Air Motor Reprocessing Instruction
2 Pages
Preview
Page 1
Reprocessing Instruction in accordance with ISO17664
Air Motor 1 Preparation
20sec. Cleaning Wipes
2 Cleaning, Disinfection and Drying Manual 38°C
60sec.
3.5L/min 500lx Minuten Wipes (ALPRO)
M4
B2
WL-Adapter 01 (ALPRO)
Minuten Wipes (ALPRO)
WL-clean (ALPRO)
WL-Adapter 01 (ALPRO)
M4
B2
WL-dry (ALPRO)
M4 WL-Adapter 01 (ALPRO)
B2
WL-Adapter 01 (ALPRO)
WL-cid (ALPRO) WL-Blow (ALPRO)
Automated ISO 15883-1 500lx Minuten Wipes (ALPRO)
・Automated cleaning can only be performed on the FX205m / S-Max M205. ・After cleaning remove any remaining moisture using a dry cloth or compressed air ensuring it is dry internally and externally. ・Cover the product with a cloth when drying with compressed air to prevent water aerosol from being produced. ・Use a nationally approved disinfectant with proven bactericidal, fungicidal, and virucidal properties.
3 Lubrication REF
2~3sec.
Z020010
REF
Y900084
NSKPANASPRAYPlus
REF
M4
Z182100
B2
4 Packaging, Sterilizing and Drying
ISO 11607-1 Use a sterilization pouch that conforms to ISO 11607-1. Seal the pouch. Maximum temperature for NSK instruments is 135°C.
Type
Class N
Class B
Temperature
132°C
134°C
FullCycleTime
≥15min
3-18min
DryingTime
≥30min
≥30min
Perform steam sterilization using the conditions above. Use a steam sterilizer to perform sterilization.
Steam sterilization
CAUTION Plasma
EO
5 Storage
OM-DM0904EN 000
2020.07.03 00001 N