THE
Clifton RANGE
Stirred Digital Baths NE4-D and NE4-D/CT Series Circulator Digital Baths NE4-P and NE4-HT Series
This series of water baths features a low liquid level float switch and if the liquid drops below the recommended level heating and stirring is automatically switched OFF and a “FIL”message appears on display - top up liquid level for heating and stirring to resume.
IMPORTANT: ALWAYS FILL the water bath before connecting to the power supply.
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Contents About this Manual General Notes Amendments Symbols Location Unpacking Safety Assembly Power Supply Lead and Connection to Electrical Supply Front and Rear Views - NE4D and NE4-D/CT Front and Rear Views - NE4P and NE4-HT Liquid Level Suitable Liquids NE4-P and NE4-HT Series Connecting Circulated Supply Operating Instructions - Switching On and Off - Heating Control Modes Control Panel -Keyboard Description - LED Indicators Single Point Temperature Control - Setting Temperature Set Point 1 Process Control with Two Temperature Settings and Timer - Setting Temperature Set Point 2 Setting Time - Process Control Operation Explanation of Temperature Control Terms Clifton Optimized Control When best to use Optimization Optimization Setting Under and Over-temperature Alarms Safety experiment protector Low Liquid level Float switch - all models - “FIL” message Periodical check General Cleaning Cleaning external painted surfaces with Anti-bacterial finish Cleaning the stainless steel tank Descaling the stainless steel tank Decontamination of Equipment Virucidal Disinfectant Disinfectant/Sterilant 3 Year Warranty Out of Warranty End of Life Portable Appliance Testing Accessories Fitting Accessories Spares/Service Diagram - NE4-D and NE4-P Spares/Service Diagram - NE4-HT Specification NE4-D Specification NE4-D/CT Specification NE4-P Specification NE4-HT Specification NE4-25D Declaration of Conformity Final Inspection and Electrical Safety Test Report
The Clifton Range - 3 Year Warranty
2
3 3 3 3 4 4 4 4 4 5 6 7 7 7 8 8 8 8 8 8 9 9 9 9 10 10 11 12 13 14 15 15 15 15 16 16 16 17 17 17 17 18 18 18 18 19 20 21 22 23 24 25 26 27 28 Enclosed Rear
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
About this Manual This user Manual contains instructions which must be followed in order that the product is operated correctly. General Notes 1. Fill the tank prior to connection to power supply. 2. Connect only to a power supply with the corresponding voltage to that specified on the rating label positioned on the rear of the unit. 3. Ensure the power supply has a safety earth (ground) terminal. 4. Ensure the mains switch and power supply connector are accessible during use. 5. The mains supply cord fitted to this products is a heat resistant type and should be replaced by an equivalent type. 6. Do not block ventilation slots during use. 7. Always follow good laboratory practice by ensuring substances being heated present no risk of a hazard (explosion, implosion or release of toxic or flammable gases) or that these have been addressed. When heating substances where liberation of gases occurs suitable extraction should be used. 8. Use only liquids specified in this Instruction Manual within their specified temperature range. If the alarm lamp is illuminated the liquid temperature may be above its recommended maximum - do not inhale the vapours. 9. Use caution when topping up or draining the tank as the liquid in the tank may be very hot or cold. 10. Drain before moving the bath. Allow the liquid to cool to 50°C or heat up to 10°C or above before draining. 11. In the event of the over or under temperature alarm being illuminated do not touch the liquid as it may be very hot or cold. Use caution when refilling as a hot heating element can create scalding steam. 12. Always use a lid or layer of polypropylene spheres when operating at temperatures above 60°C. Take care when lifting the lid. Steam and hot vapours can cause scalding. 13. Always use the display or a thermometer to check the temperature - do not touch liquid. 14. If this product is not used in accordance with these instructions, then basic safety protection afforded by the water bath may be affected. 15. Check the operation of the over temperature device regularly. 16. If pumping to an external device ensure connecting pipe or hoses are suitable for the liquid and operating temperature. Ensure connections are secure. 17. Ensure pipes or hoses are empty before disconnecting. 18. Before using any cleaning or decontamination method except those recommended, check with your distributor that the proposed method will not damage the equipment. Amendments Issue 2 Issue 3 Issue 4 Issue 5 Issue 6 Issue 7 Issue 8 Issue 9 Issue10 Issue 11 Issue 12 Issue 13 Issue 15 Issue 16 Issue 17 Issue 18 Issue 19 Issue 20 Issue 21
September June April May May July June September December January August September August January November December March November June
1992 1993 1995 1995 1996 1996 1997 1997 1997 1998 1998 2001 2002 2004 2004 2004 2006 2006 2008
MK2 Thermostirrer, revised instruction book. Digital display version introduced. Revised range DC and D merged. EMC approved. Address update. NE4-MT added to range. Bridge assy modified and SSR. NE4-25 added and pump ver included. NE4's common bridge assy. Adaptor plate fitted making common fitting point. Multi-lingual instruction book. Addition NE4-D/CT range+layout updates. Improvements D+T, cover revision. Product Improvement PID controller: time and temperature control Product Improvement: Liquid Level Amendment to service dagram WEEE and Product improvement: Opt feature Anti-bacterial paint finish Improvement with n ew “FIL” message feature.
