NSK NAKANISHI
SurgicPro Surgery System Operation Manual May 2012
Operation Manual
113 Pages
Preview
Page 1
Surgery System
Surgic Pro
OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.
OM-E0560E
Original Operatrion Manual
English
Thank you for purchasing the NSK Surgic Pro series surgical unit. We recommend that prior to use, you carefully read this document regarding instructions for use, handling method, or maintenance check so that you can carry on using the unit in the future. In addition, keep this operation manual in a place where a user can refer to it at any given time. یIntended to Use Surgic Pro series is intended for use in dental oral surgery by qualified personnel.
Contents 1. Safety precautions prior to use 2. Package Contents 3. Control Unit and Foot Control 4. Installation 5. Setting 6. Operation 7. Protection Circuit 8. Error Code & Troubleshooting 9. Fuse Replacement 10. Maintenance 11. Sterilization 12. Contra Angle Handpieces and Accessories 13. Specification 14. Disposing Product 15. Symbols 16. EMC Information
2 4 5 7 10 12 16 16 18 18 19 20 20 20 21 21
یClassification of equipment • Type of protection against electric shock : – Class l equipment • Degree of protection against electric shock : – Type BF applied part • Method of sterilization or disinfection recommended by the manufacture : – See 11. Sterilization • Degree of protection against ingress of water as detailed in the current edition of IEC 60529 : – Foot Control: IPX8 (Protected against the effects of continuous immersion in water) • Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide : – Foot Control: Category AP Equipment • Mode of operation : – Intermittent operation
1
Cautions for handling and operation Read these safety cautions thoroughly before use and operate the product properly. These indicators are to allow you to use the product safely, prevent danger and harm to you and others. These are classified by degree of danger, damage and seriousness. All indicators concern safety, be sure to follow them.
Classification
Degree of Danger or Damage and Seriousness
W A R N I N G Explains an instruction where personal injury or physical damage may occur. C A U T I O N Explains an instruction where minor to medium injury or physical damage may occur. N O T I C E Explains an instruction that should be observed for safety reasons.
1. Safety precautions prior to use Read Handpiece’s Operation manual carefully before use. WARNING • TO PREVENT ELECTRIC SHOCK Do not unplug the AC Power cord with wet hands. • TO PREVENT ELECTRIC SHOCK Be sure to prevent water on the Control Unit. • TO PREVENT ELECTRIC SHOCK Use an electrical outlet that is grounded. • The system may present a possibility of malfunction when used in the presence of an electromagnetic interference wave. Do not install the system in the vicinity of the device which emits magnetic waves. Turn off the Main Power Switch of the Control Unit of this system when an ultrasonic oscillation device or an electrode knife is located in the vicinity is used.
CAUTION • The accuracy of the Torque Monitor depends on the high precision performance of the handpiece attached to the micromotor. Real torque value may not otherwise be correctly displayed. To guarantee handpiece torque accuracy, that matches the monitor display system, USE ONLY the NSK contra angle handpieces listed on “12. Contra Angle Handpieces and Accessories”. When other handpice had connected, it may not output accurate torque which indicated on handpiece. • Surgic Pro series needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information. (Refer to 16. EMC Information) • Portable and mobile RF (Radio Frequency) communications equipment can affect Surgic Pro series. Do not use RF equipment outskirts for the product. • The use of accessories, Motors and cables other than those specified, with the exception of Motors and cables sold by the manufacturer of Surgic Pro series as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the Control Unit. • Surgic Pro series should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the Control Unit should be observed to verify normal operation in the configuration in which it will be used. • When operating this system always consider the safety of the patient. • Do not attempt to disassemble the Control Unit /Foot Control / Micromotor nor temper with the mechanism. • Check for vibration, noise and overheating before use and if any abnormalities are found in use, stop using immediately and contact dealer. • Use an electrical outlet that is grounded. • Do not drop, hit, or subject to excessive shock. • Do not bend the Irrigation Tube while the irrigation pomp is operating. It could cause tube breakage. • Do not use bent, damaged or sub-standard burs or drill. The Shank could bend or brake. • Do not exceed the recommended speed and torque. 2
NOTICE • Turn off the Main Power Switch after each use. • For service requirements and spare parts contact dealer. • Use of NSK genuine pre-sterilized, disposable Irrigation Tube is recommended. • Total recording time is about 100 minutes. (Surgic Pro+) • Do not insert a USB flash drive during operation. • Do not remove USB flash drive while LED lamp is on or blinking. • Do not use data stored in USB flash drive for diagnostic purposes. • Do not use a USB flash drive other than USB 2.0. • This device is not guaranteed to work with all USB flash drive. It may not work with USB flash drive with security functions or those requiring a great deal of electricity.
"#%&#
&%(
%!$"#"#$$&#
$
%' ) )
%'
%'
%!## $"!#%%!
