2 Pages
Preview
Page 1
Care & Maintenance instructions 1. Preparation • Remove handpiece from tubing • Remove bur • Do not submerge handpiece in any solution
2a. Automatic cleaning and disinfection using a thermo/washer disinfector • Always ensure that handpieces are placed on the correct adaptors • Cleaning agent pH level must be between 2.5-10.5 • Rinse cycle must be activated OR (if automatic method is not available see 2b below)
2b. Manual cleaning using tap water, brushes and cloths • Clean handpiece under cold or tepid running water using a soft to medium brush
6. Packaging • Pack in accordance with current applicable standards/legislation if using a B or S type autoclave. If using an N type autoclave, pack AFTER sterilisation.
7. Sterilisation • Sterilise in a steam steriliser (autoclave) at a temperature not exceeding 138°C in accordance with current applicable standards. Always remove from autoclave as soon as cycle is complete.
8. Storage • Store handpieces protected from dust, dirt and microorganisms, ideally in an upright position (head up) • Handpieces MUST be stored dry • Always comply with expiry dates of sterilised items in accordance with current applicable standards/legislation
3. Inspect • Inspect handpieces for debris and repeat above process if necessary
4. Drying • The drying procedure is normally part of the cleaning programme of the thermo/washer disinfector alternatively dry manually using a tissue
PANA SPRAY Plus Nozzles Always choose the right one for your handpiece
5a. Automatic lubrication using NSK iCare • Check machine is connected and working correctly • Connect handpiece to correct port • Select appropriate cycle (refer to NSK iCare user manual) • Use air cycle as required • Always lubricate straight handpieces with closed chuck and blanking piece inserted • Always complete steps 2,3,4,5 in quick succession • Lubricate the Chuck Mechanism by using the chuck cleaning function on the iCare
5b. Manual lubrication using NSK PANA SPRAY plus • Select correct nozzle for handpiece (see list below) • Shake spray can • Keep spray can in upright position • Insert nozzle into handpiece • Always lubricate straight handpieces with closed chuck and blanking piece inserted • Spray handpiece for 2-3 seconds holding a clean tissue over the head end • Check tissue for debris and if oil is not running clean repeat process until handpiece is clean and free of debris • Lubricate chuck mechanism using PANA SPRAY plus with tip nozzle weekly • Always complete steps 2,3,4,5 in quick succession
KV Nozzle – Z090-051 For KaVo® MULTIflex Coupling SR Nozzle – Z090-052 For Sirona® Quick Coupling BA Nozzle – Z090-053 For Bien-Air® Unifix® Coupling WH Nozzle – Z090-054 For W&H® Roto Quick® Coupling For Phatelus Air Motor Heads – M070-090 To be used with P183-090 QD Nozzle – Z096-090 For QD Couplings Standard E-Type Nozzle – Z019-090 For E-Type attachments NSK Surgical Head Nozzle – Z114-090 For E-Type attachments and surgical heads: SGM-I, SGMS-I and SGQ-H Endo-Mate Head Nozzle – Z256-090 For ENDO-MATE DT, TC, TC2 & TASKAL Wizard heads Tip Nozzle – Z020-201 For Borden 2/3 and Midwest 4 hole turbines and Borden 2/3 and Midwest 4 hole air motors and chuck mechanism NSK Nozzle – P183-090 For high speed instruments which attach to NSK Flexi Quick Coupling
PURGE out excess lubricant from the handpiece after manual lubrication
Call NSK on 1300 44 33 21 (Australia) or 0800 44 33 21 (New Zealand) for more information or contact your Preferred Authorised Dental Dealer Product Information | Technical Support 1300 44 33 21 (Australia) or 0800 44 33 21 (New Zealand) Email: [email protected] www.nskoceania.com.au
Automatic cleaning and lubrication using NSK iCare is simple A
A. Check oil level indicator light. Add more oil if required (light will come on when oil comes down to 150ml or less) B. Open the door and connect 1, 2, 3 or 4 handpieces to the correct handpiece adaptor. Close the door (iCare will not operate if the door is open)
C. Select the desired purge programme Mode. By selecting the selection button (1, 2 or 3), B until the mode light is illuminated - S Mode for High Speed handpieces (air turbines) (Cycle time 28 seconds) - S Mode for Blue band Low Speed handpieces and NSK heads (Cycle time 50 secs) S: Short / M: Middle / L: Long - M (Middle) Mode for High Speed handpieces (air C turbines) (Cycle time 30 seconds) - M (Middle) Mode for Blue band Low Speed handpieces and NSK heads (Cycle time 64 secs) - L (Long) Mode for High Speed handpieces (air turbines) (Cycle time 33 Start button seconds) D - L (Long) Mode for Blue band Low Speed handpieces and NSK heads (Cycle time 84 secs) D. Press the ‘Start’ button. The door should not be opened until the cycle is finished.
Always select the correct adaptor before cleaning and lubrication E-Type Handpiece For All Brands ISO E-Type and E-Type LUX Low Speed handpiece attachments – Z280-035 E-Type NSK Coupling For High Speed handpieces that connect to NSK couplings – Z257010 NSK Borden 2 hole For Borden 2 hole High Speed handpieces – T904 Borden® Kavo Coupling For High Speed handpieces that connect to Kavo® Multiflex LUX® couplings – Z257020 Kavo® Sirona Coupling For High Speed handpieces that connect to Sirona® Quick couplings – Z257023 Sirona® W&H Coupling For High Speed handpieces that connect to W&H® Roto Quick® couplings – Z257021 W&H®
Always select the correct spray nozzle before cleaning & lubrication It is preferable to purge any excess solution by briefly operating the handpiece. When purging cover the handpiece head with a tissue. Firmly attach the nozzle onto the valve at the top of the Pana Spray Plus can. Keep the spray can in an upright position High Speed Optic & Non Optic handpieces that connect direct to: NSK type nozzle P183-090
Kavo® type nozzle Z090-051
W&H® type nozzle Z090-054
Sirona® type nozzle Z090-052
Bien-Air® type nozzle Z090-053
QD type nozzle Z096-090
Contra-angles, Handpieces, Shanks & Heads:
E Type (all brands) Z019-090
ISO E Type (all brands) and NSK Surgical Heads SGM-I, SGMS-I & SGQ-H Z114-090
NSK HEADS used with Endo-Mate DT, TC, TC2 and TASKAL Wizard Z256-090
Tip nozzle for Borden 2/3 and Midwest 4 hole turbines and Borden 2/3 and Midwest 4 hole air motors Z020-201
Tip nozzle for chuck mechanism maintenance Z020-201
Bien Air Coupling For High Speed handpieces that connect to Bien-Air® Unifix® couplings – Z280022 Bien-Air® F-Type Head For ENDO-MATE DT, TC / TASKAL Wizard Heads – Z280-042 F-Type
Product Information | Technical Support 1300 44 33 21 (Australia) or 0800 44 33 21 (New Zealand) Email: [email protected] www.nskoceania.com.au Kavo® and MULTIflex LUX® are registered trademarks of Kaltenbach & Voigt GmbH & Co., Germany. Sirona® is a registered trademark of Sirona Dental Systems GmbH, Germany. W&H® and Roto Quick® are registered trademarks of W&H Dentalwerk Bürmoos GmbH, Austria. Bien-Air ® and Unifix® are registered trademarks of Bien-Air Dental S.A., Switzerland.