PAGE 1 OF 12
THE SPIRIT ULTRA LOW BED Pasa Indeminty Number IFA - 371 It is the duty of all staff who handle the Spirit bed to read the enclosed :-
Guidance Notes For Safe Use
Updated April 07
1
PAGE 2 0F 12
SPIRIT BED – QUICK REFERENCE GUIDE TO OPERATE THE BED: • Use the rail control, patient hand control or the footboard control panel to raise or lower the head section, the knee section, or the entire bed. • The patient controls can be locked or unlocked (to restrict patient use). Lockouts for each bed function are located on the footboard control underneath each function. When the “lock” light is on, that particular function is locked on the hand control. If the “lock” light is off, then the hand control is active. TO MOVE AND LOCK THE BED: • Use the red and green pedal: RED = All four wheels locked GREEN = Foot-end wheels lock into ‘steer’ mode (head-end wheels swivel) NEUTRAL = Middle position unlocks all four wheels
TO OPERATE TRENDELENBURG: • Push and hold the red “Trendelenburg” or green “Reverse Trendelenburg” button on the footboard control to achieve either function. To automatically level the bed, push and hold the opposite function until the bed stops. TO OPERATE ‘CARDIAC CHAIR’: • Push and hold the ‘chair’ button. Green light illuminates when bed angle is attained, then head and knee will raise to full extension. Push and hold button again to remove chair. (Bed will reverse its position each time the ‘Chair’ button is pressed). TO OPERATE ‘INSTANT CPR’: • Push and release the red ‘CPR’ button : head and knee sections will automatically flatten. TO OPERATE ‘AUTO CONTOUR’: • Press the ‘Auto Contour’ button so that the green light is on. Ensure “lock-out” lights are off. Knee section will raise automatically when head section is raised. TO OPERATE ‘QUAD RAILS’: • To lower the rails gently push the black knob and then rotate the rail down. Head end rails rotate toward the headboard; foot end rails rotate toward the footboard. The rail will automatically lock in the ‘up’ or ‘secure’ position. SPIRIT GUIDANCE NOTES
2
PAGE 3 0F 12
PRIOR TO HANDOVER The person supplying and setting up the bed will carry out the following checks:
¾
Visual check of frame and welds
¾
Mattress platform
¾
Castors / Brakes
¾ ¾ ¾ ¾ ¾ ¾ ¾ ¾ ¾ ¾ ¾
Motors which operate Hi-low positions Head section Knee section Trendelenburg / reverse Trendelenburg Mains lead / plug Embedded Patient Controls in Rails Nurse Station (controls at foot of bed) Moulded Quad Rails Moulded Head and Foot board Drop-in Connections Load test (Carried out in the manufacturing plant)
SPIRIT GUIDANCE NOTE
3
PAGE 4 OF 12
THE SPIRIT ULTRA LOW BED Before using the bed:
¾ ¾ ¾ ¾ ¾
Familiarise your self with the various functions Check for obstructions before raising the bed; (i.e. Windowsills) Check for obstructions before lowering the bed. Make sure it is clean Make sure all mattress deck sections are flat before putting a patient on it
Safe working load = 225kgs or 35 stone. (Throughout its travel range) Maximum load on either Quad-Rails = 77kgs or 12 stones. Ensure that mattress used conforms to the bed – RECOMMENDED SIZE IS 35.5 INCHES / 900MM WIDE 80 INCHES/2030MM LONG MINIMUM 6 INCHES HIGH Do not stand on the bed.
THIS BED IS WATERPROOF:
• IT IS ESSENTIAL THAT THE POWER CABLE DOES NOT GET TRAPPED BETWEEN MOVING PARTS. THE LIKELIHOOD OF THIS HAPPENING MAY BE GREATER WHEN THE BED IS PLACED IN ITS LOWEST POSITION.
SPIRIT GUIDANCE NOTES
4
PAGE 5 OF 12
AT THE INSTALLATION The person supplying and setting up the bed will familiarise staff with how it operates. All staff that may come into contact with the bed whether nurses, domestics or porters for example, will need to understand how the bed operates. A FURTHER TRAINING SESSION MAY BE REQUIRED FOR THE ABOVE STANDARD TO BE MET AND IN ORDER TO CONSOLIDATE STAFF COMPETENCE REMEMBER: TRAINING IN THE USE OF EQUIPMENT IS A LEGAL REQUIREMENT The prerequisites for using the Spirit Ultra Low bed are that it has been electrically tested and that the person operating it has had instruction in its use. The bed must also have an up to date inspection record.
