NSK NAKANISHI
VarioSurg3 Ultrasonic Bone Surgery System Operation Manual Feb 2022
Operation Manual
145 Pages
Preview
Page 1
OM-DE1032MA 000
- Contents -
1. User and Intended Use... 1 2. Precautions for handling and operation... 1 3. Package Contents... 4 4. Component Names... 5 5. Installation... 8 6. Check before treatment... 13 7. Operating method... 13 8. Various functions... 14 9. Maintenance... 15 10. Periodical Maintenance Checks... 17 11. Error code... 17 12. Troubleshooting... 18 13. Specifications... 20 14. Classification of Equipment... 20 15. Operation Principle... 20 16. Symbol... 21 17. Warranty... 21 18. Option Parts List... 22 19. Spare Parts List... 22 20. Disposal of the product... 22 21. EMC Information (Electromagnetic Compatibility Information)... 22
1 User and Intended Use User: Qualified Professionals Intended use: Dentistry and oral surgery (implant site preparation, bone cutting, maxillary sinus floor (mucus membrane) elevation, osteoplasty, bone resection in radectomy, periodontal operation, prosthesis maintenance, surgical endodontic procedure
2 Precautions for handling and operation • Please read these precautions carefully and use only as intended or instructed. • Safety instructions are intended to avoid potential hazards that could result in personal injury or damage to the device. Safety instructions are classified as follows in accordance with the seriousness of the risk. Class
WARNING CAUTION NOTICE
Degree of Risk A hazard that could result in serious injury or damage to the device if the safety instructions are not followed. A hazard that could result in light or moderate injury or damage to the device if the safety instructions are not followed. General product specification information highlighted to avoid product malfunction and performance reduction.
1
English
English Thank you for purchasing the VarioSurg 3. Please read this Operation Manual carefully before use to become familiar with operation instructions and care & maintenance. Keep this Operation Manual for future reference.
WARNING • Do not handle the power cord with wet hands. Wet hand contact with electricity may result in an electric shock. • Keep away from explosive substances and flammable materials. • If the product overheats or smells of burning, immediately turn off the power and disconnect the main power plug. Contact your Authorised NSK Dealer. • TO PREVENT ELECTRIC SHOCK, use a main electrical outlet that is protective earthed. • Do not use on the following patients. -Those with medical complications or allergies -Those who have pre existing conditions (Eg Cardiac, Pulmonary, Renal disturbance or High blood pressure) -Those who are pregnant or lactating -Patients with cardiac pacemakers and infants • Be careful not to get water or liquid disinfectant on the control unit. This could cause short circuits and lead to fire and/ or electric shock. • Repeatedly turning the Main Power Switch ON and OFF may blow a fuse. • When installing the product, provide space of approximately 10cm around the product for easy access to the inlet and the Power Cord. • The Irrigation Tubes, included in the product package, are sterile. When using these products, follow the instructions below. - The irrigation tube is a single-use item. Do not reuse nor sterillize as product breakage or infection could occur. - Check before use that the sterile package is NOT open nor damaged. Do not use products with open or damaged packages as sterility will be negated and infection could occur. - Observe the use-by date written on the sterile package label. Do not use expired products as sterility may be negated. • Do not point the handpiece LED light directly to the eyes of the patient, nor the operator, as it may be harmful to eyes.
