NSK Care & Maintenance instructions in accordance with ISO/EN 17664 and HTM 01-05 1.
Preparation • Remove rotary instrument from tubing • Remove bur • Do not submerge rotary instrument in any solution
2a. Automatic cleaning and disinfection using a thermo/washer disinfector • Always ensure that rotary instruments are placed on the correct adaptors • Cleaning agent pH level must be between 2.5-10.5 • Rinse cycle must be activated
• Spray rotary instrument for 2-3 seconds holding a clean tissue over the head end • Check tissue for debris and if oil is not running clean repeat process until rotary instrument is clean and free of debris • Lubricate chuck mechanism using PANA SPRAY plus with tip nozzle weekly • Always complete steps 2,3,4,5 in quick succession
6.
Packaging
• Pack in accordance with current applicable standards/legislation if using a B or S type autoclave. If using an N type autoclave, pack AFTER sterilisation.
OR (if automatic method is not available see 2b below)
7.
2b. Manual cleaning and disinfection using tap water, brushes, cloths and detergent
• Sterilise in a steam steriliser (autoclave) at a temperature not exceeding 138°C in accordance with current applicable standards. Always remove from autoclave as soon as cycle is complete.
• Clean rotary instrument under cold or tepid running water using a soft to medium brush • Disinfect externally using approved detergent wipes. DO NOT USE SPRAYS • When using external disinfectant rinse of residues afterwards under running water • Disinfect internally using an approved product followed by ALPRO WL-Dry or WL-Blow
8.
Sterilisation
Storage
• Store rotary instruments protected from dust, dirt and micro organisms, ideally in an upright position (head up) • Rotary instruments MUST be stored dry • Always comply with expiry dates of sterilised items in accordance with current applicable standards/legislation
• You MUST finish by using ALPRO WL-DRY or WL-Blow at all times if using internal disinfectant
3.
Inspect
• Inspect rotary instruments for debris and repeat above process if necessary
4.
PANA SPRAY Plus Nozzles Always choose the right one for your handpiece KV Nozzle - Z090-051 For KaVo® MULTIflex Coupling
Drying
• The drying procedure is normally part of the cleaning programme of the thermo/washer disinfector alternatively dry manually using a tissue
5a. Automatic lubrication using NSK Care3 Plus • Check machine is connected and working correctly • Connect rotary instrument to correct port • Select appropriate cycle (refer to NSK Care3 Plus user manual) • Use air cycle as required • Always lubricate straight handpieces with closed chuck and blanking piece inserted • Always complete steps 2,3,4,5 in quick succession • Lubricate chuck mechanism using PANA SPRAY plus with tip nozzle at least 2-3 times per week
5b. Manual lubrication using NSK PANA SPRAY plus • Select correct nozzle for rotary instrument (see list below) • Shake spray can • Keep spray can in upright position • Insert nozzle in to rotary instrument
SR Nozzle - Z090-052 For Sirona® Quick Coupling BA Nozzle - Z090-053 For Bien-Air® Unifix® Coupling WH Nozzle - Z090-054 For W&H® Roto Quick® Coupling For Phatelus Air Motor Heads - M070-090 To be used with P183-090 QD Nozzle - Z096-090 For QD Couplings Standard E-Type Nozzle - Z019 - 090 For E-Type attachments NSK Surgical Head Nozzle - Z114-090 For E-Type attachments and surgical heads: SGM-I, SGMS-I and SGQ-H Endo-Mate Head Nozzle - Z256-090 For ENDO-MATE DT, TC, TC2 & TASKAL Wizard heads Tip Nozzle - Z020-201 For Borden 2/3 and Midwest 4 hole turbines and Borden 2/3 and Midwest 4 hole air motors and chuck mechanism NSK Nozzle - P183-090 For high speed instruments which attach to NSK Flexi Quick Coupling