Symbols HOT SURFACES Paragraphs marked by this symbol indicate that a potential hazard to your personal safety exists from heated surfaces or other appendages on the outside or inside of the equipment. CAUTION This icon accompanies text and/or other international symbols dealing with potential damage to equipment. When present, it indicates that there is a potential danger of equipment damage may occur if information stated within the "CAUTION" paragraph is not adhered to or procedures are executed incorrectly. PROTECTIVE EARTH OR GROUND TERMINAL Protective earth conductor terminal.
The Clifton Range - 3 Year Warranty
3
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Location The product must be placed on a smooth, level and sturdy work surface, preferably near a sink or drain for emptying. Use in a ventilated room. Suitable for use in ambient temperatures 5°C to 40°C with a maximum humidity 80% (temperature 31°C) decreasing to 50% (temperature 40°C). The product is designed for laboratory use. Unpacking Remove the product from its packaging. Any damage please notify your dealer immediately. Retain the packaging over the warranty period. NE4-8D,NE4-14D,NE4-22D, NE4-28D, NE4-38D, NE4-8D/CT,NE4-14D/CT NE2-22D/CT NE4-25D
NE4-8P,NE4-14P,NE4-22P NE4-28P NE4-8HT, NE4-14HT NE4-22HT NE4-D, NE4-P and NE4-HT
Contents consist of a thermostirrer mounted on a stainless steel bridge unit to fit the selected tank, stainless steel or clear acrylic tank (8,14,22 or 28 litre capacity), a stainless steel shelf, black nylon propellor, fasteners for securing the thermostirrer to the tank, power lead and an instruction manual. Contents consist of an NE4-D thermostirrer mounted on a stainless steel bridge unit, deep chambered bath, stainless steel shelf, black nylon propellor, fasteners for securing the thermostirrer to the tank, power lead and an instruction book. Contents consist of a thermocirculator mounted on a stainless steel bridge unit to fit the selected tank, stainless steel tank (8,14,22 or 28 litre capacity), stainless steel shelf, black nylon propellor, fasteners for securing the thermocirculator to the tank, power lead and and instruction manual. When purchased as a separate item an accessory NE4-MB mounting bracket is required. This allows a thermostirrer/thermocirculator to be fitted to any tank or vessel with a maximum wall thickness of 30mm.
Assembly 1. Turn the thermostirrer/thermocirculator upside down and screw the nylon propellor hand tight onto the spindle. The spindle is threaded in the opposite direction to the propellor to ensure the propellor stays secure IN USE. 2. Place the stainless steel shelf into the tank. 3. Locate the thermostirrer/thermocirculator onto the tank aligning the slots in the bridge unit with the holes in the side of the tank. 4. Secure the thermostirrer/thermocirculator to the tank using the the M4 screws and washers supplied. 5. Fill with liquid to level as described on following page. Safety Do not touch any electrical contacts or open any closure panels. RISK OF ELECTRICAL SHOCK! NE4-D, NE4-D/CT and NE4-P are Class 1 (IEC519 - Part 2) = low liquid level protection, reference to over temperature condition providing product over temperature protection. NE4-HT Class 2 = Adjustable over temperature protection device and low liquid level protection.
Power Supply Lead and Connection to Electrical Supply Fit the power lead by plugging it into rear of the Thermostirrer/Thermocirculator and then to mains supply. Before connecting the product to the electrical supply, check the information on the rating label is compatible. IF IN DOUBT CONSULT AN ELECTRICIAN. THE PRODUCT MUST BE EARTHED! Where the mains supply or plug connection differs refer to local regulations or qualified electrician.
The Clifton Range - 3 Year Warranty
4
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
NE4-D and NE4-D/CT Series Stirred Digital PID Baths Thermostirrer
Front View
Stainless steel construction for durability. Red LED display of actual or set bath liquid temperature or time. Indicators: heating, temperature alarm, timer and set temperature. Keyboard: function, up and down arrow adjustment, run buttons. Splash proof controls, all over smooth wipe clean surfaces. Water bath, stainless steel tank with False base shelf fitted inside.
