Ohio Medical Corporation
Continous Vacuum Regulators Instructions for Use Rev 5 Nov 2012
Instructions for Use
10 Pages
Preview
Page 1
Continuous Vacuum Regulators Instructions for Use MED
100 120 -mmHg
140
40
160
HIGH
20 180 200
F U L L V AC
ME D
120
mm Hg
140
H
40
100
Low Vacuum
G HI
80 60 L OW
LOW
80 60
160
20
6700-0371-000 (Rev.5) 11/2012
Table of Contents Safety Instructions... 2 Intended use... 2 Receiving/Inspection... 2 User Responsibility... 2 Definitions... 3 Regulator Identification... 3 Operation... 4 Cleaning... 8 Troubleshooting... 8 Warranty... 9 Safety Instructions This manual provides you with important information about the Continuous Vacuum Regulators (CVR). To ensure the safe and proper use of this device, READ and UNDERSTAND all of the safety and operating instructions. IF YOU DO NOT UNDERSTAND THESE INSTRUCTIONS, OR HAVE ANY QUESTIONS, CONTACT YOUR SUPERVISOR, DEALER OR THE MANUFACTURER BEFORE ATTEMPTING TO USE THE DEVICE. Intended Use: The vacuum regulator is intended to be used in the medical facility as a means to evacuate media (i.e. fluids) from the body. Receiving / Inspection: Remove product from package and inspect for damage. If product is damaged, DO NOT USE and contact your dealer or equipment provider. WARNINGS w This device is to be used only by persons who have been adequately instructed in its use. Do not use this device in the presence of flammable anesthetics. Static charges my not dissipate and a possible explosion hazard exists in the presence of these agents. User Responsibility: This Product will perform as described in this operating manual and accompanying labels and/or inserts, when assembled, operated, maintained and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product should not be used. Parts that are broken, missing, plainly worn, distorted or contaminated should be replaced immediately. Should such repair or replacement become necessary, Ohio Medical recommends that a telephone or written request for service advice be made to the nearest Ohio Medical Regional Service Center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Ohio Medical and by Ohio Medical trained personnel. The Product must not be altered without the prior written approval of Ohio Medical’s Quality Assurance Department. The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Ohio Medical. AAA A 12345 This alpha character indicates the year of product manufacture and when the serial number was assigned; “Y” = 1995, “Z” = 1996, “A” = 1997, etc. “I” and “O” are not used. CAUTIONS Only competent individuals trained in the repair of this equipment should attempt to service it. 2
6700-0371-000
2
Definitions Detailed information for more extensive repairs is included in the service manual solely for convenience of users having proper knowledge, tools and test equipment, and for service representatives trained by Ohio Medical. Not for field or transport use.* * The categories of Field and Transport Use are specifically defined in ISO® 10079-3, “Field” means accidents or emergencies outside the hospital. “Transport” means use in ambulances, cars and airplanes. These situations may expose the equipment to uneven support, dirt, water, mechanical shock and temperature extremes. Ohio Medical suction equipment has not been tested to comply with the specific requirements of these categories.
WARNING CAUTION Note Important MAX
= possible injury to patient or operator = possible damage to equipment = Provides additional information to clarify a point in the text. = Similar to a note but of greater emphasis = Attention. Alerts you to a warning or caution in the text. = maximum
High Flow High Vacuum
= high flow, high vacuum
High Flow Low Vacuum | (On) O (Off) kPa mL mm
= high flow, low vacuum = on = off = Kilopascal = Milliliter = Millimeter
Regulator Identification Continuous Vacuum Regulator 1 Figure 1 Mode Selector Switch Vacuum Gauge Suction Control Knob
Probe/Adapter Port Fitting Port O (Off)
I (On)
MED
400
60
600
700
70
300
100
50 120 60 -mmHg 40
140
30
80
High Flow High Vacuum 160
90
18010
40 20
80
20
100
200
HIGH
LOW
500
100
200 - kP
F U L L V AC
a
Hg -mm
3
6700-0371-000
3
Operation Equipment Set-up Insert the probe into the vacuum wall outlet. If the regulator is mounted elsewhere, connect a vacuum supply hose between the regulator’s probe adapter and the wall outlet.
WARNING Connection to positive pressure sources such as oxygen and medical air, even momentarily, could injure the patient or operator.
CAUTION: Connection to positive pressure sources such as oxygen and medical air, even momentarily, could damage the equipment.
