> Insert the drip chamber of the feeding set into the drip detector slot of the pump.
2
> Slowly open the roller clamp on the feeding set and allow the total set to fill with feed. Make sure the drip chamber is filled up about one-third.
3
> Stretch the tubing around the pump wheel and insert the plastic retainer into the bracket on the pump.
4
> Pull the tubing through the tubing guide.
Connect the feeding set to your tube and continue to operate the pump.
Operating instructions > Turn on the pump by pressing the “POWER/ON” key. The pump is in “HOLD” mode.
Turning on the pump Programming Continuous Feeding
INC
DEC
RUN
Programming Dose Feeding
DOSE INC
DEC RUN
Changing Programming
Viewing total volume delivered Turning off the pump
> Set the desired flow rate (ml/h) by pressing the “INC” or “DEC” keys. > Flow rates can be set from 1 to 300 ml/h, in 1 ml increments. > Press the “RUN” key to start the pump. > After programming the flow rate do not immediately press the “RUN” key. > Press the “DOSE” key to set the desired dose. > Press “INC” or “DEC” to program the correct dose (in ml). > Press the “RUN” key to start the pump.
RUN
> Press “HOLD” key, pump will stop, all memory and programming is retained. Use this to: change flow rate, alter dose setting, change feed, administer medication. > Press “RUN” to restart the pump.
VOL FED
> Press “VOL FED” key, display will indicate total volume delivered.
HOLD
> Turn on the pump by pressing the “OFF” key for ca 3 seconds.
Flocare 800 Problem solving instructions Error messages and instructions Please refer to detailed Instructions for Use for additional information Problem/Alarm
Likely cause
Instructions
HOLD
> The pump has been left in “hold” mode for 3.5 minutes.
> Reset the alarm by pressing the “HOLD” key, or > Continue feeding by pressing the “RUN” key.
SET
> The feeding set is incorrectly placed.
> Press the “HOLD” key, insert the plastic retainer into the bracket on the pump, and then “RUN”.
DOSE COMPLETE
> The pump has finished administering the programmed dose.
> Switch the pump off by pressing “OFF” key, or > Program the pump to administer a repeat dose by pressing “HOLD” and then “RUN”.
> The feeding container is empty.
> Change the feeding container or switch off the pump.
> The liquid level in the drip chamber is too high.
> Remove the set from the pump; hold the set upside-down so that feed runs back in to the container, until the chamber is ca. one-third full. Reload the set.
> The sensor area is dirty.
> Remove the drip chamber from the slot and clean the drip detector sensors with a damp cotton swab.
> There is an obstruction to flow, either in the feeding set or in the tube.
> Check the feeding set, ensure that the roller clamp is fully opened and that there are no kinks in the tubing, and/or > Remove the set from the pump and check if feed is flowing from top to bottom (with the roller clamp open), and/or > Check if your tube is clogged by flushing the tube.
LOW BATTERY
> Ca. 30 minutes of battery time remain.
> Connect the pump to the mains and charge the pump. During charging the pump can be used.
MALFUNCTION (E-1 to E-11)
> The pump needs repair.
> Contact your healthcare practitioner.
OCCLUSION NO FLOW