The Spirit Plus Ultra Low Bed is designed with a unique Central Lock-and-Steer System. The bed can only be moved when you have disengaged the Central Lock System. ALWAYS UNPLUG THE POWER CABLE FROM THE WALL SOCKET BEFORE MOVING THE BED AND REMEMBER THAT A TRAILING CABLE CAN PRESENT A TRIPPING HAZARD The power cable can also be unplugged from the wall socket at night if for example there were concerns over cables and patient safety. If this were done, the bed could still be operated from the Nurse’s Control Panel at the foot of the bed because it has a back-up power supply. Should staff need at anytime to immobilise the bed completely there is a lockout feature. (see page 8 for details) SPIRIT GUIDANCE NOTES
5
PAGE 6 0F 12
OPERATION
The nurse or patient can operate the bed with the built in patient controls in the Quad-Rails, or only by the nurse from the Nurse’s Control Panel at the foot of the bed. The embedded patient controls in the Quad-Rails, will only function when the orange lights (LED’s) beside each function button (touch pad) at the Nurse’s Control Panel are off. When an orange light underneath each function button is on, then that function cannot be operated by the patient controls. In this way the patient can be prevented from operating the bed. Alternatively (and for safety reasons) it may be necessary to immobilise the bed altogether. The inner patient controls (on Quad-Rails) operate the head section (as in sitting up), the foot section (which slightly flexes the knees) plus a Nurse call alarm button. The embedded outer Nurse control station (on the Quad-Rails) allows for the raising and lowering of the whole bed frame, plus profiling head and knee section.
THE NURSE’S STATION HAS MORE FUNCTIONS
• It has all the functions of the embedded patient controls and can also raise the head, knee and foot sections simultaneously.
• It can be used to immobilise the patient control functions. • The bed can be placed into cardiac chair positioning. • Trendelenburg and reverse Trendelenburg positions can be achieved of 18 decrees tilt.
SPIRIT GUIDANCE NOTES
6
PAGE 7 OF 12
RELEASE THE BUTTON WHEN THE DESIRED TILT POSITION HAS BEEN ACHIEVED. THE BED WILL SELF-LEVEL FROM THE TRENDELENBURG POSITIONS BY PRESSING THE REVERSE TRENDELENBURG BUTTON UNTIL THE BED STOPS. SIMILARLY, TO LEVEL THE BED FROM THE REVERSE TRENDELENBURG POSITION, SIMPLY PRESS AND HOLD THE TRENDELENBURG BUTTON (THE BED WILL STOP MOVING ONCE IT ACHIEVES A LEVEL POSITION.
.
There is also a Cardiopulmonary (CPR) mode, which flattens all deck panels with one touch.
The Quad-Rails The moulded Quad-Rails on either side of the bed DO NOT have the same function as cot-sides / bed-rails. Full-length bed rails can be purchased and fitted to the Spirit bed if required and subject to an appropriate assessment. The Quad-rails are released by gently pressing the black knob, then rotating the rail down. The head end rail rotates toward the headboard, while the foot end rail rotates towards the footboard. The rail will automatically lock into place at its uppermost position. Quad-rails have two positions: 1. 2.
Guard position (horizontal) Transfer position (swing down below level of bed frame)
SPIRIT GUIDANCE NOTES
7
PAGE 8 OF 12
USING THE NURSE’S STATION TOUCH PADS (CONTROLS) To Raise The Head Section To Lower The Head Section To Raise The Knee / Foot Section To Lower The Knee / Foot Section
= = = =
press the top left arrow on the panel press the bottom left arrow on the panel press the top right arrow on the panel press the bottom right arrow on the panel
NOTE: It may be beneficial to elevate the knee / foot sections when the head section has been elevated in order to prevent the occupant from sliding down the mattress. TO RAISE THE HEAD, KNEE AND FOOT SECTIONS SIMULTANEOUSLY = first press the bottom right “Contour” pad / button whereupon a green light will illuminate and then press the diagonally opposite top left hand button (described further on)
TO LEVEL THE BED AGAIN = Press the lower left arrow or press the CPR mode button (Described further on)
TO LOWER THE WHOLE BED FRAME = Press the lower middle arrow on the panel
TRENDELENBURG POSITION = Press the second to the top right RED pad / button on the control panel
REVERSE TRENDELENBURG = Press the GREEN pad / button just below the above button
CARDIOPULMONARY RESUSCITATION (CPR) MODE = Press the top right pad / button just once. Pressing any other button while the CPR function is still lowering the deck panels will cause the CPR to stop.
