O-Two Medical Technologies Inc
e700 User Manual Rev 14 Aug 2017
User Manual
64 Pages
Preview
Page 1
e700 Automatic Transport Ventilator User Manual
e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
Table of Contents Chapter 1
Safety………………………………………………………………..…………………………….
1
Chapter 2
Intended Use………………………………………………..…………………………………
4
Chapter 3 Overview ……………………………………………………..………….……………………. 3.1 Control and Display Layout …………………..…………………..……….. 3.2 Function Keys ……………………………………………………..…..………..…… 3.2.1 ON/OFF…………………………………………………………..………..…… 3.2.2 Control Selection Knob………………………………..………..…… 3.2.3 Lock…………………………………………………………….……………..…… 3.2.4 Alarm Silence…………………………………………………………..…… 3.2.5 Waveform ………………………………………………………………..…… 3.2.6 Screen Brightness ……………………………………..…………..…… 3.2.7 Cancel ……………………………………………………..……..………..…… 3.2.8 Pause/Resume…………………………………………….………..……… 3.2.9 Manual / Inspiratory Hold …………………………..………..…… 3.3 External Connectors …………………………………………………………….. 3.4 Patient circuit ………………………………………….……………………………... 3.5 Display ……………………………………………………………………..…………….. 3.5.1 Screen Layout ……………………………………………..………..…… 3.5.2 Live Monitoring Parameters ………………………………..…… 3.6 Symbols & Notations ………………………………………………………..….. Chapter 4 Preparation For Use ……………………………………………………………..……..
5
4.1
5 6 6 6 7 7 7 7 8 8 9 10 10 11 11 13 15 16 16 16 16 17 18 18 19 21
Set up ………………………………………………………………………………..……… 4.1.1 Connecting electrical power supply ………………….……. 4.1.2 Installing / replacing the battery …………………….……… 4.1.3 Connecting the gas supply ………………………………..…….. 4.1.4 Connecting patient circuit ………………………………..……… 4.1.5 Turning Ventilator ON ………….…………………………………… 4.2 Pre-use check ……………………………………………….………………………… Chapter 5 Operating Instructions ………………………………………………………………… 5.1 Startup & setting the ventilation parameters ………………….. 21 5.2 Ventilation modes………………………………………………………..…………. 25 5.2.1 A/C V (Assist Control Ventilation) ……………………….….………. 26 5.2.2 SIMV (Synchronized Intermittent Mandatory 28 Ventilation) ……………………………………………………………………..……. 5.2.3 BiLVL (Biphasic Positive Airway Pressure). ………….………
5.2.5 CPR mode ……………………………………………………………..…….
30 32 34
Turning Ventilator OFF …………………………………………………….…..
37
5.2.4 CPAP (Continuous Positive Airway Pressure)………..……..
5.3
e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
Chapter 6 Post use ………………………………………………………………….……………………
38
6.1 Disconnect device after use …………………………………………………..
38
6.2 Storage ………………………………………………………….……………..………….…..
38
Alarms and Indicators …………………………………….…………………….……
39
7.1 Ventilation Alarms ………………………………………….………………….……..
39
7.2 Battery status indicator …………………………………………………….…….
41
7.3 LEDs ………………………………………………………………….…………………….…..
42
Chapter 8 Cleaning, Preventive Maintenance & Servicing …………….….………
43
Chapter 7
8.1 Cleaning & Disinfection ………………….………………………………………. 43 8.2 Charging the Battery
………………………………..………………………….
43
8.3 Ambient Air entrainment Filter …………………….………………………..
44
8.4 Preventive Maintenance and Servicing ……..…………………………
44
Chapter 9 Technical Data ………………………………………………………………………….…..