CAUTION • To be confident in using this product, please read this Operation Manual before use. We suggest that to fully understand the product functions and file for future reference. • When operating the product always consider the safety of the patient. • The end user shall be responsible for any judgment that relates to the application of this product to a patient. • This product does not consider patient’s age (except infants), gender, weight or nationality. • This product does not consider operator’s age (mature person), height, weight, gender, or nationality. • Users are responsible for the operational control, maintenance and continual inspection of this product. • This device is for indoor use only. • Keep the Control Unit on a level surface. • Do not attempt to disassemble the product or tamper with the mechanism except as recommend by NSK in this Operation Manual. • Do not allow any impact on to the handpiece. Do not drop the handpiece. • Operators and all others in the area must wear eye protection and a mask when operating this handpiece. • Should the product function abnormally, cease operation immediately and contact your Authorized NSK Dealer. • If the Irrigation Pump gets wet, wipe it dry. If it remains wet, the roller in the pump may slip and the pump may not operate properly. • Do not bend or fold the Irrigation Tube while the Irrigation Pump is operating as the Tube may be damaged or may detach. • If abnormality is detected in irrigation, it may be due to wear of the Irrigation Tube or leaking of saline from the Tube. In such cases, replace the Irrigation Tube. • Do not use high acid water or sterilizing solutions to wipe, immerse or clean the product. • The following products are delivered in a non-sterile condition and must be autoclaved prior to use. (Handpiece, Tip, Tip Wrench, Tip Holder, Tube Holder) • Perform regular function and maintenance checks. • If the product is not used for a long period check it is functioning correctly before using on a patient. • To avoid clinical downtime it is recommended that a spare be kept on hand in case of a breakdown during surgery. • This product may be affected by an electric scalpel. Turn OFF the Power Switch when an electric scalpel is used. 2
NOTICE • No special training is required for this device. • The handpiece or handpiece cord during operation could interfere with computers, LAN cables in close vicinity, or could cause noise in radio receivers nearby. *Emission: Unnecessary energy generated by the device. *Immunity: Tolerance to unnecessary energy received by the device.
3
English
• This product is rated Medical Electrical equipment. EMC (Electromagnetic compatibility) is described in the documentation included. • Installation and use of this product requires special precautions regarding EMC according to the EMC information. • Portable and mobile RF communications equipment can affect Medical Electrical equipment. Do not use RF equipment near the product. • The use of ACCESSORIES such as handpieces and cables, other than those specified by the manufacturer, with the exception of handpieces and cables sold by the manufacturer of this product as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of this product. • Do not place this product adjacent to, or on top of, another device. This product and NSK Surgic Pro can however be stacked for use with the NSK Link connection. • The system may present a possibility of malfunction when used in the presence of an electromagnetic interference wave. Do not install the system in the vicinity of any device which emits magnetic waves. Turn OFF the Main Power Switch of the system an ultrasonic oscillation device or an electrode knife is located close to the vicinity of use. • The handpiece may get hot when some types of tips are used at high power level or in an overloaded state. (The temperature of the handpiece surface might exceed 41˚C.) If such abnormality is found, stop using for about 5 minutes to allow the handpiece to cool. • Surface temperature of tip shall be more than 65 degree without using coolant water. To avoid this event, be sure to use coolant water. Power plug below is used in North America area. • For added safety install the Control Unit in a place where the AC POWER Cord can be easily removed. (It is possible to disconnect the control unit from the power source by removing the AC POWER Cord.) • Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle marked "Hospital Only" or "Hospital Grade". Plug Type NEMA 5-15P (Hospital Grade Type) • U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
3 Package Contents 1
6
2
5
7
8
3
4
9
10
11
No. Part Name 1 Foot Control 2 Control Unit 3 AC Power Cord 4 LED handpiece with cord (Unshielded 2m) 5 Irrigation Tube 6 Irrigation Pole 7 Handpiece stand 8 Tube Holder 9 Tip Wrench 10 Tip 11 Tip Holder 12 Sterilization Cassette
4
12
Quantity 1 1 1 1 5 1 1 7 1 6 1 1
4 Component Names 4-1 Control Unit
Front Panel
Irrigation Pump
Coolant Solution Hanger Post Holder AC Power Cord Connector Socket
Main Power Fuse Box Switch
Handpiece Cord Connector Socket
SG Link Cable Socket
English
LCD Display
Foot Control Cord Connector Socket
Operation Panel
(11) POWER key (10) PROGRAM key (8) AUTO CLEANING key (9) MEMORY key MODE key (7) SELECT key
(1) LIGHT key
(3) P key (5) S key (PERIO Mode) (SURG Mode) (2) COOLANT (4) E key (6) BURST key key (ENDO Mode)
4-1-1 Keys on the Operation Panel (1) LIGHT key To select the Handpiece illumination intensity: OFF / LOW / HIGH (Default setting : HIGH) (2) COOLANT key To select the Coolant Flow from 5 levels. (3) P key To set the PERIO Mode (the mode suitable for maintenance, etc.). (4) E key To set the ENDO Mode (the mode suitable for the root canal treatment). (5) S key To set the SURG Mode (the mode suitable for cutting bones).
