NSK NAKANISHI
Surgic XT Plus Operation Manual
Operation Manual
11 Pages
Preview
Page 1
SURGERY SYSTEM
OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.
OM-E0454E 003 ’11.04.03 S
Classification of equipment • Type of protection against electric shock : – Class l equipment • Degree of protection against electric shock : – Type BF applied part • Method of sterilization or disinfection recommended by the manufacture : – See 8. Sterilization • Degree of protection against ingress of water as detailed in the current edition of IEC 60529 : – Foot Control: IPX8 (Protected against the effects of continuous immersion in water) • Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide : – Motor, Foot Control: Category AP Equipment •Mode of operation : – Intermittent operation
Cautions for handling and operation Read these safety cautions thoroughly before use and operate the product properly. These indicators are to allow you to use the product safely, prevent danger and harm to you and others. These are classified by degree of danger, damage and seriousness. All indicators concern safety, be sure to follow them. Classification WA R N I N G
Degree of Danger or Damage and Seriousness Explains an instruction where personal injury or physical damage may occur.
CAUTION
Explains an instruction where minor to medium injury or physical damage may occur.
NOTICE
Explains an instruction that should be observed for safety reasons.
• Surgic XT Plus should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the Control Unit should be observed to verify normal operation in the configuration in which it will be used. • When operating this system always consider the safety of the patient. • Read this Operation Manual before use, and fully understand the functions of each part for operation key. • Do not attempt to disassemble the Control Unit /Foot Control / Micromotor nor temper with the mechanism. • Check for vibration, noise and overheating before use and if any abnormalities are found in use, stop using immediately and contact your dealer. • Use an electrical outlet that is grounded. • Do not drop, hit, or subject to excessive shock. • Do not bend the Irrigation Tube while the water pomp is operating. It could cause tube breakage. • Do not use bent, damaged or sub-standard burs. The Shank could bend or brake. • Do not exceed the recommended speed. • After each Operation, be sure to lubricate and sterilize the handpiece as soon as it’s cleaned. Blood coagulation can cause corrosion and rusting. However, do not lubricate the Micromotor. Oil could generate excessive heat and cause damage. • The Control Unit can be cleaned with a moist cloth. Disconnect the power supply before cleaning. The Control Unit and the Foot Control cannot be sterilized by any method. • When the product is very frequently used, consider the maintenance of a small stock of replaceable parts. • Do not disconnect the Motor Cord from the motor. • Do not wipe with or, clean or immerse in, high acid water or sterilizing solutions. • Follow the table below for the motion time of motor and handpiece. Continuous Continuous time use over a long time may cause the overheating of the handpiece, which in turn ON OFF may result in an accident. 3 min. 10 min. • Applied parts for patient and/or operator is Handpiece.
NOTICE • Turn off the Main Power Switch after each use. • For service requirements and spare parts contact your dealer. • The use of NSK genuine pre-sterilized, disposable Irrigation tube Kit is recommended.
Temperature
Humidity
Atmospheric pressure
Use
Between 0-40˚C (32-104˚F)
Between10-85% RH
Between 700- 1060hPa
Store Transportation
Between -10-60˚C (14-140˚F)
Between 10-85%RH
Between 500-1060 hPa
1. Safety precautions prior to use Read Handpiece’s Operation manual carefully before use.
Intended to Use Surgic XT Plus is intended for use in dental oral surgery and surgical procedures by qualified personnel.
WARNING
** No moisture condensation in the Control Unit. ** Use at outside of these limits may cause malfunction.
The system may present a possibility of malfunction when used in the presence of an electromagnetic interference wave. Do not install the system in the vicinity of the device which emits magnetic waves. Turn off the power switch of the Control Unit of this system when an ultrasonic oscillation device or an electrode knife is located in the vicinity is used.
CAUTION • Surgic XT Plus needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information. • Portable and mobile RF communications equipment can affect Surgic XT Plus. Do not use RF equipment outskirts for the product. • The use of accessories, Motors and cables other than those specified, with the exception of Motors and cables sold by the manufacturer of Surgic XT Plus as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the Control Unit. 1
2
English
Thank you for purchasing the NSK Surgic XT Plus surgical unit. We recommend that prior to use, you carefully read this document regarding instructions for use, handling method, or maintenance check so that you can carry on using the unit in the future. In addition, keep this operation manual in a place where a user can refer to it at any given time.
