Operators Manual
78 Pages
Preview
Page 1
Operating Table MOT-VS600 Operator’s Manual
The operating table is designed to support a patient during surgical procedures. Using it for other purposes may result in damage or injury. The operator and the person in charge of the maintenance of this operating table MUST read this Operating Table Operator’s Manual carefully and understand the contents before operating, inspecting, adjusting, or servicing the table. Keep this Manual where it is readily accessible for reference by those concerned in case of need.
MES-CK18-600-20EN Ver.8 2021-07-05
Contents
1
Introduction
2
1. Introduction... 1
Before use
2. Before use... 7
Component identification
3. Component identification... 15
Operation
4. Operation... 21
Maintenance and inspection
Installation and battery charging...22 Switch On/Off power...31 In the case of emergency...33 Lock and unlock the operating table...34 Brake Release...35 Lateral tilt...36 Trendelenburg...38 Tilt the back plate...40 Change height of the tabletop...41 Slide the tabletop...42 Flexing...44 Change height of the lift-up unit...46 (MOT-VS600DK/MOT-VS600DHK only)...46 Return to level...47 Adjustment of head plate...48 Adjustment of back plate...50 Adjustment of leg plate...53
5. Maintenance and inspection... 57 Periodic replacement parts...59 Version information of the software...59
6. Specifications... 61 7. Troubleshooting... 65 8. Before contacting for repairs... 67 App.-1 Electromagnetic Compatibility ... 70 App.-2 Glossary ... 73
3 4 5 6
Specifications
7
Troubleshooting
8
Before contacting for repairs
1
Introduction
1. Introduction
2 1
Introduction
3 1
Introduction
Introduction
This Manual is intended to support safe and effective operation of the operating table. Read this Manual carefully and familiarize yourself thoroughly with its contents (operation, inspection, adjustment, servicing) before operating the table. Using the table without following the instructions contained in this Manual may cause a serious accident. Safety information in this Manual and on the labels attached to the table is given in the following forms for easy and full understanding:
WARNING
A WARNING notice indicates a situation which can result in death or serious injury to a person if the notice is ignored and the table is handled improperly.
CAUTION
A CAUTION notice indicates a situation which can result in injury to a person and/or damage to property if the notice is ignored and the table is handled improperly.
NOTE
Description under this heading gives supplementary explanation of the functions of the table.
The WARNING and CAUTION notices in this Manual relating to the operation, inspection, adjustment, or servicing of the table apply in using the table for surgical procedures. The user is solely responsible for safety if the table is operated, inspected, adjusted, or serviced in a way not specified in this Manual.
About this product This product is designed to support patients during surgical procedures. This product has functions to adjust the height of the tabletop and to change a patient's body position. Both a medical grade AC power and a battery are available as a power source. This product is to be used by health care professionals, including but not limited to surgeons, nurses and biomedical technicians.
4 1
Introduction
Accessories Standard compornents and accessories Mattresses (for head, back, seat and leg)
Control unit
Power cord
Operator's manual
Main unit
5 1
Optional parts zzCordless control unit
Introduction
zzFoot switch
MOT-VS600DK
REV. TREND.
TREND.
TILT LEFT
TILT RIGHT
BACK UP
BACK DOWN
TABLE UP
TABLE DOWN
SLIDE FOOT
SLIDE HEAD
KIDNEY UP
KIDNEY DOWN
FLEX
REFLEX
BRAKE UNLOCK
LEVEL
ENERGIZE
ON/OFF
PRESS TO LOCK BRAKES
zzMizuho Original Pad
for Head Section Type A
for Back Section Type A
for Seat Section Type A
for Seat Section Type B
for Back and Seat Section Type A
for Double Leg Section Type A
6 1
Introduction
2
Before use
2. Before use
8 2
Before use
Please read carefully before use Misuse of the operating table could cause damage, electric shock or fire. Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority in which the user and / or patient is established.
Head plate or leg plate
WARNING - Do not sit on the head plate or the leg plate(s). It could cause a person to tumble down and get injured.