5
(6) BURST key To set the Burst Output Level (only for the SURG Mode) Each time the key is pressed, the setting changes as follows: OFF B1 B2 B3 OFF. *Burst function: Increases and decreases vibration at regular intervals. Select an appropriate setting according to the hardness of the patients’ bones (bone density). *Frequencies for Burst are B1: 10Hz, B2: 30Hz and B3: 60Hz. (7) SELECT key This key is used when two systems (NSK VarioSurg 3 and NSK Surgic Pro) are operated by a single Foot Control. *Surgic Pro (sold separately) and the Link Set (sold separately) are required in order to use this key. (8) AUTO CLEANING key For cleaning the irrigation circuit in the Handpiece and tube. (9) MEMORY key To set the program memory for the values which are displayed in the Panel. (10) PROGRAM key (- +) For calling up the set values memorized in the program. (11) POWER key (- +) For setting the Power Level. By pressing (+) and (-) keys, the set value increases and decreases respectively. *When the key is held down, the value increases or decreases continuously. However, when (+) key is held down, the value stops at 100% for the sake of safety. If you want to further increase the value, press the key again. NOTICE • Program numbers are allocated to each Mode as shown in the following table. Each program number can be set only for the allocated Mode. Program No. 1 2 3 4 5 6 7 8 9
Mode
SURG
Burst function*1
Coolant flow (5 levels)
Available (Off, B1: 10Hz, B2: 30Hz, B3: 60Hz)
10 – 150% In multiples of 10%
Not available
5 – 100% Below 50%, multiples of 5%. 3 - 55 ml/min*2 50% and larger, multiples of 10%.
ENDO PERIO
Power range
Lighting brightness
17 - 95 ml/min*2 OFF/LOW/HIGH
*1 Burst function: Vibration levels can be changed at certain intervals. Levels can be selected depending on patient’s bone hardness (density). *2 The amount of water flow may vary slightly depending on the condition of the Irrigation Tube. • While the Handpiece is vibrating, the following keys cannot be operated. (3) P key (4) E key (5) S key (7) SELECT key (8) AUTO CLEANING key (9) MEMORY key (10) PROGRAM key (- +)
6
4-1-2 LCD Display on the Operation Panel (B) POWER Display
(C) POWER Level Display
English
(A) Surgic Pro Display
(D) Program Number
(E) POWER Bar Graph
(F) Brightness (G) Coolant Flow Level
(H) Mode
(I) Burst Output Level
(A) Surgic Pro Display This is displayed only when the this VarioSurg 3 is connected to NSK Surgic Pro via the NSK Link Set and Surgic Pro has been selected by SELECT key. *For the Link function, refer to “8-3 Link function with Surgic Pro”. (B) POWER Display Displays when vibration is active from the Handpiece. (C) POWER Level Display Displays the set Power Level (5 – 150%). (D) Program Number Displays the selected program number. (E) POWER Bar Graph The relevant POWER level is indicated by the number of indicator bars illuminated. Each lower horizontal bar represents 50 %. Each upper vertical bar represents 10%. The maximum number of bars is 15. In Auto Cleaning mode, the countdown of the remaining time is displayed. (F) Brightness Displays the selected illumination intensity. (G) Coolant Flow Level The selected Coolant Flow is displayed on a scale of 1 - 5 (1 is the least flow). (H) Mode Displays the selected Mode. (I) Burst Output Level Displays the Burst Output setting. Displays only in SURG Mode. NOTICE • When the transparent protective sheet on the Operation Panel is peeled off or when an object charged with static electricity is placed near the LCD Display, fine lines may appear in the display area. This is normal and the lines will disappear after a short time.
7
4-2 Foot Control
Foot Control Cord Plug
(c) Program Selection Button
(d) Burst Output Mode Selection Button
(b) Coolant Flow Level Selection Button
(a) Ultrasonic ON-OFF button
(a) Ultrasonic ON-OFF Button To activate Tip Piezo vibration depress this control. (b) Coolant Flow Level Selection Button The Coolant Flow can be selected from 5 levels. Each time the button is depressed the flow will increase by 1 level. If the button is depressed when the Flow is in the level 5 it returns to level 1. (Level 0 cannot be selected for the Coolant Flow.) (c) Program Selection Button A specific Program Number can be selected. Each time the button is depressed the Program Number increases by one. When the button is held down for approximately 1 second the program returns to the previous number. (d) Burst Output Mode Selection Button Burst Output Mode setting can be selected.
