Operation Manual
15 Pages
Preview
Page 1
English
Ultrasonic Bone Surgery System
OPERATION MANUAL
0197
Please read this Operation Manual carefully before use, and file for future reference.
Classification of Devices • Classification by type of protection against electric shock: - Class I devices • Classification by degree of protection against electric shock: - Applied part type BF • Classification by sterilization or disinfection method allowed by the manufacturer: - Refer to Sterilization. • Classification by degree of protection against harmful intrusion of liquid based on IEC60529: - Foot Control ··· IPX8 (protected against immersion) • Classification by safety level of use in air, flammable anesthetic gas or dinitrogen monoxide (laughing gas), flammable anesthetic gas - Not suitable for use in the presence of a flammable anesthetic mixture with air or oxygen or nitrous dioxide • Classification by mode of operation: - Continuously operating device
Warning • Do not turn the power ON/OFF switch without reason, It might blow out a fuse. • If you feel any abnormality such as vibration, heat generation, abnormal noise, etc., prior or during the use of the unit, immediately stop using it. • This product is Medical Electrical equipment. EMC (Electromagnetic compatibility) is described in the accompanying documentation. • Portable and mobile RF communications equipment can affect Medical Electrical equipment. Do not use RF equipment outskirts for the product. • Do not place anything within 10 centimeters around the unit. Cautions
• Do not plug or unplug the power cord with a wet hand. You may get an electric shock. • Do not use the unit in any room where there is the risk of explosion or in the vicinity of combustible material. In particular, you must not use the unit for a patient who has been anesthetized with flammable anesthetic gas (laughing gas). • In use, be careful so that no water or antiseptic solution spills onto the control unit, which might cause fire or have you get an electric shock. If any water or antiseptic solution spills, immediately wipe it off with a dry cloth, etc. • Be careful not to give strong shock (dropping, in particular) to a handpiece. This might break the optics (Optic Handpiece only) or give you an electric shock. • You must not touch the connections (electric pins) of the handpiece or of the handpiece cord. • You must neither disassemble nor make alterations to the handpiece/control unit. • Do not use the unit on any patient who has a cardiac pacemaker (the unit may affect operation of a cardiac pacemaker). • Make sure to use the specified fuse.
• Only qualified personal is allowed to use the device in dentistry or oral surgery : Qualified doctors in medicine and surgeons. Assistants and paramedical personal, under the responsibility of qualified doctors in medicine or surgeons. • When using the unit in surgery, make sure to take action for contingency, such as preparing 2 sets of the unit in case of failure. • Give the priority to patient’s safety, and pay adequate attention so that there will be no accident. • This product does not consider patient’s age (except infants), gender, weight or nationality. • This product does not consider operater’s age (mature person), hight, weight, gender, or nationality. • Do not use this product if the patient has the any of the following conditions: · Patient complications · Patient allergic reactions · Patient has preexisting condition e.g. (heart, lung, renal, high-blood pressure) · Patient is Pregnant, patient of child-bearing potential or lactating patient · Patient is an infant (Age) · Patient has a heart pace maker fitted •The end user shall be responsible for any judgment that applies this product to a patient. • Place the unit on a flat and horizontal area. Be careful not to give any strong shock (such as dropping) to the control unit. • Before using the unit, make sure to operate it outside the oral cavity for checking. Then, if you find any abnormality, immediately stop using it and contact the dealer. • Turn the power switch OFF prior to disconnecting the power cord or handpiece cord by holding their plugs. Do not unplug by pulling on cords: they may be broken. • Do not unplug the power cord or handpiece cord while pressing on the foot control. • When the pump gets wet, wipe well and dry it. If the pump remains wet, it may not work normally as the roller inside the pump may slip. • If you bend or fold the irrigation tube while the pump is in operation, the irrigation tube may be broken or disconnected. • If you feel any abnormality in the irrigation flow, there is the possibility that the irrigation tube is worn and saline solution leaks out. Thus, replace the irrigation tube. • The irrigation tube is disposable. After using it once, dispose of it as medical waste. • Do not separate the handpiece from its cord, except when you replace the lamp (Optic Handpiece only). • Neither turn the power switch ON with the handpiece cord unconnected, nor turns it ON while pressing on the foot control. • Always remove the tip prior to disconnecting the handpiece cord, install or remove the irrigation tube. Otherwise, you might get your hand, etc. injured by the tip. • Saline solution is used for irrigating and cooling down. If the irrigation flow is not adequate, bone cells might necrotize or tooth surface might be damaged. • In use, never allow the vibrating tip to touch any prosthesis such as ceramics. Touching might cause break or chipping. • In use, never allow the vibrating tip to touch a metal crown, porcelain crown, etc., which may cause loosening or breakage, etc. • You must use the tip within the power range described on the tip case. If you use it out of the power range, the tip might break or damage an operative site. • Do not use a tip that is scratched, twisted or rusted, etc. In use, the tip may break.
2
3
Safety Precautions and Indications of Dangerous Items We recommend that prior to using the device; you read carefully the safety precautions in order to operate it correctly. Indications of Dangerous Items is designed so that you can use the product safely, thus obviating any safety hazard or damage to you or others. They are classified according to their magnitude and emergency degree of safety hazard or damages. As each and every item concerns safety, observe them. Classification
Degree of Danger or Danger and Seriousness
WARNING
Explains an instruction where personal injury or physical damage may occur.
CAUTION
Explains an instruction where minor to medium injury or physical damage may occur.
NOTICE
Explains an instruction that should be observed for safety reasons.
