Nonin Medical Inc
Model 9843 Operators Manual 4963-001-06
Operators Manual
43 Pages
Preview
Page 1
Operator’s Manual Model 9843 Pulse Oximeter and Carbon Dioxide Detector
English
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician. Caution: Read this manual carefully before using the Model 9843. The information in this manual has been carefully checked and is believed to be accurate. In the interest of continued product development, NONIN reserves the right to make changes and improvements to this manual and the products it describes at any time, without notice or obligation.
Nonin Medical, Inc. 13700 1st Avenue North Plymouth, MN 55441-5443 USA (763) 553-9968 (800) 356-8874 (USA and Canada) Fax (763) 553-7807 [email protected] www.nonin.com
References to “NONIN” in this manual shall imply Nonin Medical, Inc. Nonin is a registered trademarks of Nonin Medical, Inc. ©2008 Nonin Medical, Inc. 4963-001-06
Guide to Symbols ATTENTION: See Instructions for Use Not for Continuous Monitoring (No Alarm for SpO2)
Type BF Equipment (Patient isolation from electrical shock.)
Indicates separate collection for electrical and electronic equipment (WEEE).
C
SSIFIE LA
D
C
Do Not Reuse
UL US
IPX1
Protected Against Vertically Falling Water Drops per IEC 60529.
IPX4
Protected Against Splashing Water per IEC 60529.
Guide to Symbols
1
UL Mark for Canada and the United States with respect to electric shock, fire, and mechanical hazards only in accordance with UL 60601-1 30EM and CAN/CSA C22.2 No. 601.1.
1
Introduction Indications for Use The NONIN Model 9843 Pulse Oximeter and Carbon Dioxide Detector is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and approximate carbon dioxide (CO2) changes in the airway of intubated patients. These functions may be used separately or simultaneously.
Figure 1. The Model 9843 Pulse Oximeter and Carbon Dioxide Detector
Pulse Oximeter Intended Use The pulse oximeter is intended to be used for noninvasively monitoring oxygen saturation and pulse rate for adult, pediatric, and neonatal patients in hospital, ambulatory, and Emergency Medical Services (EMS) environments. The pulse oximeter may be used for spot checking and/or continuous monitoring when attended by a healthcare professional.
Carbon Dioxide Detector Intended Use The CO2 detector is a mainstream device intended to be used for semi-quantitative detection of CO2 levels in intubated patients during patient transport, and for short-term hospital use (e.g., emergency rooms or crash carts), and where gaseous anesthetic is not present. The CO2 detector may be used to initially confirm proper placement of the endotracheal tube and to provide continued confirmation of correct endotracheal tube placement and patient respiration status. The CO2 detector is not intended for prolonged CO2 monitoring. The CO2 detector is not intended for longterm monitoring of end-tidal CO2. The CO2 detector is not intended for use in patients younger than 3 years old and weighing less than 10 kg.
2
1
Introduction
Precautions for Use Contraindications • •
Do not operate the Model 9843 in an MRI environment.
•
Do not use the Model 9843 CO2 detector for patients less than 3 years old and under 10 kg due to the dead space introduced by the airway adapter tube.
•
Do not use the Model 9843 CO2 detector during mouth-to-tube ventilation. The presence of CO2 in the exhaled breath from the person performing resuscitation will cause inaccurate readings.
•
The Model 9843 CO2 detector cannot distinguish between oropharyngeal tube placement and endotracheal tube placement if the airway is patent. Standard clinical assessment must be used.
Do not use the Model 9843 in a situation where alarms are required. The Model 9843 has no audible alarms.
Warnings •
Explosion hazard. Do not use the Model 9843 in the presence of flammable anesthetics.
•
The Model 9843 is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of assessing clinical signs and symptoms.
•
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
•
General operation of the Model 9843 may be affected by the use of an electrosurgical unit (ESU).
• •
Do not use a damaged sensor.