46
9.1 Specifications …………………………………………….……………..…….……..
46
9.2 Circuit Description …………………………………..………………..………….
48
9.3 Battery and Power Supply ………………………………………..…….……..
49
9.4 Battery Operating time ………………………………………………..….……..
50
9.5 Electromagnetic Compatibility ……………………………………….…….
51
9.6 Oxygen consumption ………………………………………………………….…..
53
9.7 Oxygen Concentration delivered against different back pressures …………………………………………………………………..………….…..
54
Chapter 10
Trouble Shooting ……………………………………………………………………….….
55
Chapter 11
Abbreviations and Acronyms………………………………………………….……
56
Chapter 12
Accessories ……………………………………………………………….……..……….……
57
Chapter 13
Warranty …………………………………………………………………………………….…..
58
e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
Chapter 1 Safety Warnings Federal Law restricts this device to sale by or on the order of a physician. The ventilator shall only be used for purposes specified under “Intended Use”. The ventilator should only be used by qualified personnel trained in its use. Strict adherence to all instructions contained within this manual is essential for safe use. During use, the patient must be constantly monitored by qualified personnel. Alternate means of ventilation must be available in case of power failure or malfunction. Keep away from open flames, sparks and grease/oil. To avoid the risk of fire or explosion this ventilator must not be used with flammable gases or anaesthetic agents. Operating the unit in a confined space will elevate oxygen levels. The ventilation setting will be interrupted during battery replacement while the device is in operational mode and the external power supply is not connected. Only use O-Two specified hoses, patient circuits and external power supplies to avoid affecting the output performance of the ventilator. Antistatic or conductive hoses or tubing are not used in the Ventilator Breathing System. Unauthorized modification of this medical device is prohibited. Do not disassemble or modify any part of the ventilator except where described in this manual. Any unauthorized disassembly will void the warranty. Do not use this ventilator in toxic environments as entrainment of ambient room air during spontaneous breathing or air mix mode may permit toxic gases to be delivered to the patient. Do not use this ventilator within a Magnetic Resonance Imaging (MRI, NMR, NMI) suite. Do not use this ventilator in hyperbaric (high pressure) chambers.
1 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
Do not use the external electrical power supply outdoors as moisture may affect its function. Do not contact the connector from the power supply and the patient simultaneously. The performance of this ventilator may be affected if used around portable and mobile RF telecommunication devices (cell phones) within the minimum distance specified in section 9.5 of this user manual. Intake port on the patient valve must be left unobstructed for proper function & safety. When using a bacterial filter or HME (Heat Moisture Exchanger) connect the bacterial filter or HME to the patient connection between the elbow and the endotracheal tube or face mask, note that this action will increase the dead space. Bacterial filters or HME may increase breathing resistances. Use pressure regulators maintaining a minimum output flow of 120 L/min at a dynamic pressure of 45 PSI to ensure the proper functioning of the ventilator. Operation of this ventilator outside the environmental condition range specified in this manual may result in reduction and or failure in the ventilator’s performance. In extreme temperature conditions the effect is not noticeable in terms of delivered ventilations but may cause excessive wear in ventilator or components over time. Extreme low temperature reduces operation time of the battery (Refer to 9.4 battery operating Time). Operation of this ventilator outside the supply pressures range detailed in this manual may result in reduction in the ventilator’s performance, component failure, low pressure alarm or possible loss of automatic cycling. Operation of this ventilator outside the electrical power range detailed in this manual may result in reduction or failure in the ventilator’s performance. Component failure, inadequate power of internal components may occur. Operation of this ventilator below sea level or above 4,000 m (13,000 feet) may result in reduction or failure in the ventilator’s performance, low pressure alarm or possible loss of automatic cycling. The ventilator is intended for use in the prehospital, intrahospital, interhospital and ground transport environments only. Connecting the patient circuit to a patient prior to turning on the ventilator may cause a calibration error. 2 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
Cautions
When the ventilator is not in use, always turn off the gas supply.
Never allow oil or grease to come into contact with any part of the cylinder, regulator, or ventilator.
After use, always ensure that all components are cleaned in accordance with the instructions provided in this manual. (See section 8.1 - Cleaning and Disinfection).