%' ) )
%'
%'
No moisture condensation in the Control Unit. Use at outside of these limits may cause malfunction.
3
English
• After each Operation, be sure to lubricate and sterilize the handpiece as soon as it’s cleaned. Blood coagulation can cause corrosion and rusting. However, do not lubricate the Micromotor. Oil could generate excessive heat and cause damage. • The Control Unit can be cleaned with a moist cloth. Disconnect the power supply before cleaning. The Control Unit and the Foot Control cannot be sterilized by any method. • When the system is very frequently used, consider the maintenance of a small stock of replaceable parts. • Do not disconnect the Motor Cord from the motor. • Do not wipe with or, clean or immerse in, high acid water or sterilizing solutions. • Follow the table below for the motion time of motor and handpiece. Continuous use over a long time with high torque may cause the overheating of the handpiece, which in turn may result in an accident. • Applied parts for patient and/or operator is Handpiece. • Do not fixate the eyes of LED light. • If any abnormalities (dark, do not light on or flashing) are found for LED, stop using the it immediately and contact dealer. • Using the Micromotor with light ‘ON’ for more than 3 minutes continuously, ! ! motor will become hot. (The maximum surface temperature might reach 45˚ C by the situation used) In such cases, stop using the Micromotor till the light part cools down, or use the Micromotor without turning on the light. • Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle marked "Hospital Only" or "Hospital Grade". " " • For safety’s sake, install the Control Unit in a place where the AC POWER Cord can be easily removed. (It is possible to disconnect the control unit from the power source by removing the AC POWER Cord.)
2. Package Contents
1
2
3
6
9
11
12
4
5
7
8
10
13
14
15
16
17
18
19
*Figure shows Optic Motor and Optic Handpiece
3(/
(2&1+03+/.
3(/
4$.3+36
(2&1+03+/.
/5(1/1'
$.'0+(&(3$.'
//3/.31/,,(15+3*/1' -
.3(1.$,11+)$3+/./77,(
/.31/,!.+3
/77,(/,'(1
11+)$3+/. 4%(
,($.+.)"+1(
//,$.3/,43+/.$.)(1/23
$,+%1$3+/.41
03+&$.'0+(&(/.03+&$.'0+(&(
360(01$6/77,(
03+&/3/1/.03+&/3/1"+3*/3/1/1'
1/3(&3+/.,4)
#/..(&3/1
1+.)1/3(&3+/.,4)
0$1(42(
1+.)/3/1
4%(/,'(1
* Handpiece will be attached suitable one (Applied part) * USB flash drive not included.
4
4$.3+36
3. Control Unit and Foot Control 22/%162/76-21 %1+)42562/()4
!<%5,(4-8)2'.)6
44-+%6-21703
-+,6); 22/%16); #"); ); $); );
English
%/-&4%6-212%(
262424( 211)'6242'.)6
); !); ); !
%-129)49-6',
75)2/()4
29)424( 211)'62452'.)6 22621642/24( 211)'6242'.)6
*The item which does not have number is explained in context.
3-1 Keys on the Control Unit (1) Light Key Select the light level. (OFF -> LOW->HIGH) Repeat press Key to cycle through light levels. (Default:HIGH) (2) Coolant Key Use to select 6 levels of coolant solution flow from 0 to 5, repeat press Key to cycle through coolant flow levels. (3) FWD/REV Key Use to select the direction of rotation. Press this Key once to change the rotational direction. (Once in Reverse the Control Unit audibly warns with a ‘beep’) (4) GEAR Key Use to set gear ratio of the attached handpiece before use. Repeat press this Key until the LCD display indicates the correct gear ratio of the handpiece. (5) MEMORY Key Use to memorize the program parameters set by the operator. Press & hold this Key for approx. 1 second to memorize parameters. When beep sounds, the new program parameters have been memorized. (6) CAL Key Use to activate calibration cycle of the handpiece before use. (7) SPEED Key (- +) Use to set the Micromotor speed. Press [+] Key to increase, and [-] Key to decrease speed. When keep pushing the key to change the speed quickly. (When the upper and lower speed limits are reached an audible alarm ‘beep’ is sounded). (8) TORQUE Key (- +) Use to set the torque range. Press [+] Key to increase and [-] Key to decrease torque range. When keep pushing the key to change the torque range quickly. (When the upper and lower torque range limits are reached an audible alarm ‘beep’ is sounded). The torque range setting should be selected according to the attached handpiece gear ratio. (9) PROGRAM Key (- +) Select the available programs. (1 to 8) Press [+] Key to ascend and [-] Key to descend program number. You will find all program numbers sequentially by pressing Keys. (10) USB REC (Surgic Pro+ ONLY) Press this Key to record the Surgic Pro series operation to USB flash drive. A short push will save data being memorized to USB flash drive. A long push (about 2 seconds) will erase data saved on the control unit memory.