•
The use of accessories, sensors, and cables other than those specified in this manual may result in increased emission and/or decreased immunity of this device.
•
Use only NONIN manufactured pulse oximeter sensors. These sensors are manufactured to meet the accuracy specifications for NONIN Pulse Oximeters. Using other manufacturer's sensors may cause improper pulse oximeter performance.
This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device should be observed carefully to verify normal operation.
Precautions for Use
1
3
•
Check the pulse oximeter sensor application site frequently to determine the circulation, positioning, and skin sensitivity of the patient. Each patient's sensitivity to NONIN sensors may vary depending on their medical status or the condition of their skin.
•
Discontinue use of the NONIN double-backed adhesive tape strips or the Hydrogel tape strips if the patient exhibits allergic reactions to the adhesive material.
•
Do not stretch the adhesive tape while applying the pulse oximeter sensor. This may cause inaccurate readings or skin blisters.
• •
Do not reuse the Model 9840AAT Airway Adapter Tube.
•
The Model 9840AAT Airway Adapter Tube will increase dead space by approximately 6 cubic centimeters; this may adversely affect ventilation for patients with small tidal volumes.
•
Do not use the airway adapter tube if the airway adapter tube is below 5 °C. An airway adapter tube that is below 5 °C may frost, causing a false reading. Warm the airway adapter tube to above 5 °C by putting it in a warm place (for example in your hands or in a vehicle) before use.
•
If the Model 9843 CO2 detector results are inconclusive, the correct anatomic location of the endotracheal tube must be confirmed by other methods.
•
Do not use the Model 9843 CO2 detector with a humidifier or nebulizer in the breathing circuit, as the fine mist may cause erroneous readings.
If the airway adapter tube becomes contaminated or damaged, discard it and replace it with a new one. Cleaning the interior will damage the anti-fog coating and cause inaccurate readings.
Cautions
4
1
•
Federal law (USA) restricts this device to sale by or on the order of a physician.
• • •
Read this manual carefully before using the Model 9843. Before use, carefully read the package insert provided with the sensors. Model 9843 Pulse Oximeter and Carbon Dioxide Detectors are sensitive electronic instruments and must be repaired by knowledgeable and specially trained personnel only.
•
Ensure that the Model 9843, the airway adapter tube, and the sensors have stabilized at the specified environmental operating conditions before use.
•
Verify that all audible and visible indications are operating properly during the power on self-test. If any self-test event does not occur or functions incorrectly, do not use the Model 9843. Consult the troubleshooting guide or contact NONIN Customer Support. Precautions for Use
• •
Do not immerse the Model 9843 or NONIN sensors in liquid.
•
Do not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at the same time. These actions may cause the batteries to leak.
•
Do not remove any covers other than the battery cover when replacing batteries. There are no user serviceable parts inside other than the replaceable batteries.
•
Follow local governing ordinances and recycling instructions regarding disposal or recycling of device components, including batteries.
•
Alkaline batteries may leak or explode if used or disposed of improperly.
•
Do not block the audible indicator speaker holes. Blocking the speakers will significantly reduce the sound volume.
•
This equipment complies with International Standard EN 60601-12:2001 for electromagnetic compatibility for medical electrical equipment and/or systems. This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However, because of the proliferation of radiofrequency transmitting equipment and other sources of electrical noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be installed and put into service according to the EMC information specified in this manual.
•
Portable and mobile RF communications equipment can affect medical electrical equipment.
•
The Model 9843 pulse oximeter is intended for spot checking or monitoring by an attending healthcare professional. Because the Model 9843 has no audible alarms, international labeling requirements (EN 865) dictate it be labeled “Not for Continuous Monitoring.”
•
The Model 9843 pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that nothing is hindering the pulse measurement before relying on the SpO2 measurement.
•
The Model 9843 pulse oximeter may not work on all patients. If you are unable to achieve stable readings, discontinue use.
Do not use caustic or abrasive cleaning agents on the Model 9843 or the sensors.