The use of this device in a carrying case may result in an increase in oxygen concentration or lower than intended ventilation volume when used in the 60 % air mix mode. When air mix mode is being used it is recommended that the ventilator be placed in its normal operating orientation and the air intake on the side of the ventilator is not obstructed.
Never operate the ventilator without an intake filter being fitted, otherwise particles may contaminate the ventilator and affect its function.
This ventilator must only be serviced by the manufacturer or its authorized service centers.
Device, single use patient circuits and battery packs should be safely discarded in accordance with local state and institutional laws and procedures. .
Spare O-Two patient circuits are recommended at all times.
The design of this Ventilator does not incorporate a negative pressure phase during either automatic or manual ventilation.
The external power supply and battery are component parts of the medical electrical equipment system.
The ventilator is considered as a high flow device as its maximum flow output at pressure of 40.6 PSI is around 75 L/min. It should only be connected to a pipeline system that allows for the indicated high flow to avoid interfering with the operation of adjacent equipment.
Always ensure that all components are assembled correctly and ready for use.
Exhaled volume of the patient can differ from the measured exhaled volume due to leaks around the mask.
When selecting very small tidal volumes during ventilation of infants with, take into consideration the dead space in the patient circuit. 3 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
Chapter 2
Intended Use
The e700 is a time-cycled, volume-constant and pressure-controlled emergency and transport ventilator designed for use in the pre-hospital, intrahospital, inter-hospital and transport settings. It is intended for use with Adult, child, infant patients with a tidal volume from 50 ml upwards who are in respiratory and/or cardiac arrest or respiratory distress and who require the ventilatory support.
4 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
Chapter 3 Overview 3.1 Control and Display Layout N M
A
B
C L D E
K F
G
I
H
J
Figure 1
A - Warning indicator B – Battery operation indicator C – External power indicator
H – Pause/Resume button I – Lock button J – Waveform selection button
D – Battery charging indicator E – Manual / Hold button F – Screen brightness button G – Alarm Silence button
K – Cancel button L – Confirmation indicator light M – Control Selection Knob N – ON/OFF button
5 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
3.2 Function Keys While all ventilation parameter settings are controlled by the Control Selection Knob (M) in Figure 1, there are a number of key membrane buttons which control additional ventilator functions: 3.2.1 ON/OFF To turn on the ventilator, Press the ON/OFF button (N) in Figure 1 for one second, during that second the associated green LED will start flashing at a high frequency. After 1 second the ventilator will turn on but without ventilation until the appropriate patient size symbol is selected. If the button is pressed and released for less than a second, the ventilator will remain OFF. To turn off the ventilator, Press and hold the ON/OFF button for 4 seconds, the green power LED will start flashing at a high frequency. After 4 seconds the ventilator will turn off and all ventilation will stop. If the button is pressed and held for less than 4 seconds, the ventilator will stay ON. Warning At approximately 2% of full battery capacity, the ventilator will not start when in the off position or will shut down when operating. 3.2.2 Control Selection Knob The Control Selection Knob (M) in Figure 1 is used to navigate between parameters, change modes, select primary function change when rotated, and to confirm function changes when pressed. Diagram below illustrates path of cursor when Control Selection Knob is rotated clockwise.
Control Selection Knob rotation contour Direction
for
clockwise
knob
rotation
6 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
3.2.3 Lock Lock function will disable all buttons and the Control Selection Knob except for the ON/OFF, Alarm Silence and Day/Night buttons, which are enabled at all times. To lock the key membrane or cancel the lock function: 1- Press Lock button (I) in Figure 1. The Lock symbol will be displayed on the screen. 2- To cancel the Lock function, press the Lock button (I) again. Note During lock function, if any locked button is pressed, the Lock symbol will flash while the locked button is pressed.