CAUTION - When transferring a patient from the bed (Fig a) or changing a patient’s body position (Fig b), do not apply excessive force on the head plate or leg plate. The operating table may get deformed or damaged. Fig a when transferring a patient from the bed
Fig b when changing a patient’s body position
CAUTION - Before lowering the tabletop or placing it in a reverse Trendelenburg position, check if there are any devices under the leg plates. If the leg plates come in contact with such devices and excessive force is applied to the leg plates, the leg plate inserting shaft may be damaged.
head plate
leg plate
head plate
leg plate
Power cord
WARNING - Do not place any heavy object on the power cord. - Do not ride over the power cord with a castered device. - Do not pull the power cord forcefully. - Do not place any objects in the place where the power cord is to be unplugged from the medical grade outlet, which would obstruct it from being unplugged.
Base
CAUTION - Do not place any objects on the base. The object(s) may get caught in the slide cylinder and damage the operating table.
Installation of the operating table
WARNING power cord
- Do not install the operating table on an uneven floor. - Do not put a pad under the column base for raising the table. It may cause the table falling down and injury.
9 Patient's position during surgical operation
WARNING - Make sure to attach the provided mattress securely. Improper attachment may cause a patient to get injured. - Position a patient's body at least 10 mm (0.39 in) away from the side rails. When using the electric scalpel, they may become hot and if the body touches the rails, it may cause the patient to get burned. Positioning the patient - Follow the steps below to position the patient. 1. Attach the accessory mattress onto the tabletop with their Velcro fastenings aligned with each other. 2. Put the patient on the mattresses. 3. Position the patient according to the purpose of the surgical operation.
Operator's position
CAUTION - Have the person who operates the operating table at the position where the emergency stop switch can be immediately pressed, and the patient’s condition can be constantly observed. Operator's position 180°
Head plate ” /16 -11 19
m) 0m (50
Back plate Head of the patient
Control unit
CAUTION - Do not pull the control unit cord forcefully. - Do not apply strong shock to the control unit. It may get damaged.
uneven floor surface
pad
2
Before use
Others
WARNING Prohibition of disassembling and modification - Do not attempt to disassemble or modify the table. It may cause a malfunction. Patient’s body position - When using accessories to secure a patient's body position, make sure to monitor the condition of the patient. Improper body position can lead to paralysis or pressure injuries. When uses together with other medical electrical equipment - When using a high-frequency surgical equipment and/or a cardiac defibrillator etc. with the operating table, refer to their operator's manual provided by the manufacturers and observe the precautions and usage. Improper precautions and usage may cause the operator or the patient to get burned and/or devices to malfunction. - Before using together with other equipment, make sure to check that it is not affected by electromagnetic interference. An electromagnetic interference from other medical electrical equipment may cause malfunction. Allowable patient body weight - Do not apply loads which exceed the allowable patient body weight (total of the patient and accessories)*. The table may become inoperable and cause malfunction. *Lift: 450kg (992lbs)/ Trendelenburg, Lateral tilt: 360kg (794lbs)
Preventive maintenance and inspection - Make sure to perform inspection before and after use, and appropriate preventive maintenance inspection. The operating table may require replacement of the parts due to significant wear, deterioration, and/ or breakage depending on the length of service and frequency of use. - For preventive maintenance inspection, please contact your distributor or us directly. Antistatic measure - Do not use the operating table on floors and/or with accessories that do not possess static electricity countermeasures. This may impede surgical operations.
10 2
Before use
Others
CAUTION When used together with other medical equipment or accessories - Before using another devices or accessories, read carefully the instruction manual of the device and make sure that the operating table will not be affected adversely. Before fitting an accessory of third party company, make sure to contact your distributor or us. Some accessories cannot be fitted. - When using the operating table, make sure to check the positions of other equipment. They may come in contact with each other during the operation, and may get damaged. - For hygiene, be sure to use sterile drapes on the areas on this product where the patient comes into contact with it. Disinfection - The following procedure should be followed when disinfecting an operating table. 1. Power off and remove the power cord from medical grade wall outlet. 2. Remove all table pads from the operating table. 3. Apply a proper quantity of disinfectant on a clean and lint-free cloth, and wipe the top and sides and bottom of the pads with the cloth. A proper quantity of disinfectant means a quantity such that the cloth with disinfectant does not dry when wiping. 4. Disinfect the table board of the table top and the surface of the side rail in the same way. 5. Wipe all disinfected parts with dry, clean and lint-free cloth within 15 minutes after disinfecting.