Tank View
Heating element coil. False base/shelf for standing work/ samples onto - removable for cleaning. Low liquid Safety cut-out Float switch
Stirrer shaft and prop.
Rear View
Mains power switch “I” ON, “O” OFF.
Serial number and electrical supply details. IEC connector, for power supply lead and plug. Fuse draw fitted underneath. To operate from -20°C to ambient temperatures, pop the plastic cover out from rear and immerse the Cooling Coil from an accessory DC1-300 Immersion Cooler. The Clifton Range - 3 Year Warranty
5
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
NE4-P and NE4-HT Series Circulator Digital PID Baths Thermocirculator.
Front View
Stainless steel construction for durability. Red LED display of actual or set bath liquid temperature or time. Indicators: heating, temperature alarm, timer and set temperature. Keyboard: function, up and down arrow adjustment, run buttons. Splash proof controls, all over smooth wipe clean surfaces. Water bath, stainless steel tank with False base shelf fitted inside.
Tank View Circulation pump - white. Heating element coil. False base/shelf.
Rear View Mains power switch “I” ON, “O” OFF. NE4-HT only - Safety Experiment Protector, can be set just above working temperature, protecting work or samples. IEC connector, for power supply lead and plug. Fuse draw fitted underneath. External Liquid Circulation, inlet and outlet ports, ‘push fit and lock’ fittings. OUTLET port - circulation INLET port - circulation To operate the Circulatorfrom -20°C to ambient temperatures, pop the plastic cover out from rear and immerse the Cooling Coil from an accessory DC1300 Immersion Cooler. The Clifton Range - 3 Year Warranty
6
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Liquid Level 8,14, 22 Litre Baths 28 Litre Baths 38 Litre Bath NE4-25D
Minimum liquid level - must cover the top of the false base by 70mm Maximum liquid level - must not exceed the ridge in the tank. Minimum liquid level - must cover the top of the false base by 130mm. Maximum liquid level - must not exceed the ridge in the tank. Mininum liquid level - must cover the top of the false base by 130mm. Minimum liquid level - 50mm from the top lip bath/level mark on tank. Maximum liquid level - must not exceed the 3 round indentations in the tank.
When filling or emptying the bath disconnect the unit from the mains electrical supply. Suitable Liquids Operating temperatures and recommended liquid options: - 20°C to Ambient
100% Heat transfer liquid. The LB range is formulated for tempera tures from -45°C to 90°C and provides complete protection from freezing and algae growth and safeguards against corrosion. See accessories for the full range available. 100% car antifreeze (ethylene glycol).
Ambient to 99°C
Distilled water. Virkon dissolved in distilled water. Proven efficacy against bacteria, viruses, spores a nd fungi in a variety of independant tests using different protocols. Heat transfer liquid. The LB range is formulated for temperatures from -45°C to 90°C and provides complete protection from freezing and algae growth and safe guards against corrosion. See accessories for the full range available.
99°C to 130°C
For use above 99°C we recommend [NE4-HT series]: Silicone oils [up to 10 centistokes]. Synthetic thermal liquids. Above 60°C or below room temperature it is recommended that to achieve optimum performance the bath should be covered with SL4 lid or polypropylene spheres.
