Quick Reference Guide
12 Pages
Preview
Page 1
G U I D E
Description and Operation Garment Models
R E F E R E N C E
The Quick Reference Guide is a guideline only and does not replace the User's Manual. Please refer to the User’s Manual for full instructions.
Garment Troubleshooting
Q U I C K
Controller Troubleshooting
COOLING THERAPY SYSTEM
Step-by-Step Operation FRONT VIEW
Handle Control Panel Surface Sensor Socket
Core Sensor Socket
Water-in Quick Coupling Connector
Water-out Quick Coupling Connector
Water Feeder Tube
Water Level Indicator
REAR VIEW
Power Switch
Thumb Captive Screws (4)
3-Prong Power Socket
Air Outlet Grille
RS-232 Connector Rear Cover
6mm Threaded Hole for Equipotentiality Screw Installation Thumb Captive Screws (2)
Set: 1. Place CritiCool unit in the desired location and lock front wheels. 2. Pour cold tap water into water tank up to maximum level (between the two lines). 3. Turn power ON. • Following a short Self Test, the system automatically cools the water • Avoid further connections until patient arrival to the ICU • Place appropriate garment on bed or underneath patient (see Instructions of Use). • Connect Water tubes to garment and to CritiCool unit. The garment automatically fills up! (Lower water level in the reservoir is normal) • Check that clamps on garment are open and a clicking sound is heard.
Wrap: 4. Secure on patient
Cool: • Plug Gray sensor into Core socket and Green sensor into Surface socket of CritiCool unit.
• Once sensors are inserted to patient and valid feedback is obtained, algorithm is activated.
• The CritiCool unit will not Thermo regulate if core sensor is not properly inserted into patient
• Ensure no interference from other factors influence correct continuous temperature feedback readings.
• When feedback is unavailable predefined cooling program is set, an alarm sounds and a message appears. Re-warming: Gradual re-warming is advised (do not exceed more than 2 °C above actual temperature) * For Accidental hypothermia or fast re-warming: switch to Normothermia management mode
CONTROL PANEL
Core Temperature Surface Temperature
Mode Key
CoolingHeating Status System Function Status Energy Transfer to/from Patient Set Point Temperature
Alarm Mute Arrow Keys
Description and Operation
Step-by-Step Operation
Step-by-Step Operation 6. Monitor patient temperature and system messages. Use Arrow keys to select desired mode
Operation Mode
Temperature Graph
Stand-by Mode
Treatment Mode Selection Screen
Garment replacement / *Shut Down 1. Turn OFF CritiCool unit or switch to Stand-by mode. 2. Wait a few seconds and close clamps on garment. 3. Disconnect Water Tubes from garment. 4. Unwrap patient and dispose garment.
• Do not use garment to lift or move the patient. • Ensure garment is fully released. * Disconnect Water Tubes and remove Core and Surface sensors from patient and from CritiCool unit.
• Clean and prepare CritiCool unit and its accessories for future use. • For Disinfections / Sterilization of sensors use CIDEX / ETO only according to hospital protocol.
Controller Troubleshooting An alarm will sound to inform about any problem. The system will assist you in solving the problem.
Water leak between connecting tubes and CritiCool unit
Water tank overfilled
Possible Problem
Action to be Taken
Connecting tubes are not connected properly
Disconnect connecting tubes from machine and connect again
Damage to connecting tubes
Replace connecting tubes
Damage to quick coupling connector
Call Biomedical department
If necessary, drain water tank as follows: 1. Connect one end of the ThermoWrap connecting tube to the Water-out Quick Coupling Connector (beneath Core Sensor socket), or, Connect the gray-coded end of the CureWrap connecting tube to the Water-out Quick Coupling Connector (beneath Core Sensor socket). 2. Connect special male connector to connecting tube: For ThermoWrap, see Figure 1. For CureWrap, see Figure 2. 3. Turn CritiCool device ON. 4. Change to Operation mode. 5. Allow excess water to drain into a receptacle, pail or sink. 6. When desired water level has been reached, turn CritiCool device OFF.