Use hospital-supplied suction tubing between the end piece and the collection container, and between the patient port and the patient (minimum inside diameter is 6 mm [0.25 in.]). An Ohio Medical overflow safety trap and high flow suction filter should be used between the collection container and regulator to prevent contamination of the regulator. Hospital-supplied suction tubing must be used between the catheter and the collection bottle. Recommended minimum inside diameter is 0.25 in (6 mm). ISO 10079-3 (section 5.1.2) states that “the usable volume of the collection container shall not be less than 500 mL.” High Flow Suction Filters Hydrophilic Hydrophobic Pkg of 20 6730-0350-800 Pkg of 3 Pkg of 200 6730-0351-800 Pkg of 10 Pkg of 100
Tubing x tubing nipple 6730-0570-800 6730-0571-800 6730-0572-900
1/8” NPT x tubing nipple 6730-0580-800 6730-0581-800 6730-0582-800
Attaching the Overflow Safety Trap (OST) CAUTION
Trap fitting
To help prevent aspirate from entering the regulator, wall outlet and pipeline equipment, an overflow safety trap should be attached prior to its use. Aspirate in the regulator, wall outlet and pipeline equipment may impair their operation. The use of the overflow safety trap and suction filter will help prevent this and extend the life of suction equipment.
2 Figure 2 1. Raise the sleeve and insert the trap into the regulator fitting. 2. Turn the trap clockwise about one and a half turns to engage the threads. The trap does not need to be screwed tight; an O-ring in the regulator fitting provides a vacuum seal. The trap should rotate freely to allow the desired tubing positioning. 3. Lower sleeve to lock trap in position.
Regulator Sleeve
4
6700-0371-000
4
Operation DISS fitting 3 Figure 3 1. Insert trap into the regulator fitting. Situate the tubing in the desired position. 2. Turn the DISS wing nut clockwise to engage threads and tighten (there is no O-ring, so the vacuum seal depends on a tight connection).
Regulator Wing nut
Mode Selection 4 Figure 4 2 Mode Continuous I (On) - Suction can be adjusted with the suction control.
O (Off) - No suction supplied to the patient.
5 Figure 5 3 Mode Continuous MAX - Maximum, full-line vacuum is supplied to the patient. Note: Available on Three-mode vacuum regulators ONLY.
5
6700-0371-000
5
Intermittierende Absaugvorrichtung
16
14 12
20 22
24
180
Setting the suction level
0 0
26
Fu llV ac t o
100 Fu llV ac t o 760
80 10
60 8
40
6 4 2
20
Pa
160
18
-k
Hg
100
120 140
-m
m
Operation
(INTER)
0 (OFF)
6 Figure 6
(CONT)
1. Clamp the connective tubing to occlude the fitting port
Clamp
7 Figure 7 2. Turn the mode selector switch to I (On).
8 Figure 8 3. Rotate the suction control knob until the vacuum gauge indicates the required setting.
WARNING
The regulator fitting port or patient connector must be occluded when setting the prescribed suction level so that the patient does not receive higher than required suction.
Pre-Use Checkout Procedure WARNING
The Pre-Use Checkout Procedure must be performed before using the equipment on each patient. If the regulator fails any part of the Pre-Use Checkout Procedure, it must be removed from service and repaired by qualified service personnel.
All tests must be performed with the regulator in its normal (vertical) operating position, with a minimum supply vacuum of 500 mmHg (66.7 kPa). 1. Turn the mode selector switch to O (Off). 2. Rotate the suction control knob one full turn clockwise (increase). 3. Clamp the connective tubing to occlude the fitting port. The gauge needle should not move.
6
6700-0371-000
6
Operation 4. Unclamp the connective tubing. 5. Turn the mode selector switch to I (On). 6. Rotate the suction control knob fully counter-clockwise (decrease). 7. Clamp the connective tubing. The gauge needle should not move. 8. With the connective tubing clamped, increase the suction to 90 mmHg (12.0 kPa). 9. Slowly open and close the clamped tubing to create various flow rates through the regulator. Check that the suction level is maintained at ± 5% of full scale deflection when the tubing is clamped. For Standard Two Mode Continuous, go to step 11 10. High and Standard Three Mode Continuous ONLY a. Turn the mode selector to MAX. b. Clamp the connective tubing and check that the gauge needle indicates the maximum available suction. c. Turn the mode selector to | (On). 11. Low Continuous ONLY Clamp the connective tubing and rotate knob fully clockwise (increase) to verify the suction level does not exceed 135 mmHg ± 5 mmHg (18.0. kPa ± 0.7 kPa). 12. Reduce the suction level to zero and set the mode selector switch to O (Off).
Patient Setup
1. Make sure the Pre-Use Checkout Procedure has been performed.
9 Figure 9
O MAX
I
400 500
2. Clamp the connective tubing and turn the mode selector switch to I (On).
300 50
60 70
600
40
High Flow High Vacuum
80
200
30 20
90 700
10
100
100
- kP
a
-mm
E A
S
I
R
E
N
C
Hg
Clamp
Underwater Seal Line
2 cm H2O Made in USA
3. Set the prescribed suction level. WARNING The regulator must be occluded when setting the prescribed suction level so that the patient does not receive higher than required suction. 4. Turn the mode selector switch to O (Off). 5. Attach the connective tubing to the vacuum port of the collection container. 6. Turn the mode selector switch to | (On). Important: The Low models do not deliver suction in excess of 135 mmHg ± 5 mmHg (18.0. kPa ± 0.7 kPa). 7
6700-0371-000
7
Cleaning/Troubleshooting Routine cleaning of the regulator is recommended as a standard procedure after each use. Wipe all exterior surfaces with a solution of water and mild detergent. Should misuse occur resulting in accidental flooding of the regulator, the regulator may be sterilized using ethylene oxide (ETO). See section 5.2 (Sterilization) of the regulator service manual. After sterilization follow the service checkout procedures in section 8 (Service Checkout Procedure) of the regulator service manual. CAUTION
Do not steam autoclave or liquid sterilize the regulator. Severe impairment to the operation of the regulator will result. The only acceptable 13 method of sterilization is with gas (ethylene oxide).