2. Package Contents
3. Control Unit with an Irrigation Pump 3
5
1
6
7
Calibration Load
English
2
Irrigation Pump
Program Key Speed Key Torque Key
Micromotor Cord Connection socket Coolant Flow key Gear Ratio Key
AC Electrical Cord Connection socket
Memorize Key
System Key
Fuse Holder Main Power Switch Coolant Solution Hanger Post Holder
FWD-REV Key
Foot Control Cord connection socket
Description of Operation Keys on the Control Unit
8
9
10
11
12
4
13
14
15
16
17 18
19
* Figure shows the Optic Motor, Optic Handpiece Item No.
Description
Quantity
Item No.
Description
Quantity
1
Control Unit
1
11
Internal Irrigation Nozzle
1
2
Foot Controller with Cord (2m)
1
12
Y-Connector
1
3
AC Electrical Cord (2m)
1
13
Calibration Bur
1
4
Optic Motor / Non-Optic Motor (With Motor Cord) *
1 (*1)
14
E-type Spray Nozzle
1
5
Optic Handpiece / Non-Optic Handpiece *
1 (*1)
15
Cleaning Wire
1
6
Irrigation Tube
5
16
Autoclave Plug
1
7
Coolant Solution Hanger Post
1
17
O-ring (Autoclave Plug)
2
8
Handpiece Stand
1
18
TA Bulb (For Optic Motor)
1
9
Tube Holder
7
19
Spare Fuse
2
10
Nozzle Holder
1 (*2)
(1) Program Key Use to Cycle through available programs. Press [+] key to ascend and [-] key to descend program number. You will find all program numbers sequentially by pressing keys. (2) Speed Key Use to set the Micromotor speed. Press [+] key to increase, and [-] key to decrease speed. (When the upper and lower speed limits are reached an audible alarm ‘beep’ is sounded). (3) Torque Key Use to set the torque range. Press [+] key to increase and [-] key to decrease torque range. (When the upper and lower torque range limits are reached an audible alarm ‘beep’ is sounded). The torque range setting should be selected according to the attached handpiece gear ratio. (4) System Key Use to activate calibration cycle of the handpiece before use. (5) Gear Ratio Key Use to set gear ratio of the attached handpiece before use. Repeat press this key until the LCD display indicates the correct gear ratio of the handpiece. (6) Coolant Flow Key Use to select 6 levels of coolant solution flow from 0 to 5, repeat press key to cycle through coolant flow levels. (7) FWD-REV Key Use to select the direction of rotation. Press this key once to change the rotational direction. (Once in reverse the Control Unit audibly warns with a ‘beep’) (8) Memorize Key Use to memorize the program parameters set by the operator. Press & hold this key for approx. 1 second to memorize parameters. When beep sounds, the new program parameters have been memorized.
(*1)
Handpiece will be attached suitable one, (*2) Handpiece Dependent * One of that handpieces/motor should be packed. 3
4
4. Foot Control
Program Number
Forward / Reverse
English
LCD display on the Control Unit Console
Speed
Foot control Cord and Plug
Torque PRG (Program) Button
Coolant Flow Level
Gear Ratio
Torque Bar Graph
(1) Coolant Flow Level Displays the selected coolant solution flow level; the selected Flow level is indicated by 1 to 5 levels of illuminated indicators. No illumination indicates when the Coolant Solution Flow is off. (2) Program Number Displays the selected program number. (3) Gear Ratio Displays the selected gear ratio of the handpiece. (4) Forward/Reverse Indicator Displays the selected direction of the Micromotor. (5) Speed Displays the selected speed. Depressing Foot Control:Rotation speed of the motor displayed on the LCD (Assuming the correct gear ratio is selected) Without pressing Foot Control : Max Rotation preset setting displayed on the LCD (6) Torque: Displays the selected torque. * When using the 1:1 direct drive or Speed increasing handpiece, the torque is not displayed. (This function is appropriate for 2:1 increase speed handpiece or more) (7) Torque Bar Graph Display the actual operating torque range. When all bars illuminate, the operating torque is at maximum. When bars are half illuminated then the operating torque is approximately 50% of the preset torque.