Precautions for Use
1
5
6
1
•
The Model 9843 pulse oximeter may interpret motion artifact as good pulse quality.
•
The Model 9843 pulse oximeter is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Significant levels of dysfunctional hemoglobin such as carboxyhemoglobin or methemoglobin may affect the accuracy of the measurement.
•
Cardiogreen and other intravascular dyes, depending on the concentration, may affect the accuracy of the SpO2 measurement.
•
Some fingernail polish colors (particularly blues and blacks) may reduce light transmission and thereby affect SpO2 accuracy.
•
Ear Clip and Reflectance pulse oximeter sensors are not recommended for pediatric or neonatal use. The accuracy of these sensors has not been established for pediatric or neonatal use.
•
Remove earrings from the patient’s ear before applying the Ear Clip Sensor.
•
Water or other liquid between the airway adapter tube and the CO2 sensor may cause erroneous readings.
•
Ensure that all connections to the airway adapter tube are tight and leak-free, and that the airway adapter tube is properly attached to the CO2 sensor.
•
An airway adapter tube that is between 5 °C and 10 °C may cause inaccurate readings due to fogging of optical surfaces. It is recommended that the airway adapter tube be warmed to above 10°C before use.
•
Gastric distention with air prior to intubation may introduce CO2 into the stomach and esophagus and yield false results. Observe six breaths before interpreting results.
•
This device has not been tested for immunity to electromagnetic disturbances.
•
The Model 9843 CO2 detector should not be used with gaseous anesthetics.
•
In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your distributor’s contact information.
Precautions for Use
Manufacturer’s Declaration Refer to the following tables for specific information regarding this device’s compliance to IEC Standard 60601-1-2. Table 1: Electromagnetic Emissions Emissions Test
Compliance
Electromagnetic Environment- Guidance
This device is intended for use in the electromagnetic environment specified below. The customer and/or user of this device should ensure that it is used in such an environment. RF Emissions CISPR 11
Group 1
This device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF Emissions CISPR 11
Class B
Harmonic Emissions IEC 61000-3-2
N/A
This device is suitable for use in all establishments, including domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Voltage Fluctuations/ Flicker Emissions IEC 61000-3-3
N/A
Precautions for Use
1
7
General Description The Model 9843 Pulse Oximeter and Carbon Dioxide Detector is a handheld, battery-operated, noninvasive monitoring device that has visible and audible indicators for tracking patient status. The Model 9843 typically will operate for 90 hours continuously between battery replacements when used for pulse oximetry alone, or for 20 hours continuously when used for both CO2 detection and pulse oximetry. The Model 9843 requires no routine calibration or maintenance. Serial Data Infrared Link
SpO2 Display (Percent Oxygen Saturation)
Pulse Oximeter Sensor Connector
%SpO
Tri-Color Pulse Quality Indicator
CO CO2 Display
Pulse Rate Display
Low Battery Indicator
On / Off Switch CO2 Sensor Connector Advance Switch
Audible Indicators (Speakers)
Setup / Touch Print Switch
Figure 2. Model 9843 Controls and Indicators 8
1
General Description
Pulse Oximeter The Model 9843 determines functional oxygen saturation of arterial hemoglobin (SpO2) by measuring the absorption of red and infrared light passed through perfused tissue. Changes in absorption caused by pulsation of blood in the vascular bed are used to determine arterial saturation and pulse rate. Oxygen saturation and pulse rate values are indicated on light-emitting diode (LED) digital displays. On each detected pulse, the pulse quality indicator flashes. Patient pulse quality signals are graded as good, marginal, or inadequate and are indicated as such by the pulse quality indicator flashing green, yellow, or red respectively. This simple method gives the user a pulse-by-pulse visual indication of waveform signal quality without requiring the user to perform complex waveform analysis during critical patient care situations. If the pulse oximeter sensor is disconnected or malfunctions, or an adequate pulse oximeter sensor signal is not detected, then a dash will appear to the left of the SpO2 value on the SpO2 display and the displayed values will freeze for 10 seconds. If, 10 seconds after the first dash appeared, adequate pulse signals still are not detected, the SpO2 and pulse rate numerical values will be replaced by dashes in the middle digit of each display.
