Ohmeda
Giraffe Incubator Operators Manual Rev 102
Operators Manual
56 Pages
Preview
Page 1
Giraffe® Incubator Operator’s Manual
User Responsibility This Product will perform in conformity with the description thereof contained in this operating manual and accompanying labels and/or inserts, when assembled, operated, maintained and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product should not be used. Parts that are broken, missing, plainly worn, distorted or contaminated should be replaced immediately. Should such repair or replacement become necessary, Ohmeda Medical recommends that a telephone or written request for service advice be made to the nearest Ohmeda Medical Regional Service Center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Ohmeda Medical and by Ohmeda Medical trained personnel. The Product must not be altered without the prior written approval of Ohmeda Medical’s Quality Assurance Department. The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Ohmeda Medical. w U.S Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
CAUTION AAA A 12345
This alpha character indicates the year of patient probe manufacture and when the serial number was assigned; “A” = 1997, “B” = 1998, “C” = 1999, “D” = 2000, etc. “I” and “O” are not used.
0086
Ohmeda Medical has declared that this product conforms with the European Council Directive 93/42/EEC Medical Device Directive when it is used in accordance with the instructions provided in the Operation and Maintenance Manual. This symbol indicates that the waste of electrical and electronic equipment must not be disposed as an unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacture for information concerning the decommissioning of your equipment.
Table of Contents General Information
Control Modes...1-1 Controls and Displays...1-2 Operating with Two Temperature Probes...1-4 Accessory Options...1-5 Alarms ...1-12 Options Alarms...1-17 Cable Connections and Mechanical Controls...1-19 Alarm Table...1-21
Checkout
Mechanical Checks...2-1 Controller Checks...2-2 Humidifier pre-use checkout...2-3 Servo Oxygen pre-use checkout...2-3 Accessory Checks...2-3
Operating the Incubator
Basic Operating Procedure...3-1 Additional Operating Procedures...3-4 Accessing the Baby...3-4 Raising and Lowering the Bed...3-4 Tilting the Bed...3-5 Raising or Lowering the Hood...3-5 Using the X-ray Tray...3-6 Administering Oxygen...3-6 Installing Wristlets...3-6 Using the Humidifier...3-7 Thermalink Option...3-7 Using the Serial Data interface...3-7 Using the Nurse Call System interface...3-8 Using the In-Bed Scale...3-9 Mounting Accessories...3-9 Gas Cylinder Holder...3-10 Tubing Management Arm...3-10 Oscillator Port Cover...3-10
Maintaining the Incubator
Repair Policy...4-1 Maintenance Schedule...4-2 Disassembling the Incubator for Complete Cleaning...4-3 Cleaning and Disinfecting Individual Components...4-5
Accessory Options
Giraffe Accessories...5-1
Appendix
iii
Specifications... A-1 Performance... A-2 Accessories... A-3
6600-0354-000 102
iii
Table of Contents Definitions wWarnings Before using the Incubator, read through this entire manual. As with all medical equipment, attempting to use this device without a thorough understanding of its operation may result in patient or user injury. This device should only be operated by personnel trained in its operation under the direction of qualified medical personnel familiar with the risks and benefits of this type of device. Additional precautions specific to certain procedures are found in the text of this manual. Complete the Checkout procedures in this manual before putting the unit into operation. If the unit fails any portion of the checkout procedure it must be removed from use and repaired. Do not use the Incubator in the presence of flammable anesthetics; an explosion hazard exists under these conditions. Always disconnect the power before performing service or maintenance procedures detailed in this manual. Apply power only if you are specifically instructed to do so as part of the procedure. Thoroughly air dry the Incubator after cleaning it with flammable agents. Small amounts of flammable agents, such as ether, alcohol or similar cleaning solvents left in the incubator can cause a fire.
wCautions Only competent individuals trained in the repair of this equipment should attempt to service it as detailed in the Service Manual (6600-0356-000). Detailed information for more extensive repairs is included in the service manual solely for the convenience of users having proper knowledge, tools and test equipment, and for service representatives trained by Ohmeda Medical. Note: A Note provides additional information to clarify a point in the text. CAUTION: A Caution statement is used when the possibility of damage to the equipment exists. WARNING: A Warning statement is used when the possibility of injury to the patient or the operator exists.