NE4-P and NE4-HT Series Connecting Circulated Supply This range of water baths are deisgned for external circulation of temperature controlled water in a closed circuit. 1. Turn unit off. 2. Inlet and Outlet ‘push and lock’ connectors are located on the rear of the bridge unit behind the thermocirculator. These are fitted with a factory fitted link which needs to be removed to fit the desired pipes/hoses. 2. Depress the collar on the ‘push and lock’ connectors to release the factory fitted link. 3. Push the desired pipes/hoses into the ‘push and lock’ connectors. Please note 6mm diameter hose is suitable. The pipe/hose will automatically be locked in place for secure external circulation. 4. To check the pipe/hoses for leakage prior to normal operation we suggest the following: - Fill the tank with water. - Switch unit on. - Any leaks will now become apparent and the pipe/hose should be replaced. - Switch off and refit replacement hose. Please see page 11 for accessories hoses. The Clifton Range - 3 Year Warranty
7
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Operating Instructions Switching ON and OFF Switching ON - the unit may be turned ON (I) at the mains switch located at the rear. When ON (I) the switch is illuminated and unit performs a self test where all segments of the 3 digit LED display and indictors illuminate. Switching OFF - the unit may be turned OFF(O) at the mains switch located at the rear. All temperature and time values remain in memory. Heating Control Modes The Clifton NE4 range feature an advanced PID temperature controller that can be used in either of the following ways: - single point temperature control referred to as 'Set Point 1'. - process control with two temperatures and a time setting, 'Set Point 1' and 'Set Point 2'. Control Panel
Key Pad Description FUNCTION - Used to scroll through parameters. - Used to set timer. DOWN ARROW - Used to decrease a value. Hold continously to scroll. - When pressed for more than 1.5 seconds displays set point 1 value. UP ARROW - Used to increase a value. Hold continously to scroll. - When pressed for more than 1.5 seconds displays set point 1 value. RUN - When pressed for more than 1.5 seconds will activate/deactivate timer function. - Used to turn off buzzer. LED Indicators HEATING INDICATOR When LED is illuminated bath is being heated. OVER AND UNDER TEMPERATURE ALARM INDICATOR LED is illuminated when bath temperature is either 4°C above or 4°C below set temperature. TIMER INDICATOR - Continous illumination indicates timer is set. - Flashing illumination indicates timer is running back/counting down. SET TEMPERATURE INDICATOR - Continous illumination indicates Set Point 1 value is shown on display. - Flashing illumination indicates Set Point 2 value is shown on display. The Clifton Range - 3 Year Warranty
8
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Single Point Temperature Control Setting Temperature - 'Set Point 1' 1. Press and hold up or down arrow for more than 1.5 seconds to display value at Set Point 1. The 'set temperature' indicator will illuminate. 2. Use up and down arrow keys to select required temperature.
3. After 2 seconds the display will revert to show actual liquid temperature. Heater indicator will illuminate. The controller features an over or under temperature alarm which is automatically set 4°C above and 4°C below Set Point 1. If the actual temperature rises or falls beyond this value the alarm is automatically activated. Heating Mode - The Clifton stirred bath is now set and will heat and control the liquid to set temperature. +4C
Set point e.g. 60.0 C
Auto setting Over temperature alarm cutting out heating
Single Point Temperature Control Operation
Actual liquid temperature Auto setting Under temperature alarm
Process Control with Two Temperature Settings and Timer Setting Temperature - 'Set Point 2' 1. Follow instructions for setting temperature at Set Point 1. 2. Press and hold RUN button for more than 1.5 seconds to show value at Set Point 2. The timer indicator will illuminate. 3. Use up and down arrows to select required temperature for Set Point 2. The 'set temperature' indicator will illuminate. 4. After a brief delay display will revert to show actual liquid temperature. Heater indicator will illuminate. The controller features an over or under temperature alarm which is set 4°C above and 4°C below Set Point 2.If the actual temperature rises or falls beyond this value the alarm is activated. 5. Press RUN buttton for 1.5 seconds to exit set point 2 mode. The Clifton Range - 3 Year Warranty
9
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Setting Time 1. Repeatably press FUNCTION button until "t" appears on the display. It will then alternate between showing "t" and time in minutes. 2. Press either up or down arrow to select desired number of minutes. The maximum time setting is 999 minutes. 3. Press function button to save settings. Process Control Operation 1. Operation can be started once bath temperature is at set point 1. To start proc ess (using set point 1,set point 2 and run back timer) press and hold RUN button for more than 1.5 seconds. Heating indicator will illuminate indicating heater and control activity raising the temperature of the liquid to 'set point 2'. Timer indicator will illuminate. 2. On reaching 'set point 2' the timer indicator starts to flash showing the timer is running. The display will automatically switch back to show actual bath tempera ture. 3. To view time remaining, momentarily press UP ARROW button until 't' appears and time remaining is displayed. 4. An audible beeping and "End" message indicates timed period has finished. Press RUN button to deactivate buzzer and clear "End" message. 5. The bath temperature will fall to 'set point 1'. 6. The process can be re-started at any time and values are retained in memory.
Auto setting Over temperature alarm cutting out heating
Set point 2 e.g. 38.0C
Actual liquid temperature
+4C Auto setting Under temperature alarm
Set point 1 e.g. 30.0C
Activate set point 2 "Run" selected
The Clifton Range - 3 Year Warranty
Temperature set point 2 achieved, timer running.
Process Control with Two Temperature Settings and Timer
End timed period, audible buzzer.
