Instruction on the Inspection of Reusable Medical Devices
1. INTRODUCTION The purpose of this document is to describe the inspections the user needs to perform to guarantee the proper functioning of reusable medical devices from NOVASTEP. Refer to the label and package leaflet to establish whether the instrument is reusable. In general, these dysfunctions may be related to poor maintenance or inappropriate use of the device, which may continue until the end of its useful life. Because a high number of factors which are dependent on the user may impair the devices, NOVASTEP does not recommend a maximum number of uses. Rigorous inspection, associated with functional tests if applicable, will make it possible to check the performance of the reusable device and its proper operation throughout its entire useful life.
2. INSPECTION Take precautions when handling blunt and sharp instruments. If necessary, disassemble the device to perform the inspection and reassemble it according to the information provided in the Surgical Technique.
IIDM-Ed1-08-22-EN
2/11
Criteria Condition of sterilization trays
Instructions for visual and functional inspection of the whole device.
Example of defect requiring replacement of the device
Example of the non-defective device
Appearance of the boxes and marking Operation of fasteners and handles
Disassembled handles
Damaged fastener Condition of identifiers and markings
IIDM-Ed1-08-22-EN
Readability of markings No corrosion
3/11
Criteria
Condition of cutting tools (drill, countersink, etc.)
Instructions for visual and functional inspection of the whole device.
Example of defect requiring replacement of the device
The teeth and/or tips have not been knocked or deformed No blunt tip The cylindrical part is not deformed (concentricity)
IIDM-Ed1-08-22-EN
Example of the non-defective device
4/11
Criteria
Instructions for visual and functional inspection of the whole device.
Example of defect requiring replacement of the device
Example of the non-defective device
Condition of the forceps
IIDM-Ed1-08-22-EN
Closing and opening of the forceps Functioning of the springs Appearance of the active zones of the forceps (no burrs, deterioration or deformation) The screws are present and retain their holding ability Functioning of the racks
5/11
Criteria
Instructions for visual and functional inspection of the whole device.
Example of defect requiring replacement of the device
Example of the non-defective device
Condition of the screwdrivers
The screwdrivers have not been knocked, deformed or broken
IIDM-Ed1-08-22-EN
6/11
Criteria
Instructions for visual and functional inspection of the whole device.
Example of defect requiring replacement of the device
Example of the non-defective device
Condition of the gauge
IIDM-Ed1-08-22-EN
Holding ability of the slide The screws are present Gauge reading Screwing of the gauge shank 7/11
Criteria
Instructions for visual and functional inspection of the whole device.
Example of defect requiring replacement of the device
Example of the non-defective device
Holding ability of the gauge bracket on the body
Condition of the cannulated or pierced instruments
IIDM-Ed1-08-22-EN
Use the appropriate pin to check that the cannula is not blocked Check that the teeth have not been knocked or damaged
8/11
Criteria
IIDM-Ed1-08-22-EN
Instructions for visual and functional inspection of the whole device.
Example of defect requiring replacement of the device
Example of the non-defective device
9/11
Criteria Condition of the threaded instruments
IIDM-Ed1-08-22-EN
Instructions for visual and functional inspection of the whole device. Check the condition and functioning of the thread
Example of defect requiring replacement of the device
Example of the non-defective device
10/11
Condition of the wires
The tip has not been knocked No deformation, bending or deterioration on the cylindrical part (darkening, grooves, high roughness, etc.)
Condition of the handles and AO connections
IIDM-Ed1-08-22-EN
Function of the click and lock mechanism using a connector
11/11