IMH-10 20 IMAGE MANAGEMENT HUB Service Bulletin Sept 2013
Service Bulletin
3 Pages
Preview
Page 1
SERVICE BULLETIN
OEM:
Preventative action to avoid failure of the HDD due to incorrect disconnection from the primary power supply Olympus Medical Systems Corp. TYPE: Medical
MODEL:
IMH-10/20
KM REF:
OKMSV-13-093
STATUS:
Information & Action
DATE:
05-09-2013
SUBJECT:
BACKGROUND There is a potential for the user to cause damage or failure of the internal HDD. This is caused by following the incorrect “Power Off” procedure. To reduce the chance of occurrence a Caution Label is now available and can be applied to the front panel, and an accompanying instruction sheet has been produced to provide guidance of the power off procedure. These are now supplied with all new products.
DETAILS The cause of failure is the IMH-10/20 being powered off using the main power switch of the trolley stack system or directly at the external mains supply. The power switch on the front of the IMH should always be used to power down the product. HDD failure can also occur if the power is removed before the shutdown procedure of the IMH-10/20 has completed.
continued…. Page 1 of 3
In order to determine that the IMH-10 has completed shutting down, the user will first see “FINISHING” displayed on the front panel followed by “GOODBYE”, once this is no longer displayed it is safe to remove the external power supply. To establish the IMH-20 has completed shutting down, the LCD touch panel will go blank. If the incorrect power off procedure has resulted in damage to the HDD one of the following error messages will be displayed: • • •
INITALIZE ERR (Activation error) SYSTEM ERR. 0001 (Communication error on application) SYSTEM ERR. 0302 0303 (Failure in database)
The unit will then be affected in the following ways: • • •
On power up it will display “Welcome” and fail to complete the start-up sequence. It will not be possible to store images in the external media and the remaining space will be displayed as 0KB. Any existing recordings will not be completed and the unit will freeze.
Caution labels and an instruction sheet are now supplied with all IMH-10 products from serial number 7331084 onwards, and IMH-20 products with serial number 7331572 onwards. The caution labels should be applied at installation and the end user should be made aware, highlighting the details on the instruction sheet. A copy of the instruction sheets are also attached to this bulletin for your reference. Users with models prior to this serial number should be provided with a copy of the relevant instruction sheet and the Caution Label applied to the front of the unit as soon as possible. The part numbers for the Instructions and labels are indicated as follows: IMH-10 (U), (E) IMH-20 (U), (E)
Instruction Sheet (English) Instruction Sheet (English)
GT9229 GT9232
Caution Label GT9234 Caution Label GT9234
continued…. Page 2 of 3
WARRANTY No warranty issue is applicable; Olympus will not assume any responsibility for HDD failure or any data lost.
SUMMARY OF ACTIONS Please distribute this information to your relevant personnel. All customer units prior to the change should be upgraded to have the caution label applied and the instruction sheet provided as soon as possible. Ensure caution stickers are applied and instruction sheet is highlighted to end user during installation. Attached: GT9229 - IMH-10 Instruction Sheet GT9232 - IMH-20 Instruction Sheet ISSUED BY: OLYMPUS KEYMED KeyMed House, Stock Road, Southend on Sea, Essex SS2 5QH www.olympus.co.uk Generated By: Sam Leslie – Service Engineering Approved By: Mr Martin Wallace Head of Field Service