5
3-2 LCD display on the Control Unit Console (A) Clock Display
(E) USB flash drive
(B) SPEED
(G) TORQUE
(J) FWD/REV Direction
(K) GEAR Ratio
(L) PROGRAM Number
(C) IMPLANT SYSTEM (D) Memory Time (F) TORQUE Graph (H) Light Level
( I ) Coolant Level
(A) Clock Display (Surgic Pro+ ONLY) Clock display. Time is displayed using batteries installed in the Control Unit. (B) SPEED Selected speed indication. Rotation speed is displayed during rotation. When rotation stops, the maximum rotation speed set is displayed. Depressing Foot Control: Rotation speed of the motor displayed on the LCD (Assuming the correct gear ratio is selected) Foot Control speed varies depending on how far foot is depressed. Without pressing Foot Control: Max Rotation preset setting displayed on the LCD (C) IMPLANT SYSTEM The number of the implant system being used is displayed. (1 to 8) (D) Memory Time (Surgic Pro+ ONLY) Memory time remaining is displayed (Remain Time). (E) USB flash drive (Surgic Pro+ ONLY) When USB flash drive is inserted, it is displayed on the LCD. (F) TORQUE Graph Display the actual operating torque range. When all bars illuminate, the operating torque is at maximum. When bars are half illuminated then the operating torque is approximately 50% of the preset torque. (G) TORQUE Selected torque indication. When using the 1:1 direct drive or Speed increasing handpiece, the torque is not displayed. (This function is appropriate for 2:1 reduction speed handpiece or more) (H) Light Level (For Optic ONLY) Displays the selected light level; light levels are HIGH, LOW and OFF. ( I ) Coolant Level Displays the selected coolant solution flow level; the selected Flow level is indicated by 0 to 5 levels of illuminated indicators. No illumination indicates when the Coolant Solution Flow is off. (J) FWD/REV Direction Displays the selected direction of handpiece rotation. When rotating, the circumference display rotates. (K) GEAR Ratio Selected Gear Ratio of the handpiece. (L) PROGRAM Number Selected Program Number indication. (1 to 8)
NOTICE • Right after turning on the power source, the clock may display “88:88/00:00” or the LCD of the USB REC key may blink, but these are not malfunctions. • When the power source is turned on, the device will return to its default settings. To save a setting, see 6-1 Programming the Micromotor Operation. (Default: Implant system 1, Light HIGH, Coolant Flow 3, Rotation F, Gear Ratio 20:1, Program Number 1, Torque 20N•cm, Speed 800min-1) • When packed for shipping, etc., the clock will blink on “00:00” if the clock has not been set. See 5-1 Clock and Date. • When the clock display blinks on “- -: - -,” the clock batteries have run out. Contact the retailer where device was purchased. 6
3-3 Foot Control
English
Foot Control Cord and Plug
(b) PRG (Program) Button
(a) Coolant Solution Flow Volume Button
(d) Forward/Reverse Button
(c) Speed Control Pedal
(a) Coolant Solution Flow Volume Button Use to select 6 levels of coolant solution flow from 0 to 5; each level may be increased by one step pressing this button. If you press this button in level 5, it returns to level 0. (b) PRG (Program) Button Use to select the desired program number. Program numbers will always ascend each time this button is pressed and released. To descend, press and hold for 1 second then release. (c) Speed Control Pedal Use to start and stop the Micromotor and to control the speed during operation. The Micromotor operational speed is directly proportional to the position of the foot control, up to the preset maximum. (d) Forward/Reverse Button Use to change the rotational direction of the Micromotor. Push once to change the rotational direction.
4. Installation CAUTION Install the cord/plug only power is OFF.
4-1 Connecting the Motor Cord Align the [ ] mark on the Motor Cord with [ ] mark on the Control Unit. Push the Base of Motor Cord Plug into the socket until the locking joint ‘clicks’ into place. ە
ە
To disconnect the plug; pull back the Lock Joint, pull further to disconnect the cord. Lock Joint [ ] mark ە
Fig.1 7
4-2 Connecting the Handpiece CAUTION When using the Optic Surgic Pro micromotor (SGL70M), USE ONLY Optic contra-angle handpieces. Non Optic handpieces may not precisely connect to the micromotor and it may malfunction when the micromotor is under load. 1) Push the handpiece on to the motor taking care to keep it in alignment. 2) Twist the handpiece until it locks into position with a clicking sound.
Fig.2
4-3 Connecting the Foot Control Face the screw on the Foot Control Cord Plug downward then insert the plug into the Foot Control Cord connector socket on the Control Unit. Secure the plug by fastening the Lock Nut. Refer Fig.3.