10
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Explanation of Temperature Control Terms Temperature Calibration
Accuracy We do not provide, claim or assure any form of accuracy. Accuracy is defined as "the ability of a measurement to match the actual value of the quantity being measured". For accuracy we recommend using a calibrated reference probe at the actual set point temperature and where necessary, adjust the set point accordingly. Sensitivity For an explanation of sensitivity consider a unstirred digital water bath, the PID temperature control system measures and displays the actual temperature of the water and then compares it with the ‘set point’ temperature. It automatically calculates and adjusts the required quantity of heat into the bath to make the measured temperature equal to the set temperature. As with any process there is a time delay between measuring the temperature and the heat entering the water, which causes minor fluctuations in the temperature of the bath. Heat is also distributed in an unstirred bath by convection and conduction and there are heat losses from the surface of the liquid which can cause temperature losses. These losses and heat distribution produce small fluctuations in temperature across the water in the bath. These small temperature fluctuations at any one point are defined as “sensitivity” and vary between an upper and lower limit, however occasionally a larger variation can be observed. Sensitivity as stated in DIN 58966 is the temperature difference between the upper and lower temperature level over 100 cycles after removing the largest 25% of readings. We determine sensitivity by recording the actual upper and lower temperatures of the bath using temperature loggers and is stated as plus or minus one half of the measured value. Uniformity Uniformity is calculated by measuring the temperature in opposing ends of the water bath and is the difference between the mean temperatures at these two points and stated as plus or minus half this value.
The Clifton Range - 3 Year Warranty
11
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Explanation of “Clifton Optimized Control” “Adaptive tune” provides instant control from start up and automatically adjusts PID values with actual temperature readings until it achieves set temperature. These adjusting PID values determine heat-up rate, the load and volume of liquid producing negligible overshoot. All this happens automatically and requires no setting giving you peace of mind. “Clifton Optimization Control” is a new feature, which uses SMART temperature control technology to allow for liquid and or volume variation, external environmental variables and including your work to refine control at set temperature, enhancing the range. The new “Clifton Optimization Control” is best used to fine-tune its PID values by optimizing any potential difference between set point and actual temperature readings: Unstable condition
Set point e.g. 60.0 C
+ C
Returns to "Adaptive Control" using new PID values
"Adaptive Control"
"Opt" selected
"Optimized" PID values @ 60C
Examples of “unstable conditions” where optimisation is useful include: • • •
Liquid level changes: low to high liquid levels, smaller volumes of water needing faster reacting control in comparison with larger liquid volumes. Changing liquid type: from water to oils, and oil to water. Oils suffer with hot and cold layers in the liquid, until natural convection occurs at higher temperatures unlike water where convection is immediate. External environmental conditions in the room where bath is working or circulating water baths, where piped liquid is affected by external temperatures.
Clifton Water baths can adjust to these “unstable conditions” updating the PID values, relearning how to control your process, at a specific temperature to give optimal control.
The Clifton Range - 3 Year Warranty
12
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
When it’s best to use the “Optimization Control”?
Set point e.g. 60.0 C Returns to "Adaptive Control" using new PID values
"Opt" selected
"Optimized" PID values
Heating up, the water bath measures the difference between the set point and actual temperature measurements, refining control until “Optimized”.
Unstable condition
Set point e.g. 60.0 C
+ C
"Adaptive Control"
"Opt" selected
Returns to "Adaptive Control" using new PID values
"Optimized" PID values @ 60 C
Unstable conditions with positive temperature variation, refining PID values until “Optimized”.
Set point e.g. 60.0C
Unstable condition
- C
Returns to "Adaptive Control" using new PID values
"Adaptive Control"
"Opt" selected
"Optimized" PID values @ 60 C
Unstable conditions with negative temperature variation, refining PID values until “Optimized”.
The Clifton Range - 3 Year Warranty
13
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
How to Operate the Optimization Function Clifton ‘Optimised Control’ fine-tunes the PID values by optimising any potential difference between the set point and the actual bath liquid temperature. This can be activated when necessary and forces the controller to update the PID values allowing it to cope with changing liquid types e.g. water vs. oil, changing volumes and environmental factors. The bath is calibrated for use with water. When changing from water to oil we recommend that the optimisation feature is used. This is because water and oil have different thermal properties. To start:1.
Press FUNCTION button until “OPT” appears on the display alternating with “ON.A”.
2.
Press either up or down arrow to select “OFF”.
3.
Press FUNCTION button to confirm.
4.
Press FUNCTION button until OPT appears on the screen.
5.
Press up arrow to select “ON”.
6.
Press FUNCTION button to confirm.
7.
When the “opt” and temperature value is alternately displayed the unit is self-adjusting. When only the temperature is displayed optimisation is complete. Occasionally during the “optimization” process the audible over/under temperature alarm sounds and indicator illuminates, this is quite normal; ignore this warning and allow the bath to continue to desired temperature.
8.
When only the set temperature is displayed, the optimisation process is complete.
To cancel this process at any time press FUNCTION button until “OPT” is displayed. Press up or down arrow to select “OFF”. Press FUNCTION button to confirm.
