NSK
VarioSurg Handpiece with Cord Reprocessing Instruction
2 Pages
Preview
Page 1
Reprocessing Instruction in accordance with ISO17664
VarioSurg Handpiece with Cord 1 Preparation
Auto Cleaning
Cleaning wipes Distilled or deionized water
2 Cleaning, Disinfection and Drying Manual 38°C
3.5L/min
500lx
20sec.
60sec.
WL-Adapter U-1 + U-1a (ALPRO)
20sec.
2sec. 3times
Tube
WL-clean (ALPRO)
1min
Minuten Wipes (ALPRO)
WL-Adapter
3sec.
U-1 + U-1a (ALPRO)
1time
Tube WL-Adapter U-1 + U-1a (ALPRO)
3sec. 1time
WL-dry (ALPRO)
Tube
2min
WL-cid (ALPRO)
WL-Adapter
3sec.
U-1 + U-1a (ALPRO)
1time
Tube
WL-Blow (ALPRO)
Automated
ISO 15883-1
500lx
A815 (Miele)
・Cover the product with a cloth when drying with compressed air to prevent water aerosol from being produced. ・Use a nationally approved disinfectant with proven bactericidal, fungicidal, and virucidal properties.
3 Packaging, Sterilizing and Drying ISO 11607-1
Type
20001326
Class N
Steam sterilization
Class B
Temperature
132°C
134°C
FullCycleTime
≥15min
3-18min
DryingTime
≥30min
≥30min
CAUTION Plasma
EO
Perform steam sterilization using the conditions above. Use a steam sterilizer to perform sterilization.
Use a sterilization pouch that conforms to ISO 11607-1. Seal the pouch. Maximum temperature for NSK instruments is 135°C.
4 Storage
OM-DE0944EN 000
2020.07.03 00001 N