5 Installation 5-1 Connection of the LED handpiece & cord
1) Position the Handpiece Cord so that the mark is facing up and is aligned with the mark on the Control Unit. Firmly push the Cord Coupling into the Handpiece Cord Socket until it clicks. 2) Confirm that the Cord Coupling is properly locked in position by pulling it gently. To disconnect the Coupling pull back the Lock Joint then pull back further to disconnect the cord.
Mark
Lock Joint
Fig. 1
8
1) Turn the Foot Control Cord Plug so that the screw faces down. Insert the Plug so that it correctly fits into the Foot Control Connector Socket of the Control Unit. 2) Secure the lock nut of the Foot Control Cord Plug.
Lock Nut
The screw should face down.
English
5-2 Connection of the Foot Control
Foot Control Cord Plug
Fig. 2
5-3 Connection of the AC power cord
1) Turn the Power Switch OFF. ( side) 2) Insert the AC Power Cord so that it correctly fits into the AC Power Cord Connector Socket at the back of the Control Unit. 3) Plug the AC power cord into a wall outlet specified for medical use. AC power cord
Fig. 3 • Do not tug the AC power cord by holding and pulling its cord. CAUTION • Make sure that the device has completely stopped before removing or inserting the AC power cord.
5-4 Connection of the Irrigation Tube
1) Turn the Pump Cover lever for the Irrigation Pump, located at the side of the Control Unit, clockwise by 180 degrees to open the Pump Cover. 2) After checking the irrigation direction, insert the tube into the Irrigation Pump. 3) Align the Tube Clamps of the Irrigation Tube with the tube guides and set the Tube. 4) Turn the Pump Cover Lever anticlockwise by 180 degrees to close the Pump Cover.
<When the pump cover is open>
Pump cover lever
Fig. 4 <When the pump cover is closed>
Pump cover Tube guide Align
Align Direction of Irrigation
Fig. 6
Tube Clamp
Fig. 5
9
• Before closing the Pump Cover, make CAUTION that the tube is correctly positioned on the rollers. If the tube is not correctly positioned, and the Cover is closed, the tube could be crimped or sheared. • Before fitting the Irrigation Tube, make sure the Irrigation Pump is stopped. Fig. 7
5-5 Connection of the Coolant Solution Hanger Post Align the Coolant Solution Hanger Post onto the Control Unit holder and firmly push it down.
Align
Fig. 8 • Insert the Coolant Solution Hanger Post all the way down firmly. If the Post guide and the groove guide are CAUTION not aligned, the Post cannot be inserted all the way down. • Use a pack of solution with the capacity of less than 500ml. Maximum pack size is 800g for the Coolant Solution Hanger Post.
5-6 Insertion of the Irrigation Tube
1) Close the Tube Clamp which is placed between the Irrigation Needle and the Irrigation Pump. (Fig. 9) 2) Hang the Solution Pack onto the Coolant Solution Hanger Post. 3) Insert the Irrigation Needle into the Pack entry point. (Fig. 10) 4) Insert the other Irrigation Tube tip into the handpiece Irrigation Nozzle. (Fig. 11) 5) Open the cap to let air into the Pack. (Fig. 12) 6) Open the Tube Clamp.
Tube Clamp
Fig. 9
Mouth of the pack Cap
Irrigation Needle Irrigation Nozzle
Fig. 10
Fig. 11
10
Fig. 12
English
• Always use a Pack with a sufficient quantity of saline. CAUTION • Before opening the Tube Clamp, always close the Pump Cover. If the Tube Clamp is opened while the Pump Cover is open, saline will flow out of the Irrigation Tube Tip. • When the coolant solution runs short, stop using the system and replace the Coolant Solution Pack. • Be sure to check that the coolant solution is flowing properly before using the system again.
5-7 Connection of the Tube Holder
Push the Tube Holder on to the Handpiece Cord, then push the Irrigation Tube into the Tube Holder.
Handpiece Cord
Tube Holder
Irrigation Tube
Fig. 13 • Bundle the Handpiece Cord and the Irrigation Tube at a CAUTION total of 7 locations.
