Shoulder Positioner Operator’s Manual
Table of Contents 1. 2. 3. 4. 5.
Introduction... 2 Component ... 3 Instructions for use ... 4 Maintenance / Inspection ... 5 Specifications ... 6
Be sure to read this instruction manual before using this product. Keep this manual available for reference when needed.
MES-CK08-068-21EN Prepared: 2020-05-14 (Ver.2)
1.
Introduction
1-1. Intended purpose This product is used with Ultra shoulder positioner® patient care kit with rehab sling for supporting the patient's head and back when shoulder operations are being performed. Swap the head plate and the back plate of the applicable Mizuho operating tables with this product to use it.
1-2. Intended user This product is to be used by health care professionals, including but not limited to surgeons, nurses and biomedical technicians.
1-3. Warning/Caution To ensure safe operation and proper functionality, please read the following instructions thoroughly before use and follow them carefully. Failure to follow these instructions could lead to serious injury of the patient or the user or damage to this product. If these indications are ignored and the product is incorrectly used, death and/or serious injury may result. • Do not use this product for any purposes other than the intended use. This product is designed for shoulder operations and cannot be used for other purposes. • When using this product, always pay close attention to the patient. Patients left in the same posture for an extended period of time are at risk for developing nerve paralysis and/or bedsores. • Ensure all fixing handles are tight. Loose condition can cause the product to come off or move, which may result in a patient injury. If these indications are ignored and the product is incorrectly used, injury and/or damage to property may result. • Do not use this product with other company’s operating table. This product is only validated for use with Mizuho operating tables. Do not use this product beyond the allowable load. Doing so may damage this product. (See the section “Specifications” for the allowable load.) Before attaching this product to the operating table, check that the tabletop is in the midpoint by using the sliding function. • Do not use the sliding function after this product is attached to the operating table. • Check that this product has been completely inserted to the operating table body. • When working with an operating table, take care not to allow this product to make contact or interfere with the table top or with other tools and appliances used in combination with this product. Otherwise the product could break. • For hygiene, be sure to use sterilized drapes on the areas on this product where the patient comes into contact with it.
1-4. Symbol mark for labeling MD : Medical Device
2
2.
Component
(13) (12)
(16)
(1) (14)
(2) (3)
(5)
(4) (6) (7) (8)
(11)
(15)
(9)
(10)
(17)
Option: Ultra shoulder positioner® patient care kit with rehab sling (3/case)
Main body: Shoulder Positioner Component
Material
(1)
Head rest
Thermosetting resin laminated board
(2)
Flexion positioning lever
Plastic
(3)
Head rest sliding adjustment knob
(4)
Component
Material
(13)
Forehead strap
Plastic
(14)
Head pad
Urethane foam
(15)
Jaw strap
Plastic
Plastic
(16)
Back pad
Urethane foam
Right shoulder panel
Thermosetting resin laminated board
(17)
Arm fixing strap
Plastic
(5)
Left shoulder panel
Thermosetting resin laminated board
(6)
Upper rotation lock
Plastic
(7)
Plunger
Plastic
(8)
Lower rotation lock
Plastic
(9)
Back panel
Thermosetting resin laminated board
(10)
Mounting handle
Aluminum alloy
(11)
Adapter
Aluminum alloy
(12)
Drape support rod
Stainless steel
Note: Ultra shoulder positioner® patient care kit with rehab sling (3/case) is the product of Mizuho OSI.
3
3.
Instructions for use The following describes how to attach, detach and operate this product.
Head rest
Back plate attaching pin Adapter
1. Check that the tabletop is in the horizontal position by using the operating table's automatic re-leveling function. 2. Check that the tabletop is in the center position by using the sliding function of the operating table. 3. Remove the head plate and the back plate from the operating table. 4. Attach the Adapter part of the Shoulder Positioner to the back plate attaching pin on the operating table. 5. Attach the disposable patient care kit to the Shoulder Positioner. 6. Transfer and set the patient to the Shoulder Positioner. 7. Operate the back plate up function of the operating table to adjust the angle of the Back panel. 8. Use the Flexion positioning lever and Headrest sliding adjustment knob of the Head rest to adjust the head position. 9. After use, put the position and angle of the Head rest back to in place, detach the Adapter part of the Shoulder Positioner from the back plate attaching pin on the operating table, and then attach the head plate and the back plate to the operating table.
Note: Ultra shoulder positioner® patient care kit with rehab sling is disposable. Do not reuse this product.
4
4.
Maintenance/Inspection
4-1. Inspection before and after Use • Check that this product is not damaged or broken before and after each use.
4-2. Cleaning and Disinfection • After using this product, make sure to follow the steps below to clean up and disinfect this product. a) Detach this product from the operating table. b) Detach all the Ultra shoulder positioner® patient care kit with rehab sling (3/case) from this product. c) Use a lint-free cloth soaked with proper volume of disinfectant to wipe off the upper, side, and back side of the surface of the product. d) Wipe this product dry with a clean cloth within 15 minutes after disinfecting. • Make sure to use Mizuho authorized disinfectants listed below. Disinfectants a) Sodium hypochlorite 0.1% (halogen containing compound) b) Hypo Alcohol (iodine decolorant color removing agent) c) Chlorhexidine (chlorhexidine gluconate 0.5%) d) Benzalkonium chloride (invert soap 10%) e) Povidone iodine f) Ethanol 80% g) Oxydol (hydrogen peroxide) h) Saline i) Isopropyl alcohol (IPA)
4-3. Repair and Maintenance • When this product is broken, clearly label the device as “Broken”, “Do not use”, “Need repair”, etc., and contact your local dealer or Mizuho. • Only Mizuho or its certified agents can provide repair and maintenance services of this product. Services by other agents could cause adverse events and/or deterioration of the performance and functions.
4-4. Warranty • MIZUHO Corporation will repair defective parts of this product without charge for one year from the date of delivery/installment except for cases of damage caused by a third party’s repair, act of nature, improper use or intentional damage. All other warranty terms and conditions are subject to regulations of MIZUHO Corporation.
4-5. Disposal This device must be disposed of in accordance with local regulations. Please contact your local distributor for proper disposal.
4-6. Notice Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority in which the user and/or patient is established.
5
5.
Specifications
Ultra shoulder positioner® patient care kit with rehab sling (3/case)
Shoulder positioner
Code No.
Product description
08-068-22
Shoulder Positioner without Pad
5348
® patient care kit
Ultra shoulder positioner with rehab sling (3/case)
Weight (kg)
Component
9
(1) to (12)
-
(13) to (17)
Note: Ultra shoulder positioner® patient care kit with rehab sling (3/case) is the product of Mizuho OSI.
Slide range
Max. 200 mm
Head rest Specifications
Lift range
Max. 115 mm
Side adjustment angle
Max. 30°on each right and left
Allowable load
120kg: Amount of weights of the patient and devices used together.
Operating environment Transportation and storage environment Service life
Temperature
10 to 40 °C
Relative humidity
30 to 75 %
Atmospheric pressure
700 to 1060 hPa
Temperature
-10 to 50 °C
Relative humidity
10 to 85 %
Atmospheric pressure
700 to 1060 hPa
7 years (Subject to the specified maintenance and inspection and is stored properly.)