Screw
4-4 Connecting the AC Power Cord Correctly align then insert the AC Power Cord into the AC Power Cord Connector Socket at the back of the Control Unit. (Fig.3)
Foot Control Cord Plug
AC Power Cord
Fig.3
4-5 Installing the Irrigation Tube Ensure the Pump Cover Lever is in the ‘OPEN’ position, with the Irrigation tube ‘Needle’ towards the backside of the Control Unit. Position the stoppers of the tube assembly in to the Tube Guide securely. (Fig.4) Pump Cover
<OPEN Position>
Tube Guide Fixed Position
Fixed Position Water flow direction
Pump Cover Lever Stopper
Fig.4
8
CAUTION
English
• Make sure that the tube is securely set on the rollers when closing the Pump Cover. If the tube is not correctly positioned on the Rollers and the cover is closed, the tube could be cut or sheared. (Fig.5) • Do not mount the Irrigation Tube except the person who specially educated. • Do not mount the Irrigation Tube during Power is ON.
Fig.5 <CLOSE Position>
After the tubes are correctly positioned, close the Pump Cover by turning the Pump Cover Lever to the ‘CLOSE’ position. (180 degrees counterclockwise) (Fig.6) To remove, ‘OPEN’ the Pump Cover Lever then remove the tube.
Fig.6
4-6 Mounting the Coolant Solution Hanger Post Mount the Coolant Solution Hanger Post onto the Holder on the Control Unit. (Fig.7) Align
Fig.7
4-7 Insertion of the Irrigation Tube 1) Close the Tube Clamp, between the Irrigation Tube Needle and the Irrigation Pump. (Fig.8) 2) Insert the Irrigation Tube Needle into the Bottle Cap. (Fig.9) 3) Open the Tube Cap to supply air into the bottle. (Fig.10) 4) Open the Tube Clamp. Tube Clamp Bottle Cap
Tube Cap
Irrigation Tube Needle
Fig.8
Fig.9
Fig.10
CAUTION Do not operate the Irrigation Pump if the tube is bent or the Tube Clamp is in the closed position. This could cause the tube to burst or slip out of the bottle. 9
4-8 Compatibility check of Internal Irrigation Nozzle/Drill Internal irrigation nozzles accompanied with this product; is not necessarily fitted into all the drills on the market. Follow the instructions given below for confirmation prior to use. Failure to do so or to fit the Internal Irrigation Nozzle into drills may cause a leakage of saline solution, which will result in problems such as rust or sudden stop of equipment during use. * To mount the Drill, refer to attached Handpiece Operation Manual. Instructions: 1) Attach a bottle of saline solution to the Control Unit. 2) Connect the Internal Irrigation Nozzle into the tip of the irrigation tube. 3) Insert the Internal Irrigation Nozzle into the drill from the back. (Fig.11) 4) Purge at “Maximum” for 5 seconds.
Irrigation Tube
Drill
Fig.11 Points to be checked: • Cleanliness of the saline solution coming out from the drill; if solution is colored there could be rust inside of the drill. If so renew the drill. • Water Flow; if the flow is low and/or the flow from the drill is asymmetric, renew the drill. • No water leakage between Internal Irrigation Nozzle and drill before use, ensure no water is leaking from the entry point of irrigation nozzle, a broken seal or no seal in the drill may be the cause. Replace the drill even if its new, saline solution ingress into handpiece will cause malfunction.
CAUTION If malfunction such as a leakage of saline solution from the back of Contra Head is detected during use, stop using and perform some troubleshooting.
4-9 Irrigation Nozzle Attachment X-SG20L and SG20 has 3 irrigation methods available depending on tool and application; External, Internal or both, for installation detail, refer to the attached handpiece Operation Manual.
4-10 Attaching the Tube Holder
Motor Cord
Use the Motor Cord as a strain relief for the ‘Irrigation Tube’. It is easier to insert Motor Cord first, then the Irrigation Tube. (Fig.12)
Tube Holder
4-11 USB flash drive (Not supplied) (Surgic Pro+ ONLY) 1) Open the Insertion Opening Cover on the Control Unit side. 2) Completely insert the USB flash drive. The USB mark will appear on the LCD DISPLAY when USB flash drive is recognized by the device. (Fig.13)
Irrigation Tube
Fig.12
CAUTION Do not apply strong impact to USB flash drive when it is connected. When the USB flash drive is not connected, close the Insertion Opening Cover.
USB flash drive Socket
Fig.13
5. Setting 5-1 Clock and Date By setting the date and time, data on treatment will be recorded by day. (See 6-5 USB Function to learn how to store data on a USB flash drive.) 10
• Set the Clock before you use this system. • If the clock display does not appear or is shown as “- - : - -” the internal batteries may have run out. Contact the retailer where device was purchased. • The clock runs on a 24-hour display system. • The years can be set from 2010 to 2099. 1) Push both the USB REC key and SPEED (-) for about 2 seconds. 2) The setting screen will appear on the LCD. Settings can be made for year, month, day, hour, and minute in that order. Input the data, and press the USB REC key to move to the next item. (The item you have chosen will blink.) 3) Change numbers by pressing the SPEED (+ -) key. A long push will increase numbers in increments of 5. 4) When all items are set, press the USB REC key for 2 seconds to save settings.