The Clifton Range - 3 Year Warranty
14
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Under Temperature Alarm - Automatically Set The under temperature alarm is automatically set 4°C below either 'Set Point 1' or 'Set Point 2'. When in alarm condition the ‘over and under temperature ‘ alarm indicator illuminates and actual bath temperature is shown. Please note if an accessory DC1-300 dip cooler is fitted cooling will continue. Once water temperature has risen above alarm setting then indicator clears and actual bath temperature is displayed. Always investigate the cause of the Under Temperature Alarm. Over Temperature Alarm - Automatically Set The over temperature alarm is automatically set 4°C above either 'Set Point 1' or 'Set Point 2'. When in alarm condition the ‘over and under temperature ‘ alarm indicator illuminates and actual bath temperature is shown. All heating/cooling is switched off. Please note if an accessory DC1-300 dip cooler is fitted cooling will continue. Once water temperature has fallen below alarm setting then indicator clears and actual bath temperature is displayed. When in alarm condition the motor and heater are switched off on NE4-P and NE4-HT. NE4-D models motor remains on but heater is switched off. Always investigate the cause of the Over Temperature Alarm. Safety Experiment Protector - HT Range only To set the safety device: 1. Safety device knob (located on the rear) is turned fully clockwise. 2. Set the temperature on the temperature controller at which temperature you require the device to operate, using instructions for set point 1 (see page 6). Allow the liquid to reach this temperature and settle. 3. Slowly turn the knob back anticlockwise until it clicks. 4. Adjust the temperature on the temperature controller to the actual required operatingtemperature - value less than set in 2 above. When in alarm condition heater is switched off and the motor remains on. Low Liquid Level Float Switch - all models - “FIL” meassge displayed The unit features a low liquid level float switch. This ensures heating will be switched off if the liquid level falls below the recommended level. The motor and heater are switched off in this instance. A “FIL” message appears on the display indicating Low liquid level. Refill the tank for heating and stirring to resume - caution when refilling as hot elements can spit and create scalding steam. Periodical Check The low liquid level float switch should be checked periodically by lowering the level of liquid in the tank. The unit should display “FIL” message on temperature display. In the event of the unit not displaying this message please contact our service department at the address shown above. The Safety Experiement Protector featured on the NE4-HT should be routinely checked for correct operation where used. Follow set up instructions as described above.
The Clifton Range - 3 Year Warranty
15
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Cleaning General Important - please follow these instructions to avoid possible damage to the unit, otherwise affecting its performance and warranty. Always disconnect the product from the electrical supply before cleaning. Cleaning External Painted Surfaces featuring “Anti-bacterial Paint Finish” The water bath should be cleaned at regular intervals wiping external surfaces with a cloth or sponge soaked in warm soapy water with a mild detergent. All surfaces should be cleaned using a soft cloth or sponge. Do not under any circumstances use strong solvents or solutions containi ng Chlorinated Hydocarbons, Esters, Ketones or abrasive cleaners or polish on the paint finish otherwise it damages the built in anti-bacterial properties. All painted surfaces on Clifton range products features an “Anti-bacterial paint finish” identified with this authenticating logo on the unit. This “Anti-bacterial paint finish” inhibits the growth of bacteria. It has been tested by independent specialist test houses such as the Law Laboratories (in the UK) using internationally recognized test methods and proven to be effective versus a wide range of bacteria species including Escherichia coli and Staphylococcus aureus (MRSA). We recognise hygienic coatings are part of a controlled approach to a cleaner working environment. Within its formulation an active ingredient with proven anti-bacterial properties is bound into the paint finish. The efficacy of the paint finish applied to the Clifton range is maintained over its lifetime, as the anti-bacterial agent is integral within the paint. In a laboratory environment it makes this one less source of contamination, contributing to essential clean working practices. A benefit of such a paint finish can lead to a reduction in cleaning schedules because surfaces are more protected and improves protection between cleaning. Unlike detergents “Anti-bacterial paint finish” does not offer an instantaneous action, but is intended for long-term general protection against bacterial growth. Moisture on the painted surface is necessary for the bacterium to absorb the agent and be affected by it. The coating is therefore less active in very dry conditions, but dependent upon relative humidity, moisture in the atmosphere maintains activity. Areas where moisture is trapped are also areas that normally are difficult to clean and where bacteria proliferate but these areas are most active for the anti-bacterial coating improving the defence against bacterial growth. Cleaning the Stainless Steel Tank The stainless steel crevice free tank with smooth corners should provide years of valuable service and is resistant to chloride containing solutions it is however important to avoid high concentrations of halogens - especially chloride. With such a high quality and resistant tank it may show symptoms of these halogens as rust, which are deposits from external sources in the water supply. We recommend always empty the bath of liquid after use and wipe out the internal faces of the tank with a non-abrasive cloth and allow to dry. Any deposits can be removed with nitric acid (10%) on a cloth. WEAR PROTECTIVE EQUIPMENT! It is also recommended to use an accessory lid to prevent contaminates landing in bath liquids. The Clifton Range - 3 Year Warranty