Coolant Solution Flow Volume Button
Forward/Reverse Button
Speed Control Pedal
(1) Coolant Solution Flow Volume Button Use to select 6 levels of coolant solution flow from 0 to 5; each level may be increased by one step pressing this button. If you press this button in level 5, it returns to level 0. (2) PRG (Program) Button Use to select the desired program number. Program numbers will always ascend each time this button is pressed and released. To descend, press and hold for 1 second then release. (3) Speed Control Pedal Used to start and stop the Micromotor and to control the speed during operation. The Micromotor operational speed is directly proportional to the position of the foot control, up to the preset maximum. (4) Forward/Reverse button Used to change the rotational direction of the Micromotor. Push once to change the rotational direction.
CAUTION The LCD display panel is produced from liquid crystal and should always be treated with care.
5
6
5. Installation
CAUTION Make sure that the tube is securely set on the rollers when closing the Pump Cover. If the tube is not correctly positioned on the Rollers and the cover is closed, the tube could be cut or sheared. (Fig. 7)
5-1 Connecting the Motor Cord
English
Align the [ ] mark on the Micromotor cord with [ ] mark on the Control Unit; push the Base of Micromotor Cord Plug into the socket until the locking joint ‘clicks’ into place. To disconnect the plug; pull back the Lock Joint, pull further to disconnect the cord (Fig. 2).
Case Marking
Fig. 1
Rollers
Fig. 7
Base of Micromotor Cord Plug Lock Joint
Fig. 2
Cable Marking
After the tubes are correctly positioned, close the Pump Cover by turning the Pump Cover Lever to the ‘CLOSE’ position (180 degrees counterclockwise) (Fig. 8 & 9)
5-2 Connecting the Foot Control Face the screw on the Foot Control Cord Plug downward then insert the plug into the Foot Control Cord connector socket on the Control Unit. Secure the plug by fastening the Lock Nut. Refer Figs. 3 & 4.
Pump Cover Lever
Fig. 8 Fig. 3
5-5 Mounting the Coolant Solution Hanger Post
Lock Nut
Mount the Coolant Solution Hanger Post onto the Holder on the Control Unit, the post will only fit into one position (Fig10). Place the bottle as shown in Fig. 11.
Screw
Fig. 9
Fig. 4
5-3 Connecting the AC Electrical Cord Correctly align then insert the AC Electrical Cord into the AC Electrical Cord Connection at the back of the Control Unit (Fig. 5). Fig. 5 Align Pump Cover
5-4 Installing the Irrigation Tube
Fixed Position
Tube Guide
Mount the Irrigation tube. Ensure the pump cover lever is in the ‘OPEN’ position, with the Irrigation tube ‘Needle’ towards the backside of the Control Unit. Position the stoppers of the tube assembly in to the Tube Guide securely. (Fig. 6)
Fig. 10 Fixed Position “Needle” Direction Stopper
“Control Unit” Direction
Fig. 11
Fig. 6 7
8
5-6 Insertion of the irrigation Tube
CAUTION
Tube Clamp
5-8 Irrigation Nozzle Attachment 2) Insert the Irrigation Tube Needle into the Bottle Cap. (Fig.13) Fig. 12 3) Open the Tube Cap to supply air into the bottle. (Fig.14)
Bottle Cap
X-SG20L/SG20 has 3 irrigation methods available depending on tool and application; External, Internal or both, For installation detail, refer to the attached handpiece Operation Manual.
5-9 Attaching the Tube Holder
Motor Cord
Use the Motor Cord as a strain relief for the ‘Irrigation Tube’. It is easier to insert Motor Cord first, then the Irrigation Tube (Fig.16).
Irrigation Tube Needle
4) Open the Tube Clamp.
Tube Holder Irrigation Tube
Fig. 16 Fig. 13
6. Operation Tube Cap
6-1 Programming the Micromotor Operation
CAUTION Do not operate the Irrigation Pump if the tube is bent or the Tube Clamp is in the closed position. This could cause the tube to burst or slip out of the bottle.
The Control Unit can memorize 8 programs. Each program includes the following functions which will be automatically performed when the appropriate program number is selected. Fig. 14
5-7 Compatibility check of Internal Irrigation Nozzle/Drill Internal irrigation nozzles accompanied with this product; is not necessarily fitted into all the drills on the market. Follow the instructions given below for confirmation prior to use. Failure to do so or to fit the internal irrigation nozzle into drills may cause a leakage of saline solution, which will result in problems such as rust or sudden stop of equipment during use.