Symbols
w m y x
~ MAX
Attention! Consult the manual for more information. Type B equipment. Functional Earth Terminal Protection Earth Terminal Alarm Silence Alternating Current Maximum European Union Representative
iv
6600-0354-000 102
iv
General Information Giraffe® Incubator
CI.31.084
The Giraffe Incubator combines state of the art microprocesor control with enhanced thermal performance and a number of unique new features. These include doors that allow access from either side, highly visible displays, controls that can be accessed from both sides of the incubator, and a much larger bed surface with a rotating mattress option.
Control Modes The Giraffe Incubator has two control modes: the incubator can control the inside temperature comparing air temperature readings from the probe on the compartment wall to a setting you enter using the control panel (air mode) or from a probe attached to the baby’s skin (baby mode).
1-1
6600-0354-000 102
1-1
General Information Controls and Displays
1 10 9
3
8 7
4 CI.31.092
The controls and displays, located at the top of the uprights at the head of the bed, can be divided into two areas: on the left are the buttons and numeric temperature displays that control the temperature regulation of the infant, on the right are the graphics display screen and control knob that control all the accessory options.Temperatures can be shown in Celsius or Fahrenheit degrees: the default setting is Celsius.
2
5
6
1. Alarm Light 2. Alarm Silence Button 3. Options Control Knob 4. Graphics Screen 5. Options Displays and Controls 6. Temperature Regulation Controls 7. ControlModeButtons 8. Temperature/Power controls 9. Numeric Displays 10. Power Failure Indicator 2
Temperature Regulation
3
1
At start up, the Incubator prompts you to select one of the control modes. If the unit was used in the last 2 hours, it will ask if you would like to retain the settings you used previously. Select Baby or Air Mode.
4
10 9 5 8 6 7
1. Greater than 37°C Indicator 2. Greater than 37°C Button 3. Baby Mode Indicator 4. Baby Mode Button 5. Air Mode Indicator 6. Air Mode Button 7. Boost Air Curtain Button 8. Boost Air Curtain Indicator 9. Decrease Temperature 10. Increase Temperature
1-2
6600-0354-000 102
1-2
General Information Air Control Mode Select this control mode by pushing the Air Mode Button at the bottom of the temperature regulation controls. The Air Mode Indicator lights when air mode is selected. Two temperature values are shown in the lower third of the display: the value shown in the larger numerals at the top is the air temperature measured by the compartment probe, the value shown below in smaller numerals is the air set temperature. The set temperature can be adjusted using the temperature buttons located at the left of the mode buttons. Temperature settings from 20° to 37°C can be entered in 0.1°C increments. Temperatures from 37° to 39°C can only be entered after pressing the >37°C Button. Note: When operating in the air mode with a probe attached to the infant’s skin, the infant’s temperature will also be shown in the baby temperature section at the top of the display.
CI.31.103
4 3 2
1
1. Air Set Temperature 2. Air Temperature 3. Air Mode Button 4. Air Mode Indicator
2
Baby Control Mode 4
The Baby Mode Indicator lights when the baby mode is selected. Two temperature values are shown in the upper third of the display: the value shown in larger numerals at the top is the temperature measured by the probe on the infant’s skin, the value shown below in smaller numerals is the baby set temperature, which can be adjusted by using the temperature buttons located at the left of the mode buttons. Temperature settings from 35° to 37°C can be entered in 0.1°C increments. Temperatures settings from 37° to 37.5°C can only be entered after pressing the >37°C Button.
3
Note: When operating the in the baby mode, the compartment temperature will also be shown in the air temperature display section at the bottom of the display.
1-3
6600-0354-000 102
1
CI.31.104
Select this control mode by pushing the Baby Mode Button at the top of the temperature regulation controls
1. Baby Mode Button 2. Baby Temperature 3. Baby Set Temperature 4. Baby Mode Indicator
1-3
General Information Boost Air Curtain If the doors will be open for an extended length of time, pushing the Boost Air Curtain Button increases fan speed and improves open door thermal performance.
