UltraPro S100 Hardware Installation
UltraPro S100 System Hardware Installation Instructions These instructions describes how to assemble and set up your UltraPro S100 system
Contacts Technical Support Natus Neurology Incorporated 3150 Pleasant View Road Middleton, WI USA 53562-3530 608-829-8500 1 800-356-0007 Fax: 608-829-8589 www.Natus.com Natus EU Authorized Representative Natus Manufacturing Limited IDA Business Park Gort, Co.Galway, Ireland
Domestic Natus Neurology Incorporated 3150 Pleasant View Road Middleton, WI USA 53562 1-800-356-0007 [email protected] www.Natus.com International Natus Neurology Incorporated Phone: 0049 (0) 180 501 5544 Fax: 0049 (0) 89 83942777 Email: [email protected] www.Natus.com
269-653600 Rev 05 © 2013 - 2018 Natus Medical Incorporated or one of its subsidiaries. All rights reserved. Natus is a registered trademark of Natus Medical Incorporated. All product names appearing on this document are trademarks or registered trademarks owned, licensed to, promoted or distributed by Natus Medical Incorporated, its subsidiaries or affiliates. All other trademarks are the property of their respective owners.
12/11/18
1
Natus Neurology Incorporated
Blank page.
2
12/11/18
UltraPro S100 Hardware Installation
Labels and symbols The following labels and symbols may be affixed to the UltraPro S100 system: When applied on device: Attention: Consult Accompanying Documentation (ISO 7000-0434A) When used in documentation: Caution, Warning or Precaution follows. Consult Operating Instructions. Failure to follow operating instructions could place the patient or operator at risk (ISO 7010 M002). Image on blue background. Consult Operating Instructions. (ISO 7000-1641) European Authorized Representative
Manufacturer
Disposal at end of operating life instructions. When the equipment comes to the end of its operating life, it should be disposed of in accordance with local waste regulation authority, which is typically within the local government office. CE Mark and Notified Body (Compliant to Medical Device Directive 93/42/EEC) Type BF equipment.
RX Only
CAUTION: USA Federal law restricts this device to sale or on the order of a licensed medical practitioner.
Safety summary In these instructions, two labels identify potentially dangerous or destructive conditions and procedures:
The WARNING label identifies conditions or practices that may present danger to the patient and/ or user.
The CAUTION label identifies conditions or practices that could result in damage to the equipment. NOTE: Notes help you identify areas of possible confusion and avoid potential problems during system operation.
12/11/18
3
Natus Neurology Incorporated
System Components Base unit and laptop
3 and 4 channel amplifiers
3 channel amp
4 channel amp
Electrical stimulator probe
HSLink (High Speed Link) cable
USB cable
Power adapter and IEC power cable
System power cable
4
12/11/18
UltraPro S100 Hardware Installation
Cabling the UltraPro S100 System 1.
Connect all of the signal interface cables (USB / HS-Link) as shown in the Version #1/Version #2 illustrations on the following two pages.
Version #1: Amp port on right side panel.
Version #2: Amp port on rear panel.
2.
Connect the system power interface cables.
3.
Connect the power cord to the MAINS (wall outlet).
Use only optional devices specified by Natus Neurology Incorporated in order to comply with IEC 60601-1 (IEC 60601-1-1). NOTE: Always use shielded power line cables from Natus Neurology Incorporated to avoid hum line interference, especially near the patient or the amplifier. NOTE: Make sure the device connected to the MAINS (wall power) outlet is positioned for easy disconnection. NOTE: Unplugging the power cable from the MAINS input on the Base Unit or optional ISO Box disconnects the mains power from the entire UltraPro S100 system.
12/11/18
5
Natus Neurology Incorporated
Version #1 cabling
The Version #1 amplifier port is located on the right side of the Base Unit.
Version #1a: UltraPro S100 system without the optional ISO box
Version #1b: UltraPro S100 system with the optional ISO box
Item Description 1 USB 2.0 cable (2M) 2 High-Speed Link cable (proprietary) 3 Power adapter cable, IEC, (1ft) 4 Advanced Stimulus probe or optional Comfort probe RS10 (not shown) 5 Hospital grade power cable, shielded 6 Power jumper cable, IEC (2M)
6
12/11/18
UltraPro S100 Hardware Installation
Version #2 cabling
The Version #2 amplifier port is located on the rear panel of the Base Unit.
Version #2a: UltraPro S100 system without the optional ISO box
Version #2b: UltraPro S100 system with the optional ISO box
Item Description 1 USB 2.0 cable (2M) 2 High-Speed Link cable (proprietary) 3 Power adapter cable, IEC (1ft) 4 Advanced Stimulus probe or optional Comfort probe RS10 (not shown) 5 Hospital grade power cable, shielded 6 Power jumper cable, IEC (2M)
12/11/18
7
Natus Neurology Incorporated
Mounting the components on the optional carts Two optional models of the carts are available for use with the UltraPro S100 system. Note: Cart options may not be available in all regions..
UltraPro UB3 cart.
8
UltraPro S100 cart.
12/11/18
UltraPro S100 Hardware Installation
Mounting the components on the optional UltraPro UB3 cart
UltraPro S100 system with the optional UltraPro UB3 cart.
Mounting the optional amplifier arm on the UltraPro UB3 cart
12/11/18
The amplifier arm may be mounted on either side at the front of the cart. 1.
Insert the amplifier post into one of the two the mounting holes shown below.
2.
Insert the amplifier arm into the mounting post.
3.
Tighten the locking knob.
9
Natus Neurology Incorporated
Mounting the optional ISO box on the UltraPro UB3 cart
1.
Place the ISO box on the top shelf of the cart, just below the work surface.
2.
Secure the ISO box with 2 screws and washers from under the shelf.
Connecting the ground strap on the UltraPro UB3 cart without the optional ISO box
1.
Connect the smaller connector of the ground cable to the UltraPro UB3 cart.
2.
Connect the larger connector of the ground cable to the Base Unit.
10
12/11/18
UltraPro S100 Hardware Installation
Mounting the components on the optional UltraPro S100 cart
UltraPro S100 system with the optional UltraPro S100 cart.
Mounting the optional amplifier arm on the UltraPro S100 cart
12/11/18
The optional amplifier arm can mount at the rear of the cart. 1.
Insert the amplifier arm into the amplifier arm holder on the cart.
11
Natus Neurology Incorporated
Mounting the ISO box on the UltraPro S100 cart (optional)
12
1.
Insert the ISO box into the ISO box cage from the rear of the cart.
2.
Secure the ISO box with 4 thumb screws from the bottom of the cart.
12/11/18