Gear ratio of contra angle handpieces Speed Direction of rotation Torque upper limit Coolant Solution Flow
No water leakage between Internal Irrigation Nozzle and drill Before use, ensure no water is leaking from the entry point of irrigation nozzle, a broken seal or no seal in the drill may be the cause. Replace the drill even if its new, saline solution ingress into handpiece will cause malfunction.
Power Switch (1) Turn on the power by pushing the Main Power Switch toward [-]; on power up Symbol Mark program #1 is displayed by default. (2) Select a program number by using either step (a) or step (b): OFF ON Function (a) Press the [Program] Key on the Control Panel until the desired program number is displayed. (b) Press the [Program] button on the Foot Control until the desired program number is displayed. (3) Selecting the Gear Ratio of the handpiece relevant to the program; Press the [Gear Ratio] Key to select the gear ratio of the handpiece (Gear Ratio will display on the LCD). (4) Set the required max operating speed by pressing the [Speed] Key. Each time this Key is pressed display changes to the next speed level. By pressing this Key for more than 1 second brings the speed quickly to the next level until the speed display reaches its upper or lower limit. - When the speed setting reaches the upper or the lower limit, an audible beep is heard and the speed setting cannot be changed any further. (5) Set the torque upper limit by pressing the [Torque] Key on the Control Panel. Each time this Key is pressed display changes to the next torque level. By pressing this Key for more than 1 second brings the torque quickly to the next level until the torque display reaches its upper or lower limit. - When the torque setting reaches the upper or the lower limit, an audible beep is heard and the torque setting cannot be changed any further. (6) Set the rate of the Coolant Solution Flow volume by pressing the [Coolant Flow] Key. The rate of Coolant Solution Flow volume has 6 Flow rates (0-5) (0 = no coolant Flow).
9
10
Instructions: 1) Attach a bottle of saline solution to the Control Unit. 2) Connect the Internal Irrigation Nozzle into the tip of the irrigation tube. 3) Insert the Internal Irrigation Nozzle into the drill from the back (Fig 15). 4) Purge at “Maximum” for 5 seconds.
Internal Irrigation Nozzle
Irrigation Tube
Drill
Fig. 15 Points to be checked : Cleanliness of the saline solution coming out from the drill; if solution is colored there could be rust inside of the drill. If so renew the drill. Water Flow; if the flow is low and/or the flow from the drill is asymmetric, renew the drill.
English
If malfunction such as a leakage of saline solution from the back of Contra Head is detected during use, stop using and perform some troubleshooting.
1) Close the Tube Clamp, between the Irrigation Tube Needle and the Irrigation Pump, as shown in Fig.12.
* Repeat the above steps 2 - 7 to program any one of the 8 available programs.
6-2 Calibration Function The rotation resistance of a handpiece depends on the handpiece model, condition, and internal wear of the handpiece gears. This Control Unit incorporates an automatic function to recognize the level of the resistance of the attached handpiece and Micromotor.
(7) Speed calibration: “CAL Lmin-1” is displayed. Ensure the calibration bur is removed and the handpiece is free to rotate (Fig. 20) ; Press the [SYSTEM] Key after a notification sound, the calibration is automatically started, the handpiece will illuminate (Optic Handpiece only) and start to rotate, it will automatically cycle through to the highspeed test. On completion the motor stops “donE” is displayed the calibration is complete, normal operation mode will continue.
English
(7) Memorize setting; after completing steps 2 - 6 press and hold [Memorize] key until beep is heard. The beep confirms that the programming is completed. If you hear a short beep when the [Memorize] Key is first pressed ignore this sound and keep the [Memorize] Key depressed until a long beep is heard.
The calibration bur is removed.
Fig. 20
CAUTION During the process, the rotational speed is automatically increased and finally rotates at high speed. Significant care should be taken to avoid danger.
CAUTION • This equipment is optimized to obtain the highest accuracy at a gear ratio of 1:20 (X-SG20L/SG20) when using another gear ratio, note that the accuracy decreases with an increase in the ratio relative to 1:20. • Micro saw handpiece should not be calibrated. • Calibration should be performed only on NSK handpiece.