16
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Descaling the Stainless Steel Tank Descale the stainless steel tank regularly to maintain it in as new condition ensuring the corrosion resistance and normal operating conditions are maintained throughout its working life. Descale by adding 1 litre of vinegar to water and gently heating to 50°C for an hour, empty and brush the lime away. Rinse thoroughly afterwards. Decontamination of Equipment Clifton laboratory equipment can be decontaminated after spillage or contact with potentially HIV and Hepatitis infected blood samples during analysis using following recommended rapid disinfectants. Virucidal Disinfectant We recommend Virkon tablets for the safe and rapid disinfection of equipment in a wide variety of situations available from your distributor or contact Day-Impex Ltd. for more details. Telephone: 44+(0)1787 223232 or http://www.day-impex.co.uk The ultimate high level surface disinfectant, dissolve VIRKON in water, providing a safe working solution with a faint lemon odor. It has proven efficacy against bacteria (including mycobacteria), viruses, spores and fungi in a variety of independent tests using different protocols. Presents no serious long term health risks to staff – obviating the need for costly ventilation equipment and health monitoring. Also provides high level disinfection of laboratory equipment and instruments where autoclaving is neither practical nor necessary. For more detailed information relating to how Virkon should be used with access to test reports www.relyon.dupont.com Is Virkon solution corrosive? Virkon solution requires only 10 minutes contact time to be effective so long-term exposure is not necessary and therefore will not corrode most materials. Care should be taken with Stainless steel water bath tanks, these surfaces should not be affected however, it is important that generally you do not leave Virkon solution in contactwith metal surfaces “FOR LONGER THAN IS NECESSARY”. Virkon is Registered in accordance with the requirements of the Medical Devices Directive, (93/42/ EEC) as a Medical Device. Disinfectant/Sterilant We recommend PeraSafe a powder product for the safe and rapid chemical sterilant of equipment in a wide variety of situations available from your distributor or contact Day-Impex Ltd for more details. Telephone: 44+(0)1787 223232 or http://www.day-impex.co.uk PeraSafe has a proven safety profile for end-users with none of the undesirable properties of skin sensitisation, toxic fumes or unpleasant odours that are associated with aldehyde solutions. Leading UK and USA microbologists have proven PeraSafe to be active against viruses, mycobacteria and fungi. It is microbiologically superior to glutaraldehyde, destroying sporing bacteria in one minute. It has also been independently proven that PeraSafe sterilises in just 10 minutes. For more detailed information relating to how PeraSafe should be used with access to test reports www.relyon.dupont.com
The Clifton Range - 3 Year Warranty
17
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
3 Year Warranty Our service engineers are fully trained in the assembly, calibration and servicing of all Clifton instrumentation. Products can be returned to our comprehensively equipped service centre where a fast and efficient turnaround is guaranteed: Service Department, Nickel Electro Limited, Oldmixon Crescent, Weston-super-Mare, North Somerset BS24 9BL, UK. Tel +44 (0)1934 626691 Fax +44 (0)1934 630300. Out of Warranty Our Service Department has comprehensive stock of charegeable spare parts maintaining working life of equipment or units can be returned for quotation before repairs are undertaken. End of Life This symbol indicates that this product should not be disposed of with your waste. Instead, dispose waste electrical equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment. The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment. For more information about where you can drop off your waste equipment for recycling, in UK please contact Service Department, rest Europe contact your Distributor. Health & Safety, unless in receipt of a Decontamination Notice or Report the unit cannot be returned or accepted for disposal. Clifton electrical and electronic equipment has been designed for recycling and takes into account the dismantling and recovery its components and materials. Clifton products are easily recycled with majority of the product constructed from stainless or mild steels, which can readily be re-used or recycled. In excess of 78% of this product range can be easily re-cycled economically. Portable Appliance Testing When conducting testing, ensure it is conducted by a qualified person. DO NOT PAT TEST THE BATH UNLESS IT CONTAINS WATER. THIS EQUIPMENT MUST NOT BE FLASH TESTED!