CI.31.105
An indicator on the button lights when the Boost Air Curtain is activated. Running the fan at higher speed, however, will raise the noise level in the infant compartment. To deactivate the Boost Air Curtain, push the button a second time to resume Whisper Quiet™ mode. The Boost Air Curtain will automatically shut off after twenty minutes and the indicator light will go out. The button must be pressed again to continue for another twenty minutes at high speed. Note: The fan speed may increase automatically due to other requirements without pushing the Boost Air Curtain button, and in these cases the button indicator will not light.
1
Operating with Two Temperature Probes
CI.31.106
The Incubator is equipped with two patient temperature probe jacks for co-bedding of twins, or to monitor a single baby’s temperature from two anatomical sites. They are identified as probe 1 and probe 2 on the jack panel behind the bed. Temperature readings for probe jack 1 are displayed in Baby Temperature at the top of the numeric displays. Temperature readings from probe jack 2 are displayed in the upper right corner of the graphic display labeled Probe 2. The Incubator will only Baby Control from a probe attached to jack 1. When both jack 1 and jack 2 are used, the unit will not operate in the Baby Control mode. 1. Probe 2 Temperature
1-4
6600-0354-000 102
1-4
General Information Accessory Options The Incubator may be equipped with a number of accessory options. Menu icons of these options are shown on the graphics display at the right of the control area. Pressing the knob activates the screen. Rotating the control knob either direction scrolls through each menu icon; highlighting each icon as it is selected. Pressing in the control knob selects the highlighted option and brings up its menu screen. Note: Some of the features shown are optional. These icons will not appear on your display if your Giraffe is not equipped with that particular option.
2
3
4
14
13
5
12
6
11
7
10
8 9
1-5
6600-0354-000 102
1. Help 2. Timer 3. Trending 4. Comfort Zone 5. Setup/Customize 6. Scale 7. Day 8. Time 9. Date* 10. Baby Information 11. SpO2 12. Servo Oxygen 13. Humidity 14. Options CI.31.107
1
* The date may be set using the service screen; see the service manual.
1-5
General Information
CI.31.090
Help: Select the Help icon (question mark) to bring up the “Help” screen. The screen will present a menu of topics on how your Incubator unit works. “Alarms” list all the alarm messages on the left and more detailed descriptions of the alarm and how to react to a given alarm on the right. “Controls” identifies the various button and switches on the control panel. “Operation” lists basic procedures on the left and step by step instructions on the right. Select Done/Exit to return to the option icon screen.
CI.31.109
CI.31.091
CI.31.108
If you select the Help icon during an alarm condition, the description of that particular alarm will automatically appear on the screen; it is not necessary to scroll through the help menu options to find it.
1-6
6600-0354-000 102
1-6
General Information
CI.31.110
Timer: Select the Timer icon (clock face) to bring up the “Timer” menu screen. Select Start/Stop to start timer; select it again to stop the timer. Select Pause/Resume to pause display number while timer continues in back-ground. Scroll down and select Reset to return the timer to zero. Select the Apgar Tone symbol (musical note) to activate or deactivate a tone after one minute and every 5 minutes after the timer was started. Select Done/Exit to return to the option icon screen.
CI.31.111
Trend: Select the Trending icon (graph) to bring up the “Trending” screen. The graph will plot temperatures from two patient skin probes (if they are installed in the jacks), the air temperature read at the compartment probe, and the set temperature entered during the time period. The graph also shows what control modes the Incubator was in during the trended period. By selecting “Humid” in the “Display” option the “Modes” graph is replaced by a graph plotting the percent relative humidity read by the compartment humidity sensor. Turn the knob to scroll through 2, 8, 24, or 96 hours and press it to select the desired trending time period.
CI.31.096
CI.31.095
Note: The trending screen does not retain data that is not currently shown on the display. For example, if you have selected that only 2 hours of data be trended, after the display has been running for 3 hours, the first hour of data is lost.