(1) Connect the Micromotor to the Control Unit, Attach the handpiece to the Micromotor. Turn the power ‘ON’. (2) Attach the calibration bur to the handpiece. (Fig. 17) (3) Press the [Gear Ratio] Key and select the gear ratio of the handpiece. (4) Press & hold the [SYSTEM] Key for two seconds initiates the calibration mode. The LCD displays “CAL 0N-cm.” (5) Torque Calibration (no load): Ensure the calibration bur is inserted and free to rotate (Fig. 18). Press the [SYSTEM] Key; after a notification sound the handpiece will illuminate (Optic Handpiece only) and the bur will slowly rotate. On completion the motor will stop the calibration cycle will continue to the next test (Under load). (6) On the screen “19.6N-cm” is displayed: align and Insert the Calibration Bur to the Calibration load, Holding the bur vertically (Fig 19) press the [SYSTEM] key, after a notification sound, the calibration will automatically start, handpiece will illuminate (Optic Handpiece only) and the bur will slowly rotate the load cell, on completion the motor will stop, “donE” is displayed, the calibration cycle will continue to the speed checks. * If “FAIL” is displayed the calibration at that point has failed. Pressing the [SYSTEM] Key returns the failed process back to the start, a reattempt can be made on that point.
(8) Calibration of the handpiece is completed.
Calibration fail example • Extreme gear abrasion and mechanical loss can exceed the control limits. Correct data cannot be obtained during calibration cycle and the calibration will fail. In this case, request repair. • Calibration fail; if the calibration bur makes contact with something during no load calibration or no load is erroneously applied during load calibration.
6-3 Optical Illumination option On/Off (X-SG20L Only) Calibration Bur
* Factory initialized settings lamp ‘ON’
Fig. 17
(1) Selection 1) Lamp OFF: Hold the [System] Key and turn the power ‘ON’; Two beeps notifies the operator the lights are ‘OFF’ 2) Lamp ON: Press the [System] Key and turn the power ‘ON’; One beeps notifies the operator the lights are ‘ON’ (2) Confirmation Confirm whether the lamp of the motor lights or not by the beep issued when turning the power ON. – Long beep once: “Lights up” – Short beep twice: “Does not light up”
NOTICE Nothing makes contact with the calibration bur.
Fig. 18
• Repeat the above procedures when restoring operation. • The lamp status will remain set even if the power is switched “OFF’ and ‘ON. • This function is only applicable for an optic handpiece.
Calibration Bur
6-4 Standard Operation
CAUTION During a process, it is important to hold the Calibration Bur vertically without applying any forcible power. Making the Calibration Bur tilt or pressing it may cause a loss in accuracy.
All standard operational functions can be control at the Foot Control.
Calibration Load
Fig. 19
11
(1) Turn on the Main Power Switch: The Control Unit is ready to perform the Program. (2) Select the desired program number: Step on the Foot Control PRG (Program) Button and the program display ascends to the next program number. Pressing the PRG (Program) Button for one second more will descend the to the next program number. (3) Verify the details of the program on the display.
12
CAUTION
Error Code
Error Mode
Cause of Error
Remedy
Control Unit Interior Over-heating Error
Overheating by extended use under heavy load. Operation of the Control Unit under an extremely high temperature.
Allow it to cool down before use. Since heat is sufficiently radiated, periphery of the main Control Unit should be well-ventilated wherever possible. When an error cannot be eliminated, request repair.
E5
Breaking Error
Abnormal voltage generated in the start / stop switch circuit. Failure in the start / stop switches circuit.
When rotation and stop are repeated in short frequencies, a circuit may be activated which limits acceleration at start. Wait a few seconds and then use. When an error cannot be eliminated, request repair.
E6
Motor Rotation Failure Error
Handpiece attachment failure. Micromotor failure.
The chuck may be opened, or may not be sufficiently closed. Securely close the chuck. When an error cannot be eliminated, request repair.
E7
Pump excess voltage
E8
Pump excess current
The Irrigation tube is incorrectly fitted Check the Irrigation tube. and loading the pump roller . If an error code is displayed even when the The pump fails. Irrigation tube is normal, request repair.
E9
Foot Control abnormality
Connection failure of the connector. Failure of the inside of the Foot Control
Check the connection state of the connector. The error code is displayed even when the connector is normally connected, request repair.