Usable disinfectant solutions which we have already confirmed as compatible: 1 2 3 4 5 6 7 8 9
Sodium hypochlorite 0.1% (halogen containing compound) Sodium thiosulfate Chlorhexidine (chlorhexidine gluconate 0.5%) Benzalkonium chloride (invert soap 10%) Povidone iodine Ethanol 80% Hydrogen peroxide Saline (0.9%) Isopropyl alcohol (IPA)
Move and transportation - Do not move the table while a patient is on it. - Follow the procedure below for moving the table. * When moving the operating table, disinfect the whole operating table in order to prevent infection. 1. Power off and remove the power cord from medical grade wall outlet. 2. Check if handles and levers are in the fixed position, and each section is fixed firmly. 3. Unlock brake to transfer the table. * Do not move the operating table by pulling on the head plate or leg plate. - Transportation should be carried out with the following conditions met. 1. Disinfect the whole operating table before transportation. 2. Take measures against falling down, like moving down the table to the lowest position. 3. Actuate a brake. 4. Use cushioning materials to prevent damage during transportation. 5. Protect against dust, wind and rain. Disposal - In accordance with the European Union Waste Electrical and Electronic Equipment (WEEE) Directive, all electrical components and batteries must be disposed of in accordance with local regulations. Please contact your local distributor for proper disposal.
11 Warning label: details and position
2
Warning label is attached on the following area of the operating table. Before using it, make sure to check the contents of the warning label.
Before use
WARNING, CAUTION labels (2) (4)
(6)
(3)
(3)
(5)
(1)
(4) (8) (10)
(9)
(7)
(1) C655740
(3) C653624
(6) C656310 (MOT-VS600DK/MOT-VS600DHK only)
(2) C655752
(4) C655732
(5) C653614
(7) C656740
(8) C655803 (9) C655739
(10) C657333
12 Other labels (1/2) 2
Before use
(18) (17)
(11) (12) (14) (16)
(16) (13) (11)
(15)
(22)
(19) (20)
(21) (22) (12) C655725
(11) C655758
C655759
C655762
C655763
C655760
C655761
C655764
C655765
C655768
C655769
(14) C655722
(17) C655717
(13) C655726
(15) C655721
(19) C653515
(16) C655724
(20) C653516
(18) C600020 C655766
C655767 (21) C642002
YYYY-MM-DD Made in ***
(22) C653513
(23) C653612
13 Other labels (2/2) 2
Before use
(24)
(24)
(24)
(24)
(28) (27)
(24)
(26) (25)
(32)
(24) C653620
(28) C655734
or C655748
or C655744
(31) PIN 573A6M1
(29)
(31)
(30)
(25) C655736
(26) C655737
(27) C655755
or C655750
or C655751
or C655756
or C655746
or C655747
or C655757
(29) PIN 571B8M4
(32) PIN 50005M2
(30) PIN 571B8M3
14 Symbol mark for labeling 2
Before use
Symbol
Description
Label no.
Indicates a possibility of injury or even death if operates the table without following the warning.
(1) (2) (3) (4) (6) (7) (8) (10)
General prohibition sign
(7)
General mandatory action sign.
(7)
Emergency stop
(5)
Refer to the operators manual
(5) (9)
Indicates AC power supply.
(21)
Enclosure Class (Splash-proof)
(21)
Serial number
(21)
Catalogue Number
(21)
Indicates waste disposal information.