3.2.4 Alarm Silence Alarm Silence button (G) in Figure 1 will silence audible alarms for 120 seconds. It can also be selected when there is no alarm in order to silence potential alarms. This function is activated or deactivated by pressing the Alarm Silence button once. When selected, Alarm Silence symbol will be displayed on the screen. 3.2.5 Waveform Pressing Waveform selection button (J) in figure 1 will switch between the pressure and volume ventilation waveforms displayed on the screen. 3.2.6 Screen brightness Pressing this button will switch between 4 different screens as follows: 1- Light background with dark color text and waveform at 100 % light intensity. 2- Light background with dark color text and waveform at 35 % light intensity. 3- Dark background with light color text and waveform at 100 % light intensity. 4- Dark background with light color text and waveform at 35 % light intensity. Note: this feature is active in ventilation mode screens only.
7 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
3.2.7 Cancel The Cancel button (K) in Figure 1 allows the operator to return to the previous settings if the last unconfirmed changes in settings are not required.
3.2.8 Pause / Resume During activation of the Pause/ Resume button (H) in Figure 1, the ventilator will stop ventilating with all buttons kept active (if they are not locked) except Manual/ Hold button. To activate pause function, proceed as follows: 1. Press Pause/ Resume button (H). The Pause symbol will flash on the and the Confirmation indicator screen with the confirmation symbol (L) to guide users to activate pause function by pressing the Control Selection Knob (M). 2. The symbol will be flashing for 10 seconds and then disappear if the Control Selection Knob (M) is not selected. Users can also press the Cancel button (K) to quit this selection before 10 seconds. 3. Once activated, a flashing yellow pause symbol will be displayed on the screen and ventilator will stop ventilating. Note a. During Pause, there will be an audible alarm associated with the flashing yellow Warning indicator (A) in Figure 1 every 15 seconds. Users can press Alarm Silence button to disable the audible alarm for 2 minutes but the yellow Warning indicator will continue flashing every 15 seconds. b. During Pause, users can change and confirm new ventilation settings but no ventilation will occur unless Pause function is disabled. 4. To cancel Pause function, press Pause/ Resume button (H) again. The “Resume” symbol will flash on the screen with the confirmation symbol and the Confirmation indicator (L) to guide users to resume ventilation by pressing the Control Selection Knob (M). 5. When ventilation is resumed, the ventilator will recommence ventilation with current settings shown on the screen unless new set-up selections were made.
8 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
3.2.9 Manual/ Hold During exhalation phase, if Manual/ Hold button (E) in Figure 1 is pressed, a mandatory breath will be initiated and either the flow rate or set pressure control parameter will be delivered as long as the Manual/ Hold button is pressed or until I-time setting is achieved. After I-time if the button still pressed, the ventilator will switch to inspiratory hold function in which the ventilator will cut the flow but will keep the exhalation port closed in order to block exhaled gas from going to ambient resulting in the maintenance of lung pressure. The maximum inspiratory hold time is 6 seconds. After that time, the ventilator will switch to exhalation phase by opening airway pressure to ambient. Note Manual/ Inspiratory Hold function exists in all modes except CPAP and CPR.
9 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
3.3 External Connectors
T
O
S
R
P
O–
DC Input connector
P–
Air Intake filter
Q–
Sensor connector #1
R–
Sensor connector #2
S–
22mm gas output connector
T–
Gas supply input
Q Figure 2
3.4 Patient circuit U
W
V
X Y
Z
λ
U: Breathing control hose V: 2 pressure sensing hoses
Figure 3
W: one way intake valve X: Exhalation port Y: breathing valve Z: Flow sensor adapter λ: Elbow
10 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
3.5 Display 3.5.1 Screen Layout The screen is divided into 7 sections as shown below and each section is dedicated to display the following parameters: Section 1:
Battery status during charge and discharge.
Section 2:
Live ventilation parameters (Vte, Mve, Paw peak, Paw AV, Rate).
Section 3:
Ventilation Modes (A/C V, SIMV, BiLVL, CPAP & CPR).
Section 4:
Alarms/ Warnings.
Section 5:
Ventilation waveforms.
