MODE LOOKUP TABLE ESG-410 Quick Reference Guide
Cut Coag
MONOPOLAR
ESG-410
ESG-400
Cut Coag
ForceTriad
ForceFX
Pure
Pure
Pure
Pure
Blend
Blend
Blend
Blend
VIO 3
VIO 300 D
maXium
Mega Power
PureCut
PureCut
SwiftCoag
Swift Coag
AutoCut
Auto Cut
Blend 1, 2, 3
DryCut
Dry Cut
Super Blend
PulseCut Slow
EndoCut I/Q
Endo Cut I/Q
PulseCut Fast
PulseCut Fast
EndoCut I/Q
Endo Cut I/Q
FineCut
FineCut
PureCut
BlendCut
BlendCut
PulseCut Slow
Blend
Blend
Low
SoftCoag ForcedCoag
SprayCoag
SprayCoag
Spray
Spray
SprayCoag
Spray
PowerCoag
PowerCoag
High (Spray)
Spray
High (Spray)
Valleylab Fulgurate
Fulgurate
Fulgurate
Macro
Macro
BipolarCut
BipolarCut
PlasmaCut
SalineCut
BiSoftCoag
BiSoftCoag
Low
AutoCoag
AutoCoag
Low & “A” (“A” = start & stop)
SalineCoag
SalineCoag
Bipolar Resection Coag
HardCoag
HardCoag
LigaSure
LigaSure
Forced mixed PowerCut
Standard Laparoscopy
C-Cut Lap-C-Cut
Dry
Polyp I, II
Papillo Cut Gastro-Cut
GastroLoop GastroKnife
Polyp I, II
Papillo Cut Gastro-Cut
GastroLoop GastroKnife
Blend
ACE
Micro
SoftCoag
Soft Coag
Standard Coag
Moderate
Forced Coag
Forced Coag
Standard
Forced Coag
Forced non cutting
Spray Coag
SprayCoag
Spray Coag
Spray
Spray
Spray Coag
Spray
Spray Coag
SprayCoag
Spray Coag
Spray
Spray
Spray Coag
Pinpoint Coag
SwiftCoag
Swift Coag
Coag with Cut
Spray Forced mixed
Standard Coag
Bipolar Resection Cut Low (Precise)
ARC 400
ForcedCoag
Low (Dessicate) Spray
Coag with Cut
Precise Cut
Soft
SoftCoag ForcedCoag
Autocon III 400 C-Cut
Pure Cut
PowerCoag
PureCut
Valleylab
System 5000
Pinpoint Coag
PowerCut
Fulgurate
BIPOLAR
FT10
Micro Coag
AutoCut Bipolar
Bipolar Cut
Bipo Pure Cut
Biploar Cut
Standard
HighCut Bipolar
Bipolar Cut +
Uro Bip Pure Cut
Saline-C-Cut
Bipolar Resection (Cut)
SoftCoag Bipolar
Bipolar Soft Coag
Macro Coag
Standard Micro Macro
Bip. Soft Coag Bip. Soft Coag with Autostop
Bipolar scissors
SoftCoag Bipolar & Auto Stop
Bipolar Soft Coag & Auto stop
Macro Auto Stop
Standard Micro Macro
Bivascular-Safe
Bipolar scissors
SoftCoag Bipolar (Bipolar Resection)
Bipolar Soft Coag +
Bipol TUR Coag
Saline Coag
Bipolar Resection (Coag)
BiClamp
SealSafe
Bipolar scissors
How to choose the right settings 3-step approach: 1. Find the equivalent mode in the mode lookup table. 2. Set the same power level that was used with the previous device.* 3. Generally, start with effect 2*. (If small size electrodes are used, start with effect 1.)
ESG-410
ESG-400
FT10
ForceTriad
ForceFX
System 5000
VIO 3
VIO 300 D
maXium
Mega Power
Autocon III 400
ARC 400
* Except for PK modes or saline modes. Start with the default settings that are automatically set, when the respective instrument/cable is connected.
Manufacturer: Olympus Winter & Ibe GmbH Kuehnstr. 61
Disclaimer: The modes in this table are comparable on an application level (e.g. a pure cut mode without blend capability for smooth cutting with no or limited hemostasis). The mode lookup table should assist the user in selecting the ESG-410 mode which provides performance similar to modes experienced in a previously used
electrosurgical generator. Note that modes are never exactly the same between different generator manufacturers. The mode lookup table is merely intended to be a guideline that the surgeon can assess for applicability. Depending on individual circumstances, surgeons may need to use other modes which are different from those provided in the lookup table.