CAUTION • If the power goes off while settings are being made, the changes will not be saved. Start once more at the beginning. • If data is already stored in the internal memory, it will not be possible to set the date.
5-2 Illumination Adjustment (X-SG20L Only) Repeat press LIGHT key to cycle through light levels. Pressing LIGHT key to change illumination volume. (OFF Ѝ LOW Ѝ HIGH) (Default is HIGH) To save the changes, press the MEMORY key for at least 2 seconds.
NOTICE • Repeat the above procedures when restoring operation. • The lamp status will remain set even if the power is switched “OFF” and “ON”. • This function is only applicable for an optic motor and handpiece.
5-3 Sound Volume It is possible to change the volume for operation and warning sounds for keys. 1) Turn on the power while pressing the [CAL] key. 2) When turning ON, the volume will alternate between “Large” and “Low.” (The default is “Low”) * Sound Volume is not displayed on LCD. The actual sound changes.
NOTICE Settings will be saved even if the power is turned off.
5-4 Factory Settings When initialized, the device will revert to factory settings. Preferred settings and programs, etc. will be erased, so be sure to make memos of them. 1) Turn power on while pressing the PROGRAM (+ -) Key. 2) “Set” will come up on the LCD Display. (there will also be a short beep) By pressing [CAL] key, the device will revert to factory settings. 3) “Fin” will come up on the LCD Display. (there will also be a long beep) Initializing is complete when the screen goes back to its usual display.
CAUTION Even when the device reverts to factory settings, date and time data will not be initialized.
11
English
NOTICE
6. Operation 6-1 Programming the Micromotor Operation The Control Unit can memorize 8 programs. Each program includes the following functions which will be automatically performed when the appropriate program number is selected. Gear ratio of contra angle handpieces Speed Direction of rotation ٤ Torque upper limit Coolant Solution Flow Light Level
1) Turn on the power by pushing the Main Power Switch toward [-]; on power up IMPLANT SYSTEM Number 1 program number 1 is displayed by default. 2) Select a program number by using either step (a) or step (b): (a) Press the PROGRAM (+ -) Key on the Control Panel until the desired program number is displayed. (b) Press the PRG(Program) Button on the Foot Control until the desired program number is displayed. 3) Selecting the Rotate direction of the handpiece. Select it by using either step (a) or (b): (a) Press FWD/REV Key on the Control Panel until the desired direction is displayed. (b) Press the Forward/Reverse Button on the Foot Control until the desired direction is displayed.
NOTICE • There will be a beeping sound while “R” is pressed. • Beeping sounds heard during rotations and when the device is stopped are not the same. Different sounds do not indicate a malfunction. 4) Selecting the Gear ratio of contra angle handpieces relevant to the program; Press the GEAR Key to select the gear ratio of the handpiece (Gear Ratio will display on the LCD). 5) Set the required max operating speed by pressing the SPEED (+ -) Key. Each time this Key is pressed display changes to the next speed level. By pressing this Key for more than 1 second brings the speed quickly to the next level until the speed display reaches its upper or lower limit. - When the speed setting reaches the upper or the lower limit, an audible beep is heard and the speed setting cannot be changed any further. 6) Set the torque upper limit by pressing the TORQUE Key on the Control Panel. Each time this Key is pressed display changes to the next torque level. By pressing this Key for more than 1 second brings the torque quickly to the next level until the torque display reaches its upper or lower limit. - When the torque setting reaches the upper or the lower limit, an audible beep is heard and the torque setting cannot be changed any further. 7) The level of Coolant Solution Flow volume has 6 Flow level (0-5) (0 = no coolant Flow). Select a Coolant Solution Flow Volume using either step (a) or step (b): (a) Press the COOLANT Key on the Control Panel until the desired flow level is displayed. (b) Press the Coolant Solution Flow Volume Button on the Foot Control until the desired flow level is displayed. 8) After completing steps 2) - 7). press and hold MEMORY Key until beep is heard. The beep confirms that the programming is completed. If you hear a short beep when the MEMORY Key is first pressed ignore this sound and keep the MEMORY Key depressed until a long beep is heard. * Repeat the above steps 2) - 8) to program any one of the 8 available programs.
6-1-1 Program Skip Function When switching programs using the Foot Control, Program Numbers that are not chosen can be chosen ahead of time. 1) Choose Program Numbers that you want the Control Unit to skip. 2) Use the SPEED (-) Key to display the Minimum Rotation Speed. 3) Release the SPEED (-) Key, and then push it again. Display “---” on the LCD. 4) Push the MEMORY Key until it beeps (about 2 seconds) to memorize. 12
NOTICE
6-2 IMPLANT SYSTEM Surgic Pro Series can memorize 8 different IMPLANT SYSTEMs. Also, you can program 8 steps in one SYSTEM number which you use. When you set the program, refer to values which Implant manufacture recommends. When programming each IMPLANT SYSTEM, the SYSTEM number may be changed by pressing the PROGRAM (+) and (-) Key simultaneously.
