Quick Reference Guide
76 Pages
Preview
Page 1
TM
MVE Variō™ Pro
Quick Reference Guide Multilingual
PN 14910831 Rev A
MVE Biological Systems TM
MVE Variō™ Pro Quick Start Guide
Table of Contents
1.0 Product Identification...2 1.1 The Variō Pro Controller...2 1.1.1 Display Panel Identification...2 1.1.2 Connection Panel Identification...3 2.0 MVE Variō Plumbing Connections...4 3.0 Adjusting Temperature Settings...5 3.1 Adjusting Storage Chamber Temperature...5 3.2 Adjusting Temp A & B High / Low Alarms...6 4.0 Adjusting Display & Output Settings...7 4.1 Adjusting Temperature Units...7 4.2 Alarm Buzzer...7 4.3 Language Settings...8 4.4 Printer Settings...8 5.0 Calibration of Temperature Probes...9 6.0 Password & Security Setup...10 6.1 Password Entry Mode...10 6.2 Global Password...10 6.3 Multilevel Passwords...11 7.0 Variō Pro Alarms & Alarm Descriptions...12 8.0 Contact Information...14
NOTE: All MVE models are a Class 1, externally powered, continuous operation medical device. They are not suitable for use with flammable anesthetics. This equipment has been tested and found to comply with the limits for medical devices to IEC 601-1-2: [or EN 60601-1-1-2:2001 or Medical Device Directive 93/42/EEC]. NOTE: MVE Variō liquid nitrogen freezers should be installed by an authorized MVE Distributor per the MVE Variō Pro Technical Manual, PN 14930250.
PN 14910831 Rev A
Page 1
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
1.0 Product Identification 1.1 The Variō Pro Controller 1.1.1 Display Panel Identification
Display Panel Identification Key Front Panel
The front panel is the user interface for the Variō Pro. All displays and controls are located on the front panel.
Display
A 4 x 20 Liquid Crystal Display (LCD) provides the user with all of the pertinent information associated with the selected menu item. The display also shows any current alarm conditions that may exist.
FOG CLEAR
Used to manually clear the fog from the storage area to increase visibility.
END CLEAR
Used to manually end the fog clearing process.
Used to enter and navigate the setup menus. ALARM MUTE
Used to silence the audible alarm. Also used to reset the latching alarms after the alarm condition has been corrected. Used to increase numeric values and toggle YES/NO or ENABLED/DISABLED settings in the setup menus. Press once to increase incrementally. Hold the button down to scroll quickly. Used to decrease numeric values and toggle YES/NO or ENABLED/DISABLED settings in the setup menus. Press once to decrease incrementally. Hold the button down to scroll quickly.
ESC
Used to exit any menu or cancel any menu operation.
ENTER
Used to select any menu for editing or save any user setting.
PN 14910831 Rev A
Page 2
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
1.1.2 Connection Panel Identification
Connection Panel Identification Key Temp A
Platinum RTD connection for primary sensor
Temp B
Platinum RTD connection for auxiliary sensor
Serial I/O
RS-485 input/output connection
Manual Fill
Manual override button, forces cooling valves open
Sensor/Alarm Output
Global Remote
Main Wire Harness
PN 14910831 Rev A
Output connection for the remote monitoring of Temp A & various alarm conditions. Output connection for the remote monitoring of all alarm conditions. Any Vario Pro system alarm will cause the normally closed (NC) and normally open (NO)contacts to switch state. These connections may be wired to any remote monitoring system utilizing NC or NO contacts. Connection for main wire harness. All system components are wired through this connection, with the exception of Temp A, Temp B, and the main power supply.
Page 3
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
2.0 MVE Variō Plumbing Connections Connect a transfer line (included with freezer) from an LN2 supply tank to the fill connection at the rear of the freezer. Optimum supply tank pressure is 22 to 35 psi (1.5 to 2.4 bar). Although the plumbing assembly has a 50 psi (3.45 bar) pressure relief device, it is recommended that the supply tank be pressurized below 35 psi (2.4 bar) to reduce the LN2 “flash-off” rate during cooling cycles and to maximize the cryogenic valve life. The supply line can be insulated to minimize LN2 transfer losses. After the transfer hose is securely coupled to the freezer and supply tank, ensure all connections are leak free by opening the valve of the LN2 supply tank and applying a soap and water solution to each field joint. You should not see bubbles forming at any joint. Wipe away excess soap and water when finished. Before removing the transfer hose, ensure the LN2 supply tank valve is closed. Slowly and carefully loosen the transfer hose connection to vent any remaining pressure in the line before disconnecting the hose.