1. Safety precautions prior to use Warning
English
Thank you for purchasing our VarioSurg. This device is intended for use in dentistry and oral surgery. Implant site preparation, Bone harvesting, Sinus surgery, Osteoplasty, Osteotomy, General oral surgery, Tooth extraction, Syndesmotomy, Tooth avulsion, Periodontal surgery, Cyst surgery. We recommend that prior to use, you carefully read this document regarding instructions for use, handling method, or maintenance check so that you can carry on using the unit in the future. In addition, please keep this operation manual in a place where a user can refer to it at any given time.
• A diamond-coated tip shall be disposable. Thus, use it only once for treatment. Otherwise, cells that might have adhered to the diamond debris and still remaining even after cleaning or autoclaving might invade the mouth cavity of another patient. • Never grind the tip or bend it to change the angle, etc. In use, the tip might break or might not vibrate. • Even if the tip is used at the appropriate output level, overloading it might cause tooth chipping or tooth fracture. Use it together with a suction device and be careful not to leave any tip fragment in an operative site. • Try to keep saline solution coming out of the tip end off the unit. It might cause peeling of the sheet or discoloration of the casing. • Only Use Tips that NSK has approved for VarioSurg use. Damage, failure or an accident may occur; the following are possible failure modes (NSK Warranty Exclusions). · Vibration failure caused by using non conforming screws.(Incorrect TIP) · Patients’ accidental ingestion of Broken or damaged tip. · Damage to thread of handpiece output shaft. • The Handpiece, Handpiece cord, the Tip, Tip Wrench and Tip Holder are non sterile products, sterilize these before use. • Never sterilize the handpiece, etc., with ultraviolet radiation, which might cause discoloration. • If any water drops remain on the handpiece or handpiece cord after autoclaving, wipe them off. If you leave them, they might result in stain. • Prevent the tip from directly touching the gum, skin, or nerves, which might cause a burn or accident. In addition, do not touch mucosa membranes with any tip other than those dedicated. • The tip will abrade away as it is used. When it becomes worn, it might vibrate weakly or not vibrate, or its coating might peel off. Whenever you have observed abrasion of the tip or peeling of the coating, or felt a weakening of the vibration, replace it with a new one. For any tip other than those diamond-coated, we recommend that you replace it after using it five times, to achieve optimum cutting performance. • When replacing the tip, make sure to stop vibrating it and to securely fasten it using the special tip replacing wrench. Poorly fastened, it might vibrate weakly or not vibrate at all. • If you install the tip with dirt attached to its threading, it might vibrate weakly or not vibrate. When the threading is dirty, clean it. • If you feel the tip is not vibrating, remove it from an operative site, and press the foot control again. If this does not improve the condition, re-fasten the tip or turn the power off once and restart it. • This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information. • The use of ACCESSORIES, transducers and cables other than those specified, with the exception of transducers and cables sold by the manufacturer of this product as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of this product. • This product should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this product should be observed to verify normal operation in the configuration in which it will be used.
2. Features of the product Key feature: High cutting ability can shorten time for treatment. Visibility: Light embedded within the handpiece provides the surgeon with a clear sight for a safe surgery. (Optic Handpiece only) Safety: Even if the unit accidentally touches an operative site, it does not damage the soft tissue. Cavitation Effect: The cavitation effect washes away any blood from an area under treatment, thus ensuring a clear view of the operative field. Minimally invasive: Irrigation flow suppresses heat generation during bone cutting or peeling. Operation Principle A sinusoidal electrical signal, at ultrasonic frequency, is delivered by the generator. This signal is applied to the piezoelectric ceramic located inside the transducer. Piezoelectric ceramic converts this signal into mechanical vibrations. These vibrations are at the same ultrasonic frequency as the electrical signal. The mechanical vibrations are propagated towards the distal end of the transducer. The insert, which is attached at the distal end of the transducer, vibrates at ultrasonic frequency and makes it possible to achieve the aimed purpose.
3. Specifications Type Rated power supply Resonance frequency Maximum output Power supply input Lighting Dimensions Weight Fuse Use environment Transportation and Storage environment
NE214 AC100V 50/60Hz, AC120V 50/60Hz, AC230V 50/60Hz 28~32kHz 17W 50VA Halogen lamp (Optic Handpiece (VA-SG-HPSC) only) W268 x D230 x H103(mm) 3.1kg 100/120V Specification...T2AL 250V 230V Specification...T1AL 250V Temperature 0 - 40 ˚C Humidity 30 - 75 % Atmospheric pressure 70 - 106 kPa Temperature -10 - 60 ˚C Humidity 10 - 85 % Atmospheric pressure 50 - 106 kPa
Notice • When the vibrating handpiece or handpiece cord is in the vicinity of any computer or LAN cable, it might affect them. If it is in the vicinity of a radio receiver, the receiver might be affected. If you use any device such as a digital video, digital camera, etc. in its vicinity, electromagnetic interference might affect these devices. • Turn the power switch off after use. In addition, if you do not use the unit for a long time, disconnect the mains plug. • Clean/sterilize the unit immediately after using it. Then store it. Leaving it with blood, etc. attached thereto might lead to a failure. • When you have not used the unit for some time and use it again, you must check before using that the unit can work normally and safely. • Responsibility for operation and maintenance check of a medical device lies with the user. • No special training is required for this device. • Applied parts for patient and/or operator are/is tip and Handpiece.
4
5
English
Cautions
Handpiece Stand:
4. Package Contents
Place the handpiece horizontally on the handpiece stand. 1
3
English
2
CAUTION • Do Ensure the Tip does not touch the handpiece stand. • Do Ensure the handpiece is placed level into the handpiece stand (Do not incline it). • Do Ensure Handpiece stand is clean and free form contaminants, Handpiece Stand may be cleaned with an alcohol damped cloth. Straight
Incline
Tip had been touched to handpiece stand.