1-7
6600-0354-000 102
1-7
General Information
Comfort zone: Select Comfort Zone icon (thermometer and face) to bring up the “Comfort Zone” screen.
CI.31.112
Important: The comfort zone screen is only a reference information screen used for calculating suggested temperature ranges* during air control mode from data you supply. You must manually enter the temperature settings using the buttons on the temperature regulation controls. Push the knob to select weight then enter the patient’s weight range in grams by turning the knob in 100 gram increments. Push the knob to enter the weight value. Scroll to Gest age. Push the knob to select Gest age then turn the knob to find the patient’s gestational age in weeks. Push the knob to enter the gestational age value and scroll down to Post natal age. Push the knob to select Postnatal age then turn the knob to find the age in days. Push the knob to enter the postnatal age.
CI.31.113
The suggested incubator air set temperature range will now appear in degrees Celsius (or Fahrenheit if you have changed the temperature unit default on the Setup Screen). If you wish to use this suggested temperature setting, enter it using temperature control buttons to the left of the screen. Select done/exit to return to the option menu screen. Note: The natal day, or day of birth, is considered day of life number one. * Adapted from Sauer, Dane and Visser, Archives of Disease in Childhood, 1984. 59, 18-22.
1-8
6600-0354-000 102
1-8
General Information
Setup/Customize: Select Setup/Customize icon (wrench) to bring up the setup screen. Select “Temp Units” to change temperature displays from the default Celsius setting to Fahrenheit. Select “Alarm Volume” to choose from 4 volume levels. CI.31.114
Select “Man Temp Alarm” to choose from 3 settings: choose “Off” to deactivate the alarm, choose “.5” to set the alarm to activate when a 0.5°C difference is read between a set temperature and the baby probe temperature, or choose “1.0” to set the alarm to activate when a 1.0°C difference is read between a set temperature and the baby probe temperature. “Off” is the default setting. The “Man Temp Alarm” can be used in Air mode, however, to select a set temperature you must first select the Baby mode, input the set point, then select the Air mode. This optional alarm provides you audiovisual notification in the Air control mode if the skin probe temperature drifts outside the range you select. For example, if you select 0.5°C on the setup screen, and then set 36.5°C as the skin temperature threshold in the Baby mode before switching to the Air mode, the alarm will sound if the baby’s skin temperature falls below 36°C or rises above 37°C while in the Air mode. If this alarm is not required, or if the skin temperature probe is not being used, “Man Temp Alarm” should always be set to “Off” on the setup screen. Select “Patient Alarm” to choose between 2 settings: choose “.5” to set the alarm to activate when a 0.5°C difference is read between a set temperature and the baby probe temperature, or choose “1.0” to set the alarm to activate when a 1.0°C difference is read between a set temperature and the baby probe temperature.”1.0” is the default setting. The “Patient Alarm” is used in the Baby mode. Select “Set Clock” to change the time setting in hours, minutes and seconds on a 24 hour clock. Note: Early versions of the software have no “Set Clock” menu option on the set up screen. On these units the time must be set using the service screen. See the service manual on accessing the service screen. Select “Elevate” to disable or enable the foot pedals that raise or lower the bed height. Note: If the canopy or bed elevate pedals have been disabled on the service screen they cannot be enabled using the setup screen. Select “Baby Info” to bring up the Baby Info screen. By turning the knob until the desired character appears and then pushing the knob, you can enter up to a 9-character name. Blanks and periods can also be selected. Selecting “ ” allows you to go back and change the previous character. Selecting “ ” enters the letters you have selected for names that are less than 9 characters long. By selecting the month, day and year numbers, you can enter the baby’s birth date. These settings will be saved for 2 hours after power down Select Done/Exit to return to the option icon screen. Note: A service screen allows you to change alarm criteria and other default settings. See the service manual for a description of this screen and how it may be accessed. 1-9
6600-0354-000 102
1-9
General Information
CI.31.151
Servo Control Oxygen System is a fully integrated option available on the Giraffe® OmniBed® During operation noise levels in the infant compartment are imperceptible. Remote oxygen sensors mount inside the unit underneath the bed. Oxygen concentration is set using the OmniBed’s or Incubator’s display. Calibration, prompted once a day by the display, requires only a single push of a button. The Servo Control Oxygen System allows the administration of oxygen with virtually no intrusion to either you or the baby.