PN 14910831 Rev A
Page 4
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
3.0 Adjusting Temperature Settings The following section describes how to adjust the MVE Variō system temperature and alarm settings. At any time, the user may exit the current menu by pressing the ESC button repeatedly until the display returns to the main screen. After 30 seconds of inactivity, the controller will automatically exit the current menu and return to the main screen. 3.1 Adjusting Storage Chamber Temperature 1. From the home screen, press SETUP to enter setup menus. If Password Entry Mode is enable, the controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position. Note: The default user password is 3 4 5 6.
Higher User Level Required use ▲/▼ to ENTER Password 0000
2. After entering password, the first menu item is Temperature Menus. Press ENTER.
Press ENTER for Temperature menus or press SETUP for next menu
3. Press SETUP to scroll to the Cooling Menu display and press ENTER.
Press ENTER for Cooling menu or press SETUP for next menu
4. The first menu item will display the current temperature of the cooling temp sensor. Press SETUP to continue.
Current Cooling Temp -60.0°C
5. To adjust the maximum (warmest) allowable chamber temperature, use the ▲/▼ keys to select the desired value and press ENTER to save. Press SETUP to advance to the next screen.
Max Chamber Temp -80.0°C Use ▲/▼ to adjust Note: The Variō Pro controller will not allow the user to select a maximum chamber temp greater than Press ENTER to save
or equal to the high temperature alarm setpoint. As a result, it may be necessary to adjust the high temperature alarm prior to changing the maximum chamber temp. See the following section for further instruction.
6. To adjust the chamber temperature deadband, use the ▲/▼ keys to select the desired value and press ENTER to save. Note that the chamber deadband also defines the minimum (coldest) allowable chamber temperature. Note: The Variō Pro controller will not allow the user to select a maximum chamber temp ▲/▼ chamber deadband combination that results in a minimum chamber temp less than or equal to the low temperature setpoint. As a result, it may be necessary to adjust the low temp alarm prior to changing the chamber deadband.
7. Press SETUP repeatedly to scroll to the Cooling Time Alarm display. For the initial cooling cycle, it will be necessary to disable the alarm. Use the ▲/▼ keys to select DISABLED and press ENTER to save. Press ESC repeatedly to return to the home screen.
Chamber Deadband 10.0°C Use▲/▼ to adjust Press ENTER to save
Cooling Time Alarm ENABLED Use ▲/▼ to adjust Press ENTER to save
Note: It is recommended that the Cooling Time Alarm be re-enabled after the initial cooling cycle is complete. Follow the steps above to return to the Cooling Menu and set the Cooling Time Alarm to ENABLED.
PN 14910831 Rev A
Page 5
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
3.2 Adjusting Temp A & B High / Low Alarms 1. From the home screen, press SETUP to enter setup menus. If Password Entry Mode is enable, the controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position.
Higher User Level Required use ▲/▼ to ENTER Password 0000
Note: The default user password is 3 4 5 6.
2. After entering password, the first menu item is Temperature Menus. Press ENTER.
Press ENTER for Temperature menus or press SETUP for next menu
3. The first menu item is Temp A Menu. For Temp B, press SETUP to scroll to Temp B Menu. Press ENTER.
Press ENTER for Temp A/B menu or press SETUP for next menu
4. Press SETUP repeatedly to scroll to the Temp A/B High Alarm setpoint screen and use the ▲/▼ to select the desire value and press ENTER to save. Press SETUP to continue.
Temp A/B High Alarm -75.0°C Use ▲/▼ to adjust Press ENTER to save
5. To adjust the Low Temperature Alarm setpoint, use the ▲/▼ to select the desire value and press ENTER to save. Press ESC repeatedly to return to the home screen.
Temp A/B Low Alarm -95.0°C Use ▲/▼ to adjust Press ENTER to save
PN 14910831 Rev A
Page 6
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
4.0 Adjusting Display & Output Settings 4.1 Adjusting Temperature Units 1. From the home screen, press SETUP to enter setup menus. If Password Entry Mode is enable, the controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position.
