Flocare Infinity Short Manual
Preparation and start of the pump B A
1
2
3
4
1
> Open the pump door by pressing up on the lower wall (A) and rotating the door upwards (B) at the same time (fig. 1).
2
> Position the looped section of the silicone tubing around the rotor (fig. 2).
3
> Stretch slightly to seat the cassette into the pump (fig. 3).
4
> Close the pump door (fig. 4).
5
> The pump is now ready to be switched on
Operating instructions Clearing pump memory and priming the feeding set
ON/OFF CLR FILL SET
> When switching the pump on or off, press this key for 2 seconds. > Pressing this key will clear the pump memory. > Pressing this key for 2 seconds will automatically prime the feeding set.
> Connect the feeding set with the feeding tube and choose appropriate feeding program 1 or 2 . 1 Continuous
feeding
2 Continuous
feeding and pre-set dose
ml/h
>
ml/h
>
+ 400 ml/h – 1 ml/h
+ 400 ml/h – 1 ml/h
>
START/STOP
>
DOSE=VOL
>
+ 4000 ml – 1 ml
>
START/STOP
Flocare Infinity Problem solving instructions Error messages and instructions Please refer to detailed Instructions for Use for additional information Problem
Instructions
NO SET
> Press ON/OFF to switch off the pump > Insert the Flocare Infinity set into the pump following instructions for use > Restart the pump.
PUSH STRT
> The pump has been untouched in hold mode for over 3 minutes > Extend the hold mode by a further 3 minutes by pressing START/STOP > Program the pump and press START/STOP to start the pump.
END OF DOSE
> The programmed dose / volume has been administered > Switch off the pump by pressing ON/OFF for 2 seconds > Clear the memory by pressing . CLR
®
™
PROG
> No flow rate is installed > Make sure the correct flow rate is programmed.
DOOR
> The door is not correctly closed > Make sure the door is properly closed prior to starting a feeding program.
AIR
> Check if the set is inserted correctly and check the air sensor > Remove the air in the set by using FILL SET key.
BATT
> Connect the adaptor to the pump and mains and charge the pump for approx. 6 hours. During charging the pump can be used.
BATT E + F
OCC IN OCC OUT OCC OUT (repeated)
> Battery failure > Please refer the pump for service. > Press START/STOP > Check the feeding set (IN) or feeding tube (OUT) for blockages > Reinsert the set and start the pump, if necessary clean the sensors. > Stop the pump by pressing the START/STOP > Let it run briefly by repressing the START/STOP key > Stop the pump again with START/STOP key, ensuring there has been no OCC OUT alarm > Remove the set from the pump and reinsert it into the pump > Restart the pump by pressing START/STOP .
LOCK
> The programming feature is blocked in the set-up mode of the pump > Refer to instruction booklet to de-block.
ER01 - ER99
> Electronic error > Switch the pump off, make sure the pump door is closed and turn the pump back on > If the error persists, please refer the pump for service.
NO PLUG SYMBOL
> The wall outlet or the adaptor does not work > Connect to another wall outlet > If there is still no symbol, request replacement of the adaptor.
(while the pump is connected to the mains)