ACTIVATING TOTAL CONTROL LOCK-OUT = Press and hold for 3 seconds all three lower most lock-out pads / buttons on the control panel until the orange lights come on
IN THIS MODE THE PATIENT CONTROLS AND THE NURSE’S STATION CONTROL PANEL WILL NOT OPERATE AND THE THREE ORANGE LIGHTS WILL PERIODICALLY FLASH. THE C.P.R. MODE, HOWEVER, CAN STILL BE ACTIVATED. TO DEACTIVATE = Press and hold the same three pads / buttons for a further 3 seconds until the yellow lights stop flashing and go out. SPIRIT GUIDANCE NOTES
8
PAGE 9 OF 12
SPIRIT BED MOBILISATION This function enables the bed to be moved. 1. Using your foot, press down on Green section of Central Lock and Steer System, until the pedal is level (“NEUTRAL” position). This will unlock all four casters. (Use this position when moving the bed in small spaces, from side-to-side, etc. Please note that when the NEUTRAL mode is activated properly, the bed should move freely without any unusual noises. If any clicking noises are present in the NEUTRAL mode, stop and check that the pedal is level. Adjust if necessary.) 2. When moving the bed a long distance, press down fully on the Green section of the Central Lock and Steer (“STEER” position.) This will allow four casters to roll, but will prevent the foot end casters from swivelling. (This mode helps to keep the bed straight during transport, turn corners, etc.) 3. Once again to lock the casters, press down on fully Red section of the Central Lock and Steer System. This will prevent all four casters from rolling or swivelling.
DO NOT LEAVE A MOBILISED BED UNATTENDED, ENSURE THAT IT IS STABLE BEFORE LEAVING IT.
THE BED SHOULD BE LEFT IN ITS LOWEST POSSIBLE POSITION WHEN THE BED OCCUPANT IS LEFT UNATTENDED. SPIRIT GUIDANCE NOTES
9
PAGE 10 OF 12
QUICK FAULT FINDING AND SAFETY If the bed is not working, as it should:
• Check that it is plugged in • Check that you understand the controls • Is the Footboard connected into place, or damaged • Are any other leads damaged • Check the settings and lockouts on the Footboard Control Station IF IN DOUBT CALL THE SUPPLIER ON 01865 – 409926 FOR ADVICE. IF THE BED IS OCCUPIED TRANSFER THE PATIENT OUT OF BED UNTIL THE FAULT HAS BEEN RECTIFIED. CLEANING : THE BED CAN BE CLEANED WITH ALL USUAL HOSPITAL CLEANING AGENTS EXCEPT PHENOL- BASED CHEMICALS. REFER TO SUPPLIER IF STEAM-CLEANING IS INDICATED.
SPIRIT GUIDANCE NOTES
10
PAGE 11 OF 12
PERIODIC CHECKS BY NURSING STAFF
• Check integrity of wall plug and cable • Does the Outer Nurse Control Panel, function as it should raising / lowering head section, raising / lowering foot section, raising / lowering the bed.
Does the Nurses Station Control Panel function, as it should:
* * * * * * *
Raising / lowering head section
Raising / lowering foot section
Raising / lowering the bed
Trendelenburg and reverse Trendelenburg
Simultaneous head, Knee / foot positioning
CPR mode
Cardiac Chair Positioning
Do the bed Quad-Rails function, as they should?
. . .
Guard position (up)
Transfer position (below level of bed frame)
Are the rails clean?
SPIRIT GUIDANCE NOTES
11
PAGE 12 OF 12
IMPORTANT THINGS TO REMEMBER
+ + + + + +
Make sure you know how to operate the bed controls Check cable, rail positions and for any obstructions before lowering to floor. Safe working load = 225kgs or 35 stone / 500Lbs Do not concentrate weight on the deck panels and make sure they are level when patient gets onto bed Use the correct type of mattress The rails are not cotsides / bed rails, but full-length bed rails can be purchased and fitted.
CALL 01865 409926 IF YOU NEED ADVICE OR FURTHER INSTRUCTION