN 303A6M2
⑲ C630006 VACB-3006B 18-480-70
㉑ C657318
⑳ C657306 100-240 50-60 400 MUC-YYYY★-XXXX 3min on,7min off
㉒ C655711
㉓ C653036
12 Symbol
2
Before use
WARNING CAUTION
Referent
Possibility of injury or even death if operates the table without following the warning.
Label No.
①②③④ ⑤⑥⑧⑨
Indicates AC power supply.
⑲
Enclosure Class (Splash-proof)
⑲
Serial Number
⑲
Catalogue Number
⑲
Indicates waste disposal information
⑲
Refer to instruction manual Authorized representative in the European Community Manufacturer
Equalization terminal
General mandatory action sign
④⑩⑲ ⑲ ⑱⑲ ⑰ ④⑤⑨
Defibrillation-proof applied part (Type B)
⑱
Type B
⑫
Emergency stop
⑩
ON (Power)
⑳
OFF (Power)
⑳
14 Main unit 3
Component Identification
Side rail Emergency stop switch Auxiliary switch
Tabletop
Emergency brake release handle Auxiliary brake
Brake Caster
■ Power switch
Auxiliary switch Control unit connector
Emergency stop switch Equipotential terminal Hand console connector
Battery indicator Power switch Power connector
15 ■ Control handle
■ Control unit
3
Component Identification
⑮
②
① ③
④
⑤
⑥
⑦
⑧
⑨
⑩
⑪ ⑫ ⑬
⑭
CAUTION ATTENTION KEEP CORD CLEAR OF MOVING PARTS. MAINTENEZ LE CORDON D’ ALIMENTATION ÉLOIGNÉ DES PIÈCES EN MOUVEMENT.
⑯
① Pilot lamp ② Power lamp/Battery lamp ③ NORMAL ④ REVERSE ⑤ Reverse Trendelenburg: Head up ⑥ Trendelenburg: Head down ⑦ Lateral tilt: Left down ⑧ Lateral tilt: Right down ⑨ Table up (Lock operating table) ⑩ Table down ⑪ Brake unlock ⑫ Return to level ⑬ E button ⑭ ON/OFF switch (battery power operation) ⑮ Switch (Power On, Slide unlock) ⑯ Fixing handle
NOTE - If the switch on the control unit is pressed continuously for 3 minutes or more, the operation of the switch will be stopped. The stop will continue for about 7 minutes. Operations except for the stopped switch can be done as usual. - When
is pressed, pilot lamp will be on for 3 seconds.
16 ■ Side Rail Adaptor for Carbon Fiber Board
(Option)
3
long
Component Identification
short
Product code
Product name
08-113-10 08-113-10-R1 08-113-10-R2 08-113-11 08-113-11-R1 08-113-11-R2
Side Rail Adaptor II, Long for Carbon Fiber Board Side Rail Adaptor II, Long for Carbon Fiber Board (R1) Side Rail Adaptor II, Long for Carbon Fiber Board (R2) Side Rail Adaptor II, Short for Carbon Fiber Board Side Rail Adaptor II, Short for Carbon Fiber Board (R1) Side Rail Adaptor II, Short for Carbon Fiber Board (R2)
■ Radiolucent Head Extension for VACB with
Pad (Option)
Radiolucent Head Extension for VACB
Product code
Pad for Radiolucent Head Extention for VACB
Product name
08-075-47
Radiolucent Head Extension for VACB with pad
■ Foot Switch
Elevation: Up Elevation: Down Trendelenburg: Head down
Lateral tilt: Right Lateral tilt: Left Reverse Trendelenburg: Head up
4
Operation
4 Operation ON
FIX
RELEASE
Installation and battery charging P18
Switch On/Off power P24
Emergency Shutdown P26
Lock/unlock the operating table P27
Brake Release
Auxiliary brake
P29
Lateral tilt
Trendelenburg P32
Change height of the tabletop P33
Return to level P34
Slide the tabletop
P35
P30
Accessory installation P37
P28
18 Installation and battery charging ■ Installation 4
Operation
CAUTION
1. Move to a flat area.
Equipotential grounding
2. Check the battery lamp. If the battery lamp (orange) on the control unit flashes, battery charging is required. battery lamp
3. If the battery indicator shows empty (red), battery charging is required.
red
Ground the equipotential terminal and earth ground terminal on the wall by means of an equipotential wire (not included).
19 Installation and battery charging ■ Battery charging
CAUTION
1. To start charging, insert the power cord into a power connector and into a medical grade wall outlet and turn on the power switch. medical grade wall outlet
- Connect the power cord to a power source with protective grounding to prevent the risk of an electric shock. - Make sure to use the dedicated power cord with "MIZUHO" logo.
power switch
- Make sure that the connector of the power source is not wet and is not dusty before plugging the power cord into the connector of the power source.
power cord connector of the power source
2. During the battery charge, the battery indicator of the operating table flashes yellow.
3. The battery indicator flashes yellow to green in order when charging is completed.
- Pull out the power cord from the medical grade wall outlet to cut the power supply completely. - If the battery deteriorates, it will not be available for the operating table when the AC power is not supplied due to power outage and etc.
NOTE green
- When you first purchase the product or have not used it for long time, make sure to charge the battery before use. - Battery will get discharged naturally even when the table is not used. Make sure to charge the battery before use.
yellow
- If the battery lamp on the control unit flashes after using the operating table or the battery indicator on the lower part of the operating table shows empty (red) when using, charge the battery immediately.
4
Operation