E10
Lamp excess voltage (Optic Motor)
Failure of lamp circuit.
Request repair.
Check if the saline solution hasn’t penetrated the motor lamp.
Remove the saline solution completely.
E11
Lamp excess current (Optic Motor)
Failure of lamp circuit.
Request repair.
E4
Using the Micromotor with light ‘ON’ for more than 15 minutes continuously, the light will become hot (The maximum surface temperature might reach 45˚C by the situation used.). In such cases, stop using the Micromotor till the light part cools down, or use the Micromotor without turning on the light. (5) Activation of the torque limiter: During use, when the drilling load reaches the preset torque upper limit, the integrated torque limiter automatically activates to prevent over torque. When the torque limiter activates, the motor stops after beeping for 1 second. To reactivate the Micromotor, release the speed control pedal and depress it again. (6) Stopping the Micromotor: Release the Foot Control Pedal, and the Micromotor will automatically stop. (7) Reversing the Micromotor rotational direction: To reverse direction of the Micromotor (and bur) simply step on the Foot Control Forward/Reverse button. A warning beep can be heard when the rotational direction is in reverse mode.
7. Care and Maintenance 7-1 Protection Circuit An electronic circuit breaker automatically functions to protect the Micromotor and the Control Unit if the Micromotor is ever overloaded. Power supply to the Micromotor will automatically be terminated an Error code will be displayed on the Control Unit. Resetting the Protection Circuit To reset the Protection Circuit, release and then depress the Speed Control Pedal.
7-2 Error Codes If an operational problem occurs the display will show an error code allowing immediate problem diagnosis. Error Code
Error Mode
Cause of Error
E0
System Error
Erroneous memory. Memory failure.
E1
Excessive Current Detected
Extended use under heavy load. Short circuit in the micromotor.
Excessive Voltage Detected
Micromotor cord failure.
E2
Request repair.
Micromotor sensor failure (Hall IC). Micromotor Cord failure. E3
Remedy
Motor Sensor Error Ingress of water into a Motor.
13
7-3 Fuse Replacement Fuse Box
Electrical contact may be insufficient. Securely re-connect the Motor Cord. When an error cannot be eliminated, request repair.
Request repair. Make sure to put an Autoclave Plug when autoclaving it.
If the Control Unit does not function, check the fuses ( Fuse Box lock located on the rear of the Control Unit ). To access the Fuse, use a pointed tool push on the fuse locking latch and the drawer will spring open (Fig. 21). Fuse Ratings 120V
T3.15AL 250V
230V
T1.6AL 250V
14
Fuse
Fig. 21
English
(4) Operating the Micromotor : Step on the speed control pedal in the middle of the Foot Control the Micromotor will start to run. The Coolant Pump will also run (if programmed). Speed increases as the pedal is depressed. When the Speed Control Pedal is fully depressed the speed reaches the maximum preset value.
7-4 Replacement of the Bulb (Optic Motor ONLY)
Bulb Socket
A
Connector Motor Housing
Bulb
View from A
Bulb
Fig. 22
Implant Handpiece can be cleaned and disinfected with a Thermo-Disinfector.
[Autoclaving] 1) Remove blood and debris from the handpiece. 2) Clean inside the handpiece; by using the spray lubricant (refer to "7-6. Maintenance of the Handpiece Attachment"). • Do not attempt to spray lubricant into the Micromotor. • Attach the Autoclave Plug to the Micromotor. (Fig. 23) 3) Place those in autoclave pouch (not included in the package) and seal it. 4) Autoclave up to max. 135˚C. (Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.) 5) Keep the handpiece in the autoclave pouch to keep it clean until you use it. Autoclave Plug
Micromotor
* Sterilization at 121 ºC for more than 15 minutes is recommended by EN13060 or EN ISO17665-1.
Fig. 23
CAUTION • Make sure the power ‘OFF’ (remove the AC Electrical cord) • Do not touch the glass of new bulb.
7-5 Maintenance of the Control Unit and Foot Control If blood or saline solution has stained the Control Unit or Foot Control, remove the AC Electrical Cord; wipe Control Unit or Foot Control with a damp cloth, then with an alcohol-absorbed cloth.