6-3 Calibration Function The rotation resistance of a handpiece depends on the handpiece model, condition, and internal wear of the handpiece gears. This Control Unit incorporates an automatic function to recognize the level of the resistance of the attached handpiece and motor.
NOTICE • This equipment is optimized to obtain the highest calibration accuracy at a gear ratio of 20:1. For calibration USE ONLY the 20:1 reduction handpiece listed on “12. Contra Angle Handpieces and Accessories”. • Micro saw handpiece should not be calibrated. • Calibration should be performed only on NSK handpiece.
6-3-1 Calibration preparation 1) Attach Gear Ratio (Reduction Ratio) 20:1 handpiece to the motor. 2) Attach the Calibration Bur to the handpiece. 3) Press the [CAL] key for about 2 seconds. There will be a beep and the display will switch to Calibration Mode.
Calibration Bur
Fig.14
6-3-2 Conducting Calibration
CAUTION Do not apply load during no-load calibration. If load is applied, “FAIL” will appear on the LCD, and calibration will stop. 1) No-load calibration … torque display “L” Hold the motor in your hand and make sure that “L” is displayed on the torque display. Push [CAL] key without applying load. The calibration bur will begin to rotate. Calibration is complete when there is a beep and “donE” appears on the display. 2) Load calibration…torque display “H” Make sure “H” is displayed on the torque display, plug the calibration bur into the Torque Lord Push [CAL] key (The calibration is done while applying torque load, so securely grasp the handpiece. The calibration bur will begin to rotate. Calibration is complete when there is a beep and “donE” appears on the display). 3) Speed calibration Remove the bur from the Torque Lord then press the [CAL] key (Calibration bur should not touch anywhere). The bur will begin to rotate at a slow speed. After rotating slowly for 8 seconds, the bur will rotate at high speed for 8 seconds. There will a beep when calibration is complete. Calibration mode will finish automatically.
13
English
It is possible to choose programs that have been skipped using the Control Unit.
CAUTION During a process, it is important to hold the Calibration Bur vertically without applying any forcible power. Making the Calibration Bur tilt or pressing it may cause a loss in accuracy.
Calibration fail example • Extreme gear abrasion and mechanical loss can exceed the control limits. Correct data cannot be obtained during calibration cycle and the calibration will fail. In this case, request repair. • If nonconformities such as extreme abrasion of the gear and mechanical loss exceed the control limit, correct data cannot be obtained during calibration will fail. In this case, request repair. • Calibration fail; if the calibration bur makes contact with something during no load calibration or no load is erroneously applied during load calibration.
6-4 Standard Operation 1) Select the desired program number. Control Unit: Press PROGRAM (+ -) Key. Foot Control: Step on the Foot Control PRG (Program) Pressing the PRG (Program) Button for 1 second more will descend the to the next program number. 2) Step on the Speed Control Pedal in the middle of the Foot Control the motor will start to run. The Coolant Pump will also run (if programmed). Speed increases as the pedal is depressed. When the Speed Control Pedal is fully depressed the speed reaches the maximum preset value. 3) Reversing the motor rotational direction: To reverse direction of the motor (and bur) simply step on the Foot Control Forward/Reverse Button. A warning beep can be heard when the rotational direction is in reverse mode.
CAUTION • Using the Micromotor with light ‘ON’ for more than 3 minutes continuously, motor will become hot (The maximum surface temperature might reach 45˚C by the situation used). In such cases, stop using the Micromotor till the light part cools down, or use the Micromotor without turning on the light. • Activation of the torque limiter: During use, when the drilling load reaches the preset torque upper limit, the integrated torque limiter automatically activates to prevent over torque. When the torque limiter activates, the motor stops after beeping for 1 second. To reactivate the Micromotor, release the speed control pedal and depress it again. 4) Release the Foot Control Pedal, and the Micromotor will automatically stop.
6-5 USB Function (Surgic Pro+ ONLY) NOTICE When the clock is not set, the time and date will not be recorded on data. It is recommended that the clock be set before using the device. The internal clock is used along with the different operations of this device (date used, starting time, length of time, Gear Ratio,level of Coolant Solution Flow, Rotation Speed, FWD/REV, Torque, Program Number), making it possible to store data in chronological order on USB flash drive (CSV files, BMP files). Stored data can be viewed on a personal computer.