Fig. 14
11
5-8 Mounting the Tip
1) First, lightly screw in the Tip with by hand. (Fig. 15) 2) Match the square shape of the Tip with the hole of the Tip Wrench, then insert the Tip into the hole. (Fig. 16) 3) Rotate the wrench in the (clockwise) direction, (Fig. 17 and 18) until it makes a click sound and cannot rotate any further. * To remove the Tip, rotate the wrench in the (counter-clockwise) direction shown in Fig. 17 and Fig. 18.
Fig. 15
Fig. 16
Tighten
Loosen
Tighten Loosen
Fig. 17
Fig. 18
• When attaching or removing a Tip, which is longer than the Tip Wrench, be careful that the protruding Tip CAUTION does not cause an injury. • Do not use a Tip which has been damaged, bent or corroded as it will be weakened and may break while in use. • To avoid bending of the Tip thread or disengaging of the Tip, always make certain that the Tip is inserted at the correct angle and firmly tightened. • If debris is found on the Tip thread clean the Tip immediately so as to avoid possible weak vibration • Do not attempt to sharpen or bend the Tip as this may cause the Tip to break during operation, or the vibration may be weakened. • When Tip become worn vibration may weaken or the Tip may suddenly break. If a weak Tip is detected replace with a new one. • Tips need to correctly inserted and tightened using the Tip Wrench. If the Tip is not tightened sufficiently vibration may be weak. • When inserting a Tip always wear surgical gloves and ensure the Tip, Handpiece and Tip Wrench are in an hygienic condition. • Before connecting or disconnecting the Handpiece Cord or Irrigation Tube make sure to remove the Tip first to avoid injury by the Tip. • Tip Wrench will wear with frequent use time and may need annual replacement. • Use only genuine NSK Tips. Other brand Tips may cause malfunction, damage or injury: - Poor vibration when an incompatible thread has been forcibly fitted. - The patient may swallow pieces of broken non genuine Tips. - Damage or premature wear of the Handpiece thread.
12
5-9 Handpiece Stand
When the Handpiece is not in use, place the Handpiece horizontally on the Handpiece stand.
Horizontal
English
• Make sure that the Tip does not rest on the Handpiece Stand. CAUTION • Make sure that the Handpiece rests horizontally on the Handpiece stand, and not at an angle. • To maintain an hygienic condition, the Handpiece Stand should frequently be cleaned with alcohol. Not horizontal
No Tip contact Fig. 19
6 Check before treatment Before use, operate the product without patient contact and check the following points. If any abnormality, such as low vibration, noise, heat, is detected at anytime disconnect the product and contact your Authorized NSK Dealer. -Is irrigation solution flowing correctly at the Tip? -Is the Tip vibrating correctly? -Are there any abnormalities at the Tip, Eg. vibration, noise and heat? -Is the Handpiece illumination operating correctly?
7 Operating method 7-1 Check Irrigation
1) Confirm that the Irrigation Tube correctly connected to the Coolant Solution Pack, the Irrigation Pump and the Handpiece. 2) Turn ON (-side) the Power Switch. 3) Open the Tube Clamp which is placed between the Irrigation Needle and the Irrigation Pump. Then, depress the Ultrasonic ON/OFF Button on the Foot Control and confirm that the irrigation solution correctly flows from the Tip. When a new Irrigation Tube is used, it may take several seconds before irrigation solution flow commences.
7-2 Set the Power Level
1) Change the Mode – Use MODE key. 2) Select the Program No. – Use PROGRAM key. 3) Adjust the Power Level – Use POWER key. 4) Adjust the Coolant Flow Level – Use COOLANT key. 5) Adjust the Handpiece illumination intensity – Use LIGHT key. SURG 6) Select the Burst Output Level – Use BURST key. (In SURG Mode only).
10%
ENDO
* Power outputs at various Modes are compared in Fig. 20.
5%
PERIO
5%
150%
100% 00% 100% OUTPUT
Fig. 20
13
• Use the tip ONLY in the correct mode and never exceed the maximum power defined for the tip by the tip CAUTION power guide. If used outside the prescribed power range the Tip may break, and the tooth surface or soft tissue may be damaged. • ONLY tips specified for maintaining prostheses can be applied to prostheses. Contact of other types of tips with prostheses may cause Tip breakage.
7-3 Activation
The Tip will commence vibrating when the ON-OFF Foot Control button is depressed.