7-6 Maintenance of the Handpiece Attachment
Caution for Autoclaving • Clean and lubricate the handpiece before autoclaving. Autoclaving a handpiece stained with blood or debris could cause damage to the handpiece. • Do not lubricate the Micromotor. • Do not disconnect the Motor Cord from the Micromotor. • The Irrigation tube is a single use disposable type and cannot be autoclaved. • Do not use Autoclave Plug without the mounting O-Ring. It may cause a malfunction. If O-Ring had damaged, replace it immediately. * O-Ring (Autoclave Plug): Order code 0312457102 • You can hang Autoclave Plug on the Hunger Post to prevent loosing it. • Do not hung the Autoclaving Plug with Micromotor. It may drop the motor and cause damage.
For detail, refer to the attached handpiece Operation Manual.
9. Optional Accessories
8. Sterilization
Part No.
• Autoclave sterilization is recommended. • Autoclave sterilization is required for the first time use and after each patient as noted below. The following items can be autoclaved. • Implant Handpiece • Micromotor with Motor Cord (Including the Motor Housing) • TA Bulb • Handpiece Stand • Internal Irrigation Nozzle • Tube Holder • Nozzle Holder • Autoclave Plug • Calibration Bur
Description
C823752
Y-Connector
Y900113 20000396 (for X-SG20L) 20000357 (for SG20) 10000324 Y900083 Z182100
Irrigation tube (Pack of 5)
Remarks Used for branching the internal and external coolant irrigation. For replacement tube use.
Nozzle Holder
Supplied as standard accessory items.
Internal Irrigation Nozzle Tube Holder PANA SPRAY Plus
Supplied as standard accessory. Supplied as standard accessory for Irrigation Nozzle. For High & Low speed handpieces.
CAUTION Only Items Specified above can be autoclaved
15
16
English
1) Loosen Motor Housing and detach from the motor, and remove the old light bulb using the small pin, like the precision screwdriver. (Fig. 22) 2) Align and insert the connector of the new bulb into the Socket Hole securely. Screw the motor housing securely to the motor. * Optional Bulb: TA Bulb (Pack of 3) Order No. Y900132
Symbols
10. Specifications
10-2 Micromotor Model Type Speed Range Input Voltage Dimensions 10-3 Foot Control Type Code length
TUV Rhineland of North America is a Nationally Recognized Testing Laboratory (NRTL) in the United States and is accredited by the Standards Council of Canada to certify electro-medical products with Canadian National Standards.
Surgic XT Plus NE179 AC120 / 230V 50/60Hz 52VA 75mL / min. (.02 gal/min) W268 x D230 x H103mm
This product is designed not to become the ignition source in air and flammable anesthetic gas. The EU directive 93/42/EEC was applied in the design and production of this medical device.
Protected against the effects of continuous immersion in dust and water.
SGL50M (with Optic) SG 50MS (without Optic) E295-050 E290-050 200 - 40,000min-1 (rpm) DC30V ø24 x L105mm (Without the Motor Cord)
Dispose of this device and its accessories via methods approved for electronic device and in compliance with the Directive 2002/96/CE Type BF applied part
See Operation Manual
Marking on the outside of Equipment or Equipment parts that include RF transmitters or that apply RF electromagnetic energy for diagnosis or treatment.
FC-51 2M Guidance and manufacturer's declaration - Electromagnetic Emissions.
10-4 Handpiece Model Max. Rotation Speed Chuck Type Gear Ratio Spray Type Bur/Drill Type Bur/Drill Length Max. bur length Max diameter Optic
X-SG20L
SG20 40,000min-1 Push Bottom Chuck 20:1 Reduction External, Internal* Ø2.35Surgical bur-drill (ISO3964) 11.6mm 36mm Ø4.7mm Glass Rod Optic –
Do not exceed the rotation speed that bur manufacture recommends. Adjust the rotation speed which bur manufacture recommend, if you use the allowable rotation speed is less than 120,000min-1. * If you use internal irrigation system drills.