CI.31.152
Note: Calibration is required only if the Servo Control Oxygen System is activated. The Giraffe Servo Control Oxygen System uses fuel cell type sensors that generate specific voltages depending on the oxygen concentrations that are present. The microprocessor stores this measurement and compares it with the setting you enter using the display, opening and closing the valves that allow supply oxygen into the baby compartment. Fluctuations in fuel cell temperature and humidity are compensated for by the microprocessor. The temperature of the oxygen sensor housing under the bed is always slightly higher than the baby compartment, so condensation that would affect accuracy will not occur. Note: A FiO2 > 26% message indicates a leak or higher oxygen levels from other equipment. For more information see Option Alarms later in this chapter. Select the O2 icon to bring up the Servo Control Oxygen System screen. The actual concentration of oxygen, from the sensor underneath the compartment, is displayed in large numerals. The set percentage of oxygen by volume, ranging from 21 to 65%, is shown in smaller numerals. Adjust the set percentage to the desired level by pushing the knob to select the setting. Turn the knob to increase or decrease percentage then push the knob to enter the new setting. Exit the menu. The actual and set oxygen concentrations will remain displayed.
1-10
w The servo-control system must be calibrated at the same atmospheric pressure in which it is to be used. Operation at atmospheric pressures other than that present during calibration may result in readings outside the stated accuracy for the unit.
6600-0354-000 102
CI.31.148
CAUTION
CI.31.147
The Servo control system should be calibrated before each patient and at least once during every 24 hours. The system must be connected to an oxygen source during calibration. The display will prompt you when the system requires calibration. Calibration is initiated from the Set Up (wrench icon) screen by selecting Cal Oxygen. Once you start it, calibration is automatic and takes less than five minutes. A bar graph indicates progress toward completing calibration. If for any reason you wish to discontinue calibration before it is completed, turning the control knob in either direction will cause the word Cancel to appear on the calibration screen. Pushing in the control knob will discontinue calibration.
1-10
General Information
CI.31.116
CI.31.115
Scale: Available if the in-bed scale is installed. Select the Scale icon (balance scale) to bring up the scale menu screen. Select Weigh to start the weighing procedure. An audio prompt sounds and the screen will prompt you to lift the baby from the bed surface while the scale zeros. It then tells you to replace the baby on the bed and calculates the infant’s weight. Select Trend to plot a series of weights on a graph. Select Modify to change to smaller increments on the Y axis, and so enlarge the differences shown between weights plotted on the graph. Select Done/Exit to return to the option icon screen.
1-11
6600-0354-000 102
Options Done/Exit
? Humidifier
set
65%
65%
O2 O2
6/23/00
1-11
13:45
CI.31.188
Servo-Humidity: Select the Humidity icon to bring up the humidity menu screen. The actual relative humidity measured inside the infant compartment is displayed in large numerals. The set relative humidity is shown in smaller numerals. Adjust the relative humidity percentage to the desired level by pushing the knob to highlight the setting. Turn the knob to increase or decrease the humidity setting then push the knob to enter the new setting. Exit the menu. The actual and set relative humidity will remain displayed. To deactivate the humidifier, set the humidity level to “off”.
General Information 1 2
Alarms
3
An alarm message appears across the top graphics display screen identifying the various alarm conditions. In addition, the alarm light at the top of the control panel housing activates and an audio alarm sounds.
With the exception of the system failure and power failure alarms, all audio alarms can be silenced by pressing the alarm silence button located above the graphics screen. At the end of the alarm silence period the alarm reactivates if the alarm condition has not been resolved. If two or more alarm conditions occur at the same time, the audio signal for the highest priority alarm will sound while all the alarm messages will alternately appear at the top of the graphics screen. If you select the Help icon during an alarm condition, the description of that particular alarm will automatically appear on the screen.