• Always lubricate straight handpieces with closed chuck and blanking piece inserted
Call NSK on 0800 6341909 for more information or contact your preferred dental dealer NSK United Kingdom Ltd
Office 5 Gateway 1000 Arlington Business Park Whittle Way Stevenage Hertfordshire SG1 2FP
www.nsk-uk.com
Tel: 0800 6341909 +44 (0)1438 310670 Fax: 0800 6341910 +44 (0)1438 310671
Automatic cleaning and lubrication using NSK Care3 Plus is simple A. C heck oil level indicator. Add more oil if required
A
B. Open the door and connect 1, 2 or 3 handpieces to the correct handpiece adaptor. Close the door (Care3 Plus will not operate if the door is open)
Maintenance oil tank
C. Select the correct programme Mode for each handpiece type by pressing the relevant Mode selection button (1, 2 or 3), until the required mode light is illuminated
B
Oil level indicator
Instrument connector
C
Mode selection buttons
D
Start button
- S Mode for High Speed handpieces (air turbines) (Cycle time 35 seconds)
Always select the correct spray nozzle before cleaning & lubrication It is preferable to purge any excess solution by briefly operating the handpiece. When purging cover the handpiece head with a tissue. Firmly attach the nozzle onto the valve at the top of the Pana Spray Plus can. Keep the spray can in an upright position High Speed Optic & Non Optic handpieces that connect direct to: NSK type nozzle P183-090 Kavo® type nozzle Z090-051 W&H® type nozzle Z090-054 Sirona® type nozzle
- L Mode for Blue band Low Speed handpieces and NSK heads (Cycle time 45 secs)
Z090-052
- E Mode for all speed reducing, speed increasing and surgical contra-angles and straight handpieces (Cycle time 50 secs)
Z090-053
D. Press the ‘Start’ button. The door should not be opened until the cycle is finished. When the cycle is finished, before removing the handpieces, flush with air by pressing the ‘air’ button - as required
Always select the correct adaptor before cleaning and lubrication E-Type Handpiece For All Brands ISO E-Type and E-Type LUX Low Speed handpiece attachments - Z280-035 E-Type NSK Coupling For High Speed handpieces that connect to NSK couplings Z257-010 NSK Borden 2 hole For Borden 2 hole High Speed handpieces T904-001 Borden® Kavo Coupling For High Speed handpieces that connect to Kavo® Multiflex LUX® couplings - Z257-020 Kavo® Sirona Coupling For High Speed handpieces that connect to Sirona® Quick couplings - Z257-023 Sirona® W&H Coupling For High Speed handpieces that connect to W&H® Roto Quick® couplings - Z257-021 W&H®
Bien-Air® type nozzle
QD type nozzle Z096-090
Contra-angles, Handpieces, Shanks & Heads:
E Type (all brands) Z019-090
ISO E Type (all brands) and NSK Surgical Heads SGM-I, SGMS-I & SGQ-H Z114-090 NSK HEADS used with Endo-Mate DT, TC, TC2 and TASKAL Wizard Z256-090 Tip nozzle for Borden 2/3 and Midwest 4 hole turbines and Borden 2/3 and Midwest 4 hole air motors Z020-201
Tip nozzle for chuck mechanism maintenance Z020-201
Bien Air Coupling For High Speed handpieces that connect to Bien-Air® Unifix® couplings - Z280-022 Bien-Air® F-Type Head For ENDO-MATE DT, TC / TASKAL Wizard Heads Z280-042 F-Type
NSK United Kingdom Ltd
Office 5 Gateway 1000 Arlington Business Park Whittle Way Stevenage Hertfordshire SG1 2FP
www.nsk-uk.com
Tel: 0800 6341909 +44 (0)1438 310670 Fax: 0800 6341910 +44 (0)1438 310671
Kavo® and MULTIflex LUX® are registered trademarks of Kaltenbach & Voigt GmbH & Co., Germany. Sirona® is a registered trademark of Sirona Dental Systems GmbH, Germany. W&H® and Roto Quick® are registered trademarks of W&H Dentalwerk Bürmoos GmbH, Austria. Bien-Air® and Unifix® are registered trademarks of Bien-Air Dental S.A., Switzerland.