Carbon Dioxide Detector The Model 9843 determines approximate CO2 changes in the airway of intubated patients by measuring the absorption of mid-infrared light passed through the airway adapter tube. The approximate CO2 concentration change is indicated by an 8-segment LED bar graph display (the CO2 display). The CO2 detector relies on the assumption that the inhaled air contains negligible amounts of CO2. Breaths are indicated when the CO2 level increases by approximately 5 mmHg during exhalation. A detected breath is indicated on the CO2 display and by an audible breath beep.
General Description
1
9
If the CO2 sensor is initially connected and operating and then:
• • • •
the CO2 sensor is unplugged, or the airway adapter tube is removed from the CO2 sensor, or the light path is blocked, or a CO2 sensor failure occurs,
then the third and sixth bars on the CO2 display will be steadily illuminated. This visible indication is defined as a CO2 sensor fault.
Unpacking Your Model 9843 Contact the carrier immediately if the shipping carton for the Model 9843 is damaged. Carefully unpack the instrument and its accessories. Confirm that the items listed below are packed with the Model 9843 Pulse Oximeter and Carbon Dioxide Detector. The Model 9843 complete system includes: ■
1 Model 9843 Pulse Oximeter and Carbon Dioxide Detector
■
1 Operator’s Manual for the Model 9843
■
1 Model 8000AA-1 Adult Articulated Finger Clip Sensor
■
1 Model 9840SA Carbon Dioxide Sensor
■
3 Model 9840AAT Airway Adapter Tubes
■
6 AA Size Alkaline Batteries
If any item on this list is missing or damaged, contact your local distributor.
10
1
General Description
Basic Operation Batteries The Model 9843 Pulse Oximeter and Carbon Dioxide Detector is powered by six AA size alkaline batteries that will typically provide 20 to 90 hours of continuous operation, depending on how the 9843 is used. The 9843 will typically operate for 90 hours continuously between alkaline battery replacements for pulse oximetry alone, or 20 hours continuously with both CO2 detection and pulse oximetry operating concurrently. The low battery indicator steadily illuminates when the battery power is marginal. The batteries should then be replaced as soon as possible. When the batteries reach a critically low level the low battery indicator flashes, and no patient data is displayed. The batteries must be replaced before using the 9843. In order to conserve battery life, the 9843 will automatically power off after 10 minutes of both no SpO2 readings and no breath detection. If the batteries are critically low at power on, setup mode will be disabled and the displays will be blank. If the battery level then returns to a marginal level (after removing a sensor or after the power on self-test, for example), the Model 9843 will be functional but setup mode will remain disabled. Notes: ■
To conserve battery life, NONIN recommends disconnecting the CO2 sensor from the 9843 when CO2 detection is not in use. The flashing lamp in the CO2 sensor consumes a significant amount of energy.
■
Setting the month to “00” disables the clock function and helps conserve battery life. Refer to “Calendar Settings” on page 19 for more information.
■
The memory of the 9843 may be erased when the batteries are removed.
■
Replacing batteries may erase the clock settings of the 9843. Refer to “Clock Settings” on page 20 for more information.
Basic Operation
1
11
Replacing Batteries in the Model 9843: 1. Remove the battery door on the bottom of the 9843. 2. 3.
Remove all six used batteries. If necessary, gently tap the device against the palm of your hand to dislodge all of the batteries. Replace all six used batteries with fresh AA size batteries. Be sure to use the correct battery orientation when installing the new batteries. IMPORTANT: Insert these two batteries first.
