FLOCARE INFINITY PUMP SHORT MANUAL Preparing and setting up the pump
1
3
Open the pump door by pressing gently up on the lower wall (A) and rotating the door upwards (B) at the same time.
Stretch slightly to seat the cassette into the pump.
2
Position the looped section of the silicone tubing around the rotor.
Operating instructions Clearing pump memory and priming the feeding set
ON/OFF INFO
1 Continuous feeding
2 Pre-set dose
ml/h
ml/h
+ 400ml/h – 1ml/h + 400ml/h – 1ml/h
4
Close the pump door - you should hear a small click when it is closed properly. The pump is now ready to be switched on.
When switching the pump on or off, press this key for 2 seconds.
CLR
This will clear the previous volume. The rate and dose will remain unchanged.
FILL SET
Pressing this key for 2 seconds will automatically fill the feeding set. Connect the feeding set to the feeding tube and choose appropriate feeding program 1 or 2 .
DOSE=VOL
DOSE=VOL
CLR + 4000ml/h – 1ml/h
START/STOP
START/STOP
PROBLEM SOLVING INSTRUCTIONS If your pump alarms: STEP 1 STEP 2 STEP 3
Look at the pump and note the error message displayed on the screen. Push the START/STOP button to stop the alarm noise. Refer to the chart below for steps to resolve the problems.
Error messages and instructions Please refer to the detailed Instructions for Use booklet that comes with your pump for more information. Problem
NO SET
Instructions Press ON/OFF to switch off the pump. Insert the Flocare® Infinity™ set into the pump following instructions for use. Press ON/OFF to restart the pump.
PUSH START The pump has been untouched in hold mode for over 3 minutes. Extend the hold mode by a further 3 minutes by pressing START/STOP . Check the pump programming is correct the pump and press START/STOP to start the pump.
END OF DOSE The programmed dose/volume has been administered. Switch off the pump by pressing ON/OFF . PROG
This could be a result of holding your finger down on the CLR button for too long, or no flow rate has been installed. Make sure the correct flow rate is programmed.
DOOR
The door is not correctly closed. Make sure the door is properly closed prior to starting a feeding program.
AIR
Press START/STOP . Check if the set is inserted correctly and check the air sensor. Remove the air in the set by pressing the FILL SET key once. Once all visible air has been removed from the feeding set, press FILL SET to pause the pump movement. Recommence feeding by pressing START/STOP .
BATT
Connect the charging cable to the pump, and plug the adaptor in to an electrical outlet. Ensure that the adaptor (the part that connects to the electrical outlet) is clicked in to place, sitting flat and flush - a green light should display when properly connected. A full charge takes approximately 6 hours. The pump can be used whilst on charge.
BATT E + F
Battery failure. Please contact the pumps team to organise a pump service (contact information below).
OCC IN OCC OUT
Press START/STOP to silence the alarm. Check the feeding set for any kinks in the tubing or visible blockages. If required, remove the set from the pump and clean the pump sensors before reinserting the cartridge back in to place. Press START/STOP to recommence feeding.
LOCK
The programming feature is locked in the set-up mode of the pump. Refer to instruction booklet to unlock.
ER01 - ER99 Electronic error. Switch the pump off, make sure the pump door is closed and turn the pump back on. If the error persists, please refer the pump for service by contacting Nutricia.
NO PLUG SYMBOL
First check the adaptor is firmly pressed down on the area housing the two pins, you should hear a click when this is in place. Reconnect to the wall. If there is still no plug symbol on the pump and no green light on the adaptor, try using another wall outlet. If the pump is still not able to charge, request a replacement of the adaptor.