5 4
6
9
8
5. Control Unit with an Irrigation Pump
10
12
11
13
Pump Cover
* Picture shows Optic Handpiece 7
No.
LCD display
Part Name
Operation Panel
Quantity
1
Control Unit with Irrigation Pump
1
2
Foot Control
1
3
AC Power Cord
1
4
Autoclavable Handpiece with Cord attached (Optic / Non-Optic)
1*1
5
Irrigation Tube
5
6
Tube Holder
7
7
Saline Solution Hanger Post
1
8
Tip Wrench
1
9
Tips Holder
1
10
Spare VA lamp (For Optic Handpiece)
1
11
Tips
12
Handpiece Stand
1
13
Sterilization Cassette
1
6 (Basic Kit)
Front Panel
Handpiece Cord Jack
AC Power Cord Connection Jack
Main Power Switch
Foot Control Cord Jack
*1 One of these handpiece should be packed. 6
7
2SHUDWLRQSDQHO
'HVFULSWLRQRI2SHUDWLRQ
8OWUDVRQLF2XWSXW'LVSOD
$GMXVWWKHDPRXQWRIFRRODQWZDWHU7KLVGLVSODVWKHFRRODQWZDWHURQ WKH/&'SDQHO7KLVLVDGMXVWDEOHHYHQGXULQJRVFLOODWLRQRIXOWUDVRQLF ZDYH
9LEUDWLRQ/HYHO
$XWRPDWLF&OHDQLQJ0RGH 3URJUDP1R
$IWHUWKHEXWWRQLVORQJKHOGGRZQIRUDERXWVHFRQGVWKHSXPSZLOO URWDWHDWWKHPD[LPXPZDWHUDPRXQWIRUVHFRQGV WRFOHDQWKH LQWHULRURIWKHKDQGSLHFH
)ORZ/HYHO
3URJUDP.H
9LEUDWLRQ0RGH
0(025,=(.H
6HOHFWVDSURJUDP 7KLVGLVSODVWKHSURJUDP1RRQWKH/&'SDQHO NHDOORZVWRLQFUHDVHWKHOHYHOZKLOHŎNHDOORZVWRGHFUHDVHLW ,WFDQQRWEHDFWLYDWHGGXULQJRVFLOODWLRQRIXOWUDVRQLFZDYH
3RZHU.H
6HOHFW0RGH.H
$GMXVWVLQWHQVLWRIYLEUDWLRQ 3RZHUOHYHOLVGLVSODHGRQWKH/&'SDQHO NHDOORZVWRLQFUHDVHWKHOHYHOZKLOHŎNHDOORZVWRGHFUHDVHLW 7KLVLVDGMXVWDEOHHYHQGXULQJRVFLOODWLRQRIXOWUDVRQLFZDYH
8OWUDVRQLF2XWSXW'LVSOD 'LVSODLQJDQLPDWLRQGXULQJRVFLOODWLRQRIXOWUDVRQLFZDYH
3URJUDP1R 3URJUDP1R 'LVSODLQJVHOHFWHGSURJUDP1R
9LEUDWLRQ/HYHO 7HQEDUVLQGLFDWHWKHOHYHORIYLEUDWLRQ
)ORZ/HYHO 'LVSODVWKHVHOHFWHGFRRODQWVROXWLRQƃRZYROXPHOHYHO
7KHSURJUDP1RVDUHDVVLJQHGDVOLVWHGEHORZ 685*PRGHř3URJUDP1R1R1R1R1R (1'2PRGHř3URJUDP1R1R 3(5,2PRGHř3URJUDP1R1R 3URJUDP1RDQG1RDQG1RKDYHEXUVWIXQFWLRQ KDPPHUGULOOHIIHFW 3URJUDP1Rř+] 6HWDSSURSULDWH3URJUDP1RGHSHQGLQJRQERQHGHQVLW 3URJUDP1Rř+] 3URJUDP1Rř+] 3URJUDP1RDQG1RSURYLGHFRQWLQXRXVYLEUDWLRQZLWKRXWWKHEXUVWIXQFWLRQ
9LEUDWLRQ/HYHO 685*PRGHř<RXFDQVHOHFWWKHPRGHLQLQFUHPHQWVRIIURPWR (1'2PRGHř<RXFDQVHOHFWWKHPRGHLQLQFUHPHQWVRIIURPWRDQGLQLQFUHPHQWVRI WKHUHDIWHU
9LEUDWLRQ0RGH 'LVSODLQJVHOHFWHGPRGHRQO
$PRXQWRI&RRODQW)ORZ
0(025,=(.H 0HPRUL]LQJRSHUDWLRQFRQGLWLRQVGLVSODHGRQWKHOLTXLGFUVWDOGLVSOD /&' SDQHO
6HOHFW0RGH.H
<RXFDQDGMXVWWKHƂOOHGZDWHUDPRXQWLQVWDJHV7KHUDQJHRI>)/2:@ZDWHULVDVIROORZV 685*PRGHř<RXFDQVHOHFWLQVWDJHVIURPPOPLQWRPOPLQ (1'2DQG3(5,2PRGHVř<RXFDQVHOHFWLQVWDJHVIURPPOPLQWRPOPLQ +RZHYHUGHSHQGLQJRQWKHFRQGLWLRQRIWKHLUULJDWLRQWXEHWKHUHPDEHVRPHWROHUDQFHLQWKHZDWHUDPRXQW :KHQRXVHWDORZYDOXHRIWKH>)/2:@ZDWHUWKHSXPSVWDUWVIDVWDQGVORZVGRZQWRWKHVHWYDOXH
6HOHFWDQPRGHDFFRUGLQJWRDSXUSRVHRIWUHDWPHQW7KLVFKDQJHVWKHUDQJHRISRZHU
English
)ORZ.H
)URQWSDQHODQG/&'GLVSOD
