Reprocessing Instruction in accordance with ISO17664
Coupling 1 Preparation 20sec.
Cleaning wipes
2 Cleaning, Disinfection and Drying Manual 38°C
3.5L/min
500lx
15sec.
5sec.
15sec. Coupling
Coupling Wrench / Wrench
*PTL-CL-LED and KCL-LED are shown as typical models.
2sec.
M4
B2
3times 1min
WL-Adapter 01 (ALPRO)
Minuten Wipes (ALPRO)
WL-clean (ALPRO)
WL-Adapter 01 (ALPRO)
3sec. 1time
3sec.
M4
B2
WL-Adapter 01 (ALPRO)
1time
WL-cid (ALPRO)
WL-dry (ALPRO) 2min
M4
B2
WL-Adapter 01 (ALPRO)
3sec. 1time
WL-Blow (ALPRO)
・After cleaning remove any remaining moisture using a dry cloth or compressed air ensuring it is dry internally and externally. ・Cover the product with a cloth when drying with compressed air to prevent water aerosol from being produced. ・Use a nationally approved disinfectant with proven bactericidal, fungicidal, and virucidal properties.
3 Packaging, Sterilizing and Drying
ISO 11607-1 Use a sterilization pouch that conforms to ISO 11607-1. Seal the pouch. Maximum temperature for NSK instruments is 135°C.
Type
Class N
Class B
Temperature
132°C
134°C
FullCycleTime
≥15min
3-18min
DryingTime
≥30min
≥30min
Steam sterilization
CAUTION Plasma
Perform steam sterilization using the conditions above. Use a steam sterilizer to perform sterilization.
EO
4 Storage
OM-DT0926EN 000