CI.31.120
The audio tone varies with the alarm’s priority: an alternating two tone alarm indicates a critical alarm that triggers automatic heater shutdown; an intermittent single tone alarm is used for other alarms. The alarm light remains on continuously for over temperature and system failure alarms; it blinks for all other alarms.
1. Alarm Light 2. Alarm Silence Button 3. Alarm Display
High Air Temperature
CI.31.118
Active only in the air mode, the High Air Temperature alarm triggers when the temperature measured by the compartment probe exceeds the air temperature you set previously by more than 1.5°C. The audio alarm is suspended for a ten minutes when you push the alarm silence button. This alarm usually occurs when you have just lowered the air set temperature, or when you have changed from baby to air control mode. Monitor the unit closely to assure the air temperature decreases.
Low Air Temperature Active only in the air mode, the Low Air Temperature alarm triggers when the temperature measured by the compartment probe falls more than 3.0°C below the air temperature you set previously. The audio alarm is suspended for a ten minute period when you push the alarm silence button. This alarm usually occurs when the doors are open without using Air Curtain Boost, or when both doors are open at the same time. Monitor the unit closely to assure the air temperature increases.
1-12
6600-0354-000 102
CI.31.119
Note: The Low Temperature alarm is suspended for 60 minutes when the unit is first turned on, and 15 minutes after the set point is increased.
1-12
General Information Baby Hot - Check Probe 1
CI.31.121
Active whenever the unit is in the baby mode, the Baby Hot alarm activates when the temperature measured by the probe on the baby skin is 1.0°C above the baby temperature you set previously (this can be adjusted to 0.5°C using the Setup/ Customize screen). Monitor the patient carefully and check the attachment of the probe on the baby’s skin. The audio alarm is suspended for a ten minute period when you push the alarm silence button. This alarm can also be activated in the Air mode. See setup screen.
Baby Cold - Check Probe 1
CI.31.122
Active whenever the unit is in the baby mode, the Baby Cold alarm activates when the temperature measured by the probe on the baby skin is 1.0°C below the baby temperature you set previously (this can be adjusted to 0.5°C using the Setup/ Customize screen). Monitor the patient carefully and check the attachment of the probe on the baby’s skin. The audio alarm is suspended for a ten minute period when you push the alarm silence button. This alarm can also be activated in the Air mode. See setup screen.
Air Temperature Greater Than... Alarms These alarms warn of abnormally high temperatures read by the compartment probe in all control modes and automatically shut off the heater. They are not self resetting; once the alarms are activated they must be manually reset by pushing the alarm silence button, even if the alarm condition is resolved. Computer independent circuitry continuously measures and compares the compartment temperature with factory set levels. Visual and audible alarms are activated and the heater shuts off if the compartment temperature exceeds these preset levels, independent of the software. The functionality of all other alarms is continually checked by the microprocessor software during normal operation. If a fault occurs in any of this circuitry, an indicator lights and a tone is sounded. Additionally, a computer independent “watchdog” circuit will reset the computer, turn off the heater, and activate the alarms in the event of a microprocessor failure or software error.
Air Temperature >38 C
1-13
6600-0354-000 102
CI.31.123
Active only when the set temperature is below 37°C, the Air Temp >38°C alarm activates in the air mode when temperatures greater than 38°C are read by the compartment probe. This alarm may occur when you have just lowered the air set temperature, or when you have changed from baby to air control mode. Monitor the unit closely to assure the air temperature decreases.
1-13
General Information Air Temperature > 40 C In the air mode, when the air temperature is set above 37°C by using the >37°C key, the Air Temp >40°C alarm activates when temperatures greater than 40°C are read by the compartment probe.
CI.31.124
In the baby mode, the Air Temp >40°C alarm activates whenever temperatures greater than 40°C are read by the compartment probe. Check for external heat sources on the baby (sunlight, phototherapy lamp, etc.)
