Quick Guide
2 Pages
Preview
Page 1
1
Enter administrative mode
Switch Off pressure with the “Standby” button [ > 3 sec.].
2
HQ004591002
Standby
Change Settings
Change Canister and Tubing 1
Clamp tubing
2
3
Avance Negative Pressure Wound Therapy TM
- Avance Pump changes into the standby mode
+
Quick Guide
Press 3 Sec.
press simultaneously
Switch Off 3
4
5
6
7
8
9
Off (>3 sec.)
To change setting, choose with
or “selection buttons“
4
Press
to select
Basic Double Key Functions
“OK“
Administrative mode or
6
Press
to confirm
Log file Press
change setting
Change setting
Change pressure
to exit settings “Back“
or
Change pressure
Press
+
Press and hold
“OK“
7
press simultaneously
5
+ press simultaneously
On
Unclamp tubing
Please see IFU for warnings and contraindications. Mölnlycke Health Care AB (publ.), Gamlestadsvägen 3C, Box 13080, SE-402 52 Göteborg, Sweden. Phone + 46 31 722 30 00. www.molnlycke.com. The Mölnlycke Health Care name and logo and AvanceTM are registered trademarks of Mölnlycke Health Care AB. ©2009 Mölnlycke Health Care. All rights reserved.
Switch On How to enter administrative mode
Enter administrative mode
2. On
1. Press and hold
2
or
Acknowledge disclaimer
change pressure
Troubleshooting on the display
Remarks/potential cause of fault
301
Leak in system
Check system for leaks
• Tubing not fixed • Missing or defective seal between the canister and Avance Pump • Dressing not sealed properly
beep
✘
beep beep beep
1
Problem description on the display
Example: +
2
+ Therapy number Therapy time
2
press simultaneously
Mute Alarm 60 Sec.
3
• Check that tubing is clear/not twisted • Check canister filling level
305
Battery empty
Charge battery
306
Canister full
Change canister
Pressure is maintained if Avance Pump is in upright position
✔
313
Filter clogged
Change canister and switch Avance Pump on
Filter in the canister is clogged
✘
401
Battery low
Charge battery
Remaining time of battery approx. 30 min.
✔
405
Standby mode
Switch on or off
In standby mode after 5 minutes
✔
XXX
Internal fault
• Switch Avance Pump off and on • Inform Mölnlycke Health Care Customer Service
If fault message repeats, report fault number to Mölnlycke Health Care Customer Service
Fault number
3
Intermittent mode
Constant mode
On
On Graph mode
All further instructions will be the same for constant and intermittent mode
Troubleshooting on the display
✘
System clogged
Enter administrative mode Warning
Same patient
Log File
1
• Tubing is kinked or clogged • Filter in canister contaminated
Pressure
302
Internal fault
New patient
3
Standby Mode
Problem description on the display
3
OK
4
(Operation continues)
✔
Fault number
Alarm (Operation stops)
Warning
1
2
Warnings and Alarms
Change Pressure 1
✘ Operation stops
✔ Operation continues
✘ ✘