Higher User Level Required use ▲/▼ to ENTER Password 0000
Note: The default user password is 3 4 5 6.
2. After entering password, press SETUP to scroll through the menu items. When the Display and Output screen appears, press ENTER.
Press ENTER for Display and Output or press SETUP for next menu
3. The first menu item is the Temperature Units screen. Use the ▲/▼ keys to select either °C, °F, or K. Press ENTER to save.
Temperature Units °C Use ▲/▼ to adjust Press ENTER to save
4.2 Alarm Buzzer 1. From the home screen, press SETUP to enter setup menus. If Password Entry Mode is enable, the controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position.
Higher User Level Required use ▲/▼ to ENTER Password 0000
Note: The default user password is 3 4 5 6.
2. After entering password, press SETUP to scroll through the menu items. When Press ENTER for the Display and Output screen appears, press ENTER. Display and Output or press SETUP for next menu 3. Press SETUP to scroll to the Advanced Display and Output menu and press ENTER.
Press ENTER for Advanced Display and Output or press SETUP for next menu
4. The first menu item is the Alarm Buzzer menu. Use the ▲/▼ keys to enable/disable the audible alarm buzzer. Press ENTER to save.
Alarm Buzzer ENABLED Use ▲/▼ to adjust Press ENTER to save
PN 14910831 Rev A
Page 7
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
4.3 Language Settings 1. From the home screen, press SETUP to enter setup menus. If Password Entry Mode is enable, the controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position.
Higher User Level Required use ▲/▼ to ENTER Password 0000
Note: The default user password is 3 4 5 6.
2. After entering password, press SETUP to scroll through the menu items. When the Display and Output screen appears, press ENTER.
Press ENTER for Display and Output or press SETUP for next menu
3. Press SETUP to scroll to the Advanced Display and Output menu and press ENTER.
Press ENTER for Advanced Display and Output or press SETUP for next menu
4. Press SETUP scroll to the Language menu. Use the ▲/▼ keys to select either English, French, Italian, German, or Spanish. Press ENTER to save.
Language ENGLISH Use ▲/▼ to adjust Press ENTER to save
4.4 Printer Settings 1. From the home screen, press SETUP to enter setup menus. If Password Entry Mode is enable, the controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position.
Higher User Level Required use ▲/▼ to ENTER Password 0000
Note: The default user password is 3 4 5 6.
2. After entering password, press SETUP to scroll through the menu items. When the Display and Output screen appears, press ENTER.
Press ENTER for Display and Output or press SETUP for next menu
3. Press SETUP to scroll to the Advanced Display and Output menu and press ENTER.
Press ENTER for Advanced Display and Output or press SETUP for next menu
4. Press SETUP to scroll to the Printer menus and press ENTER.
Press ENTER for Printer menus or press SETUP for next menu
5. The first menu item is the Print Interval menu. Use the ▲/▼ keys to adjust setting to the desire value and press ENTER to save.
Print interval 30 min Use ▲/▼ to adjust Press ENTER to save
PN 14910831 Rev A
Page 8
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
5.0 Calibration of Temperature Probes Each of the temperature sensor probes used with the Variō Pro have been calibrated at the factory using the Low Temperature Range method. This calibration method provides a level of accuracy of +/-1.8°F or +/-1°C when operated at an altitude range of 1000ft – 1500ft (305m - 457m). Further calibration should not be required unless desired by the user. Refer to the Variō Pro Technical Manual (PN 14930250) for information on calibration methods and procedures.
PN 14910831 Rev A
Page 9
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
6.0 Password & Security Setup 6.1 Password Entry Mode 1. From the home screen, press SETUP to enter setup menus. If Password Entry Mode is enable, the controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position.
Higher User Level Required use ▲/▼ to ENTER Password 0000
Note: The default user password is 3 4 5 6.
2. After entering password, press SETUP to scroll through the menu items. When the Password Menus screen appears, press ENTER.
Press ENTER for Password menus or press SETUP for next menu
3. The controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position.
Higher User Level Required use ▲/▼ to ENTER Password 0000
4. The first menu item is the Password Entry Mode display. Use the ▲/▼ keys to enable/disable password entry mode and press ENTER to save.