3UHVVLQJ>0(025,=(@NHVWRUHWKHSDUDPHWHUVGLVSODHGRQWKH/&'LQWRWKHVHOHFWHG3URJUDP1R&KHFNWKH YDOXHVRIWKHSDUDPHWHUVWREHVWRUHGDQGSUHVVWKHNHIRUVHFRQGRUORQJHU$œEHHSŔLVSURGXFHGZKHQ SDUDPHWHUVDUHVDYHG ,Q685*PRGHRXFDQRQOVDYHVHWWLQJVRI685*PRGHVHWWLQJ<RXFDQQRWVDYHDVHWWLQJVWDWHRIDQRWKHU PRGH7KLVDOVRDSSOLHVWRWKHRWKHUPRGHV :KHQRXKDYHPDGHDFKDQJHWRVHWWLQJDQGSUHVVHG>0(025,=(@NHDJDLQWKHVHWWLQJVDYHGLQWKDW3URJUDP 1RLVHUDVHGDQGWKHQHZVHWWLQJLVVDYHG
)RRW&RQWURO )RRW&RQWUROSOXJ )RRW&RQWUROFRUG
6HOHFWLQJ0RGH 6HOHFWPRGHE>6HOHFW0RGH@NH
$XWRPDWLF&OHDQLQJ0RGH $IWHUKROGLQJWKH$XWR&OHDQLQJNHGRZQIRUDURXQGVHFRQGVWKHSXPSURWDWHVDWWKHPD[LPXPFRRODQWƃRZRI POPLQWRFOHDQWKHLQWHULRURIWKHKDQGSLHFHLUUHVSHFWLYHRIWKHPRGHVHOHFWHG
3URJUDP.H
212))6ZLWFK IRU:DWHU)ORZ
352*5$0 6ZLWFK
7KHNHVHOHFWVDSURJUDP3URJUDP1RLVGLVSODHGRQWKH/&'SDQHOLVLQDVFHQGLQJRUGHUZKLOHŎLVLQ GHVFHQGLQJRUGHU3URJUDPVFDQQRWEHFKDQJHGGXULQJRVFLOODWLRQRIXOWUDVRQLFHPLVVLRQ
3RZHU.H 7KHNHDGMXVWVLQWHQVLWRIXOWUDVRQLFYLEUDWLRQ7KHSRZHUOHYHOLVGLVSODHGRQWKH/&'SDQHONHDOORZVWR LQFUHDVHWKHOHYHOZKLOHŎNHDOORZVWRGHFUHDVHLW<RXFDQDGMXVWLQWHQVLWRIWKHYLEUDWLRQHYHQGXULQJXOWUDVRQLF HPLVVLRQ8VHORZOHYHOVZKHQRXVWDUWDQGDGMXVWWKHSRZHUOHYHODFFRUGLQJWRWKHRSHUDWLQJFRQGLWLRQV
9LEUDWLRQ 212))6ZLWFK
352*5$0 6ZLWFK
&$87,21 7KH/&'GLVSODSDQHOLVPDGHIURPOLTXLGFUVWDODQGVKRXOGDOZDVEHKDQGOHGZLWKFDUH
212))6ZLWFKIRU:DWHU)ORZ :KLOHRXDUHSUHVVLQJRQWKDWVZLWFKLUULJDWLRQLVGHOLYHUHGDWWKHVHWWLQJRI
352*5$0 6ZLWFK (YHUWLPHRXVWHSRQWKHVZLWFKWKHSURJUDPQXPEHULQFUHDVHVE
352*5$0 6ZLWFK (YHUWLPHRXVWHSRQWKHVZLWFKWKHSURJUDPQXPEHUGHFUHDVHVE
9LEUDWLRQ212))6ZLWFK :KLOHRXDUHSUHVVLQJRQWKHVZLWFKYLEUDWLRQDUHJHQHUDWHGDWWKHVHWWLQJOHYHOGLVSODHGRQWKHSDQHO
&$87,21 (YHQZKHQRXHUURQHRXVOWXUQWKHSRZHU21ZLWKWKHYLEUDWLRQ212))VZLWFKSUHVVHG21YLEUDWLRQGRHV QRWRFFXU RXRQOKHDUZDUQLQJVHTXHQFHVRIEHHSV
English
0HPRU)XQFWLRQ
7. Installation
7-5 Installing the Irrigation Tube
A side
(1) 1 Turn the Pump Knob to OPEN position.
7-1 Connecting the Handpiece Cord
Front Side of the Unit
2 Hook the Irrigation tube (Bottle Side) on to (A) side. (Fig. 6)
English
With mark of the cord plug facing upward, insert it into the socket of the control unit.(Fig.1). Then, make sure to hold the rear of the connector and press it till it clicks. Then, try to pull it in order to check that it has been locked. To remove the plug, hold the lock joint and pull it out.
Pump Knob Lock Joint
Fig. 6 Fig. 1
3 Hook the Handpiece side Irrigation tube on to (B) side. (Fig. 7)
B side
7-2 Connecting the Foot Control Face the screw on the foot control cord plug downward then insert the plug into the foot control cordjack on the control unit.Screw the plug by fastening the lock unit.
Fig. 7
Lock Ring
Fig. 2
7-3 Connecting the AC Power Cord
4 Check that it has been pushed deeply to
the back.
Back Side of the Unit
Make sure that the power switch is turned OFF ( side). Securely insert the power cord into the inlet located on the rear of the control unit, according to the shape of the power cord. Insert the plug of the power cord into a wall outlet of rated power supply from which earthing can be taken.
