Surgic Pro Motor with Motor Cord Reprocessing Instruction
2 Pages
Preview
Page 1
Reprocessing Instruction in accordance with ISO17664
Surgic Pro Motor with Motor Cord 1 Preparation Max.
38°C
20sec. Cleaning wipes
=15sec.
=20mm
38°C
2 Cleaning, Disinfection and Drying Manual 38°C
3.5L/min
500lx
15sec.
15sec.
60sec.
20sec.
15sec.
Minuten Wipes (ALPRO) 2min
3sec. 1time
2min
WL-cid (ALPRO)
Automated ISO 15883-1 500lx
・After cleaning remove any remaining moisture using a dry cloth or compressed air ensuring it is dry internally and externally. ・Cover the product with a cloth when drying with compressed air to prevent water aerosol from being produced. ・Use a nationally approved disinfectant with proven bactericidal, fungicidal, and virucidal properties.
3 Packaging, Sterilizing and Drying
ISO 11607-1 Use a sterilization pouch that conforms to ISO 11607-1. Seal the pouch. Maximum temperature for NSK instruments is 135°C.
Type
Class N
Class B
Temperature
132°C
134°C
FullCycleTime
≥15min
3-18min
DryingTime
≥30min
≥30min
Perform steam sterilization using the conditions above. Use a steam sterilizer to perform sterilization.
Steam sterilization
CAUTION Plasma
EO
4 Storage