• Start with a low power setting within the power range, observing the patient’s reaction and according to CAUTION the treatment plan. • Always use the irrigation. If the irrigation is not sufficient, the Handpiece may excessively heat, or oral injury may occur. • Do not apply excessive pressure to the Tip. • Do not place the active Tip anywhere other than the specific treatment area, as the ultrasonic vibration may cause oral injury. • Even if the power level is correctly set for the Tip in use application of excessive pressure to the Tip may cause Tip damage. Should Tip breakage occur, immediately evacuate the oral cavity of any Tip pieces using strong suction. • If Tip vibration reduces during operation, release the Foot Control activation button, and then again depress the Foot Control activation button. If the vibration level still does not return to normal check that the Tip is correctly and firmly positioned. Incorrectly inserted Tips may loosen during operation and power loss may be experienced. • Do not connect or disconnect the AC power or the Handpiece Cord while Foot Control is depressed.
8 Various functions 8-1 Memory function
The set values displayed in the panel (Eg. Mode, Program No., Power Level, Irrigation Flow Level, illumination intensity) can be memorized in each program. Once memorized, the programs can be called up whenever required. 1) Press the PROGRAM key and select the Program No. for memory. 2) Check the values of the parameters to be stored. 3) Hold down the MEMORY key for approximately one second. When the notification tone beeps, memorizing the setting is complete.
8-2 Volume selection function for the notification tone
Sound volume of the notification tone can be selected between high & low. 1) Make sure the power is off. Then turn ON the power while holding down the BURST key. 2) The notification tone in the selected volume sounds. 3) To change the sound volume, repeat 1) and 2) above.
8-3 Link function with Surgic Pro
This product has a link function whereby two systems (NSK VarioSurg 3 and NSK Surgic Pro) can be operated via one Foot Control. * To use this function, Surgic Pro (sold separately) and the SG Link Set (sold separately) are required. For the Link function, refer to the Operation Manual included with the SG Link Set.
14
When initialized, the device will revert to factory settings. Preferred settings and programs, etc. will be erased, so be sure to make memos of them. 1) Turn power on while pressing the PROGRAM (+ -) key. 2) “SEt” will come up on the LCD Display. (there will also be a short beep) By pressing BURST key, the device will revert to factory settings. 3) “FIn” will come up on the LCD Display. (there will also be a long beep) Initializing is complete when the screen goes back to its usual display.
9 Maintenance After each patient maintain the product as follows.
9-1 Auto Cleaning
1) Remove the Irrigation Needle from the solution pack. 2) Place the Irrigation Needle in distilled or deionized water in an open container. 3) Place the Tip of the Handpiece into the water. 4) Start Auto-Cleaning by holding down the AUTO CLEANING key for approximately one second. While Auto-Cleaning is active “CLn” appears in the POWER Display on the LCD Display. The Bar Graph indicates the time remaining (Auto Cleaning lasts for approximately 30 seconds). *Auto Cleaning can be stopped anytime by pressing AUTO CLEANING key. 5) Once Auto Cleaning has completed, the LCD Display returns to normal display.
Fig. 21
• Perform Auto Cleaning after every use. If Auto Cleaning is omitted, the Handpiece may clog or malfunction. CAUTION
9-2 Cleaning the LED handpiece with cord
1) Detach the Handpiece Cord from the Control Unit. 2) Remove the Irrigation Tube from the Handpiece and dispose. 3) Remove any debris from the product. Do not use a wire brush. 4) Wipe clean with alcohol.
This icon denotes that the product can be washed via Thermo Disinfector. At the print date of this Operators Manual, the Miele (Model:G7882) is the only Thermo-Disinfector verified for use with the VarioSurg 3. Refer to the Thermo-Disinfector manual. • When using in a Thermo-Disinfector, ensure that the Handpiece and Handpiece Cord are stable by placing CAUTION them in a basket or container. • Never use solvents, such as benzene and thinner for cleaning. • The Handpiece can not be removed from the Handpiece Cord.
15
English
8-4 Factory Settings
9-3 Cleaning the Glass Rod
Wipe clean the Glass Rod tip with an alcohol immersed cotton swab. Remove all debris. (Fig. 22)
Fiber end face
• Do not use needles or blades when cleaning the CAUTION Glass Rod as the glass surface may be scratched and illumination intensity will be reduced.
Cotton swab
Fig. 22
9-4 Cleaning the Control Unit and the Foot Control
1) Turn OFF the power and disconnect the AC power cord from the wall power socket. 2) Wipe the surface, first with a moist cloth, then with an alcohol soaked cloth.