*Files and graphs saved
Folders are created in order by date CSV file
14
BMP file
NOTICE A BMP file can only be created only when Gear Ratio is 201 and the number of rotations is 50min-1 or less. Time
Date
Torque Irrigation Volume Rotation Speed Gear Ratio Program Number
English
*BMP File Torque
Set torque index
Time
CAUTION • Total recording time is about 100 minutes. • Do not insert a USB flash drive during operation. Only insert after operations are completed in order to store data. • Do not use data stored in USB flash drive for diagnostic purposes. • Do not use a USB flash drive other than USB 2.0. (To 32GB verification being complated) • Do not remove USB flash drive while the LED lamp is on or blinking. • The LED lamp and their meanings are given below: lamp on = internal memory is copying data, blinking = USB flash drive is storing data. • This device is not guaranteed to work with all USB flash drive. It may not work with USB flash drive with security functions or those which use a great deal of electricity. • Do not turn the device on or off with the USB flash drive inserted. • Be careful to insert the USB flash drive in the correct direction. • When a USB flash drive is not connected, close the cover of the connector. • If power is turned off, the internal memory will be erased. Turn off power after storing data on USB flash drive.
6-5-1 Internal memory After power is turned on and the motor begins to rotate, memorizing will begin automatically (A USB flash drive need not be connected).
6-5-2 Transmitting stored data 1) Insert a USB flash drive. (When correctly recognized, the USB flash drive mark will appear on the LCD). 2) Push the USB REC key to copy stored data onto the USB flash drive (During the control unit copying the data. LED lamp blink).
NOTICE • Distinguish rotating direction and torque that has been set using line types. • While in operation (memorizing), the LED lamp will light up. About 10 seconds after operation stops, the LED lamp will turn off and memorizing will stop temporarily (The time counter will also stop temporarily).
6-5-3 Resetting Press the USB REC key for at least 2 seconds. The amount of remaining recording time will be reset after the beep goes off (Initialized time: 100 min). 15
7. Protection Circuit An electronic circuit breaker automatically functions to protect the motor and the Control Unit if the motor is ever overloaded. Power supply to the motor will automatically be terminated an Error code will be displayed on the Control Unit. When torque exceeds the set figures, “SAFE” will appear on the LCD, and the device will stop. *Resetting the Protection Circuit To reset the Protection Circuit, release and then depress the Speed Control Pedal.
8. Error Code & Troubleshooting If an operational problem occurs the display will show an Error Code allowing immediate problem diagnosis. When an error occurs, there will be a beep and an Error Code will appear on the LCD. The Error Code will blink until the error is alleviated. The USB REC LCD will go on for about 10 seconds. 22/2$/%&
22/2/%&
"53&/'22/2
&-&%9
934&-22/2
&15&342&0"*2 &-/29'"*,52& 52(*$2/ 22/.&/53-&-/29/4 ;"3)%2*6&/4)&2 4)".4)&490&34*05,"4&% -/53&02*.4&2&4$*3 $/..&$4&%
8$&33*6&522&.4 &4&$4&%
84&.%&%53&5.%&2)&"69,/"% )/24$*2$5*4*.4)&-/4/2
8$&33*6& /,4"(& &4&$4&%
/4/2$/2%'"*,52&
,&$42*$",$/.4"$4-"9#&*.35':$*&.4 &$52&,92&$/..&$44)&/4/2/2% !)&.".&22/2$"../4#&&,*-*."4&%2&15&34 2&0"*2
/4/23&.3/2'"*,52&",, /4/2/2%'"*,52&
/4/2&.3/222/2 .(2&33/'7"4&2*.4/"/4/2
&15&342&0"*2 "+&352&4/054"2/4&$4*/.,5(7)&. )&2-/%*3*.'&$4*.(*4
/.42/,.*4.4&2*/2 6&2)&"4*.(22/2
,,/7*44/$//,%/7.#&'/2&53& 6&2)&"4*.(#9&84&.%&%53&5.%&2 *.$&)&"4*335':$*&.4,92"%*"4&%0&2*0)&29/' )&"69,/"% 4)&-"*./.42/,.*43)/5,%#& 0&2"4*/./'4)&/.42/,.*45.%&2 7&,,6&.4*,"4&%7)&2&6&20/33*#,&!)&.". ".&842&-&,9)*()4&-0&2"452& &22/2$"../4#&&,*-*."4&%2&15&342&0"*2
2&"+*.(22/2
#./2-",6/,4"(&(&.&2"4&%*.4)& 34"2434/037*4$)$*2$5*4 "*,52&*.4)&34"2434/037*4$)&3 $*2$5*4
!)&.2/4"4*/.".%34/0"2&2&0&"4&%*. 3)/24'2&15&.$*&3"$*2$5*4-"9#&"$4*6"4&% 7)*$),*-*43"$$&,&2"4*/."434"24!"*4"'&7 3&$/.%3".%4)&.53& !)&.".&22/2$"../4#&&,*-*."4&%2&15&34 2&0"*2
/4/2/4"4*/. "*,52&22/2
".%0*&$&"44"$)-&.4'"*,52& /4/2'"*,52&
)&$)5$+-"9#&/0&.&%/2-"9./4#& 35':$*&.4,9$,/3&%&$52&,9$,/3&4)&$)5$+ !)&.".&22/2$"../4#&&,*-*."4&%2&15&34 2&0"*2
16
*"%-((+$!) '')$#)*(#$'')!. /))#!$#)%*"%'$!!' *"%-((*''#) %*"%!(
)'')$#)* #''$' $((%!.+#,#) '')$#)*(#$'"!'&*()'%'
$$) $#)'$! #$'"!).