Battery Orientation
Battery Door
Figure 3. Replacing Batteries in the Model 9843
IPX2 UL 60601-1
30EM
SN
Serial Number
NONIN MEDICAL, INC. PLYMOUTH, MN USA MODEL 9843
PULSE OXIMETER/CO2 DETECTOR SEE ACCOMPANYING DOCUMENTS
Battery Orientation Battery Door
6x1.5V AA IEC LR6 US PATENT # Re. 33,643 5808-001-03
Figure 4. Rear View of the Model 9843
12
1
Basic Operation
Connecting the Sensors Pulse Oximeter Sensor Connect the pulse oximeter sensor (with the NONIN logo facing up) to the top of the 9843 as shown in Figure 5. Ensure that an appropriate sensor is firmly plugged in, and position the sensor on the patient. Pulse Oximeter Sensor
CO2 Sensor and Airway Adapter Tube
Figure 5. Connecting Sensors to the Model 9843
Carbon Dioxide Sensor and Airway Adapter Tube Connect the CO2 sensor (with the NONIN logo facing up) to the side of the 9843 as shown in Figure 5. Ensure that the sensor is firmly plugged in. Refer to “Carbon Dioxide Sensor” on page 29 for more information.
Turning On the Model 9843 Turn on the Model 9843 by pressing the on/off switch on the front of the unit. Refer to “Detailed Operation” on page 17 for detailed information about entering setup mode and general operation of the 9843.
Basic Operation
1
13
Power On Self-Test When the 9843 is turned on for normal operation, it cycles through a self-test before displaying valid data. When entering setup mode at power on, the 9843 cycles through the same self-test routine. Verify that all audible and visible indications are operating properly during the power on self-test. If any self-test event does not occur or functions incorrectly, do not use the 9843. Consult the troubleshooting guide or contact NONIN Customer Support.
•
The audible breath beep will sound three times.
•
The low battery indicator will display for approximately 2 seconds, then will remain off unless the batteries are low.
•
The pulse quality indicator will first flash red, then green, then will either continue to flash if a pulse oximeter sensor is connected to the unit, or will turn off if no pulse oximeter sensor is connected.
•
The SpO2 and pulse rate ♥ displays will cycle through the following sequence:
•
14
1
–
“888 888”;
–
the current time (if set) or “00 00” if the time is not set;
–
the software revision number;
–
a single dash (-) in the middle digit of both displays if no pulse oximeter sensor is connected to the 9843, or
–
a single dash (-) in the middle digit of both displays if a pulse oximeter sensor is connected to the 9843 but is not detecting an adequate signal. (The 9843 will then begin to display values when the pulse oximeter detects an adequate signal.) or
–
“Prn” and the current printer mode if entering setup mode.
The CO2 display will cycle through the following sequence: ramp up and then down; turn all bars off; and finally either remain blank (until a CO2 sensor is connected to the unit, and after exiting the setup mode, if applicable) or illuminate the bottom bar when ready for use.
75 m 50 m H 30 g 20
10
10
1.3
6
0.8
2
0.3
7 k P 4 a 2.6
Basic Operation
Overview of Monitoring Pulse Oximeter Verify that the sensor is properly positioned, and ensure that the pulse oximeter is sensing adequate pulse quality by:
• • •
verifying that the pulse quality indicator (
) is flashing green and
verifying that the pulse rate (♥) and SpO2 (%SpO2) displays are showing readings and verifying that the pulse quality indicator flashing is correlated to the pulse rate for at least 10 seconds
If the pulse quality indicator light is flashing red or yellow or is flashing erratically, reposition the pulse oximeter sensor or try a different sensor. If no pulse oximeter sensor is attached to the pulse oximeter after system initialization (a few seconds after powering on), only a dash ( - ) will be displayed in the middle segment of both the SpO2 and pulse rate ♥ displays.