9-5 Cleaning other components
Remove surface debris with a bristle brush (NOT METAL BRISTLES) then wipe the surface with an alcohol soaked cloth.
9-6 Sterilization
Autoclave sterilization is recommended. Autoclave sterilization is required for before the first use and then after each patient as noted below. The following items can be autoclaved. LED Handpiece with cord, Tip, Tip Wrench, Tip Holder, Handpiece Stand, Tube Holder, Sterilization Cassette. 1) Place the Handpiece and Tip Wrench in the Sterilization cassette. (Fig. 23) 2) Place the Tips onto the Tips Holder before putting the Tips Holder in the Sterilization Cassette. 3) Commence Autoclave sterilization. Autoclavable under the conditions below. Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or 3 min. at 134°C. 4) To maintain sterility, keep the product sealed in the autoclave pouch until required for use. LED Handpiece with cord
Tip Holder
Tip Wrench
Tip holder alternative location.
Tip Replacement Wrench (for V10-S) and VS Tip Wrench (Option)
Tube Holder
Sterilization Cassette locations for VarioSurg 3 components
Fig. 23
(c) Program Selection Button
• Do not place Maintenance Tips in the Tip Holder, as these Tips are too long. Use an autoclave pouch when CAUTION sterilizing Maintenance Tips. • Do not autoclave the product with other instruments even when it is in a cassette. This is to prevent possible discoloration and damage to the product from chemical residue from other instruments. • Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The 16
NOTICE • NSK recommends Class B Sterilizers as stated in EN13060.
10 Periodical Maintenance Checks Perform periodical maintenance checks every three months, referring to the check sheet below. If any abnormalities are found, contact your Authorized NSK Dealer. Points to check Vibration Irrigation Illumination Display Foot Control
Details Operate the Handpiece and confirm that there are no vibration, noise, or heat abnormalities Confirm that there are no abnormalities in the Coolant Flow Level or no leaks Confirm that the Handpiece LED illumination is correctly functioning Immediately after AC power is switched ON, all display lights should illuminate. Confirm that all display lights are working Confirm that all the buttons operate normally
11 Error code If the device stops operating due to excessive pressure, Tip or cord disconnection or incorrect usage, the LCD Display will show an error code. If this happens, release, then depress the Foot Control to repeat the error check. If the problem self corrects the error display will cancel and operation can be resumed. If the error display is repeated refer to the table below and take appropriate action. Error code
E2
E4
E8
Description of error
Cause of error Check / Remedy The end of the Tip is under too Avoid applying excessive pressure. If the Tip much pressure. is caught during a cutting operation, press the Ultrasonic ON-OFF Button and gently extract the Tip while vibrating it. (Never use force) The Tip is not correctly Insert the Tip and tighten it using the Tip Wrench Vibrator error inserted. The Tip is not until it clicks into position. tightened sufficiently. Check the connection of the Handpiece Cord. The Handpiece Cord is not If the problem persists, contact your Authorized properly connected. NSK Dealer. Breaking in the Handpiece Cord. Increase in temperature Turn OFF the power, allow time to cool, then Overheating inside the inside the Control Unit due operate the device again. Control Unit to prolonged use under high If this error code is displayed frequently, contact pressure. your Authorized NSK Dealer. The Irrigation Tube is caught Check the Irrigation Tube is correctly connected Pump error in the Pump Roller. If the problem persists, contact your Authorized Failure of the Pump. NSK Dealer. 17
English
air should be free from fine dust, salt and sulphur. • CAUTION Do not heat or cool the product too quickly. Rapid temperature change could cause damage to the product. • If the sterilizer chamber temperature exceeds 135°C during the dry cycle then delete the dry cycle. • Autoclave sterilization is recommended for the product. The validity of other sterilization methods is not confirmed. • Do not touch the product immediately after autoclaving as it will be very hot and it must remain in a sterile condition until used. • After autoclave sterilizarion if there are water drops on the LED Handpiece with Cord, wipe them away, as they may cause discolouration.
12 Troubleshooting If a problem is detected, check the following list before requesting service. If none of the listed remedies corrects the fault contact your Authorized NSK Dealer. Trouble Cause Control Unit cannot be turned AC Power Cord is not connected to Control ON (LCD display does not turn Unit. on) AC Power Cord is not plugged into wall outlet. Fuse has blown. Control Unit turns ON, Foot Control is being operated. Ultrasonic Output is displayed, and beep sound is heard.