Rollers
Power Switch
Fig. 8 Fig. 3
5 Turn the Pump Knob to closed. (Fig. 9)
7-4 Mounting the Saline Solution Hanger Post
CAUTION
Insert the saline solution hanger post into the encircled area on the unit. Match the guiding pin of the hanger to the slot of the hole. Then, check for correct placement and stability of the saline solution hanger post.
At any time, insure that there is enough saline solution inside the bag. Fig. 9
Fig. 4
Fig. 5 12
13
6 Mount the irrigation tube onto the pump as shown in Fig.10.
3 Open the tube cap to supply air into the bottle.
English
Tube Cap
Fig. 10 (2) Insert the free end of the irrigation tube deep around the water pipe of the handpiece. Water pipe
Fig. 14
7-6 Attaching the Tube Holder Use the Tube Holder to assemble together the Handpiece Cord and the Irrigation Tube. It is easier to fasten them onto the Handpiece Cord first and, next, onto the Irrigation Tube.
Water pipe
Handpiece Cord
Tube Holder
Irrigation Tube
Fig. 15 Optic handpiece
Fig. 11
Non-Optic handpiece
CAUTION
CAUTION Shallow insertion might unfasten the tube. In addition, except when you replace the tube, always keep the end of the tube connected to the water pipe. (3) 1 Close the tube clamp located between the irrigation tube needle and the irrigation pump, as illustrated in Fig.12.
• Assemble the handpiece cord and irrigation tube by means of the 7 tube holders. (Fig. 16) • If you let the irrigation pump operate with the tube bent or with no liquid flowing, the tube might be ruptured or fall off from the bottle.
Tube clamp
Fig. 16
Fig. 12 2 Hook the bottle to the hanger and insert the irrigation tube
needle into the bottle cap.
CAUTION
Bottle Cap
Use only 500ml saline solution bags. Never hang more than 800 grams to the hanger post.
Irrigation Tube Needle
Fig. 13
14
15
8. Operation
8-2 Checking and Adjustments
To tighten the tip, screw it first gently by hand (Fig. 17). Next, introduce it from the end of the tip to the hole of the wrench (Fig. 18).
(2) Switching the mains supply ON Turn ON the power switch located at the rear of the unit.(See Fig. 3).
CAUTION • To avoid any wound, particularly take care that the end of the tip is not contacted internally of tip wrench inside (Fig 19). • If the wrench is turned whereas it is excentric, tightening is likely to be insufficient and the ultrasonic vibrations can be strongly weakened.
power OFF
power ON
(3) Priming irrigation circuit Immediately before using the unit, release the tube clamp from the irrigation tube, step on the ON-OFF switch for water flow of the foot control, and check that the saline solution flows normally out of the tip end. With a new tube, it takes several seconds for the saline solution to reach the tip. (4) Setting Output Press [ Select Mode ] key on front panel for select suitable application. Adjust the amount of water, select a program, and adjust power on the Operation panel. Fig. 21 shows a comparison of outputs of respective modes:
Turn the Tip clockwise by using Tip Wrench (Fig.19, 20) (Turn it till you hear the click sound.) To remove the Tip; Place tip Wrench over the tip and turn the Tip Wrench counter-clockwise.
Set to the power mode described on the tip case and use it within the power range (Fig. 22). The abbreviations for the selection modes mean the following: S … Surg mode E … Endo mode P … Perio mode SURG
Fig. 17 Loosen (CCW)
Fig. 18
Tighten (CW)
ENDO PERIO
Tighten
0
20
40
60
Output (Relative Comparison % )
80
100
Selection Mode
Fig. 21
Power Range
Fig. 22 Loosen
Be careful! Fig. 19
Fig. 20
CAUTION If you set out of the Power Range described on the tip case, the tip might break, damage a tooth surface, gum or mucous membrane.
CAUTION • When you install the tip, make sure to use the sterilized one. • Check the Tip before use. (Flush, Damages, Bending or Rust) • During cutting, Tip will gradually wear away, as the Tip wears the stroke will get smaller and cutting power level will decrease. When level drops too far, change the Tip. • Tip Wrench is consumable For reliable operation replace annually.
16
17
English
(1) Checking Irrigation circuit Ensure that the irrigation tube is correctly connected to the saline solution pack, pump, and handpiece.
8-1 Attaching or detaching the Tip
CAUTION • Similarly after replacing the bottle, turn the ultrasonic vibration ON outside the oral cavity of a patient, and check that the saline solution flows out of the tip end. • Press Adjust Power key within the power range and use while checking appropriate vibration. • Do not turn ON the ultrasonic vibration with no saline solution. No water filling would lead to heat generation at the handpiece. It might also damage an operative site or tooth surface. • In use, do not apply too much pressure to the tip end. • The unit is designed to display an error (E-P) and to stop vibrations if the pump does not work due to some failure.
9. Care and Sterilization The following items can be autoclaved. · Handpiece (Optic/Non-Optic) · Handpiece cord (Optic/Non-Optic) · Tip Wrench · Tips Holder · Tube Holder · VA lamp
· Tip · Handpiece Stand · Sterilization Cassette
English
(5) Activation When you step on the foot control, the tip is activated. Then the light of the handpiece also turns on. When you use it for the first time and thus the irrigation tube is not filled with saline solution, check the following: • Let it ultrasonic vibrate outside the oral cavity of a patient, rotate the pump, and check that the saline solution flows out of the tip end. • Check that the light turns on. (VA-SG-HP only) • Check that there is no unpleasant feeling with the handpiece and tip.