Password Entry Mode ENABLED Use ▲/▼ to adjust Press ENTER to save
6.2 Global Password 1. From the home screen, press SETUP to enter setup menus. If Password Entry Mode is enable, the controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position.
Higher User Level Required use ▲/▼ to ENTER Password 0000
Note: The default user password is 3 4 5 6.
2. After entering password, press SETUP to scroll through the menu items. When the Password Menus screen appears, press ENTER.
Press ENTER for Password menu or press SETUP for next menu
3. The controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position.
Higher User Level Required use ▲/▼ to ENTER Password 0000
4. Press SETUP to scroll to the Change Global Password display and press ENTER.
Press ENTER to Change global Password or press SETUP for next menu
5. Use the ▲/▼ keys to scroll to the desired number and press ENTER to advance the cursor to the next position. When complete, press ENTER to save.
6. At the Confirm New Password? display, use the ▲/▼ keys to select either YES to confirm, or NO to cancel. Press ENTER to save.
PN 14910831 Rev A
Global Password Use ▲/▼ to adjust Press ENTER for next XXXX Confirm new Password? NO Use▲/▼ to adjust Page 10
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
6.3 Multilevel Passwords 1. From the home screen, press SETUP to enter setup menus. If Password Entry Mode is enable, the controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position. Note: The default user password is 3 4 5 6.
Higher User Level Required use ▲/▼ to ENTER Password 0000
2. After entering password, press SETUP to scroll through the menu items. When the Password Menus screen appears, press ENTER.
Press ENTER for Password menus or press SETUP for next menu
3. The controller will prompt for a password. Use the ▲/▼ keys to scroll to the appropriate number and press ENTER to advance the cursor to the next position.
Higher User Level Required use ▲/▼ to ENTER Password 0000
4. Press SETUP to scroll to the Change Password 1 display and press ENTER.
Press ENTER to Change password 1 Password or press SETUP for next menu
5. Use the ▲/▼ keys to scroll to the desired number and press ENTER to advance the cursor to the next position. When complete, press ENTER to save
Password 1 level Use ▲/▼ to adjust Press ENTER for next XXXX
6. At the Password 1 Level display, use the ▲/▼ keys to select the desired access level for password 1.
Password 1 level Use ▲/▼ to adjust Press ENTER for next XXXX
7. At the Confirm New Password? display, use the ▲/▼ keys to select either YES to confirm, or NO to cancel. Press ENTER to save.
Confirm new Password? NO Use▲/▼ to adjust
Security Access Levels Level 1
- Fog Clear - End Clear - Alarm Mute - Change Display Units
PN 14910831 Rev A
Level 2
Level 1 + : - Temp Settings - Time & Date Settings - Calibration Menus - Language Selection - Hot Gas Bypass Menus
Level 3
Level 4
Level 2 + : Level 3 + : - Communications Settings - Password Settings - Programming
Page 11
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
7.0 Variō Pro Alarms & Alarm Descriptions The MVE Variō Pro controller is equipped with a total of 15 system status alarms, all of which may be customized by the user to suit the needs of the application. The various alarm indication screens and the corresponding descriptions are shown below:
Alarm Display
Description
ALARM High Temp A Started at: 01/01/2011 12:00 PM
The Temp A sensor is reading above the user defined high temperature alarm setting.
ALARM High Temp B Started at: 01/01/2011 12:00 PM
The Temp B sensor is reading above the user defined high temperature alarm setting.
ALARM Low Temp A Started at: 01/01/2011 12:00 PM
The Temp A sensor is reading below the user defined low temperature alarm setting.
ALARM Low Temp B Started at: 01/01/2011 12:00 PM
The Temp B sensor is reading below the user defined low temperature alarm setting.
ALARM Temp A Sensor Fail Started at: 01/01/2011 12:00 PM
The Variō Pro has lost communication with the Temp A sensor. The sensor has either been disconnected from the controller or has been damaged.
ALARM Temp B Sensor Fail Started at: 01/01/2011 12:00 PM
The Variō Pro has lost communication with the Temp B sensor. The sensor has either been disconnected from the controller or has been damaged
ALARM Inlet Sensor Fail Started at: 01/01/2011 12:00 PM
The Variō Pro has lost communication with the inlet sensor. The sensor has either been disconnected from the controller or has been damaged.