Disconnected Baby Probe 1
CI.31.125
Active in both modes, the Disconnected Baby Probe alarm activates when the baby skin probe jack is unplugged at the jack panel. Check that jack is fully inserted. Check that the Man Temp Alarm has not been set in error.
Fan Failure CI.31.126
The Fan Failure alarm triggers when the fan is missing or malfunctioning and there is inadequate air flow. The heater is automatically turned off. Remove the baby from the unit. Service required.
Fan Always in High Speed CI.31.127
The Fan Always in High Speed alarm activates when the heater temperature sensor malfunctions.
Air Probe Disconnect CI.31.128
The Air Probe Disconnect alarm activates when the compartment air temperature probe is disconnected. The heater is automatically turned off. Confirm connection of compartment probe to unit.
Air Probe Failure
CI.31.129
The Air Probe Failure alarm activates when the compartment air temperature probe malfunctions. The heater is automatically turned off. Remove the baby from the unit. Service required.
Baby Probe 1 Failure CI.31.178
Active in all modes, baby Probe 1 Failure alarm activates when the difference in the two thermistors in the skin probe in jack one is greater than 0.5˚C for over 6 minutes. Replace probe.
Baby Probe 2 Failure
1-14
6600-0354-000 102
CI.31.179
Active in all modes, baby Probe 2 Failure alarm activates when the difference in the two thermistors in the skin probe in jack two is greater than 0.5˚C for over 6 minutes. Replace probe.
1-14
General Information Bed Height Pedal Disabled CI.31.181
This alarm activates if the bed height pedals have been disabled on either the service screen or the set up screen, and one of the pedals is pressed. When the pedal is released, the alarm deactivates.
Baby Mode Disabled CI.31.183
This alarm activates if Patient Control is set to manual on the service screen and the Baby Mode button on the control panel is pressed. To deactivate the alarm, press the alarm silence button.
Disconnect Probe 2 for Baby Mode CI.31.184
This alarm activates when either the Giraffe is in the Baby Mode and a probe is inserted in Jack 2 on the probe panel, or when the Giraffe has 2 probes connected and the Baby Mode button is pushed. If the Giraffe is in the Baby Mode and a probe is inserted in Jack 2, the Giraffe will automatically switch to the Air Mode. To deactivate the alarm, remove probe from Jack 2. Note: the unit will continue to operate in the Air Mode until the Baby Mode button is pressed. If the Giraffe is in the Air Mode with probes in Jack 1 &2, and the Baby Mode button is pressed, you can deactivate the alarm by purshing the alarm silence button.
Temperature Out of Calibration
1-15
6600-0354-000 102
CI.31.185
This alarm activates if there is a temperature calibration error while turning on the unit; the unit will now use a default value to control temperature. Pressing the alarm silence will cancel the audio alarm, but the Temperature Out of Calibration message will remain on the display. You can continue to use the Giraffe while a baby occupies the bed, but the unit should be serviced before usining it with another patient.
1-15
General Information WARNING
CI.31.130
System Failure w Do not use the Incubator if the system failure alarm is activated. Remove the unit from service and refer to qualified personnel for repair.
If an electrical failure is detected, the system failure alarm automatically shuts off the heater and triggers a two tone audio alarm that can not be silenced.
Bad Membrane Switch 1
CI.31.137
One of the electronic switches behind the control panel buttons has failed. This alarm can not be silenced.
Power Failure If power to the unit is cut off, an audio alarm sounds and a LED lights. 1. Power Failure Indicator
Service Alarms The following alarm messages indicate a repair is required, but the Incubator can be used in another operating mode, the unit need not be taken immediately out of service. •
Bed Up Pedal Failure- A switch in the pedal is shorted. Continue to use at the current height.
•
Bed Down Pedal Failure- A switch in the pedal is shorted. Continue to use at the current height.
•
Motor Drive Failure- The drive system for raising and lowering the bed has failed. Continue to use the Incubator in its current operating mode. The humidifier will not operate when this alarm is activated.
1-16
6600-0354-000 102
1-16