NSK

Air Motor Reprocessing Instruction

2 Pages

Reprocessing Instruction in accordance with ISO17664  Air Motor 1 Preparation  20sec. Cleaning Wipes  2 Cleaning, Disinfection and Drying Manual 38°C  60sec.  3.5L/min 500lx Minuten Wipes (ALPRO)  M4  B2  WL-Adapter 01 (ALPRO)  Minuten Wipes (ALPRO)  WL-clean (ALPRO)  WL-Adapter 01 (ALPRO)  M4  B2  WL-dry (ALPRO)  M4 WL-Adapter 01 (ALPRO)  B2  WL-Adapter 01 (ALPRO)  WL-cid (ALPRO) WL-Blow (ALPRO)  Automated ISO 15883-1 500lx Minuten Wipes (ALPRO)  ・Automated cleaning can only be performed on the FX205m / S-Max M205. ・After cleaning remove any remaining moisture using a dry cloth or compressed air ensuring it is dry internally and externally. ・Cover the product with a cloth when drying with compressed air to prevent water aerosol from being produced. ・Use a nationally approved disinfectant with proven bactericidal, fungicidal, and virucidal properties.