11. Disposing Product Consult with dealer from whom you purchased it about waste disposal.
Surgic XT Plus is intended for use in the electromagnetic environment specified below. The customer or the user of the Surgic XT Plus should assure that is used in such an environment. Guidance and manufacture’s declaration electromagnetic emissions. Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR11
Group 1
Surgic XT Plus uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR11
Class B
Harmonic emissions IEC61000-3-2
Class A
Surgic XT Plus is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Voltage fluctuations/flicker emissions IEC61000-3-3
Complies
Guidance and manufacturer's declaration - Electromagnetic Immunity Surgic XT Plus is intended for use in the electromagnetic environment specified below. The customer or the user of the Surgic XT Plus should assure that it is used in such an environment. Immunity test
IEC60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD) IEC61000-4-2
± (2, 4) 6 kV contact ± (2, 4) 8 kV air
± (2, 4) 6 kV contact ± (2, 4) 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst IEC61000-4-4
± 2kV for power supply lines ± 1kV for input/output lines
± 2kV for power supply lines ± 1kV for input/output lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC61000-4-5
± 1kV differential mode ± 2kV common mode
± 1kV differential mode ± 2kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC61000-4-11
<5% Ut (>95% dip in Ut) for 0.5 cycle
<5% Ut (>95% dip in Ut) for 0.5 cycle
40% Ut (60% dip in Ut) for 5 cycles
40% Ut (60% dip in Ut) for 5 cycles
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Surgic XT Plus requires continued operation during power mains interruptions, it is recommended that the Surgic XT Plus be powered from an uninterruptible power supply or a battery.
70% Ut (30% dip in Ut) for 25 cycles
70% Ut (30% dip in Ut) for 25 cycles
<5% Ut (>95% dip in Ut) for 5 sec
<5% Ut (>95% dip in Ut) for 5 sec
3A/m
3A/m
Power frequency (50/60Hz)magnetic Field IEC61000-4-8
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE ‘Ut’ is the AC mains voltage prior to application of the test level.
17
18
English
10-1 Control Unit Model Type Power Supply Voltage Frequency Power Consumption Max. Pump Output Dimensions
Guidance and manufacturer's declaration - electromagnetic immunity Surgic XT Plus is intended for use in the electromagnetic environment specified below. The customer or the user of the Surgic XT Plus should assure that it is used in such an environment. Immunity test
IEC60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted RF IEC61000-4-6
3V rms 150 kHz to 80MHz
3V rms
Portable and mobile RF communications equipment should be used no closer to any part of the Surgic XT Plus, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Radiated RF IEC61000-4-3
3V/m 80MHz to 2.5 GHz
3V/m
Recommended separation distance d = 1.2 P d = 1.2 P 80MHz to 800MHz d = 2.3 P 800MHz to 2.5GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer, and (d) is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters as determined by an electromagnetic site survey(a) should be less than the compliance level in each frequency range(b). Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1
At 80MHz and 800MHz, the higher frequency range applies.
NOTE2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Surgic XT Plus is used exceeds the applicable RF compliance level stated above, the Surgic XT Plus should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Surgic XT Plus.
b
Over the 150kHz to 80MHz frequency range, the field strength should be less than 3V/m.
Cables and accessories
Maximum length
Shield
Complies with
Micromotor with Motor Cord
2m
Unshielded
RF emissions, CISPR11:
Foot Controller with Cord
2m
Unshielded
Harmonic emissions:
IEC61000-3-2
AC Electrical Cord
2m
Unshielded
Voltage fluctuations/ flicker emission:
IEC61000-3-3
Electrostatic discharge (ESD):
IEC61000-4-2
Class B/ Group 1
Surge:
IEC61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines:
IEC61000-4-11
Power frequency(50/60Hz) magnetic field:
IEC61000-4-8
Conducted RF:
IEC61000-4-6
Radiated RF:
IEC61000-4-3
Recommended separation distances between portable and mobile RF communications equipment and the Surgic XT Plus. The Surgic XT Plus is intended for use in an electromagnetic environment in which radiated RF disturbances are control LCD. The customer or the user of the Surgic XT Plus can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Surgic XT Plus as recommended below, according to the maximum output power of the communications equipment. Separation distance according to frequency of transmitter m Rated maximum output power of transmitter W 0.01 0.1 1 10 100
150kHz to 80MHz
80MHz to 800MHz
3.5
3.5
d=
V1
P
d=
0.12 0.38 1.2 3.8 12
E1 0.12 0.38 1.2 3.8 12
800MHz to 2.5GHz 7
P
d=
E1
P
0.23 0.73 2.3 7.3 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance ‘d’ in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where ‘P’ is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
19