Remedy Check the connection. Check the connection. Service only by Authorized NSK Dealer.* To prevent accidents, the Handpiece will not work if the Main Power is turned ON while the Foot Control is being operated. Activate the Foot Control again.
Foot Control is not connected. Check the connection. The Tip does not vibrate (no indication of Ultrasonic Output Handpiece Cord is not connected to Control Check the connection. Display). Unit. The Tip does not vibrate Faulty circuit. Service only by Authorized NSK Dealer.* (Indication of Ultrasonic Output Display). Weak vibration The Tip is worn or broken. Replace with a new Tip. Wrong mode. Change to the correct mode. The Tip is not correctly tightened. Tighten the Tip until the Tip Wrench clicks. Incorrect power setting. Change to the correct mode and power level for the Tip as indicated in the tip power guide or tip operation manual. Failure inside the Handpiece. Service only by Authorized NSK Dealer.* Failure inside the Foot Control. Service only by Authorized NSK Dealer.* The Tip breaks easily. Power Level is not right for the Tip Change to the correct mode and power mounted. level indicated in the tip power guide or tip operation manual. The Tip comes off. The Tip is not correctly tightened. Tighten the Tip until the Tip Wrench clicks. Handpiece makes a loud Power Level is not correct for the selected Change to the mode and power level noise. Tip. indicated in the tip power guide or tip operation manual. Tip is not correctly tightened. Tighten the Tip until the Tip Wrench clicks. Failure inside the Handpiece or the Control Service only by Authorized NSK Dealer.* Unit. Handpiece becomes hot. Power Level is not right for the selected Change to the mode and power level Tip. indicated in the tip power guide or tip operation manual. Tip is not tightened sufficiently. Tighten the Tip until the Tip Wrench clicks. Failure inside the Handpiece or the Control Service only by Authorized NSK Dealer.* Unit. Debris is clogging the Handpiece and Place air syringe against Water Pipe and blocking irrigation flow. blow air through it. If this does not clear it, Contact your Authorized NSK Dealer. Irrigation is not supplied. Make sure the pack contains solution and the Irrigation Tube is not leaking nor blocked.
18
Spray depends on the combination of flow level, power level, and Tip shape.
With some flow levels and Tip shapes, it is more difficult to form a spray. This is not a malfunction. Water does not exit, or flow Foot Control is not connected. Check the connection. level is low. Irrigation Tube is not connected to Irrigation Check the connection. pack or Handpiece. Irrigation Tube is not properly mounted to Check the connection. the pump. The Pump Cover is open. Close the cover. The Irrigation Tube is broken (leaking). Replace the Irrigation Tube. Irrigation leak. Leak from the connection between Insert Irrigation Tube Needle all the way Irrigation pack and Irrigation Tube. into the Irrigation pack. Leak from the connection between Insert the end of the Irrigation Tube Handpiece and Irrigation Tube. completely into the Handpiece water pipe. Leak from Irrigation Tube. Replace the Irrigation Tube. Irrigation does not stop. Mode is set to Auto Cleaning. To stop irrigation, press the Flow key. Faulty circuit. Service only by Authorized NSK Dealer.* Handpiece LED does not LED is burned out. Service only by Authorized NSK Dealer.* illuminate. Failure inside the Control Unit or inside the Service only by Authorized NSK Dealer.* Handpiece Cord. Failure or disconnection of the circuit. Service only by Authorized NSK Dealer.* LCD display does not properly Overheating. Heat can cause LCD display malfunction. display. Turn the Main Power OFF to cool down. Part of display is missing. Error Code is displayed. Refer to the Error Code table. Failure of the LCD display or driving circuit. Turn the Main Power OFF and then ON. If part of initial display is still missing Contact your Authorized NSK Dealer. Settings are not memorized Figures have been temporarily changed. Increase or decrease the Program No. by Wrong settings are one, and then return to the original setting. memorized. Foot control is being operated. Settings cannot be saved while LED Handpiece with cord is operating. An Error Code is displayed. Refer to the Error Code table. * Repairs cannot be made by the customer.
19
English
Irrigation does not exit as spray.