ALARM Cooling Sensor Fail Started at: 01/01/2011 12:00 PM
The Variō Pro has lost communication with the cooling sensor. The sensor has either been disconnected from the controller or has been damaged.
ALARM Supply Time X min Started at: 01/01/2011 12:00 PM
The Supply Time alarm is triggered when the temperature of the incoming nitrogen fails to reach the Inlet Temp Setpoint within the time specified by the Supply Alarm Delay. Once activated, the Variō Pro will terminate the Hot Gas Bypass cycle and begin supplying the freezer with N2.
ALARM Valve Stuck Open Started at: 01/01/2011 12:00 PM
The Valve Stuck Open alarm is triggered when the Cooling Cycle has terminated and the Inlet Temp Sensor fails to warm above the Inlet Temp Setpoint within the time specified by the Stuck Open Delay time setting.
ALARM Valve Stuck Closed Started at: 01/01/2011 12:00 PM
The Valve Stuck Closed alarm is triggered when the Variō Pro has begun feeding N2 to the freezer, and the Inlet Temp Sensor fails to cool to the Inlet Temp Setpoint in the time specified by the Stuck Closed Delay time setting. Note that although this alarm is similar to the Supply Time Alarm, it will remain active regardless of the ENABLED/DISABLED status of the Inlet Temp Sensor.
PN 14910831 Rev A
Page 12
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
Alarm Display
Description
ALARM Lid Open X min Started at: 01/01/2011 12:00 PM
This alarm is triggered when the Lid Switch Installed setting is set to YES and the freezer lid remains open for the duration of the time specified by the Lid Open Alarm Delay. Note that by default, the Lid Switch Installed setting is set to NO.
ALARM Cooling Time X min Started at: 01/01/2011 12:00 PM
The Cooling Time Alarm is triggered when the Variō Pro has initiated a Cooling Cycle and the Chamber Temp fails to reach the minimum chamber temp setpoint in the amount of time specified by the Cooling Time Alarm Delay.
ALARM Power Failure Started at: 01/01/2011 12:00 PM
The Power Failure Alarm is triggered when the Variō Pro is equipped with battery backup and has been running on battery power for 30 minutes. With typical use, a battery backup equipped Variō Pro will retain functionality for 48 hours after the loss of its primary power source.
ALARM Low Battery Started at: 01/01/2011 12:00 PM
The Low Battery Alarm is triggered when the voltage of the battery backup powering the Variō Pro drops below 21VDC.
Communications Loss Unit 1 Started/Ended at: 01/1/11 12:00
The Communications Loss screen is displayed in the event that the Display Screen has been disconnected from the Variō Pro control board. This screen does not necessarily indicate a loss of system functionality, but it does impair the user’s ability to monitor the system status.
If any alarms occur, contact your authorized MVE Distributors or customer/technical service. North & South America Toll Free 1-800-482-2473 Fax: 1-888-932-2473 Japan
Phone: +(03) 5776 2670 Fax: +(03) 5776 2676
Asia, Australia, & Pacific Rim Phone: +61 (2) 974 94333 Fax: +61 (2) 974 94666 Europe
PN 14910831 Rev A
Phone: +44 (0) 1344 403100 Fax: +44 (0) 1189 799245 Page 13
MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM
8.0 Contact Information Customer / Technical Service: North & South America Chart BioMedical 2200 Airport Industrial Drive, Suite 500 Ball Ground, GA 30107 Phone: 770-721-7759 Toll Free 1-800-482-2473 Fax: 770-721-7758 or 1-888-932-2473 Email: [email protected] [email protected] Japan Chart Japan Co., Ltd. Phone: +(03) 5776 2670 Fax: +(03) 5776 2676 Email: [email protected] Asia, Australia, & Pacific Rim Chart Australia Pty Ltd. Phone: +61 (2) 974 94333 Fax: +61 (2) 974 94666 Email: [email protected] Europe Chart BioMedical Ltd. Phone: +44 (0) 1344 403100 Fax: +44 (0) 1189 799245 Email: [email protected] [email protected]
PN 14910831 Rev A
Page 14