Special Male Connector for Draining Water Tank
Figure 1
Special Male Connector for Draining Water Tank
Figure 2
Controller Troubleshooting
Observation
Controller Troubleshooting Message HALT - PLEASE RESTART
Probable Cause
Action to be Taken
Error during normal function
Turn OFF system for three seconds and then ON again. If problem persists, turn OFF CritiCool unit and contact your local service representative. Note the number that appears on screen
ATTENTION NO WATER – PLEASE ADD WATER
No water in tank
Refill water tank
Water tank float is jammed
Open water tank cap and insert long object to release float
ATTENTION OUT OF NORMOTHERMIA PRESS OK TO CONFIRM
Cooling Therapy Desired set-point is set to exceed 38°C (100.4oF) Normothermia Management Desired set-point is set above 38°C (100.4oF) or below current set point
Approve action
ATTENTION CHECK CORE SENSOR
Misplacement of sensor in core socket
Connect core sensor to appropriate socket
Core adapter is connected to CritiCool unit without disposable temperature sensor
Connect disposable sensor
11
Misplacement of sensor in surface socket
Connect appropriate surface sensor to socket
Surface adapter is connected to CritiCool unit without disposable temperature sensor
Connect disposable sensor
XX
ATTENTION CHECK SURFACE SENSOR 12
- Indicates alarm activated
Disapprove action
Controller Troubleshooting
ATTENTION CONNECT CORE SENSOR*
ATTENTION WATER TEMP TOO LOW
Probable Cause No Core Sensor inserted in socket
Water Temperature < 10oC (50.0F)
16 ATTENTION WATER TEMP TOO HIGH
Water Temperature > 42oC (107.6F)
17 ATTENTION CORE READ OUT TOO LOW 18 Patient temp ABOVE 38.5ºC (101.3ºF)**
Action to be Taken Connect Core Sensor
Turn OFF system for three seconds and turn ON again. If problem persists, turn OFF CritiCool unit and contact your local service representative Turn OFF system for three seconds and turn ON again. If problem persists, turn OFF CritiCool unit and contact your local service representative
Cooling Therapy: Core temperature is more than 2°C below set point Normothermia management: Core temperature is less than 27.0ºC (80.6ºF)
Re-warming: Do not attempt to increase more than 2 ° C above actual core temperature Insert core sensor into patient
Core temperature reading above 38.5°C (101.3°F)
Inform physician
- Indicates alarm activated * The alarm is activated again after 30 minutes if no action is taken. ** Pressing Alarm Mute silences buzzer for this present condition and any re-occurrence of this condition for thirty minutes. ATTENTION ELECTRO-MAGNETIC INTERFERENCE 22
The CritiCool unit detects electromagnetic field higher than allowable standard
The message will disappear after ten seconds If message appears for more than ten seconds, replace Core Sensor
ATTENTION CONNECT SURFACE SENSOR
No Surface Sensor inserted into socket
Connect Surface Sensor to socket
Controller Troubleshooting
Message
Garment Models
°CureWrap
Adult Model
Patient Height cm (in.)
3500
Fits All Adult Sizes
Pediatric ThermoWrap
Pediatric Model
Patient Height cm (in.)
3131
79 – 91 (31 – 36)
3136
91 – 104 (36 – 41)
3141
104 – 122 (41 – 48)
3148
122 – 135 (48 – 53)
Garment Models
Infant ThermoWrap Infant °CureWrap
Infant Model
Weight Kg (Lb)
3118 / 3518*
2.5 – 4.0 (5.5 – 8.8)
3121 / 3521*
4.0 – 7.0 (8.8 – 15.4)
3125 / 3525*
7.0 – 11.0 (15.4 – 24.2)
*NEW Infant °CureWrap models
Garment Models
POSITIONING THE PATIENT
°CureWrap
Pediatric ThermoWrap Indicates Head Positioning
Infant ThermoWrap
-------- Indicates Armpit Line
•The garment should be filled with water prior to wrapping and securing to patient. •Do not use garment to lift or move patient. •Avoid handling garment with sharp objects. •Make sure garment is not folded or clamped. •Before moving patient from the operating table, ensure garment is fully released.
Garment Troubleshooting An alarm will sound to inform about any problem. The system will assist you in solving the problem. Observation Garment leak
Possible Problem
Action to be Taken
Accidental puncture of garment during course of operation
Turn OFF CritiCool unit and allow water to return to reservoir Replace garment if possible
Water leak from connector between garment and connecting tube
Connecting tubes are not connected properly
Close garment clamps. Disconnect connecting tubes and connect again Ensure click sound is heard
ATTENTION Check ThermoWrap
Damage to connecting tubes
Replace connecting tubes
Connecting tubes are not connected
Connect connecting tubes
Garment is blocked due to incorrect wrapping
Check for snarls, folds, or objects that obstruct water flow in garment
Garment clamps are closed
Check clamps
10
- Indicates alarm activated
QUICK REFERENCE GUIDE Catalog No. 014-00082