WARNINGS
Following sterilization with ethylene oxide, parts should be quarantined in a well ventilated area to allow dissipation of residual ethylene oxide gas absorbed by the material. Aerate parts for 8 hours at 54°C (130°F). Follow your hospital sterilization procedure.
Clean and sterilize all suction equipment before shipment to ensure transportation personnel and service personnel are not exposed to any hazardous contamination. After patient use, regulators may be contaminated. Handle in accordance with your hospital’s infection control policy.
Troubleshooting If the regulator does not operate and you have performed the Pre-Use Checkout Procedure, the following procedures may be used to attempt to correct the problem. Problem
Possible Cause
No suction
Mode selector is in the O (Off) position or between positions
Remedy
Switch to I (On)
Leak in system
Check that lid is secure on the collection container
Check tubing connections
Suction control knob at full decrease
Rotate the suction control knob in the increase direction (clockwise)
Important: If the above actions do not correct the problems or other problems exist, refer to CVR service manual or refer servicing to qualified service personnel.
8
6700-0371-000
8
Warranty This Product is sold by Ohio Medical Corporation, a Delaware corporation (the “Company”) under the express terms of the warranty set forth below. For a period of SIXTY (60) MONTHS from the date the Company ships this Product to the customer, but in no event for a period of more than five years from the date of original delivery by the Company to an authorized dealer, this Product, other than its expendable parts is warranted to be free from functional defects in materials and workmanship and to conform in all material respects to the description for the Product contained in this operation manual, if this Product is properly operated under conditions of normal use, regular periodic maintenance and service is performed and repairs are made in accordance with this operation manual. The warranty period for all expendable parts of the Product is sixty (60) days from the date the Company ships the Product to the customer. The Company’s sole and exclusive obligation and customer’s sole and exclusive remedy under the above warranty is limited to repair or replacement, at the Company’s option, of the defective Product. The foregoing warranty shall not apply if the Product has been repaired or altered by anyone other than the Company or an authorized dealer; or if the Product has been subjected to abuse, misuse, negligence, or accident. The Company reserves the right to stop manufacturing any product or change materials, designs, or specifications without notice. This warranty is extended to only the initial customer with respect to the purchase of this Product directly from the Company or an authorized dealer as new merchandise. Dealers are not authorized to alter or amend the warranty of any Product described in this agreement. Any statements, whether written or oral, will not be honored or be made part of the agreement of sale. THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. THE COMPANY SHALL NOT BE LIABLE FOR INCIDENTAL, COLLATERAL, CONSEQUENTIAL, OR SPECIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOST PROFITS, OR LOSS OF USE. THE COMPANY’S LIABILITY, IN THE AGGREGATE, SHALL NOT EXCEED THE PURCHASE PRICE OF THE PRODUCT. In order to file a warranty claim, customer is required to return Product prepaid to the Company at 1111 Lakeside Drive, Gurnee IL, 60031 USA. As determined at the sole discretion of the Company, Products which qualify under the warranty will be repaired or replaced, at the Company’s option, and returned to customer via ground delivery at the Company’s expense. All claims for warranty must first be approved by Ohio Medical Corporation Customer Service Department. For US Domestic customer returns: [email protected] or 1-800662-5822 (Option 3). For International customer returns: [email protected] or 1-800-662-5822 (Option 3). Upon approval the customer service department will issue a Return Goods Authorization (RGA) number. An RGA must be obtained prior to commencement of any warranty claim.
Form No. 550022 (Rev.1) 2012
9
6700-0371-000
9
Customer Service, Distribution Center Technical Support, Sales and Service Equipment Service Center Ohio Medical Corporation 1111 Lakeside Drive Gurnee, IL 60031 USA P: 866 549 6446 P: +1 847 855 0800 F: +1 847 855 6218
© 2012 Ohio Medical Corporation. This document contains information that is proprietary and confidential to Ohio Medical Corporation. Use of this information is under license from Ohio Medical Corporation. Any use other than that authorized by Ohio Medical Corporation is prohibited. Ohio Medical, Ohio Medical Corporation, and the Ohio Medical Corporation Logo are registered trademarks of Ohio Medical Corporation.
6700-0371-000 11/2012 Rev5