9-1 Auto Cleaning (1) After using with saline solution, clean the interior of the handpiece with distilled water. Pull out the needle from the saline solution bag, plunge the needle in the container (such as a cup, etc.) in which the distilled water is contained open the tube clamp for irrigation tube, and press the [FLOW] key for about 3 seconds till the pumps starts (Auto Cleaning). During automatic cleaning, Water mark flashes. The bar of the power display indicates the remaining time, and decreases by one scale for every 3 seconds. Water Mark Flashing
Power Display Bar
NOTICE When the power is turned OFF, settings are stored with the current display. The next time the power is turned ON, the unit returns to the previous state (last memory function). However, as it does not have influence on the program memory, save the setting by pressing MEMORIZE key if you wish to save the current setting as a program.
CLN Display
Fig. 23
CAUTION
* Operation Mode: Consecutive
Enter the auto cleaning mode, and the pump rotates at the maximum water filling amount for 30 seconds. Be careful as water flows out of the end of the handpiece. If you leave the handpiece without performing this rinsing operation, sodium chloride will crystallize inside the handpiece, which might clog the irrigation hole and thus prevent saline solution from outflowing. (2)
For Optic Handpiece 1 Remove the handpiece cord and irrigation tube from the Unit. 2 Remove irrigation tube from the handpiece and discard. 3 Wipe the handpiece with alcohol-immersed cotton swab or cloth. (Do not use metal brush) 4 Put the handpiece and connected cord into the sterilization cassette. For Non-Optic Handpiece 1 Turn the “Nose cover” counter-clockwise to remove. 2 Brush any contaminants from inside the nose area with a soft brush. (Do not use metal brush) Wipe with alcohol soaked cloth. (Fig. 24) 3 Replace the “Nose Cover” onto the handpiece turn clockwise. 4 Set the handpiece with handpiece cord to Sterilization cassette.
Loosen
Tighten
Nose Area
Nose Cover
Fig. 24
18
19
Handpiece
9-2 Exchanging the Halogen Lamp (Optic Handpiece only) Remove the rear cover from the handpiece only when replacing the lamp. Turn the rear cover in the arrow direction and remove it (See Fig. 26). Prior to lamp replacement, disconnect the handpiece out from the control unit.
English
(3) Tips, Tip Wrenches, Tips Holders and Handpiece 1 Wash and brush down any stain on the surfaces with a resin brush, etc. (you should not use a metal brush), and wipe them off with cotton impregnated with rubbing alcohol. 2 Put the tips in the tips holder, and then place the tips holder in the sterilization cassette. 3 The sterilization cassette can accommodate up to 2 tips holders. 4 Put the tip wrenches in the sterilization cassette. 5 As shown in Fig. 25, put what is to be sterilized in the sterilization cassette, then close and lock it with the swivelling lever. 6 Do not put a maintenance tip (V10-S and V-P10) in the tip holder. Instead, put them in an autoclaving bag and seal the bag.
Fig. 26
Handpiece Cord
Turn the rear cover, and slide the cord backward (Fig. 27).
Handpiece
Cord
You can place another tips holder here.
Tip Wrench
Threading
Tube Holder Tip Replacement Wrench (for V10-S) (Option)
Tips Holder
*Picture shows Non-Optic Handpiece
Rear Cover
Fig. 27 Slowly pull the handpiece out (Fig. 28).
Fig. 25
(4) Sterilization Use gravity displacement Steam autoclave. Perform autoclave sterilization at 20 minutes at 121ºC or 15 minutes at 132ºC Drying time 10 minutes. Note that you cannot autoclave the control unit and the foot control.
CAUTION
Fig. 28 Slowly pull the ring out (Fig. 29).
• Never sterilize the handpiece, etc., with ultraviolet radiation, which might cause discoloration. • If you autoclave sterilize the handpiece together with any instrument to which antiseptic solution adheres, discoloration might occur. • Only handpiece can be washed via Thermo Disinfector. • If any saline solution or stain adheres to the control unit, well wipe it off with a cloth impregnated with clean water and squeezed out. Then, wipe the unit well with a dry cloth. • Do not clean the unit with any solvent such as thinner, benzene, etc.
Ring
Fig. 29
* Confirmed Thermo Disinfector is “Type: G7882” (Miele & Cie)
20
21
Insert a thin slotted screwdriver, etc. into the highlighted area, and remove the lamp (Fig. 30). Securely insert a connecting terminals of a new lamp, mating with the socket holes. Then, apply the ring, mate the projection of the cord end with the hole of the handpiece, and insert the former into the latter. Place the rear cover, and secure it to the handpiece by screwing CW.
If you suspect a failure, check the following before asking for a repair. When an error code is displayed, refer to the table of error codes. If no case applies to you, or when the condition does not improve even after taking action, this product is likely to have failed. Thus, please contact the dealer. Case
Possible Cause The power cord is not connected to the unit.
Fig. 30
WARNING
The power does not turn on (the LCD screen does not light up).
Never touch the electrical connections of the handpiece and of the lamp. Touching the connections might damage them.
CAUTION • Remove the tip before disassembling the handpiece. Otherwise, the tip might cause injury. • Then, drain any liquid from the handpiece by using a syringe and injecting air at the water pipe. Remove the rear cover only when all the liquids are drained. Any remaining saline solution in the handpiece might spill onto the lamp and cause corrosion. • Do not directly touch a glass area of a new lamp with your finger and protect it against any shock.
Although the power turns on, the unit displays Vibration Oscillator Mark icon and beeps.
The tip does not vibrate
The tip does not vibrate (Oscillator Mark displays).
The tip vibrates, but weakly. It does not cut easily.
The tip breaks easily.
22
Action to Take Check the connection.