(21)
European authorized representative
(21)
Medical Device
(21)
Manufacturer
(21)(29)(30)(31)(32)
Equalization terminal
(22)
Type B
(24)
Defibrillation - proof Type B applied part
(29) (30) (31) (32)
16 Main unit
Auxiliary switch
3
Component identification
Slide cylinder
Control unit connector Emergency stop switch
Back plate attaching/ Lift-up unit(*3) removing levers
Seat plate
Leg plate(*2)
Back plate(*2) Head plate(*2) Leg plate swinging handle Leg plate fixing lever Head plate flexing lever Side rail
Control unit
Head plate fixing handle
Leg plate flexing lever
ight receiving section of the cordless L control unit(*1)
Slide cylinder Column
Base
Power cord
Caster Power lamp Brake
Power switch *Foot switch connector (*1) Power connector
POWER CORD
Emergency brake release handle
POWER SWITCH
FOOT SWITCH
Battery indicator Equipotential terminal Light receiving section of the cordless control unit (*1)
*1: The cordless control unit and the foot switch are optional. *2: The head plate, back plate, and the leg plate are detachable. *3: Only the MOT-VS600DK and MOT-VS600DHK
17 Control unit 3
Component identification
MOT-VS600DK Power lamp Pilot lamp
Battery lamp REV. TREND.
TREND.
Lateral tilt: Left down
TILT LEFT
TILT RIGHT
Lateral tilt: Right down
Back plate bending:Up
BACK UP
BACK DOWN
Back plate bending:Down
Table up (Lock operating table)
TABLE UP
TABLE DOWN
SLIDE FOOT
SLIDE HEAD
KIDNEY UP
KIDNEY DOWN
FLEX
REFLEX
Reverse Trendelenburg: Head up
Slide foot Lift-up unit: Up (MOT-VS600DK/ MOT-VS600DHK only) Flex: Center up Brake unlock
Table down Slide head Lift-up unit: Down (MOT-VS600DK/ MOT-VS600DHK only) Reflex: Center down
BRAKE UNLOCK
LEVEL
E switch
Trendelenburg: Head down
ENERGIZE
ON/OFF
Return to level Power ON/OFF switch
PRESS TO LOCK BRAKES
NOTE
- About 7 minutes of halting will be need when operating the table continuously for about 3 minutes. If the motor gets overheated you will not be able to operate the table. When overheat occurs, about 60 minutes of rest will be need to operate the table as usual. - If
is pressed, the pilot lamp will be on for 3 seconds.
18 Cordless control unit (optional) 3
Component identification
MOT-VS600DK Pilot lamp Reverse Trendelenburg: Head up Lateral tilt: Left down Back plate bending:Up Table up (Lock operating table)
REV. TREND.
TREND.
TILT LEFT
TILT RIGHT
BACK UP
BACK DOWN
TABLE UP
TABLE DOWN
SLIDE HEAD
Lift-up unit: Up (MOT-VS600DK/ MOT-VS600DHK only) Flex: Center up
KIDNEY UP
KIDNEY DOWN
FLEX
REFLEX
Brake unlock
BRAKE UNLOCK
LEVEL
Set switch E switch
Lateral tilt: Right down Back plate bending:Down Table down
SLIDE FOOT
Slide foot
Trendelenburg: Head down
ENERGIZE
ON/OFF
Slide head Lift-up unit: Down (MOT-VS600DK/ MOT-VS600DHK only) Reflex: Center down
Return to level Power ON/OFF switch
PRESS TO LOCK BRAKES
NOTE
- About 7 minutes of halting will be need when operating the table continuously for about 3 minutes. If the motor gets overheated you will not be able to operate the table. When overheat occurs, about 60 minutes of rest will be need to operate the table as usual.
- If is pressed, the pilot lamp will be on for 3 seconds. - When using the cordless control unit, set up the set switch on the cordless control unit and the rotary switch on the operating table. Please refer to P29 for setting up the set switch. If there is anything you don't understand about setting it up, contact your distributor or us.
19 Foot switch (optional) 3
Common to MOT-VS600
Elevation: Up
Elevation: Down
Trendelenburg: Reverse Head down Trendelenburg: Head up
Component identification
Lateral tilt: Left
Lateral tilt: Right