Carbon Dioxide Detector Verify that the airway adapter tube is properly attached to the CO2 sensor and that the CO2 sensor is properly connected to the 9843. Next, verify that the lower bar of the CO2 display is illuminated and the lamp inside the CO2 sensor is flashing on and off. (When CO2 is detected, the illuminated bars on the CO2 display will indicate the approximate change in CO2 level.) Attach the CO2 sensor/airway adapter tube assembly to the patient’s endotracheal tube. The CO2 detector will reflect appropriate values and breath beeps for each breath (if the breath beep sound volume is not set to “off ”). If the CO2 sensor is initially connected and operating and then the CO2 sensor is unplugged, the airway adapter tube is removed from the CO2 sensor, the light path is blocked, or a CO2 sensor failure occurs, then the third and sixth bars of the CO2 display will be illuminated indicating a CO2 sensor fault.
Basic Operation
1
15
Cleaning the Model 9843 CAUTION! Do not immerse the 9843 or sensors in liquid. CAUTION! Do not use caustic or abrasive cleaning agents. Clean the 9843 unit separately from the sensors. Refer to the appropriate cleaning instructions for each sensor type. Notes: •
For instructions on cleaning pulse oximeter sensors, refer to the respective pulse oximeter sensor instruction inserts.
•
For instructions on cleaning the CO2 sensor refer to “Cleaning the Carbon Dioxide Sensor” on page 30.
Cleaning the 9843 ➤
16
1
Clean the 9843 (device only) with a soft cloth dampened with isopropyl alcohol. Do not pour or spray any liquids onto the 9843, and do not allow any liquid to enter any openings in the device. Allow the 9843 to dry thoroughly before reusing.
Basic Operation
Detailed Operation Controls All functions of the 9843 are controlled by switches found on the front of the unit. Powering the 9843 1. With the unit off, press the on/off
switch one time to power the unit. 2.
Press the on/off switch again to turn off the displays and put the circuitry into a low power standby mode.
On / Off Switch
Advance Switch Setup / Touch Print Switch
NOTE: In order to conserve battery life, the 9843 will automatically power off after 10 minutes of both no SpO2 readings and no breath detection. Each time an SpO2 value is displayed or a breath is detected, the 10-minute timer is restarted.
Setup Mode Setup mode is used to control the internal time-of-day clock. In setup mode, the advance switch and the setup/touch print switch are used to make the menu selections. NOTE: Setting the month to “00” disables the calendar and clock functions and helps conserve battery life. Setup mode will be disabled if the batteries are critically low at power on.
Detailed Operation
1
17
Entering Setup Mode 1. With the unit off, press and hold the setup/touch print switch while
pressing and then releasing the on/off switch. 2.
Release the setup/touch print switch when “888 888” is displayed on the SpO2 and pulse rate ♥ displays. Three brief beeps will sound, and “Prn” will appear in the SpO2 display.
Making Selections in Setup Mode 1. Upon entering setup mode, “Prn” (the first
parameter, see Table 1 on page 18) will appear in the SpO2 display. Press the advance switch (or press and hold to quickly scroll) to increment the number on the pulse rate display. The menu starts at the current value stored in memory for the parameter designated in the SpO2 display and will cycle through the range of values listed in Table 1. 2.
When the desired value appears in the pulse rate display, press the setup/touch print switch to store the value and advance the SpO2 display to the next parameter (year, month, etc.), as listed in Table 1.
3.
Continue this process until all parameters are set. When the setting sequence has been completed, the 9843 exits the setup mode and begins normal operation.
The settings can be easily checked, since the first value displayed for each parameter represents the current setting. Table 1. Printer, Calendar, and Clock Mode Parameters Parameter
1
Pulse Rate ♥ Display Range of Values
Default Value
Printer
Prn
00 - 15
00
Year
y
00 - 99a
98
Month
nn
00 - 12
00
Day
d
01 - 31
00
Hours
h
00 - 23
00
Minutes
nn
00 - 59
00
a
18
Appears in SpO2 Display:
Note: If the calendar is set properly, on 1 January 2000 the year will be displayed as “00”.
Detailed Operation