The plug is not inserted into a socket, or no electricity is supplied to the Check the connection. socket. A fuse has blown.
Contact your dealer.
An internal fuse has blown.
Contact your dealer.
You step on the foot control.
If the foot control is stepped when the power is turned on, the unit does not operate in order to prevent possible accident. Release the foot control and try again.
The foot control is not connected.
Check the connection.
The handpiece and cord are not connected. Or the cord is not connected to the unit.
Check the connection.
Circuit failure.
Contact your dealer.
Abrasion/break of a tip
Replace the old tip with a new one.
Bad mode.
Select an appropriate mode.
The tip has not been tightened correctly.
Tighten the tip once again with the tip wrench, until it clicks.
Power setting is too low.
Reset to the mode and power range described on the tip case.
The foot control is not plugged in correctly.
Plug in the foot control correctly.
Failure inside the handpiece.
Contact your dealer.
Failure inside the handpiece.
Contact your dealer.
Power setting is too high for that tip.
Reset to the mode and power range described on the tip case.
23
English
10. Troubleshooting guide
Possible Cause
The tip loosens easily.
The tip has been tightened inadequately.
Tighten the tip once again with the tip wrench, until it clicks.
Power setting is not appropriate for that tip.
Reset to the mode and power range described on the tip case.
The tip has been tightened inadequately.
Tighten the tip once again with the tip wrench, until it clicks.
You hear loud abnormal noise from the handpiece.
Action to Take
Case
Water leakage
Possible Cause
Action to Take
Water leakage from the connection between the bottle and the irrigation tube.
Securely insert the needle into the bottle (See 7-5 (3)).
Water leakage from the connection between the handpiece and the irrigation tube
Securely place the tube end around the water pipe of the handpiece (See 7-5 (2)).
Water leakage from the irrigation tube Replace the irrigation tube. Failure inside the handpiece or the control unit.
Contact your dealer.
You have not set appropriate output to Reset to the mode and power range the attached tip. described on the tip case. The tip has been tightened inadequately. The handpiece becomes hot.
Saline solution does not turn into spray.
The water does not come out. The amount is small.
The pump does not stop.
Tighten the tip once again with the tip wrench, until it clicks and it rotates.
Failure inside the handpiece or inside the control unit.
Contact your dealer.
Foreign material is clogged in the handpiece, thus preventing saline solution from outflowing.
Inject air to the water pipe with a syringe. If clogging still remains,contact your dealer.
No saline solution is supplied.
Check the remaining amount in the bottle and whether or not there is any crack in the irrigation tube.
Saline solution can’t turn into spray under certain conditions.
Some water does not tend to turn into spray, depending on the amount of water or a tip shape. However, this is not a failure.
The foot control is not connected.
Check the connection.
The tube is not connected to the bottle. The tube is not connected to the handpiece.
Check the connection.
The tube is not correctly set in the pump.
Check the set part (See 7-5 (1)).
The pump cover is not closed (the pump is spinning free).
Close the cover.
The tube is broken (it leaks somewhere).
Replace the tube.
24
It is “auto cleaning” mode.
The lamp does not light up. (Optic Handpiece only)
If you want to stop it, press [FLOW] key.
You have stepped on the Water Filling Move your foot away from the foot Flow of the foot control. control. Circuit failure.
Contact your dealer.
Halogen lamp blew.
Replace the bulb (See 9-2).
The connection terminals of the VA lamp are not inserted correctly into the socket holes.
Mate the connection terminal of the VA lamp with the socket hole, and insert it securely.
Failure inside the control unit or inside Contact your dealer. the handpiece cord. Failure or disconnection of the circuit. Contact your dealer.
The display on the LCD screen is abnormal. The display is missing.
The memory does not work. Even it works, numeric values are wrong.
Overheating.
Extremely high temperature might disable the LCD display (when the temperature goes down, it recovers).
An error code is displayed.
Refer to the table of error codes.
Failure of the LCD display or driving circuit.
After restarting the unit, if icons are missing on the display, contact your dealer.
You have fogotten to press on [MEMORIZE] key.
Check the parameters to store and press [MEMORIZE] key.
You have stepped on the foot control.
While the handpiece is operating, the unit does not memorize.
25
English
Case
Error Code E-0
E-2
Possible Error
Action to Take
Circuit failure.
Contact your dealer.
You pressed the tip end too hard, thus applying heavy load to its end, or the tip end is stunk and fixed.
Press Vibration ON/OFF switch of the foot control and slowly pull out the tip while letting it vibrate. Be careful not to apply overload to the tip end.
The tip tightening force is inadequate.
Using the tip wrench, securely attach the tip till it “clicks”.
The tip is not mounted.
Mount the tip and tighten it.
A cable is disconnected inside the handpiece or inside the unit.
Replace with a spare handpiece, or contact your dealer.
E-4
Overload occurred due to long and continuous operation.
Turn the power off, leave the unit until it cools down. Then, use the unit again. If this error occurs frequently while operated in the normal range, contact your dealer.
E-F
A cable is disconnected inside the foot control.
Replace the foot control or contact your dealer.
The rotor of the pump is locked, etc.
Check whether the tube is mounted correctly. If the rotor does not rotate even when you remove the tube and step on the foot control, it is likely to be a circuit failure. Thus, contact your dealer.