)$##)$#())$)$##)$' $##)$#!*'$)$##)$' ''$' $((%!.+#,#) !*'$)#($)$$) $##)$'(#$'"!!.$##)'&*() $#)'$! '%'
-((+$!) %)$)$'
!*'$ '*)
&*()'%'
-((*''#) %)$)$'
)(!#($!*)$#(#) %#)'))"$)$'!"%
"$+)(!#($!*)$#$"%!)!.
!*'$ '*)
&*()'%'
$$"$"
'%" $$"$
*
,%$("&$&!&&*#%&#'&%"!!& %"!!&'%!&#$"#$&*#" . .)%#*"!& %$! ,%$(
,%$( "'$'&$$"$
"&*$&$)$()%#'$%
&!!"&!%&"$"! ,%$("$&"!" &"")!$%"!% -"( "$*"! ,%$( - %" ,%$("!)"#*!%#$"& - ! %$*!'% -&$ ,%$($&$$"$
-# ,%$( - "("&&#$(!&%"#*! - !&+! -!%&$'&"!%"$ ,%$('%
!
""# !&!'( # ! !&# ! " "
" ""$""" ! !!
#!
&
#"% !# !" & ! "
"#!
%#!
#!
#"% !%"
#!"
% !%"" ! !!
$" ""
"" #"
" !!$& $ """ """ ! #" ! #
" "#"
" !"!$& $""" "#" !#" ! #
"!!
"
#"" " " "
" !"!$& ""!&!" ! #
17
English
9. Fuse Replacement If the Control Unit does not function, check the fuses (Fuse Box lock located on the rear of the Control Unit). To access the Fuse, use a pointed tool push on the fuse locking latch and the drawer will spring open. (Fig.16)
Fuse
Fig.16
10. Maintenance 10-1 Maintenance of the Control Unit and Foot Control If blood or saline solution has stained the Control Unit or Foot Control, remove the AC Electrical Cord; wipe Control Unit or Foot Control with a damp cloth, then with an alcohol-absorbed cloth.
10-2 Maintenance of the Handpiece and Motor This handpiece and motor can be cleaned and disinfected with a Thermo-Disinfector.
Protection Plug
Attach the Protection Plug to the motor. Protection Plug O-Ring
For detail, refer to the Thermo-Disinfector’s Operation Manual.
Fig.17
CAUTION • Always use the Protection Plug during Thermo-Disinfector. • Do not use Protection Plug without the mounting O-Ring. It may cause a malfunction. If O-Ring had damaged, replace it immediately.
NOTICE Protection Plug O-ring is consumable supply. If it damaged, replace it new one. Optional Protection Plug: Order Code 10001595 Protection Plug O-Ring: Order Code 0312457102
18
11. Sterilization • Autoclave sterilization is recommended. • Autoclave sterilization is required for the first time use and after each patient as noted below.
CAUTION Only Items Specified below can be autoclaved. The following items can be autoclaved. • Handpiece • Micromotor with Motor Cord • Internal Irrigation Nozzle • Tube Holder • Protection Plug • Calibration Bur
• Handpiece Stand • Nozzle Holder
*Autoclaving 1) Remove blood and debris from the handpiece and motor. 2) Clean inside the handpiece; by using the spray lubricant 3) Place those in autoclave pouch (not included in the package) and seal it. 4) Autoclavable up to max. 135˚C. ex.) Autoclave for 20 min. at 121°C, or 3 min. at 134°C. 5) Keep the handpiece in the autoclave pouch to keep it clean until you use it. * EN13060/EN ISO17665-1 recommends autoclaving for 3 minutes (minimum holding time) at 134°C or 15 minutes(minimum holding time) at 121°C. NSK recommends Class B sterilization.
CAUTION • Clean and lubricate the handpiece before autoclaving. Autoclaving a handpiece stained with blood or debris could cause damage to the handpiece. • Do not lubricate the micromotor. • Do not disconnect the motor cord from the micromotor. • The Irrigation tube is a single use and cannot be autoclaved. • We do not recommend you to sterilize on plasma sterilization or EOG sterilization.
19
English
This handpiece and motor can be autoclavable up to Max.135°C.