The Clifton Range - 3 Year Warranty
18
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Accessories for NE4 Series GRP Gable Lids GL1-14 GRP Gable Lid to suit 14 Litre Baths GL1-22 GRP Gable Lid to suit 9, 22 and 28 Litre Baths
Stainless Steel Gable Lids SL4-8 Stainless Steel Gable Lid to suit 8 Litre Baths SL4-14 Stainless Steel Gable Lid to suit 14 Litre Baths SL4-22 Stainless Steel Gable Lid to suit 22 and 28 Litre Baths SL4-38 Stainless Steel Gable Lid to suit 38 Litre Baths
Stainless Steel Flat Lids LD4-8 Flat One Piece Stainless Steel Lid to suit 8 Litre Capacity Bath LD4-14 Flat One Piece Stainless Steel Lid to suit 14 Litre Capacity Bath LD4-22 Flat One Piece Stainless Steel Lid to suit 22 and 28 Litre Capacity Bath LD4-38 Flat One Piece Stainless Steel Lid to suit 38 Litre Capacity Bath Stainless Steel Raised Shelves RS4-14 Stainless Steel Raised Shelf to suit 14 Litre Baths RS4-22 Stainless Steel Raised Shelf to suit 22 and 28 Litre Baths
Immersion Cooler DC1-300 Immersion Dip Cooler - cooling liquids down to -20°C
Stainless Steel Test Tube Racks - Dimensions 270 x 70 x 138mm (L x W x H)) 6870 Stainless Steel Test Tube Rack 26 Hole x 17mm Diameter 6871 Stainless Steel Test Tube Rack 16 Holes x 26mm Diameter 6872 Stainless Steel Test Tube Rack 36 Holes x 13mm Diameter 6873 Stainless Steel Test Tube Rack 18 Holes x 19mm Diameter/suitable for 1.5ml microtubes 6900 Stainless Steel Test Tube Rack 12 Holes x 32mm Diameter Note: 8 Litre = 1 Rack, 14 Litre = 2 Racks, 22 and 28 Litre = 5 Racks Miscellaneous LB-2.5 2.5 Litres - Lab Bath 4590 - Heat transfer fluid - Temperature range -15° to +90°C LB-5.0 5 Litres - Lab Bath 4590 - Heat transfer fluid - Temperature range -15° to +90°C TC-1 Thermometer Clip Complete With Bent Stem Spirit Filled Thermometer BX0616 Draining Syphon NE4-MB Mounting Bracket allows Thermostirrers or Thermocirculators to be mounted on any type of separate tank. HS-1 Push Fit Insulated Hoses 1m length. HS-2 Push Fit Insulated Hoses 2m length. BP0368 Polypropylene Spheres
For more information on Accessories www.nickel-electro.co.uk
The Clifton Range - 3 Year Warranty
19
High Quality Products at Affordable Prices
PDF created with pdfFactory Pro trial version www.software-partners.co.uk
Fitting Accessories for NE4 Series GRP Gable Lids Position the lid over the bath and lower into place. The lids are designed to fit snugly so that condensate runs into the bath. Stainless Steel Gable Lids Position the lid over the bath and lower into place. These lids have bo drip points, provide full working height across the bath when fitted and have a ventilation gap. Stainless Steel Flat Lids Position the lid over the bath and lower into place. Stainless Steel Raised Shelves The raised shelf provides a reduced liquid depth for shallow vessels in NE4 series water baths. With recommended minimum water level in NE4 range of 70mm with this in place it will provide a working depth of 5mm or 35mm. Stainless Steel Test Tube Racks Place the loaded test tube rack into the bath on top of either the stainless steel perforated shelf or the raised shelf. Immersion Cooler This refrigeration cooler has no temperature control, as it is designed to be used in conjunction with a stirred or circulating water bath, where its immersion coil is specially designed to fit rear access hole behind the Thermostirrer/Thermocirculator without sacrificing the working area. The immersion cooler runs continuously providing counter cooling with bath providing temperature control. Heat Transfer Fluid Use as a ‘neat’ liquid solution for best results. Thermometer Clip This can be positioned anywhere on the perimeter of the bath by cliping over the edge. The thermometer then lies flush with the top edge of the bath to prevent accidental damage. Draining Syphon Allows the easy and quick emptying of any water bath. Water temperature must be below 45°C before emptying can commence. Mounting Bracket This Mounting Bracket is designed to fit the NE4 series Clifton Thermostirrers for mounting on any number of vessels or tanks. Maximum tank thickness for the Bracket = 30mm.Thumb wheel adjuster secures the Bracket and Thermostirrer to tank. Polyproylene Spheres Place on top of the water/oil to provide a floating lid reducing evaporation and insulating the bath reducing heat losses.