E-P
11. Optional Accessorie Product Name
Items pictures
Handpiece Handpiece Cord (Optic)
Order Code Y141420
Handpiece Handpiece Cord (Non-Optic)
E377
Set of 5 Irrigation Tubings
Y900113
Set of 3 VA halogen lamps
Y900107
Set of 7 Tubing holders
Y900767
Sterilization Cassette
20001326
Tips Holder
20001327
Tip Wrench
Z221077
E Tip Replacement Wrench For V10-S
Z217399
12. Disposing Product Consult with dealer from whom you purchased it about waste disposal.
13. Warranty • The manufacturer will warrant the quality of the product for one year after you purchased it, provided that you use it in accordance with the method and procedures described in this operation manual. • Note, however, that if you did not observe what is written in this operation manual or for any single use goods, the warranty will not apply.
26
27
English
Table of Error Codes
Guidance and manufacturer's declaration - electromagnetic immunity
TUV Rhineland of North America is a Nationally Recognized Testing Laboratory (NRTL) in the United States and is accredited by the Standards Council of Canada to certify electro-medical products with Canadian National Standards.
The VarioSurg is intended for use in the electromagnetic environment specified below. The customer or the user of the VarioSurg should assure that it is used in such an environment. Immunity test
IEC60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted RF IEC61000-4-6
3Vrms150 kHz to 80MHz
3 Vrms
Radiated RF IEC61000-4-3
3V/m80MHz to 2.5 GHz
3 V/m
Portable and mobile RF communications equipment should be used no closer to any part of the VarioSurg, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
The product is designed not to become the ignition source in air and flammable anesthetic gas. *Only Foot Control is AP equipment. The EU directive 93/42/EEC was applied in the design and production of this medical device.
Recommended separation distance d = 1.2 P d = 1.2 P 80MHz to 800MHz d = 2.3 P 800MHz to 2.5GHz
Protected against the effects of continuous immersion in dust and water.
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
Dispose of this device and its accessories via methods approved for electronic device and in compliance with the Directive 2002/96/CE Type BF applied part
Follow Operation Manual for use
Authorised representative in the European community
Manufacturer
Autoclavable up to Max.135°C. *for detail see Sterilization.
This product can be cleaned and disinfected with a Thermo-Disinfector.
NOTE 1 NOTE2 a
b
Guidance and manufacturer's declaration - electromagnetic emissions The VarioSurg is intended for use in the electromagnetic environment specified below. The customer or the user of the VarioSurg should assure that is used in such an environment. Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR11
Group 1
The VarioSurg uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR11
class B
Harmonic emissions IEC61000-3-2
class A
The VarioSurg is suitable for use in all establishments other than domestic, and may be used in domestic establishments and those directly connected to public low-voltage power supply network that supplies buildings used for domestic purposes, provided the following warning is heeded:
Voltage fluctuations/ flicker emissions IEC61000-3-3
Complies
Warning: VarioSurg is intended for use by healthcare professionals only. VarioSurg may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures such as re-orienting or relocating the VarioSurg or shielding the location.
At 80MHz and 800MHz, the higher frequency range applies. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobiles radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the VarioSurg is used exceeds the applicable RF compliance level above, the VarioSurg should be observed to verity normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the VarioSurg. Over the frequency range 150kHz to 80MHz, field strengths should be less than 3 V/m. Marking on the outside of Equipment or Equipment parts that include RF transmitters or that apply RF electromagnetic energy for diagnosis or treatment.
Recommended separation distances between portable and mobile RF communications equipment and the VarioSurg The VarioSurg is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the VarioSurg can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the VarioSurg as recommended below, according to the maximum output power of the communications equipment. Separation distance according to frequency of transmitter m
Rated maximum output power of transmitter W
150kHz to 80MHz d = 1.2 P
80MHz to 800MHz d = 1.2 P
800MHz to 2.5GHz d = 2.3 P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
Guidance and manufacturer's declaration - electromagnetic immunity
1
1.2
1.2
2.3
The VarioSurg is intended for use in the electromagnetic environment specified below. The customer or the user of the VarioSurg should assure that it is used in such an environment.
10
3.8
3.8
7.3
Immunity test
IEC60601 test level
Compliance level
Electromagnetic environment - guidance
100
12
12
23
Electrostatic discharge (ESD) IEC61000-4-2
±6kV contact ±8kV air
±6kV contact ±8kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst IEC61000-4-4
±2kV for power supply lines ±1kV for input/output
±2kV for power supply lines ±1kV for input/output
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC61000-4-5
±1kV line to line ±2kV lines to earth
±1kV line to line ±2kV lines to earth
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC61000-4-11
<5% Ut (>95% dip in Ut) for 0.5 cycle 40% Ut (60% dip in Ut) for 5 cycles 70% Ut (30% dip in Ut) for 25 cycles <5% Ut (>95% dip in Ut) for 5 sec
<5% Ut (>95% dip in Ut) for 0.5 cycle 40% Ut (60% dip in Ut) for 5 cycles 70% Ut (30% dip in Ut) for 25 cycles <5% Ut (>95% dip in Ut) for 5 sec
Mains power quality should be that of a typical commercial or hospital environment. If the user of the VarioSurg requires continued operation during power mains interruptions, it is recommended that the VarioSurg be powered from an uninterruptible power supply or a battery.
Power frequency (50/60Hz) magnetic field IEC61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 NOTE2
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Cables and accessories
Maximum length
Complies with
Handpiece cord Foot Controller
2m 2m
RF emissions, CISPR11, Class B/ Group 1 Harmonic emissions, IEC61000-3-2 Voltage fluctuations/ flicker emission, IEC61000-3-3 Electrostatic discharge(ESD), IEC61000-4-2 Electric fast transient / burst IEC61000-4-4 Surge IEC61000-4-5 Voltage dips, short interruptions and voltage variations on power supply input lines IEC61000-4-11 Power frequency(50/60Hz) magnetic field IEC61000-4-8 Conducted RF IEC61000-4-6 Radiated RF IEC61000-4-3
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
28
29
English
Symbols