Section 6:
Set up parameters.
Section 7:
Confirmation request/ Patient effort/ Invalid or Conflict setting.
1
3
2 7
4
5
6
Figure 4 Note By changing the ventilation mode, section 6 of the display will change accordingly to reflect the default or set up parameters for each mode. The followings are screen layouts for each ventilation mode:
A/C screen layout
11 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
SIMV screen layout
BiLVL screen layout
CPAP screen layout
12 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
CPR screen layout
3.5.2 Live Monitoring Parameters The following Live Monitoring Parameters are displayed at the section 2 of the screen: Paw AV (cm H2O): Paw AV is the average patient airway pressure measured during the last 60 seconds. This measurement is monitored by the ventilator at all times and modes. The number on the display will be updated every 15 seconds. Mve (L): Minute volume is the total exhaled volume for the last 60 seconds as calculated using the last 8 breaths. The Mve will constantly change as the value is recalculated and displayed at the end of exhalation phase. When unit is first turned on or resumed after pause or on selecting a new mode, the Mve calculation will be based on first then second then third and so on until the 8th exhaled tidal volume when the above logic will be followed. Rate (BPM): Rate (BPM) is the rate at which breaths are delivered in one minute. It is the monitored breath rate calculated by measuring the time interval (Tb in seconds) between 2 breaths. Rate (BPM) = 60 / Tb. The number is updated after each breath. This number will be displayed for both mandatory and spontaneous breathing phases. Vte (ml): Tidal volume is the volume exhaled from the patient in mandatory, spontaneous or PSV (Pressure Support Ventilation) breaths. Vte is calculated by the measurement of the entire expired flow displayed as a volume. Vte display will be updated at the beginning of the next inspiratory phase (The end of exhalation phase).
13 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
Paw Peak (cm H2O): Peak air way pressure is the maximum pressure measured during the inspiratory phase. The displayed number on the screen represents the maximum pressure during mandatory inspiratory phase of A/CV, SIMV, BiLVL, CPR modes as well as spontaneous inspiratory phase of CPAP mode. This number will be updated at the end of each inspiratory phase.
Note Mve, Rate and Paw are not active during CPR mode and displayed with “--“
14 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
3.6 Symbols and Notations
Consult instructions for use. Warning! Risk of injury and possible negative patient outcome. Caution! Note:
Warns of material damage and negative patient outcome. Offers useful tips to assist in the proper use of the equipment. Confirmation Symbol Invalid Setting Symbol Conflict Setting Symbol Keep away from opened flames. No smoking around ventilator.
IPX4
Ingress protection rating: Splash-proof. Do not immerse. Class II equipment. Protection against electric doesn’t rely on Basic Insulation only, additional safety precautions such as Double or Reinforced insulation are provided.
Type BF applied part.
15 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version
Chapter 4 Preparation for Use 4.1 Setup 4.1.1 Connecting electrical power supply The e700 is designed to operate using one of the following power options:
Internal rechargeable battery pack AC to DC external power supply.
Caution A fully charged battery must be always installed for safety reasons, even when operating from an external power supply so that continuous ventilation is not interrupted in absence of external power. The use of batteries other than those specified may cause the ventilator to fail and/or endanger the patient and operator. 4.1.2 Installing / replacing the battery 1
Make sure the ventilator is turned off and unplugged from mains electrical supply.
2 Turn screw knob on battery compartment cover anticlockwise to open the cover downwards. 3 Disconnect the battery leads and pull out the battery pack using its stand-off. Never pull the battery pack by its leads. Caution Always use the battery stand off to pull out the battery pack, never pull the battery by its leads (Figure 5). 4 Insert the fully charged battery such that the battery stand-off is positioned upwards (as per illustration below), attach battery connectors. Close cover and turn screw knob clockwise to secure.
Battery Stand off Battery Leads
Figure 5 16 e700 User Manual (15PL1009 - Rev.14 28_8_2017)
English version